K122495 · Perkinelmer, Inc. · MQB · Nov 28, 2012 · Radiology
Device Facts
Record ID
K122495
Device Name
PERKINELMER XRD 1622 AP3 MED FLAT PANEL DETECTOR
Applicant
Perkinelmer, Inc.
Product Code
MQB · Radiology
Decision Date
Nov 28, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The XRD 1622 AP3 MED, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.
Device Story
The XRD 1622 AP3 MED is a flat panel X-ray detector designed for integration into fixed radiographic imaging systems. It utilizes an amorphous silicon (a-Si) panel with a directly deposited Cesium Iodide (CsI:Tl) scintillator to convert X-ray photons into digital signals. The device features a 41cm x 41cm active area with a 200μm pixel pitch. Data and control signals are transmitted via a Gigabit Ethernet interface to the host X-ray system. The detector is operated by qualified clinicians or technicians in clinical settings to capture diagnostic images of human anatomy. The resulting digital output is transmitted to the imaging software of the X-ray unit for display and clinical review, facilitating diagnostic decision-making. The device is intended for general radiographic procedures but is explicitly contraindicated for mammography.
Clinical Evidence
Clinical study conducted with 30 image pairs across various anatomical regions. Images were reviewed by three board-certified radiologists. Results demonstrated that the device produces images of equivalent diagnostic quality compared to the predicate device.
Technological Characteristics
Amorphous silicon (a-Si) TFT/diode array; CsI:Tl scintillator; 41cm x 41cm active area; 200μm pixel pitch; Gigabit Ethernet connectivity; external power supply (XRD-LPM or XRD-EPS); complies with UL and IEC standards.
Indications for Use
Indicated for generating radiographic images of human anatomy for diagnostic X-ray procedures in patients where conventional screen-film, digital, or computed radiography is used. Not for mammographic use.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
GE Medical Systems Digital Radiographic Detector (K042876)
Submission Summary (Full Text)
{0}------------------------------------------------
## PerkinElmer Medical Imaging XRD 1622 AP3 MED Flat Panel Detector
NOV 2 8 2012
# 510k Summary
This summary of 510(k) safety and effectiveness information is supplied in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510 (k) number is K122495
# 807.92(a)(1):
Date: November 17, 2012
| Submitted by: | PerkinElmer Medical Imaging<br>2175 Mission College Blvd.<br>Santa Clara, CA 95054 |
|-----------------------------|------------------------------------------------------------------------------------|
| Contact Person:<br>Primary: | Kay A. Taylor<br>Tele: 317 418-1735<br>Fax: 317 536-3064 |
| 807.92(a)(2): | |
| Trade Name: | PerkinElmer XRD 1622 AP3 MED Flat Panel Detector |
| Common Name: | Flat Panel Digital Detector (21 CFR 892.1650) |
| Regulation: | |
| Classification Name: | Solid State X-ray Imager |
| Classification: | 90 Radiology |
| Product Code: | (MQB) |
| 807.92(a)(3): | |
| Predicate device: | Definium 5000 X-ray system (Flat Panel Detector in system)<br>[K063283] |
| 807.92(a)(4): | |
# Device Description:
The XRD 1622 AP3 MED is a flat panel x-ray detector consisting of an amorphous silicon panel with a directly deposited CsI:Tl scintillator.
The XRD 1622 AP3 MED detector has an active area of 41cm x 41cm at a pixel pitch of 200μm. Data and control communication is accomplished via a Gigabit Ethernet interface.
1
{1}------------------------------------------------
The detector can be integrated into a fixed room x-ray system to enable digital radiography. The following accessories are available for the XRD 1622AP3 MED
- Power supply XRD-LPM (for indoor usage only) .
- Power supply XRD-EPS (for indoor or outdoor usage) .
- DC cable for XRD-EPS (in lengths of 7.6 m, 15.3 m or 30.5 m) .
- AC cable for XRD EPS (country specific AC connector) ●
- Trigger cable (in lengths of 5 m or 20 m) .
- GigE interface cable (in lengths of 7.6 m, 15.3 m, 30.5 m, or 61m) .
#### 807.92(a)(5):
#### Intended Use:
The XRD 1622 AP3 MED, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.
#### 807.92(a)(6):
### Technological Characteristics Comparison:
Comparison of the XRD 1622 AP3 MED device with its predicate.
| | PerkinElmer XRD 1622 AP3 MED | |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | Proposed Device | Predicate (K063283) |
| Intended<br>Use/Indications for<br>Use | The XRD 1622 AP3 MED, when<br>used with a radiographic imaging<br>system, is indicated for use in<br>generating radiographic images of<br>human anatomy for diagnostic X-<br>ray procedures, wherever<br>conventional screen-film (SF),<br>digital radiography (DR), or<br>computed radiography (CR)<br>systems may be used. It is not<br>intended for mammographic use. | The Definium 5000 x-ray system is<br>intended for use by a qualified /<br>trained doctor or technician on both<br>adult and pediatric subjects for<br>taking diagnostic radiographic<br>exposures of the skull, spinal column,<br>chest, abdomen, extremities, and<br>other body parts. Applications can<br>be performed with the patient sitting,<br>standing, or lying in the prone or<br>supine position.<br>Note: The predicate submission<br>(K063283) was to combine two<br>previously cleared devices; the<br>Sedecal X Plus LS Plus Universal<br>Radiographic System and the GE<br>Medical Systems Digital<br>Radiographic Detector (K042876) to<br>which substantial equivalency is<br>claimed for the XRD 1622 AP3<br>MED. |
{2}------------------------------------------------
| Panel | Single substrate a-Si TFT/diode<br>array | Same |
|---------------|------------------------------------------|------|
| Scintillator | Direct deposit CsI:Tl | Same |
| Active area | 40.96cm x 40.96cm | Same |
| Pixel pitch | 200μm | Same |
| Power | External power supply | Same |
| Communication | Ethernet | Same |
#### 807.92(b)(1):
#### Summary of Non-Clinical Studies:
The PerkinElmer XRD 1622 AP3 MED flat panel detector has successfully completed internal nonclinical testing, complies with standards and regulations such as UL and IEC. The device has completed verification and validation testing to confirm it meets the specifications and operates as planned. Tests included image quality test with internal experts. The product, manufacturing and development processes have been showen to conform to product safety, radiology and imaging standards.
### 807.92(b)(2):
#### Summary of Clinical Studies:
The PerkinElmer XRD 1622 AP3 MED flat panel detector has successfully completed external testing in actual user testing facility and was found to produce images of equivalent diagnostic quality in a study of 30 image pairs of different anatomical regions reviewed by three board certified radiologists.
#### 807.92(b)(3):
#### Substantial Equivalency:
The proposed device and predicate device (flat panel detector of the predicate) both utilize similar technology and materials, are similar in design and construction, and have been shown to produce images of equivalent diagnostic quality in a clinical setting. Both devices are intended for use in generating radiographic images of human anatomy for diagnostic Xray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. The devices are not intended for mammographic use. Both devices produce digital images which can be transmitted to imaging software of the x-ray unit. ·
## Conclusion:
The PerkinElmer XRD 1622 AP3 MED is substantially equivalent to the flat panel detector of the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol is a graphic representation of a human figure embracing or supporting another, suggesting care and assistance.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 28, 2012
PerkinElmer C/O Ms. Kay A. Taylor, MT (ASCP) RAC Vice President, Quality and Clinical Affairs 8275 Carloway Road INDIANAPLOIS IN 46236
Re: K122495
Trade/Device Name: PerkinElmer XRD 1622 AP3 MED Flat Panel Detector Regulation Number: 21 CFR 892.1650 Regulation Name: ImageOintensified fluoroscopic x-ray system Regulatory Class: II Product Code: MQB Dated: November 17, 2012 Received: November 20, 2012
Dear Ms. Taylor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Michael D. O'Hara
Janine M. Morris, M.S. Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use 2.
510(k) Number (if known): K 122495
# Device Name: PerkinElmer XRD 1622 AP3 MED Flat Panel Detector
Indications for Use:
The XRD 1622 AP3 MED, when used with a radiographic imaging system, is indicated for The AKD 1022 AF 5 MED, which ases of him in and only for diagnostic X-ray procedures, use in generating radiographic inlages of thinks and only to: Chegary of Child Stars of the Property wherever conventional served. Inni (82), angel.
(CR) systems may be used. It is not intended for manmographic use.
# Prescription Use XXXX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE OF . NEEDED)
Concurrence of CDRH, Office of Device (ODE)
Michael D. O'Hern
(Division Sign Off)
Division of Radiological Ho Office of In Vitro Diagnostics a
510(k) 5122495
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.