K111476 · Contour Technology · NGX · May 23, 2012 · Physical Medicine
Device Facts
Record ID
K111476
Device Name
CONTOUR TECHNOLOGY MUSCLE STIMULATOR
Applicant
Contour Technology
Product Code
NGX · Physical Medicine
Decision Date
May 23, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Contour Technology Muscle Stimulator is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The Contour Technology Muscle Stimulator may therefore be considered a technique or method for muscle training. The Contour Technology Muscle Stimulator Ab Belt accessory is intended for use on abdominal muscles only for strengthening and toning of abdominal muscles. The Contour Technology Muscle Stimulator BackPad accessory is intended for use on the lower back muscles only.
Device Story
Powered muscle stimulator delivers electrical stimulation to healthy skeletal muscles to improve performance, tone, and strength. Device uses 4 AAA batteries to generate monophasic, rectangular, voltage-regulated output signals. Signals transmitted via two channels to gel pad electrodes placed on target muscle regions (upper/lower back, biceps, triceps, forearms, deltoids, abdomen, gluteus, hamstrings, quadriceps, calves). Motor neuron stimulation triggers muscle fiber contraction. Operated by user in home setting; optional hands-free accessories (Ab Belt, BackPad) available. Device features pre-set programs with optimized pulse rate, duration, and rest times. User interface allows individual channel regulation. Benefits include muscle training and development of firmer muscles.
Clinical Evidence
No clinical data. Device does not diagnose, cure, mitigate, treat, or prevent disease. Substantial equivalence supported by non-clinical bench testing, including system/software verification, environmental, safety, ESD/EMI, stress, shipping, and simulated use (usability) testing.
Technological Characteristics
Powered muscle stimulator; 4 AAA battery source. Monophasic, rectangular, voltage-regulated output. 2-channel delivery. Pulse duration 340 μs; frequency 1-120 Hz. Max current density 0.55 mA (rms)/cm². Plastic housing. Reusable gel pad electrodes. Independent channel regulation. Software-controlled pre-set programs.
Indications for Use
Indicated for healthy individuals seeking to improve or facilitate muscle performance, strengthen, or tone muscles. Specific accessories indicated for abdominal muscle strengthening/toning (Ab Belt) and lower back muscle stimulation (BackPad). Contraindicated for individuals with cardiac pacemakers, implanted defibrillators, or other implanted metallic/electronic devices.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Libramedical. The logo consists of a stylized letter "O" on the left and the word "LIBRAMEDICAL" on the right. The letter "O" is made up of three concentric circles, with the inner and outer circles being thicker than the middle circle.
# 5. 510(K) SUMMARY
### 5.1 Administrative Information
MAY 2 3 2012
### 5.1.1 Name and address
510(k) Owner/Sponsor:
Mike Esson Contour Technology 10 South 5th Street, Suite 990 Minneapolis, Minnesota 54402 Phone: 612-230-3802 Fax: 612 -230-3810
### Primary Contact Person
Sara Petrie Regulatory Consultant at Libra Medical Inc. 8401 73td Ave North, Suite 63 Minneapolis, MN 55428 Phone: 612-396-9849 Fax: 763-477-6357 Email: spetrie@libramed.com
Date Prepared: 5/15/2012
### 5.1.2 Device Name
| Trade Name | Contour Technology Muscle Stimulator |
|---------------------|--------------------------------------------------------------------------|
| Common Name | Muscle Stimulator |
| Classification Name | Powered Muscle Stimulator with Limited Output for Muscle<br>Conditioning |
| Classification | 21 CFR 890.5850<br>Class II |
| Product Code | NGX |
| Model | MX9 |
### 5.1.3 Applicant's Name
Contour Technology 10 South 5th Street, Suite 1000 Minneapolis, Minnesota 54402 Telephone: 612-230-3801 Fax: 612-230-3810
{1}------------------------------------------------
## 5.1.4 Substantial Equivalence
Contour Technology Muscle Stimulator covered by this substantially equivalent to other legally marketed devices namely the following:
- K011880, Compex Sport, Compex S.A. �
- . K031611, P4-Fitness, ValMed Corporation
- � K030708, Slendertone Flex 515, Bio-Medical Research Ltd.
The Contour Technology Muscle Stimulator and the predicate devices have the same intended use and similar indications, technological characteristics and principles of operation that includes the delivery of small amount of electrical stimulation to skeletal muscles for conditioning of the muscles. There are design and engineering differences between the Contour Technology Muscle Stimulator and the predicate devices. However, these differences do not raise new questions of safety or efficacy. Specifically, the only technological difference between the Contour Technology Muscle Stimulator and its predicates is the shape of the stimulator, user interface and specific duration and pulse energy of each of the pre-set programs. These differences do not present any new issues of safety or effectiveness as the Contour Technology Muscle Stimulator is substantially similar to the three cited predicates. Further, performance testing conducted by the company demonstrates that the device meets the FDA Draft Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Powered Muscle Stimulator with Limited Output for Muscle Conditioning.
Thus, the Contour Technology Muscle Stimulator is substantially equivalent to the predicate devices.
## 5.1.5 Device Description
The Contour Technology Muscle Stimulator is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The Contour Technology Muscle Stimulator may therefore be considered a technique or method for muscle training for improvement of muscle tone, for the strengthening of the following muscles: upper back, lower/mid back, lower back, biceps, triceps, forearm-flexor, forearm-extensor, deltoids, abdominal, gluteus, hamstrings, quadriceps, and calves for the development of firmer muscles.
The electrical stimulation is delivered to the muscles through two channels. The output signal is monophasic, rectangular and based on a regulated voltage technology. The parameters (pulse rate, pulse duration, stimulation time, rest time) are optimized for each program. The power is supplied by 4 AAA batteries.
Pulses, transmitted by the Contour Technology Muscle Stimulator are sent to Gel Pad electrodes placed over the muscle region. The signal reaches out to the motor neuron that innervates the muscle fiber resulting in contraction of the muscle. The energy is delivered through 2-3 pairs of gel pads. Optional accessories are available to allow for hands free operation of the device when placed on the abdomen or back.
## 5.1.6 Indications for Use/Intended Use
The Contour Technology Muscle Stimulator is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The Contour Technology Muscle Stimulator may therefore be considered a technique or method for muscle training.
{2}------------------------------------------------
The Contour Technology Muscle Stimulator Ab Belt accessory is intended for use on abdominal muscles only for strengthening and toning of abdominal muscles.
The Contour Technology Muscle Stimulator BackPad accessory is intended for use on the lower back muscles only.
## 5.1.7 Contraindications
Do not use this device if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electric shock, burns, electrical interference, or death.
## 5.1.8 Summary of Technological Characteristics compared to predicate devices
Technological characteristics compared to predicate devices can be found in Table 1.
| ltem | Contour Technology<br>Muscle Stimulator | Predicate:<br>Compex Sport | Predicate:<br>ValMed P4-Fitness | Predicate:<br>Slendertone Flex 515 |
|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K111476 | K011880 | K031611 | K030708 |
| Indications for<br>Use | The Contour Technology<br>Muscle Stimulator is<br>intended to stimulate<br>healthy muscles in order<br>to improve or facilitate<br>muscle performance. The<br>Contour Technology<br>Muscle Stimulator may<br>therefore be considered a<br>technique or method for<br>muscle training.<br>The Contour Technology<br>Muscle Stimulator Ab<br>Belt accessory is intended<br>for use on abdominal<br>muscles only for<br>strengthening and toning<br>of abdominal muscles.<br>The Contour Technology<br>Muscle Stimulator<br>BackPad accessory is<br>intended for use on the<br>lower back muscles only. | "Compex®<br>Sport" is<br>intended to<br>stimulate healthy<br>muscles in order<br>to improve or<br>facilitate muscle<br>performance.<br>"Compex®<br>Sport" is not<br>intended to be<br>used in<br>conjunction with<br>therapy or<br>treatment of<br>medical diseases<br>or medical<br>conditions of any<br>kind. None of the<br>"Compex®<br>Sport" training<br>programs is<br>designed for<br>injured or ailing<br>muscles and its<br>use on such<br>muscles is<br>contraindicated.<br>The "Compex®<br>Sport" electrical<br>impulses allow<br>triggering action<br>potentials on | The P4-Fitness is<br>intended for<br>stimulation of healthy<br>muscles in order to<br>enhance and facilitate<br>improved muscle<br>performance. The P-4<br>Fitness is therefore<br>properly considered as<br>a technique or method<br>for muscle training.<br>The P4-Fitness is NOT<br>intended for use in any<br>therapy or for<br>treatment of any<br>medical conditions or<br>diseases. The P4-<br>Fitness training<br>programs are not<br>designed or intended<br>for injured or<br>otherwise impaired<br>muscles and use of the<br>P4-Fitness on such<br>muscles is<br>contraindicated. | The improvement of<br>abdominal muscle tone,<br>for the strengthening of<br>abdominal muscles and<br>for the development of a<br>firmer abdomen. |
| Item | Contour Technology<br>Muscle Stimulator | Predicate:<br>Compex Sport | Predicate:<br>ValMed P4-Fitness | Predicate:<br>Slendertone Flex 515 |
| | | motoneurones of<br>motor nerves<br>(excitations). The<br>excitations of<br>motoneurones are<br>transmitted to the<br>muscle fibers via<br>the motor<br>endplate hwere<br>they generate<br>mechanical<br>muscle fiber<br>responses that<br>correspond to<br>muscle work.<br>Depending on the<br>parameters of the<br>electrical<br>impulses (pulse<br>frequency,<br>duration of<br>contraction,<br>duration of rest,<br>total session<br>duration),<br>different types of<br>muscle work can<br>be imposed on the<br>stimulated<br>muscles.<br><br>The various types<br>of muscles work<br>that "Compex®<br>Sport" can<br>impose on the<br>stimulated<br>muscles are able<br>to improve or<br>facilitate muscle<br>performance.<br>"Compex®<br>Sport" may<br>therefore be<br>considered a<br>technique of<br>muscle training. | | |
| Powered<br>Muscle<br>Stimulator | YES | YES | YES | YES |
| Battery<br>operated | Four (4) 1.5V only (AAA)<br>alkaline | One (1) NIMH<br>Rechargeable<br>battery (7.2V<br>=1200mA/h) | One-9V lithium,<br>alkaline or NiMh<br>battery | Three (3) AAA 1.5V DC<br>batteries |
| Regulated | YES | YES | YES | YES |
| Item | Contour Technology<br>Muscle Stimulator | Predicate:<br>Compex Sport | Predicate:<br>ValMed P4-Fitness | Predicate:<br>Slendertone Flex 515 |
| Current /<br>Voltage | | | | |
| Plastic Housing<br>Materials | YES | YES | YES | YES |
| Maximum<br>Current Density | 0.55 mA (rms) / cm²<br>(Smallest size electrode<br>40.5cm²) | 4mAmp/cm² | 1 mA (rms)/ cm²<br>If electrode impedance<br>is <100 Ω<br>smallest size electrodes<br>ø 3.2 cm=10cm² | Information Not<br>Available |
| Independent<br>channels with<br>possibility to<br>regulate the<br>current<br>individually | YES | YES | YES | YES |
| Pulse Duration<br>(Width) | 340 μs | 200 and 400 μs | 200μs or 240μs<br>for upper extremities<br>320 μs or 340μs for<br>legs | Positive: 200-300 μs<br>Interphase: 100 μs<br>Negative: 200-300 μs |
| Frequency | 1 to 120 Hz | 1 to 120 Hz | 1 to 120 Hz | 45-77 Hz |
| Reusable Gel<br>pads | YES | YES | YES | YES |
### Table 1: Technological Characteristics comparison
{3}------------------------------------------------
.
.
{4}------------------------------------------------
The technological characteristics, features, specifications, materials, and indications for use of the Contour Technology Muscle Stimulator are substantially equivalent to the predicate devices.
Therefore, there are no new safety and efficacy issues raised with the Contour Technology Muscle Stimulation Device.
### 5.1.9 Brief Description of non-clinical tests
Non-clinical tests were performed for the Contour Technology Muscle Stimulator and a summary is provided in Table 2.
| Test |
|-------------------------------------------------------|
| System and Software Requirements Verification Testing |
| Environmental Testing |
| Safety Testing |
| Unit & Integration Testing |
| ESD / EMI Testing |
| Stress Testing |
| Shipping Testing |
| Simulated Use Testing (Usabililty Testing) |
| | Table 2: Summary of Non-Clinical Tests | | | |
|--|----------------------------------------|--|--|--|
|--|----------------------------------------|--|--|--|
{5}------------------------------------------------
The performance data demonstrates that the device meets all the product specifications and included electrical output waveforms, energy density, safety and electromagnetic compatibility testing.
### 5.1.10 Brief Description of clinical performance data
No applicable. This device does not diagnose, cure, mitigate, treat or prevent disease or affect the function of the human body.
### 5.1.11 Conclusion
The Contour Technology Muscle Stimulator is substantially equivalent to the predicate devices. Test results demonstrate that the device is safe and effective for its intended use and the results support determination of substantial equivalence.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 2,3 2012
Contour Technology % Libra Medical, Inc. Ms. Sara Petrie 8401 73rd Avenue North, Suite 63 Minneapolis, Minnesota 55428
Re: K111476
Trade/Device Name: Contour Technology Muscle Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX Dated: May 01, 2012 Received: May 02, 2012
Dear Ms. Petrie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{7}------------------------------------------------
### Page 2 - Ms. Sara Petrie
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Mark N. Melkerson | -Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# 4. INDICATIONS FOR USE STATEMENT
### Indications for Use
510(k) Number (if known): K111476
Device Name: Contour Technology Muscle Stimulator
#### Indications for Use:
The Contour Technology Muscle Stimulator is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The Contour Technology Muscle Stimulator may therefore be considered a technique or method for muscle training.
The Contour Technology Muscle Stimulator Ab Belt accessory is intended for use on abdominal muscles only for strengthening and toning of abdominal muscles.
The Contour Technology Muscle Stimulator BackPad accessory is intended for use on the lower back muscles only.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
(Division Sign-Off Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of 1
510(k) Number K111476
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.