The LungPoint Tools are endoscopic tools intended to be used with LungPoint Software guided bronchoscopes. The LungPoint Sheath is intended to be used as a working channel through which endoscopic tools may be introduced to targeted tissue. The LungPoint Dilation Balloon is intended to dilate tissue of the bronchial tree and may be inserted through the LungPoint Sheath or directly through the working channel of the bronchoscope. Not for pediatric use.
Device Story
LungPoint Tools consist of a sheath and a dilation balloon used during bronchoscopy. The sheath provides a working channel for introducing endoscopic tools to targeted lung tissue; the balloon dilates bronchial tree tissue. The bronchoscope is navigated to a predefined target using LungPoint Software (K12051). The sheath includes a stylet to improve pushability and prevent airway mucosa entry. The device is used in a clinical setting by a physician. The output is physical access to the target site or mechanical dilation of tissue, which assists the physician in reaching targets or performing interventions. Benefits include improved access to lung tissue and ability to dilate bronchial structures.
Clinical Evidence
Bench testing only. Testing included dimensional analysis, joint/tensile strength, simulated use, balloon fatigue/burst pressure, deflation time, and radiopacity. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity). All results met pre-determined acceptance criteria.
Technological Characteristics
Materials include medical-grade plastics and platinum iridium markers. Sheath includes a stylet. Dilation balloon rated for 20 atm. Sterilization via e-beam. Connectivity: None (mechanical accessory).
Indications for Use
Indicated for use with LungPoint Software guided bronchoscopes to provide a working channel for endoscopic tools or to dilate bronchial tree tissue in adult patients. Not for pediatric use.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
# 131234
## 510(k) Summary
1. Date of Summary
October 10, 2013
## 2. 510(k) Applicant
Broncus Medical, Inc. 1400 N. Shoreline Blvd, Suite A8 Mountain View, California 94043 (650) 428-1600 Phone: FAX: (650) 428-1542
OCT 15 2013
| Contact Person: | Gary Kaplan |
|-----------------|---------------------|
| Phone: | (650) 428-1600 |
| Fax: | (650) 428-1542 |
| e-mail: | gkaplan@broncus.com |
## 3. Device Overview
| Trade Name: | LungPoint™ Tools (LungPoint Sheath and LungPoint Dilation Balloon) |
|----------------------|---------------------------------------------------------------------|
| Common Name: | Sheath and Dilation Balloon |
| Classification Name: | Bronchoscope and Accessories<br>21 CFR 874.4680<br>Product Code EOQ |
# 4. Predicate Device
The predicate devices identified are as follows:
| Trade Name | 510(k) Submitter | 510(k) Number |
|----------------------------------------------|----------------------------------------|----------------------------------------------------|
| CRE Pulmonary Balloon<br>Dilatation Catheter | Boston Scientific<br>Corporation | K023337, cleared to market on<br>November 18, 2002 |
| Olympus Guide Sheath | Olympus Medical<br>Systems Corporation | K060243, cleared to market on<br>June 23, 2006 |
## 5. Device Description
The LungPoint Tools are endoscopic tools used during bronchoscopy procedures. The LungPoint Sheath is designed to be used with a bronchoscope to provide a working
{1}------------------------------------------------
channel through which endoscopic tools, such as needles, dilation balloons, or other endoscopic devices may be introduced to the targeted lung tissue within the respiratory organs. The LungPoint Dilation Balloon is used to dilate tissue of the bronchial tree and may be inserted through the sheath or directly through the working channel of the bronchoscope. The bronchoscope is advanced to a predefined target following guidance of the LungPoint Software (previously cleared under 510(k) #K 12051 to guide endoscopic tools or catheters in the lungs or to enable marker placement in soft lung tissue).
The materials used in the LungPoint Tools are commonly used medical grade materials and include platinum iridium markers.
| Catalog<br>Number | Working Length | Maximum<br>Catheter Outer<br>Diameter (OD) | Catheter Internal<br>Diameter (ID) | Minimum accessory<br>length for use through<br>sheath |
|-------------------|----------------|--------------------------------------------|------------------------------------|-------------------------------------------------------|
| 10007-1 | 900mm | 2.65mm | 2.0mm | 965mm |
The LungPoint Sheath has the following specifications:
The LungPoint Dilation Balloon has the following specifications:
| Catalog<br>Number | Balloon Size<br>(OD x length) | Rated Balloon<br>Pressure | Maximum<br>Catheter Outer<br>Diameter (OD) | Catheter<br>Length/working length |
|-------------------|-------------------------------|---------------------------|--------------------------------------------|-----------------------------------|
| 10008-1 | 4mm x 6mm | 20atm | 1mm | 1430mm/975mm |
# 6. Intended Use
The LungPoint Tools are endoscopic tools intended to be used with LungPoint Software guided bronchoscopes. The LungPoint Sheath is intended to be used as a working channel through which endoscopic tools may be introduced to targeted tissue. The LungPoint Dilation Balloon is intended to dilate tissue of the bronchial tree and may be inserted through the LungPoint Sheath or directly through the working channel of the bronchoscope. Not for pediatric use.
# 7. Comparison to Predicate Device
The LungPoint Tools are commonly used endoscopic tools with the same technological characteristics as the predicate devices. The indications for use of the LungPoint Tools for use with a bronchoscope guided by the LungPoint Software are all within the intended use of the predicate devices, which is to aid in reaching a target in the respiratory organ
{2}------------------------------------------------
either directly as is the case with the sheath or through dilating target tissue with the dilation balloon. The technological characteristics of the subject devices are the same as those of the predicate devices with the following exceptions
- LungPoint Balloon: balloon size and length, balloon burst pressure and balloon . catheter length as outlined.
| | CRE Pulmonary Balloon<br>Dilatation Catheter<br>(K023337) | LungPoint Dilation<br>Balloon |
|--------------------------------------------------|-----------------------------------------------------------|-------------------------------|
| Balloon Size and Inflation<br>(balloon pressure) | 8mm @ 3 ATM<br>9mm @ 5.5 ATM<br>10mm @ 9 ATM | 4mm @ 10 ATM |
| Catheter Length (cm) | 155 | 143 |
| Balloon Length (cm) | 3.0 | 0.6 |
| Rated Burst Pressure (atmospheres) | 9 | 20 |
- LungPoint Sheath: stylet, catheter ID/OD and catheter length as outlined. .
| | Olympus Guide Sheath<br>(K060243) | LungPoint Sheath |
|-----------------|-----------------------------------|-----------------------------------------------------------------------------------------|
| Stylet Provided | No | Yes - used to enhance<br>pushability and to prevent<br>airway mucosa entering<br>sheath |
| Catheter ID/OD | 2.1/2.7 mm | 2.0/2.6 mm |
| Catheter Length | 900 mm | 975 mm |
None of these differences raise new questions of safety and effectiveness. Performance of the subject devices has been verified by use of accepted methods.
### 8. Performance Data
The design and safety of the LungPoint Tools were verified by performing functional and performance testing. All tests were designed to subject the sheath and dilation balloon to stresses that exceed those which would be encountered during clinical use. Testing included the following:
- Dimensional testing -
- Joint/tensile test
- Simulated use -
- Balloon fatigue/burst pressure -
- Balloon deflation time -
- Radiopacity. -
All testing results met the pre-determined acceptance criteria that were established in the test protocols. Based on the testing the LungPoint Tools are as safe, as effective, and perform at least as safely and effectively as the predicate devices.
{3}------------------------------------------------
#### 9. Safety and Effectiveness
The LungPoint Tools labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the devices. The biocompatibility assessment of all patient contacting materials was performed in accordance with ISO 10993. Biological Evaluation of Medical Devices. Specifically, cytotoxicity, sensitization, intracutaneous reactivity and systemic toxicity (acute) were tested. In addition, the devices are sterilized using e-beam sterilization.
#### 10. Conclusion
Based on the testing the LungPoint Tools are as safe, as effective, and perform at least as safely and effectively as the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the United States. The logo consists of a stylized graphic element resembling an abstract human form or a bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
October 15, 2013
Broncus Medical, Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K131234
Trade/Device Name: LungPoint™ Tools (LungPoint Sheath and LungPoint Dilation Balloon) Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOQ Dated: August 26, 2013 Received: August 27, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Mark Job
""Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Eric A.Männ -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications for Use
510(k) Number (if known): K131234
Device Name:
LungPoint Tools - LungPoint Dilation Balloon and LungPoint Sheath
The LungPoint Tools are endoscopic tools intended to be Indications for Use: used with LungPoint Software guided bronchoscopes. The LungPoint Sheath is intended to be used as a working channel through which endoscopic tools may be introduced to targeted tissue. The LungPoint Dilation Balloon is intended to dilate tissue of the bronchial tree and may be inserted through the LungPoint Sheath or directly through the working channel of the bronchoscope. Not for pediatric use.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page __ of ___________________________________________________________________________________________________________________________________________________________________
Srinivaš
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.