Shiley Neonatal, Pediatric & Pediatric Long Tracheostomy Tubes with TaperGuard Cuff
Applicant
Covidien
Product Code
JOH · Anesthesiology
Decision Date
Feb 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5800
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Shiley™ Neonatal, Pediatric & Pediatric Long Tracheostomy Tubes with TaperGuard Cuff are intended for use in providing tracheal access for airway management.
Device Story
Single-use, single-cannula tracheostomy tube; provides artificial airway for airway management. Device features TaperGuard cuff; taper-shaped cuff design; medical-grade PVC construction with citrate-based non-phthalate plasticizer. Used in clinical settings for tracheotomized patients. Healthcare providers insert tube to maintain airway patency; cuff inflation provides tracheal seal. Benefits include airway access and management for neonatal and pediatric populations.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing, including cuff performance (air/fluid seal, pressure mapping, removal force) and biocompatibility testing (cytotoxicity, sensitization, irritation, acute/subchronic systemic toxicity, implantation, chemical characterization).
Indicated for neonatal and pediatric patients requiring tracheal access for airway management.
Regulatory Classification
Identification
A tracheostomy tube and tube cuff is a device intended to be placed into a surgical opening of the trachea to facilitate ventilation to the lungs. The cuff may be a separate or integral part of the tracheostomy tube and is, when inflated, intended to establish a seal between the tracheal wall and the tracheostomy tube. The cuff is used to prevent the patient's aspiration of substances, such as blood or vomit, or to provide a means for positive-pressure ventilation of the patient. This device is made of either stainless steel or plastic.
Predicate Devices
Shiley™ Cuffed Pediatric & Pediatric Long (K955680)
Shiley™ Neonatal, Pediatric & Pediatric Long Tracheostomy Tubes Cuffless (K122531)
Smiths Medical Bivona Aire-Cuf Tracheostomy Tubes (K982641)
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 13, 2015
Covidien Ms. Kelsey Lee Regulatory Affairs Product Manager 6135 Gunbarrel Avenue Boulder, CO 80301
Re: K142298
Trade/Device Name: Shiley™ Neonatal, Pediatric & Pediatric Long Tracheostomy Tubes with TaperGuard Cuff Regulation Number: 21 CFR 868.5800 Regulation Name: Tracheostomy Tube & Tube Cuff Regulatory Class: II Product Code: JOH Dated: January 9, 2015 Received: January 14, 2015
Dear Ms. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Lee
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
K142298 510(k) Number (if known):
Device Name: Shiley™ Neonatal, Pediatric & Pediatric Long Tracheostomy Tubes with TaperGuard Cuff
## Indications for Use:
The Shiley™ Neonatal, Pediatric & Pediatric Long Tracheostomy Tubes with TaperGuard Cuff are intended for use in providing tracheal access for airway management.
> Prescription Use __ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
> > (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
{3}------------------------------------------------
#### 510(k) SUMMARY
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a 510(k) Summary for the use of the Shiley™ Neonatal, Pediatric Long Tracheostomy Tubes with TaperGuard Cuff.
| Submitted By: | Covidien<br>6135 Gunbarrel Avenue<br>Boulder, CO 80301 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | February 12, 2015 |
| Contact Person: | Danielle Mueller<br>Regulatory Affairs Manager<br>(303) 305-2603 |
| Proprietary Name: | Shiley™ Neonatal, Pediatric & Pediatric Long<br>Tracheostomy Tubes with TaperGuard Cuff |
| Common Name: | Tracheostomy Tube & Tube Cuff |
| Device Classification Regulation: | 21 CFR 868.5800 – Class II |
| Device Product Code & Panel: | JOH |
| Predicate Devices: | Shiley™ Cuffed Pediatric & Pediatric Long (K955680)<br><br>Shiley™ Neonatal, Pediatric & Pediatric Long<br>Tracheostomy Tubes Cuffless (K122531)<br><br>Smiths Medical Bivona Aire-Cuf Tracheostomy Tubes<br>(K982641) |
#### Device Description
The subject devices are single use; single cannula tracheostomy tubes that are intended to provide an artificial airway for airway management. The new features on the subject devices include a new plasticizer, the addition of the TaperGuard cuff, and additional sizes.
#### Indications for Use/Intended Use
The subject and predicate Shiley™ devices have identical indications and similar indications to the BIVONA Aire-Cuf device. The indications for use are as follows:
The Shiley™ Neonatal, Pediatric Long Tracheostomy Tubes with TaperGuard Cuff are intended for use in providing tracheal access for airway management.
{4}------------------------------------------------
## Technological Characteristics Comparison
The subject device only differs from the predicate Shiley™ Neonatal, Pediatric Long Tracheostomy Tubes Cuffless (K122531) with the addition of the cuff. They have identical indications, basic design and sizes. The subject device has a similar cuff to the predicate Shiley Cuffed Tubes (K955680) and BIVONA Tubes (K083641). The difference between the cuff on the subject and predicate tubes is the overall shape, which is taper-shaped in the subject versus barrel shaped in the predicates.
| | SUBJECT DEVICE | PREDICATE DEVICE | | |
|--------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Shiley™ Cuffed | Shiley™ Cuffed [K955680] | Shiley™ Cuffless [K122531] | BIVONA Aire-Cuf [K083641] |
| Indications | Intended for use in providing tracheal access for airway management. | Identical | Identical | Intended to provide direct airway access for a tracheotomized patient for up to 29 days. It may be reprocessed for a single-patient use up to 5 times for pediatric sizes. |
| Patient population | Neonatal & pediatric | Pediatric | Neonatal & pediatric | Neonatal & pediatric |
| Sterilization | EtO | EtO | EtO | EtO |
| Cuff shape | Taper | Barrel | N/A | Barrel |
| Use | Single patient use | Identical | Identical | Identical |
| Sizes | 2.5-4.5 Neonatal<br>2.5-5.5 Pediatric<br>5.0-6.5 Pediatric long | 4.0-5.5 Pediatric<br>5.0-6.5 Pediatric long | 2.5-4.5 Neonatal<br>2.5-5.5 Pediatric<br>5.0-6.5 Pediatric long | 2.5-4.0 Neonatal<br>2.5-5.5 Pediatric |
| Materials | | | | |
| Cannula | Medical grade PVC with a citrate-based non-phthalate plasticizer | Medical grade PVC with a DEHP plasticizer | Medical grade PVC with a citrate-based non-phthalate plasticizer | Silicone |
| Flange | Medical grade PVC with a citrate-based non-phthalate plasticizer | Medical grade PVC with a DEHP plasticizer | Medical grade PVC with a citrate-based non-phthalate plasticizer | Silicone |
| ISO 15mm connector | Acrylonitrile butadiene styrene (ABS) | Medical grade PVC | ABS | ABS |
| Cuff | Medical grade PVC with a citrate- | Medical grade PVC with a DEHP | N/A | PVC |
| | based non-<br>phthalate<br>plasticizer | plasticizer | | |
| Inflation system | Medical grade PVC<br>with a citrate-<br>based non-<br>phthalate<br>plasticizer & PVC | Medical grade PVC<br>with a citrate-<br>based non-<br>phthalate<br>plasticizer & PVC | N/A | Silicone & ABS |
| Obturator | Medical grade<br>polypropylene | Medical grade<br>polypropylene | Medical grade<br>polypropylene | Medical grade<br>polypropylene |
| Neck strap | Cotton | Cotton | Cotton | Cotton |
{5}------------------------------------------------
### Substantial Equivalence - Non-Clinical Evidence
Product performance testing, including cuff air seal, cuff fluid seal, removal force, and cuff pressure mapping, were performed with the conclusion that the subject devices can be expected to perform at least as well as the predicate and can be considered substantially equivalent. Additionally, biocompatibility testing was performed per ISO 10993-1:2009 including the following: cytotoxicity, sensitization, irritation, acute systemic toxicity, subchronic toxicity implantation, chemical characterization, and risk assessment. The device met all biocompatibility requirements for its intended use.
## Substantial Equivalence - Clinical Evidence
N/A - Clinical evidence was not necessary to show substantial equivalence.
#### Substantial Equivalence - Conclusions
No new questions of safety and effectiveness have been raised. From the evidence presented in the Premarket Notification, the subject devices can be considered substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.