Medline ReNewal Reprocessed Gyrus-Diego Sinus Application Blades are designed for use in limited sinus applications involving soft tissue and thin bone such as lamina papyracea.
Device Story
Reprocessed Gyrus-Diego sinus application blades; sterile, single-use devices. Original manufacturer: OLYMPUS/Gyrus ACMI. Device functions as insertable blades for powered reusable handpiece of Diego Powered Dissector and Drill System. Used in clinical settings for sinus procedures involving soft tissue and thin bone. Medline ReNewal process includes cleaning, disinfection, inspection, refurbishment, testing, packaging, labeling, and sterilization. Output is mechanical cutting/dissection of tissue/bone during surgery. Benefits include providing reprocessed surgical instruments for sinus applications.
Clinical Evidence
Bench testing only. Testing included biocompatibility per ISO 10993-1:2009, cleaning validation, product stability, and product performance (device integrity, simulated clinical use, artificial test soil, blade sharpness, irrigation function, and outflow function).
Technological Characteristics
Insertable blades for powered dissector and drill system. Reprocessed version of predicate. Materials and design identical to original manufacturer specifications. Sterilization performed as part of reprocessing cycle.
Indications for Use
Indicated for limited sinus applications involving soft tissue and thin bone, such as the lamina papyracea.
Regulatory Classification
Identification
An ear, nose, and throat bur is a device consisting of an interchangeable drill bit that is intended for use in an ear, nose, and throat electric or pneumatic surgical drill (§ 874.4250) for incising or removing bone in the ear, nose, or throat area. The bur consists of a carbide cutting tip on a metal shank or a coating of diamond on a metal shank. The device is used in mastoid surgery, frontal sinus surgery, and surgery of the facial nerves.
Predicate Devices
Sinus Application Blades of the Diego Powered Dissector and Drill System (K020594)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three faces overlapping to suggest community and support. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18, 2014
Medline Renewal Mr. Richard D. Wynkoop Vice President Quality Assurance & Regulatory Affairs 2747 SW 6th Street Redmond, Oregon 97756
Re: K142022
Trade/Device Name: Medline ReNewal Reprocessed Gyrus-Diego Sinus Application Blades Regulation Number: 21 CFR 874.4140 Regulation Name: Ear, nose, and throat bur Regulatory Class: Class I Product Code: NLY Dated: November 12, 2014 Received: November 17, 2014
Dear Mr. Wynkoop:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Page 3 – Mr. Richard D. Wynkoop
List of models cleared in K142022:
- 7013-8000 Serrated Sinus Application Blade ●
- 7013-8001 Serrated Sinus Application Blade ●
- 7013-8002 Smooth, Round Tip Sinus Application Blade ●
- 7013-8003 Smooth, Square Tip Sinus Application Blade .
- 7013-8033 Serrated Sinus Application Blade ●
- 7013-8034 Turbinate Sinus Application Blade ●
- 7013-8035 Serrated Sinus Application Blade ●
- 7013-8100 Aggressive Sinus Application Blade ●
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142022
Device Name
Medline ReNewal Reprocessed Gyrus-Diego Sinus Application Blades
Indications for Use (Describe)
Medline ReNewal Reprocessed Gyus-Diego Sinus Application Blades are designed for use in limited sinus applications involving soft tissue and thin bone such as lamina papyracea.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline Renewal logo. The word "MEDLINE" is in blue, with a blue star-like symbol to the right of it. To the right of the Medline logo is a vertical line, followed by the word "Renewal" in green and blue. Below the word "Renewal" is the phrase "Full Circle Reprocessing" in a smaller font.
K142022 Medline ReNewal Reprocessed Gyrus-Diego Sinus Application
K142022 Medline ReNewal Reprocessed Gyrus-Diego Sinus Application Blades
## 5.0 510(k) Summary
| Submitter/<br>Owner | Medline ReNewal<br>2747 SW 6th St.<br>Redmond, OR 97756 | | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact Names | Brandi Panteleon<br>Director, Regulatory Affairs<br>P: 541-923-3310<br>F: 541-923-3375<br>E: bpanteleon@medline.com | Richard D. Wynkoop<br>VP, Quality Assurance & Regulatory<br>Affairs<br>P: 541-923-3310<br>F: 541-923-3375<br>E: rwynkoop@medline.com | |
| Date Prepared | December 15, 2014 | | |
| Device Names | Proprietary Name: Medline ReNewal Reprocessed Gyrus-Diego Sinus<br>Application Blades<br>Common Name: ear, nose, and throat shaver, reprocessed | | |
| Classification | § 874.4140 - ear, nose, and throat bur<br>Class I, nonexempt<br>Product Code NLY | | |
| Predicate<br>Device | K020594 Sinus Application Blades of the Diego Powered Dissector and<br>Drill System manufactured by OLYMPUS/Gyrus ACMI. | | |
| Device<br>Description | Medline ReNewal Reprocessed Gyrus-Diego Sinus Application Blades are<br>sterile single use devices that have been cleaned, disinfected, inspected,<br>refurbished, tested, packaged, labeled, and sterilized. The Medline<br>ReNewal Reprocessed Gyrus-Diego Sinus Application Blades are originally<br>manufactured by OLYMPUS/Gyrus ACMI. | | |
| Statement of<br>Intended Use | Medline ReNewal Reprocessed Gyrus-Diego Sinus Application Blades are<br>designed for use in limited sinus applications involving soft tissue and thin<br>bone such as lamina papyracea. | | |
| Technological<br>Characteristics | The technological characteristics of the proposed devices are substantially<br>equivalent to the predicate devices listed in this submission. The proposed<br>devices are a reprocessed version of the predicate devices. | | |
| Performance<br>Testing | The functional characteristics of the proposed devices have been evaluated<br>and found to be equivalent to the predicate devices after the specified<br>number of reprocessing cycles. Testing included:<br>biocompatibility testing performed in accordance with ISO 10993-1:2009, cleaning validation, product stability, product performance (bench) tests included: device integrity, use and operation using simulated clinical use and artificial test soil, | | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Medline Renewal logo. The Medline logo is on the left, and the Renewal logo is on the right. The Renewal logo is green and blue, and the text "Full Circle Reprocessing" is below the Renewal logo.
K142022 Medline ReNewal Reprocessed Gyrus-Diego Sinus Application
K142022 Medline ReNewal Reprocessed Gyrus-Diego Sinus Application Blades
| | blade sharpness, irrigation function, and outflow function. |
|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | Based on comparisons of the indications for use, intended use, technological characteristics, and performance data to the predicate devices, Medline ReNewal Reprocessed Gyrus-Diego Sinus Application Blades are substantially equivalent to the predicate devices. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Medline Renewal logo. The word "Renewal" is in large font, with "Re" in green and "newal" in blue. Below "Renewal" is the phrase "Full Circle Reprocessing" in a smaller font. To the left of the text is the Medline logo, which is a blue star-like shape with the word "MEDLINE" next to it.
| | Predicate | Proposed | Comparison |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Characteristics | Sinus Application Blades of the<br>Diego Powered Dissector and Drill<br>System | Medline ReNewal Reprocessed<br>Gyrus-Diego Sinus Application<br>Blades | Same device; different manufacturer |
| Predicate 510(k) | K020594 | TBD | N/A |
| Product Name | Gyrus-Diego Sinus Application<br>Blades | Reprocessed Gyrus-Diego Sinus<br>Application Blades | N/A |
| Intended Use | Gyrus Sinus Application Blades are<br>intended to be used with the powered<br>reusable hand piece of the Powered<br>Diego Dissector and Drill System and<br>are designed for use in sinus<br>applications. | Medline ReNewal Reprocessed<br>Gyrus-Diego Sinus Application<br>Blades are intended to be used with<br>the powered reusable hand piece of<br>the Powered Diego Dissector and<br>Drill System and are designed for<br>use in sinus applications. | Same intended use |
| Indications for<br>Use | Gyrus Sinus Application Blades are<br>designed for use in sinus<br>applications. Sinus applications<br>include ethmoidectomy/<br>sphenoehtmoidectomy, polypectomy,<br>septoplasty, transsphenoidal<br>procedures, and procedures such as<br>the removal of septal spurs,<br>antrostomy, and endoscopic<br>dacryocystorhinostomy (DCR). | Medline ReNewal Reprocessed<br>Gyrus-Diego Sinus Application<br>Blades are designed for use in<br>limited sinus applications involving<br>soft tissue and thin bone such as<br>lamina papyracea. | Only the sinus application blades<br>indicated for soft tissue and thin bone<br>such as lamina papyracea will be<br>reprocessed by Medline ReNewal. The<br>predicate indications for use is for the<br>whole system, not just the sinus<br>application blades. Thus, the Medline<br>ReNewal Reprocessed Gyrus-Diego<br>Sinus Application Blades have a more<br>focused indications for use than the<br>predicate system. |
| Configuration | Insertable blades for powered<br>dissector and drill system | Insertable blades for powered<br>dissector and drill system | Same configuration |
#### Predicate and Medline ReNewal Reprocessed Gyrus-Diego Sinus Application Blades device comparison chart. Table 1:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.