K013621 · Becton Dickinson Infusion Therapy Systems, Inc. · FPA · Apr 16, 2002 · General Hospital
Device Facts
Record ID
K013621
Device Name
BD SAF-T PRN LUER ACTIVATED VALVE, MODEL 385100
Applicant
Becton Dickinson Infusion Therapy Systems, Inc.
Product Code
FPA · General Hospital
Decision Date
Apr 16, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The BD Saf-T PRN Luer Activated Valve needleless connector is an accessory to an intravascular administration set that permits injection, gravity flow or withdrawal of fluids.
Device Story
BD Saf-T PRN Luer Activated Valve is a closed, needleless connector for intravascular administration sets. Device functions as an access site; opens upon insertion of standard male luer taper; closes automatically upon disconnection. Permits fluid injection, gravity flow, or withdrawal. Used in clinical settings by healthcare providers to facilitate vascular access while maintaining a closed system. Benefits include needleless access, reducing risk of needlestick injuries and microbial ingress.
Clinical Evidence
Bench testing only. No clinical data provided. Equivalence established through nonclinical performance testing including actuation force, luer torque and pull forces, flow rate, back pressure, insertion force, leak testing, hemolysis, and microbial ingress.
Technological Characteristics
Polycarbonate body with silicone seal. Needleless luer-activated valve mechanism. Compatible with standard luer slip and luer lock adapters. Standalone mechanical accessory.
Indications for Use
Indicated for use as an accessory to an intravascular administration set to permit injection, gravity flow, or withdrawal of fluids in patients requiring vascular access.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
BD Medical Systems 9450 South State Street Sandy, Utah 84070 tel: 801.565.2300 fax: 801.565.2740 www.bd.com
K013621
Image /page/0/Picture/3 description: The image shows the BD logo, which consists of a stylized sun-like symbol with a human-like figure at the bottom left. To the right of the symbol are the letters 'BD' in a bold, sans-serif font. Below the letters, there is the text 'indispensable to human health' in a smaller, sans-serif font.
# Premarket Notification [510(k)] Summary
| Submitter: | Becton Dickinson Infusion Therapy Systems Inc. |
|------------------------|--------------------------------------------------------|
| Address: | 9450 South State Street<br>Sandy, UT 84070 |
| Contact Person: | Leslie Wood<br>Manager, Regulatory Affairs |
| Telephone Number: | (801) 565-2504 |
| FAX Number: | (801) 565-2749 |
| Date Summary Prepared: | October 25, 2001 |
| Trade Name: | BD Saf-T PRN Luer Activated Valve |
| Common Name: | Luer Activated Valve |
| Classification Name: | Accessory to an Intravascular Administration Set |
| Predicate Device: | Clave® Connector C-1000<br>ICU Medical Inc.<br>K970855 |
### Description of the BD Saf-T PRN Luer Activated Valve:
The BD Saf-T PRN Luer Activated Valve is a closed system that opens when a standard male luer taper is inserted and closes automatically when the male luer taper is disconnected. The BD Saf-T PRN valve permits injection, gravity flow or withdrawal of fluids. It is compatible with standard luer slip and luer lock adapters.
### Intended Use of the BD Saf-T PRN Luer Activated Valve:
The BD Saf-T PRN Luer Activated Valve needleless connector is an accessory to an intravascular administration set that permits injection, gravity flow or withdrawal of fluids.
{1}------------------------------------------------
### Technological Characteristics Comparison:
The BD Saf-T PRN Luer Activated Valve and the Clave Connector are both IV administration set accessories that provide needleless access. Both products are (1) activated by the insertion of a luer taper, (2) built with a polycarbonate body and silicone seal; (3) have the same indications for use, and (4) similar performance characteristics.
## Nonclinical Tests Support Substantial Equivalence:
The following characteristics of the BD Saf-T PRN Luer Activated Valve and the Clave Connector C-1000 were shown to be substantially equivalent: actuation force, luer torque and pull forces, flow rate, back pressure, insertion force to access the device, leak testing, hemolysis, and microbial ingress.
#### Conclusions from Nonclinical Tests:
The BD Saf-T PRN Luer Activated Valve is substantially equivalent to the Clave® Connector C-1000 manufactured by ICU Medical.
The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without premarket approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure or a caduceus, with three abstract shapes that could represent people or snakes.
APR 1 6 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Leslie Wood Manager, Regulatory Affairs Becton Dickinson Infusion Therapy Systems, Incorporated 9450 South State Street Sandy, Utah 84070
Re: K013621
Trade/Device Name: BD SAF-T PRN Access Site, Model 385100 Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: February 27, 2002 Received: February 28, 2002
Dear Ms. Wood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 11 )) it in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration r od intilition (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and ilsung (21 et read on ) en alley systems (QS) regulation (21 CFR Part 820); and if requirences as bet relia product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and alle Free may of stance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Tim A. Ulatowsk Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE
Indications for Use:
21 CFR Part 880--GENERAL HOSPITAL AND PERSONAL USE DEVICES General Hospital and Personal Use Therapeutic Devices Sec. 880.5440:
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter. An administration set may include the needle or catheter, tubing, flow regulator, drip chamber, infusion line filter, IV set stopcock, fluid delivery tubing, connectors between parts of the set, side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an IV bag or other infusion fluid container.
The BD Saf-T PRN Luer Activated Valve needleless connector is an accessory to an intravascular administration set that permits injection, gravity flow or withdrawal of fluids.
Patricio Cucenite
(Division Sign-Off) ( Tivision Sign of Dental, Infection Control, and General Hospital Devices 510(k) Number --
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.