EVOLUTION MECHANICAL DILATOR SHEATH SET MODELS-LR-EVN-7.0, LR-EVN-9.0

K061000 · Cook Vascular, Inc. · DRE · May 10, 2006 · Cardiovascular

Device Facts

Record IDK061000
Device NameEVOLUTION MECHANICAL DILATOR SHEATH SET MODELS-LR-EVN-7.0, LR-EVN-9.0
ApplicantCook Vascular, Inc.
Product CodeDRE · Cardiovascular
Decision DateMay 10, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1310
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Cook Vascular EVOLUTION Mechanical Dilator Sheath Set is for use in the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects.

Device Story

The EVOLUTION Mechanical Dilator Sheath Set is a manual surgical instrument used to dilate tissue surrounding indwelling catheters, cardiac leads, or foreign objects. The device consists of a polypropylene dilator with a semi-threaded stainless steel tip and an ergonomic handle that functions as a drive mechanism. It is available in 7.0 and 9.0 French sizes. The device is supplied sterile for single-use. It is operated by a physician in a clinical setting to facilitate the removal or replacement of indwelling devices by mechanically dilating the surrounding fibrotic tissue.

Clinical Evidence

Bench testing only. Testing included buckling, binding, bond strength (connector/gun), pull testing (tip), handle cycling, and general device performance to ensure design conformance.

Technological Characteristics

Materials: Polypropylene dilator, stainless steel tip. Form factor: 7.0 and 9.0 French sizes. Operation: Manual mechanical dilation via threaded tip and ergonomic handle drive mechanism. Sterilization: Sterile, single-use.

Indications for Use

Indicated for percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters, and foreign objects.

Regulatory Classification

Identification

A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.

Submission Summary (Full Text)

{0}------------------------------------------------ K oblooo # 510(k) Summary # Submitted by: MAY 1 0 11 26 Jim Ferguson Quality Systems Manager Cook Vascular, Incorporated 1186 Montgomery Lane Vandergrift, Pa 15690 724-845-8621, XT 2227 April 10, 2006 #### Device: | Trade name: | Cook Vascular EVOLUTION Mechanical Dilator Sheath<br>Set | |--------------------------|--------------------------------------------------------------------| | Proposed Classification: | Dilator, Vessel, for Percutaneous Catheterization<br>870.1310, DRE | ## Predicate Devices: The Cook Vascular EVOLUTION Mechanical Dilator Sheath Set is similar in terms of intended use, materials of construction and technology characteristics to predicate devices reviewed as devices for dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects. #### Device Description: The proposed device is a polypropylene dilator with a semi-threaded stainless steel tip, This stainless steel tip is used to dilate tissue surrounding an indwelling catheter or foreign object. In addition, an ergonomic handle to be used as a "drive mechanism" will be included. The device will be made with 7.0 and 9.0 French sizes. It will be supplied sterile, intended for one-time use. ## Substantial Equivalence: This device will be manufactured to specified process controls and a Quality Assurance program. This device will undergo packaging similar to the devices currently marketed and distributed by Cook Vascular. This device will undergo sterilization similar to the devices currently marketed and distributed. Being similar with respect to indication for use, materials and physical construction to predicate devices, this device meets the requirements for section 510(k) substantial equivalence. · Cook Vascular, Inc. • Special 510(k) Premarket Submission • EVOLUTION Rotary Dilator System • Confidential • ರಿ {1}------------------------------------------------ ## Test Data: The Cook Vascular EVOLUTION Mechanical Dilator Sheath Set was subjected to the flowing tests to assure reliable design and performance under the specified testing parameters. These tests were comprised of: - Buckling Test --1. 1 - 2. Binding Test - Bond strength (Connector/gun) 3. - 4. Pull Testing (tip) - 5. Handle Cycling Test - 6. Device Performance The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use as a Dilator Sheath. · Cook Vascular, Inc. · Special 510(k) Premarket Submission · EVOLUTION Rotary Dilator System · Confidential · 10 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features an abstract eagle design, with three stylized lines forming the body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 1 0 2006 Cook Vascular, Inc. c/o Mr. Jim Ferguson, Jr. Quality Systems Manager 1186 Montgomery Lane Vandergrift, Pennsylvania 15690 Re: K061000 Trade/Device Name: EVOLUTION Mechanical Dilator Sheath Set Regulation Number: 21 CFR 870.1310 Regulation Name: Vessel dilator for percutaneous catheterization Regulatory Class: II Product Code: DRE Dated: April 10, 2006 Received: April 11, 2006 Dear Mr. Ferguson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Jim Ferguson, Jr. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement 510(k) Number: 06/000 Device Name: EVOLUTION Mechanical Dilator Sheath Set Indications for Use: . The Cook Vascular EVOLUTION Mechanical Dilator Sheath Set is for use in the percutaneous dilation of tissue surrounding cardiac leads, indwelling catheters and foreign objects. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | k061000 | |---------------|---------| |---------------|---------| | Prescription Use<br>(Per 21 CFR 801 Subpart D) | OR | Over-The-Counter Use<br>(Per 21 CFR 810 Subpart C) | |------------------------------------------------|----|----------------------------------------------------| |------------------------------------------------|----|----------------------------------------------------| · Cook Vascular, Inc. · Special 510(k) Premarket Submission · Evolution Rotary Dilator System · Confidential ·
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...