The Evolution® Mechanical Dilator Sheath Set is intended for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects. The Evolution® Shortie Mechanical Dilator Sheath Set is intended for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects. The SteadySheath™ Evolution® Tissue Stabilization Sheath is intended for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects. The SteadySheath™ Evolution® Shortie Tissue Stabilization Sheath is intended for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects.
Device Story
Mechanical dilator sheath sets used for percutaneous removal of cardiac leads/catheters; consists of coaxial sheaths with handle-actuated inner sheath rotation. Inner sheath rotates clockwise via trigger to disrupt binding scar tissue; outer sheath provides counter-traction to stabilize cardiac wall during lead detachment. SteadySheath variants include radiopaque, textured stainless steel tip for enhanced tissue stabilization and visibility. Operated by physicians in clinical settings. Benefits include improved disruption of binding scar tissue and stabilization during lead extraction procedures.
Clinical Evidence
Bench testing only. Testing included push and flex testing (50 cycles, 45°-90°), push and rotate testing (50 cycles), pull testing (minimum 10 lbs force), and sterility/bioburden/endotoxin/residual testing. All tests met predetermined acceptance criteria.
Technological Characteristics
Coaxial sheath design; inner sheath 9-13 French; working lengths 13.6cm or 40.6cm. Features handle-actuated mechanical rotation. SteadySheath variants incorporate radiopaque, textured stainless steel tips. Sterilization via ethylene oxide.
Indications for Use
Indicated for patients requiring percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters, and foreign objects.
Regulatory Classification
Identification
A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.
Predicate Devices
Evolution® Mechanical Dilator Sheath Set (K061000)
Evolution RL Controlled-Rotation Dilator Sheath Set (K141148)
Evolution Shortie RL Controlled-Rotation Dilator Sheath Set (K141148)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 12, 2014
Cook Vascular, Inc. % Thomas Kardos Vice President, Regulatory Affairs 1186 Montgomery Lane Vandergrift, Pennsylvania 15690
Re: K142301
Trade/Device Name: Evolution Mechanical Dilator Sheath Set, Evolution Shortie Mechanical Dilator Sheath Set, SteadySheath Evolution Tissue Stabilization Sheath, SteadySheath Evolution Shortie Tissue Stabilization Sheath
Regulation Number: 21 CFR 870.1310 Regulation Name: Vessel Dilator for Percutaneous Catheterization Regulatory Class: Class II Product Code: DRE Dated: August 15, 2014 Received: August 18, 2014
Dear Thomas Kardos,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, arranged in a stacked formation. The profiles are black and have a flowing, abstract design. The graphic is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrand
for
for Bram D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Traditional 510(k): Device Modification
Evolution® Mechanical Dilator Sheath Set, Evolution® Shortie Mechanical Dilator Sheath™ Evolution® Tissue Stabilization Sheath, and SteadySheath™ Evolution® Shortie Tissue Stabilization Sheath Cook Vascular, Inc.
#### Indications for Use
510(k) Number (if known): K142301
Device Name: Evolution® Mechanical Dilator Sheath Set Evolution® Shortie Mechanical Dilator Sheath Set SteadySheath™ Evolution® Tissue Stabilization Sheath SteadySheath™ Evolution® Shortie Tissue Stabilization Sheath
Indications for Use for the Evolution® Mechanical Dilator Sheath Set, Evolution® Shortie Mechanical Dilator Sheath Set, and SteadySheath™ Evolution® Tissue Stabilization Sheath:
The Evolution® Mechanical Dilator Sheath Set is intended for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects.
The Evolution® Shortie Mechanical Dilator Sheath Set is intended for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects.
The SteadySheath™ Evolution® Tissue Stabilization Sheath is intended for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects.
The SteadySheath™ Evolution® Shortie Tissue Stabilization Sheath is intended for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects.
Prescription Use X (Part 21 CFR 801 Subpart D) OR
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
| 510(k) SUMMARY: | K142301 |
|----------------------|--------------------------------------------------------------|
| Submitted By: | Thomas J. Kardos |
| | Cook Vascular, Inc. |
| | 1186 Montgomery Lane |
| | Vandergrift, PA 15690 |
| | Phone: (724) 845-8621 x2225 |
| | Fax: (724) 845-2848 |
| | Date Prepared: August 15, 2014 |
| Device: | |
| Trade Name: | Evolution® RL Controlled-Rotation Dilator Sheath Set |
| | Evolution® Shortie RL Controlled-Rotation Dilator Sheath Set |
| | SteadySheath™ Evolution® Tissue Stabilization Sheath |
| | SteadySheath™ Evolution® Shortie Tissue Stabilization Sheath |
| Common Name: | Vessel Dilator for Percutaneous Catheterization |
| Classification Name: | Dilator, Vessel, for Percutaneous Catheterization |
| | DRE (21 CFR §870.1310) |
### Indications for Use:
The Evolution® Mechanical Dilator Sheath Set, Evolution® Shortie Mechanical Dilator Sheath Set, SteadySheath™ Evolution® Tissue Stabilization Sheath, and SteadySheath™ Evolution® Shortie Tissue Stabilization Sheath are intended for use in patients requiring the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects.
### Predicate Device:
The devices, subject of this submission, are substantially equivalent to the predicate devices, the Evolution® Mechanical Dilator Sheath Set, cleared on May 10, 2006 under 510(k) number K061000 and the Evolution RL Controlled-Rotation Dilator Sheath Set and Evolution Shortie RL Controlled-Rotation Dilator Sheath Set, cleared on July 1, 2014 under 510(k) number K141148.
### Comparison to Predicate Device:
It has been demonstrated that the Evolution® Mechanical Dilator Sheath Set and Evolution® Shortie Mechanical Dilator Sheath Set are comparable to the predicate devices, and the SteadySheath™ Evolution® Tissue Stabilization Sheath™ Evolution® Shortie Tissue Stabilization Sheath are comparable to the outer sheath of the predicate devices. The Evolution® Mechanical Dilator Sheath Set and Evolution® Shortie Mechanical Dilator Sheath Set are identical in terms of intended use, principles of operation, and basic technological characteristics to the predicate devices, and also identical in materials of construction to the predicate Evolution® RL Controlled-Rotation Dilator Sheath Set and Evolution® Shortie RL
{4}------------------------------------------------
Controlled-Rotation Dilator Sheath Set (K141148). The SteadySheath™ Evolution® Tissue Stabilization Sheath and SteadySheath™ Evolution® Shortie Tissue Stabilization Sheath are identical in terms of intended use, principles of operation, and basic technological characteristics to the outer sheaths of the predicate devices, and also contains only materials found in the predicate Evolution® RL Controlled-Rotation Dilator Sheath Set and Evolution® Shortie RL Controlled-Rotation Dilator Sheath Set (K141148). The safety and effectiveness of the modifications are supported by testing.
## Device Description:
The Evolution® Mechanical Dilator Sheath Set is constructed of two coaxial sheaths connected to a handle capable of mechanically rotating the inner sheath. In use, the coaxial sheaths are advanced over an indwelling catheter or cardiac lead. The inner sheath (available in 9, 11, and 13 French diameters) is designed to rotate in a clockwise manner by pulling the trigger in the handle assembly. As the inner sheath is advanced, the rotation of the sheath assists in dilation of any binding tissue which may be anchoring the catheter or lead to the inner vascular or inner cardiac wall. The outer coaxial sheath can be used to stabilize the cardiac wall at the point of lead/catheter tip attachment to allow for detachment and removal (counter traction technique). The device has a working length of 40.6 centimeters.
The Evolution® Shortie Mechanical Dilator Sheath Set is constructed of two coaxial sheaths connected to a handle capable of mechanically rotating the inner sheath. The outer sheath has winged control tabs located at its proximal (handle) end. The sheaths are advanced over an indwelling catheter or cardiac lead. The inner sheath (available in 9 and 11 French diameters) is designed to rotate in clockwise manner by pulling the trigger in the handle assembly. As the inner sheath is advanced, the rotation of the sheath assists in dilation of any binding tissue which may be anchoring the catheter or lead to the inner vascular or inner cardiac wall. The outer coaxial sheath can be used to stabilize the cardiac wall at the point of lead/catheter tip attachment to allow for detachment and removal (counter traction technique). The device has a working length of 13.6 centimeters.
The SteadySheath™ Evolution® Tissue Stabilization Sheath and SteadySheath™ Evolution® Shortie Tissue Stabilization Sheath are designed to be used in place of the standard outer sheath that is supplied as part of the Evolution family of mechanical/controlled-rotation dilator sheath sets. They are provided for use in situations where better sheath tip visibility and/or tissue stabilization is desired. In its role as an Evolution outer sheath, the SteadySheath™ works in concert with the inner Evolution braided inner sheath to assist in the removal of cardiac leads or related catheters by bypassing and/or disrupting transvenous binding scar tissue. The SteadySheath™ has a highly radiopaque, textured stainless steel tip that helps stabilize tissue so that better binding scar tissue control and disruption can be achieved.
# Test Data:
The modifications included in the Evolution® Mechanical Dilator Sheath Set and Evolution® Shortie Mechanical Dilator Sheath Set have already been evaluated and appropriately tested, and
{5}------------------------------------------------
have been cleared by FDA, in the context of the predicate Evolution RL Controlled-Rotation Dilator Sheath Set and Evolution Shortie RL Controlled-Rotation Dilator Sheath Set (K141148). Thus, no additional tests were required to demonstrate that the Evolution® Mechanical Dilator Sheath Set and Evolution® Shortie Mechanical Dilator Sheath Set met applicable design and performance requirements and support a determination of substantial equivalence. The following tests were performed to demonstrate that the SteadySheath™ Evolution® Tissue Stabilization Sheath and SteadySheath™ Evolution® Shortie Tissue Stabilization Sheath met applicable design and performance requirements and support a determination of substantial equivalence.
- 1. Push and Flex Testing Testing was performed with the requirement that each device must remain functional after 50 cycles of flexing between 45° and 90° while within a simulated use model. The results showed that these predetermined acceptance criteria were met.
- 2. Push and Rotate Testing Testing was performed with the requirement that each device must remain functional after 50 cycles of insertion, 90° rotation, and removal from a simulated use model. The results showed that these predetermined acceptance criteria were met.
- 3. Pull Testing Testing was performed with the requirement that each device must withstand no less than 10 pounds of pull force prior to failure. The results showed that these predetermined acceptance criteria were met.
- 4. Sterility Testing The proposed devices were evaluated for bioburden levels, endotoxin levels, and ethylene oxide and ethylene chlorohydrin residuals. Their sterility was deemed acceptable.
In conclusion, the results of these tests support a determination of substantial equivalence to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.