The device is intended to be used to provide and artificial airway in order to provide access to the patient's airway for airway management. When inserted in a tracheotomy stoma, the device is secured in place with a neck strap around the patient's neck which is attached to an integral neck plate, thereby providing for a secure artificial airway for spontaneous breathing or direct hook-up to ventilation or anesthesia equipment.
Device Story
Single cannula tracheostomy tube; includes neck strap and obturator. Inserted into tracheotomy stoma; secured via integral neck plate and neck strap. Provides artificial airway for spontaneous breathing or connection to ventilation/anesthesia equipment. Used in clinical environments (neonatal/pediatric/critical care units, long-term care, home care) by healthcare professionals. Benefits patient by maintaining airway patency. Constructed from medical-grade PVC with citric-based non-phthalate plasticizer.
Clinical Evidence
No clinical data. Bench testing performed to confirm compliance with ISO 5366-3:2001/Cor:2003, ISO 5356-1:2004, and ISO 5366-1:2000. Biocompatibility testing satisfied ISO 10993-1:2009 requirements.
Technological Characteristics
Single cannula tracheostomy tube. Materials: Medical-grade PVC with citric-based non-phthalate plasticizer (cannula, neck plate, 15mm connector); medical-grade polypropylene (obturator); cotton (tie strap). Standards: ISO 5366-1:2000, ISO 5366-3:2001/Cor:2003, ISO 5356-1:2004. Sterilization: Sterile (SAL 10E-6) via validated cycle 66.
Indications for Use
Indicated for neonatal and pediatric patients requiring an artificial airway for tracheal access and airway management.
Regulatory Classification
Identification
A tracheostomy tube and tube cuff is a device intended to be placed into a surgical opening of the trachea to facilitate ventilation to the lungs. The cuff may be a separate or integral part of the tracheostomy tube and is, when inflated, intended to establish a seal between the tracheal wall and the tracheostomy tube. The cuff is used to prevent the patient's aspiration of substances, such as blood or vomit, or to provide a means for positive-pressure ventilation of the patient. This device is made of either stainless steel or plastic.
Predicate Devices
Shiley Pediatric and Neonatal Tracheostomy Tubes (K945513)
{0}------------------------------------------------
# COVIDIEN
#### K122531 510(k) Summary
Date summary prepared 21-September-2012
# 510(k) Submitter/Holder
Covidien IIc 6135 Gunbarrel Avenue Boulder, CO.80301
#### Contact
Ted Kuhn Sr. Regulatory Affairs Product Specialist Office: 303-305-2272 Mobile: 720-320-7259 Fax: 303-305-2212 email address: ted.kuhn@covidien.com
#### Name of Device
| Trade Name: | Shiley Neonatal, Pediatric, Pediatric Long Tracheostomy Tube Cuffless |
|----------------------|--------------------------------------------------------------------------------------|
| Catalog Numbers: | NEF, PEF, PELF |
| Common Name: | Tube Tracheostomy and Tube Cuff |
| Classification Name: | JOH, Tube Tracheostomy and Tube Cuff (21 CFR 868.5800) Class II, 73<br>Anestesiology |
# Purpose of Submission
The purpose of this submission is to introduce the Covidien Shiley Neonatal, Pediatric Long Cuffless Tracheostomy Tube. The subject and predicate device have the equivalent Indications for Use, Intended Use, fundamental technology, and design. This submission is to notify the FDA of the following changes:
- . Pediatric and Neonatal Tracheostomy tubes size line extension of the previously cleared Shiley Pediatric and Neonatal Tracheostomy tubes
OCT
9 2012
- Updated the materials used in the Tracheostomy tubes to a Medical Grade PVC ● with a Citric based non-phthalate plasticizer
- Update to the Manufacturing process .
#### Predicate Device
The Shiley Neonatal, Pediatric, Pediatric Long Tracheostomy Tube Cuffless was compared and found to be substantially equivalent to the following products of comparable type in commercial distribution:
| Trade Name: | Shiley Pediatric and Neonatal Tracheostomy Tubes |
|---------------------|--------------------------------------------------|
| Device Common Name: | Tube Tracheostomy and Tube Cuff |
| 510(k) Number: | K945513 |
| Manufacturer: | Covidien, formerly Tyco Healthcare |
{1}------------------------------------------------
| Trade Name: | Bivona Tracheostomy Tube Tracheostomy Tubes |
|---------------------|---------------------------------------------|
| Device Common Name: | Tube Tracheostomy and Tube Cuff |
| 510(k) Number: | K083641 |
| Manufacturer: | Smiths Medical ASD, Inc. |
#### Intended Use
The device is intended to be used to provide and artificial airway in order to provide access to the patient's airway for airway management. When inserted in a tracheotomy stoma, the device is secured in place with a neck strap around the patient's neck which is attached to an integral neck plate, thereby providing for a secure artificial airway for spontaneous breathing or direct hook-up to ventilation or anesthesia equipment.
#### Indications for Use
This device is intended for use in providing tracheal access for airway management.
### Device Description
The Shiley Neonatal, Pediatric, Pediatric Long Tracheostomy Tube Cuffless tube is a single cannula tracheostomy tube which also contains a neck strap and obturator. The device is intended to be used to provide and artificial airway in order to provide access to the patient's airway for airway management. When inserted in a tracheotomy stoma, the device may be secured in place with a neck strap around the patient's neck which is attached to an integral neck plate, thereby providing for a secure artificial airway for spontaneous breathing or direct hook-up to ventilation or anesthesia equipment.
# Technological Characteristics
The Shiley Neonatal, Pediatric, Pediatric Long Tracheostomy Tube Cuffless tube has technological characteristics that differ from its predicates. The Shiley Pediatric and Neonatal Tracheostomy Tubes (K945513) are comprised of a Medical Grade Polyvinylchloride (PVC) and BIVONA Tracheostomy Tubes (K083641) is comprised of Silicone and Acrylonitrile butadiene styrene (ABS) materials. Shiley Neonatal, Pediatric Long Tracheostomy Tube Cuffless tube is comprised of a Medical Grade PVC with a Citric Based non-phthalate plasticizer. All other fundamental technological characteristics are identical to the predicate devices.
{2}------------------------------------------------
| Subject Device: | Predicate Device | Predicate Device | |
|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Shiley Neonatal,<br>Pediatric, Pediatric<br>Long Tracheostomy<br>Tube Cuffless | Shiley Neonatal,<br>Pediatric, Pediatric<br>Long Tracheostomy<br>Tube (K945513) | BIVONA<br>Tracheostomy Tubes<br>(K083641) | |
| Indications for Use | This device is intended<br>for use in providing<br>tracheal access for<br>airway management. | This device is intended<br>for use in providing<br>tracheal access for<br>airway management. | The Bivona<br>Tracheostomy Tube is<br>intended to provide<br>direct airway access<br>for a tracheotomized<br>patient for up to 29<br>days. It may be<br>reprocessed for a<br>single-patient use up to<br>5 times for pediatric<br>sizes. |
| Intended Use | This device is intended<br>for use in providing<br>tracheal access for<br>airway management. | This device is intended<br>for use in providing<br>tracheal access for<br>airway management. | Unknown |
| Device class | Class 2 | Class 2 | Class 2 |
| Product code | JOH | JOH | JOH |
| Patient Population | Neonatal and Pediatric<br>patients requiring an<br>artificial airway | Neonatal and Pediatric<br>patients requiring an<br>artificial airway | Pediatric patients<br>requiring an artificial<br>airway |
| Environment of Use | The product shall be<br>used in a clinical<br>environment such as the<br>neonatal, pediatric or<br>critical care units of a<br>hospital, non-critical care<br>units of a hospital, long<br>term care facilities and<br>home care. | The product shall be<br>used in a clinical<br>environment such as the<br>neonatal, pediatric or<br>critical care units of a<br>hospital, non-critical care<br>units of a hospital, long<br>term care facilities and<br>home care. | Unknown |
| Invasive / Non-<br>Invasive | Invasive | Invasive | Invasive |
| Sterilization | Sterile : The SAL is 10E-6<br>utilizing the existing<br>validated cycle 66. | Sterile : The SAL is 10E-6<br>utilizing the existing<br>validated cycle 66. | Sterile with Ethylene<br>Oxide |
| Shelf Life | 1-year Minimum shelf life | 5-year shelf life | Unknown |
| Materials | | | |
| | Subject Device: | Predicate Device | Predicate Device |
| | Shiley Neonatal,<br>Pediatric, Pediatric<br>Long Tracheostomy<br>Tube Cuffless | Shiley Neonatal,<br>Pediatric, Pediatric<br>Long Tracheostomy<br>Tube (K945513) | BIVONA<br>Tracheostomy Tubes<br>(K083641) |
| Cannula | Medical Grade PVC<br>with a Citric Based non-<br>phthalate plasticizer | Medical Grade PVC | Silicone |
| Neck Plate | Medical Grade PVC<br>with a Citric Based non-<br>phthalate plasticizer | Medical Grade PVC | Silicone |
| 15 MM connector | Medical Grade PVC<br>with a Citric Based non-<br>phthalate plasticizer | Medical Grade PVC | Acrylonitrile butadiene<br>styrene (ABS) |
| Obturator | Medical Grade<br>Polypropylene | Medical Grade<br>Polypropylene | Medical Grade<br>Polypropylene |
| Tie Strap | Cotton | Cotton | Cotton |
| Biocompatibility | ISO 10993-1:2009 | ISO 10993-1:2009 | ISO 10993-1:2009 |
| Sizes | 2.5 NEF to 4.5 NEF<br>2.5 PEF to 5.5 PEF<br>5.0 PELF to 6.5 PELF | 3.0 NEO to 4.5 NEO<br>3.0 PED to 5.5 PED<br>5.0 PDL to 6.5 PDL | 2.5 to 4.0<br>Neonatal<br>2.5 to 5.5<br>Pediatric |
| Use | Single Patient Use | Single Patient Use | Single Patient Use |
| Design | Tracheostomy Tube<br>Design | Tracheostomy Tube<br>Design | Tracheostomy Tube<br>Design |
| | Subject Device: | Predicate Device | Predicate Device |
| | Shiley Neonatal,<br>Pediatric, Pediatric<br>Long Tracheostomy<br>Tube Cuffless | Shiley Neonatal,<br>Pediatric, Pediatric<br>Long Tracheostomy<br>Tube (K945513) | BIVONA<br>Tracheostomy Tubes<br>(K083641) |
| Standards met | ISO 5366-1:2000<br>Anesthetic and<br>respiratory equipment —<br>Tracheostomy tubes Part<br>1: Tubes and connectors<br>for use in adults<br><br>ISO 5366-3:2001 /<br>Cor:2003 Anesthetic and<br>Respiratory Equipment --<br>Tracheostomy Tubes --<br>Part 3: Pediatric<br>Tracheostomy Tubes<br><br>ISO 5356-1:2004<br>Anesthetic and<br>respiratory equipment -<br>Conical connectors: Part<br>1: Cones and sockets<br><br>(The key product<br>standards used to<br>develop this product<br>range are ISO 5366 Part<br>3and ISO 5356-1. The<br>PDL model encompasses<br>a 6.5mm Tracheostomy<br>tube. The 6.5mm tube is<br>beyond the scope of ISO<br>5366 Part 3. Therefore<br>ISO 5366 Part 1 is also<br>used to guide the<br>development of the<br>product range) | 5356 (1987)<br>5366 (1985) | ISO 5366-3:2001<br><br>ISO 5366-3:2001<br>Second edition 2001-<br>08-15<br><br>ISO 5356-1<br>Third<br>edition 2004-05-15 |
י
.
{3}------------------------------------------------
.
.
.
·
.
,
ﺮ
{4}------------------------------------------------
.
.
.
.
.
{5}------------------------------------------------
### Non-Clinical Data
Bench testing was performed to confirm that the Shiley Neonatal, Pediatric Long Tracheostomy Tube Cuffless meets the requirements of ISO 5366-3 [2001/Cor:2003[, ISO 5356-1:2004 and ISO 5366-1:2000 for Tracheostomy Tubes. Biocompatibility testing was performed on the Shiley Neonatal, Pediatric Long Tracheostomy Tube Cuffless which satisfied all of the requirements of ISO 10993-1:2009.
# Clinical Data
Not Required
# Substantial Equivalence
The predicate devices to which we claim equivalence are the Shiley Pediatric and Neonatal Tracheostomy Tubes (K945513) and BIVONA Tracheostomy Tubes (K083641). The basic design elements and their assemblies are identical to the cleared predicate device. The Indications for Use of the proposed devices are equivalent to the cleared predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular pattern around the bird symbol. The logo is black and white.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT
9
2012
Covidien, Limited Liability Company Mr. Ted Kuhn Senior Regulatory Affairs Product Specialist Respiratory & Monitoring Solutions 6135 Gunbarrel Avenue Boulder, Colorado 80021
Re: K122531
Trade/Device Name: Shiley Neonatal Tracheostomy Tube Cuffless Shiley Pediatric Tracheostomy Tube Cuffless Shiley Pediatric Tracheostomy Tube Long Cuffless
Regulation Number: 21 CFR 686.5800
Regulation Name: Tracheostomy Tube and Tube Cuff Regulatory Class: II Product Code: JOH Dated: September 21, 2012 Received: September 24, 2012
Dear Mr. Kuhn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): K122531
Device Name:
Shiley Neonatal Tracheostomy Tube Cuffless Shiley Pediatric Tracheostomy Tube Cuffless Shiley Pediatric Tracheostomy Tube Long Cuffless
#### Indications for Use:
This device is intended for use in providing tracheal access for airway management.
AND/OR Over-The-Counter Use × Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
> (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of cdrh, office of device evaluation (ode)
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Ancolnoons. Devices
1225
**510(k) Number:**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.