Extension Set with BD Q-Syte Luer Access Split Septum
K142527 · Acacia, Inc. · FPA · Sep 19, 2014 · General Hospital
Device Facts
Record ID
K142527
Device Name
Extension Set with BD Q-Syte Luer Access Split Septum
Applicant
Acacia, Inc.
Product Code
FPA · General Hospital
Decision Date
Sep 19, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Non-Power Injectable: Extension Sets with an attached BD Q-Syte™ Luer Access Split Septum are intended to be used with intravascular catheters to aspirate blood or administer fluids, including medications and blood products. These devices may be used with any patient population with consideration given to the procedure being performed and fluids being infused. Power Injectable: The Power Injectable Extension Set with an attached BD Q-Syte™ Luer Access Split Septurn is intended to be used with intravascular catheters to aspirate blood or administer fluids, including medications and blood or blood products. These devices may be used with any patient population with consideration given to the procedure being performed and fluids being infused. The Power Injectable Extension Set is also indicated for use with intravascular catheters indication of contrast media. The Power Injectable Extension Set is suitable for use with power injectors set to a maximum pressure of 325 psi (2,240 kPa) when the BD Q-Syte device is removed and a direct connection is made.
Device Story
Extension sets consist of non-DEHP PVC tubing with slide clamps, male Luer lock distal connectors, and female Luer lock proximal connectors with attached BD Q-Syte needleless access valves. Used in clinical settings to provide a conduit between intravascular catheters and fluid sources/power injectors. Clinicians connect the device to a catheter to facilitate fluid infusion or blood withdrawal. Power injectable models support contrast media delivery at pressures up to 325 psi after removing the Q-Syte valve. The needleless design reduces accidental needlestick risk. No software or electronic components.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on identical technological characteristics and performance data from predicate devices.
Technological Characteristics
Non-DEHP PVC tubing; ISO male/female Luer lock connections; BD Q-Syte needleless split septum valve; lengths 5-21 inches; diameters 0.035-0.110 inch ID; non-latex; power injectable up to 325 psi (2,240 kPa) when valve is removed.
Indications for Use
Indicated for use with intravascular catheters for blood aspiration or fluid/medication/blood product administration in any patient population. Power injectable versions are indicated for contrast media injection at pressures up to 325 psi when the BD Q-Syte device is removed.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
MPS Acacia Power Injectable Infusion Set (K090809)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized image of an eagle or bird with three profiles of human faces incorporated into its design. The profiles are facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2014
Acacia. Incorporated C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo MN 55313
Re: K142527
Trade/Device Name: Extension Sets with BD O-Syte™ Luer Access Split Septum Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: September 5, 2014 Received: September 8, 2014
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/5 description: The image shows the text "Mary S. Runner -S" in a simple, sans-serif font. The text is horizontally aligned and appears to be part of a name or title. The letters are clear and legible, with consistent spacing between them.
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
Image /page/2/Picture/3 description: The image contains the alphanumeric string "K142527". The text is displayed in a clear, sans-serif font, making it easily readable. The characters are uniformly sized and spaced, contributing to the overall clarity of the image. The background is plain white.
Device Name
Extension Sets with BD Q-Syte™ Luer Access Split Septum
Indications for Use (Describe) Non-Power Injectable:
Extension Sets with an attached BD Q-Syte™ Luer Access Split Septum are intended to be used with intravascular catheters to aspirate blood or administer fluids, including medications and blood products. These devices may be used with any patient population with consideration given to the procedure being performed and fluids being infused.
Power Injectable:
The Power Injectable Extension Set with an attached BD Q-Syte™ Luer Access Split Septurn is intended to be used with intravascular catheters to aspirate blood or administer fluids, including medications and blood or blood products. These devices may be used with any patient population with consideration given to the procedure being performed and fluids being infused.
The Power Injectable Extension Set is also indicated for use with intravascular catheters indication of contrast media. The Power Injectable Extension Set is suitable for use with power injectors set to a maximum pressure of 325 psi (2,240 kPa) when the BD Q-Syte device is removed and a direct connection is made.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
SE CONSTITUTION OF THE FOR FOR FOR FOR FOR FOR FOR FOR THE FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 4. 510(k) Summary
This 510(k) summary is being submitted in accordance with 21 CFR 807.92.
| APPLICANT: | Acacia, Inc. |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| ADDRESS: | 785 Challenger Street<br>Brea, CA 92821 |
| PHONE: | (714) 257-0470 |
| CONTACT PERSON: | Dan Hyun |
| DATE PREPARED: | July 31, 2014 |
| TRADE NAME: | Extension Set with BD Q-Syte™ Luer Access Split Septum |
| COMMON NAME: | Intravascular Extension Set |
| CLASSIFICATION NAME: | Intravascular Administration Set |
| DEVICE CLASSIFICATION: | Class II |
| PRODUCT CODE: | FPA |
| PREDICATE DEVICES: | MPS Acacia Power Injectable Infusion Set (K090809)<br>MPS Acacia IV Extension Sets (K895367)<br>BD Q-Syte Saf-T PRN Luer Activated Valve (K013621) |
# Substantially Equivalent To:
The subject Extension Sets with BD Q-Syte™ Luer Access Split Septum are substantially equivalent in intended use, principal of operation and technological characteristics to the MPS Acacia Power Injectable Infusion Set (K090809), MPS Acacia IV Extension Sets (K895367) and BD Q-Syte™ Luer Access Split Septum (K013621).
All materials, manufacturing methods, technological characteristics, and principles of operation are identical to the predicate devices. The reason for this submission is to clarify and align the Indications for Use for all subject devices. The differences in the indications for use are not considered significant since they do not alter the therapeutic effect of the device, which is to provide a conduit for the infusion and withdrawal of fluids through an intravascular catheter.
# Description of the Device Subject to Premarket Notification
The Extension Sets with BD Q-Syte™ Luer Access Split Septum consist of a length(s) of flexible non-DEHP PVC clear tubing with a non-removable slide clamp. The devices include an ISO male Luer lock connection with a protection cap on the distal end an ISO female Luer lock connection on the proximal end(s) with a removable BD Q-Syte device or end cap. The Extension Sets with BD Q-Syte™ Luer Access Split Septum are provided in lengths from 5 in (127 mm) to 21 in (533 mm) with tubing diameters ranging from 0.035 in ID (0.89 mm) x 0.094 in OD (2.40 mm) to 0.110 in ID (2.80 mm) x .161 in OD (4.10 mm). Select Extension Sets include injection sites with a permanently attached BD Q-Syte device. The BD Q-Syte device is a needleless access device provided to aid in reducing the risk of accidental needlestick injuries. The Extension Sets with BD Q-Syte™ Luer Access Split Septum do not contain natural rubber latex.
The Power Injectable Extension Set with BD Q-Syte™ Luer Access Split Septum is suitable for use with power iniectors set to a maximum of 325 psi (2,240 kPa) when the BD Q-Syte device is removed and a direct connection is made. This device has a label attached to the tubing stating
{5}------------------------------------------------
"Power Injectable Max 325 PSI".
# Indications for Use:
### Non-Power Injectable:
Extension Sets with an attached BD Q-Syte™ Luer Access Split Septum are intended to be used with intravascular catheters to aspirate blood or administer fluids, including medications and blood or blood products. These devices may be used with any patient population with consideration given to the procedure being performed and fluids being infused.
#### Power Injectable:
The Power Injectable Extension Set with an attached BD Q-Syte™ Luer Access Split Septum is intended to be used with intravascular catheters to aspirate blood or administer fluids, including medications and blood or blood products. These devices may be used with any patient population with consideration aiven to the procedure being performed and fluids being infused.
The Power Injectable Extension Set is also indicated for use with intravascular catheters indicated for power injection of contrast media. The Power Injectable Extension Set is suitable for use with power injectors set to a maximum pressure of 325 psi (2,240 kPa) when the BD Q-Syte device is removed and a direct connection is made.
# Technical Characteristics:
The subject Extension Sets with BD Q-Syte™ Luer Access Split Septum have the same physical and technical characteristics compared to the currently marketed predicate devices.
# Performance Data:
Performance testing was conducted for the predicate devices to demonstrate the integrity and suitability of the device for its intended use. Since the subject devices and currently marketed predicate devices are the same, the results of the predicate testing indicate that the subject Extension Sets with BD Q-Syte™ Luer Access Split Septum are substantially equivalent to the currently marketed predicate devices.
# Basis for Determination of Substantial Equivalence:
Upon reviewing the information provided in this submission and comparing intended use, principle of operation and overall technological characteristics, the subject Extension Sets with BD Q-Syte™ Luer Access Split Septum are determined by Acacia to be substantially equivalent to the currently marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.