K020594 · Gyrus Ent LLC · ERL · Mar 8, 2002 · Ear, Nose, Throat
Device Facts
Record ID
K020594
Device Name
DIEGO POWERED DISSECTOR AND DRILL SYSTEM
Applicant
Gyrus Ent LLC
Product Code
ERL · Ear, Nose, Throat
Decision Date
Mar 8, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diego Powered Dissector and Drill System's intended use is the cutting and removal of bone and tissue in general ENT, Head & Neck, and otoneurologic procedures. Sinus applications would embody ethmoidectomy/sphenoethmoidectomy, polypectomy, septoplasty, and procedures such as the removal of septal spurs, antrostomy, frontal sinus trephination and irrigation, frontal sinus drill-out, endoscopic DCR and trans-spehnoidal procedures. Nasopharyngeal/laryngeal procedures would comprise adenoidectorny, laryngeal lesion de-bulking, laryngeal polypectomy, tracheal procedures, and tonsillectomy. Head & neck procedures would encompass soft tissue shaving, rhinoplasty (narrowing of the bony vault and revision of the bony pyramid), removal of fatty (adipose) tissue (lipodebridement) in the maxillary and mandibular regions of the face, and acoustic neuroma removal. Otology procedures would include mastoidectomy and mastoidotomy.
Device Story
Powered surgical system for ENT, head/neck, and otoneurologic procedures; comprises console, footswitch, irrigation pump, and handpieces. Inputs: user-selected operating mode (forward, reverse, oscillate) and speed via footswitch. Console processes inputs via 8-bit microprocessor to control handpiece motor speed; provides digital speed indication. Outputs: mechanical rotation/oscillation of burs or blades for cutting/removing bone and soft tissue. Used in OR by surgeons. Integrated irrigation pump manages fluid delivery. System benefits include precise tissue/bone removal during complex procedures like mastoidectomy or rhinoplasty. Safety/effectiveness supported by compliance with electrical/EMC standards (UL 2601-1, IEC 60601 series).
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: Stainless steel nosecone, aluminum housing. Power: Electric. Modes: Forward, reverse, oscillate. Speeds: Up to 18,000 rpm (sinus) and 44,000 rpm (otologic). Connectivity: Console-based with footswitch. Sterilization: Autoclave or EO. Standards: UL 2601-1, CSA 22.2 No. 601, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 61000-4-2/3/4/5, IEC 529, ISO 10993-1, EN 55011 Class B.
Indications for Use
Indicated for patients undergoing general ENT, head and neck, and otoneurologic surgical procedures requiring bone or tissue cutting and removal. Includes sinus, nasopharyngeal, laryngeal, and otologic procedures.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
{0}------------------------------------------------
KO 20594
Food and Drug Administration 510(k) Notification - Diego™ Powered Dissector and Drill System February 21, 2002
MAR 8 2002
## 510(k) Summary of Safety and Effectiveness
| Trade Name:<br>Common Name:<br>Classification Name: | Diego Dissector and Drill System<br>Electrical Surgical Drill/Shaver<br>Ear, Nose and Throat electric or pneumatic surgical<br>drill (§ 874.4250) |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Greg Sredin<br>Sr. Regulatory Affairs Specialist<br>Gyrus ENT LLC<br>2925 Appling Road<br>Bartlett, TN 38133 |
| Telephone:<br>Telefax: | (901) 373-0200<br>(901) 373-0242 |
| Date Prepared: | February 21, 2002 |
The Diego Powered Dissector and Drill System's intended use is the cutting and removal of bone and tissue in general ENT, Head & Neck, and otoneurologic procedures. Sinus applications would embody ethmoidectomy/sphenoethmoidectomy, polypectomy, septoplasty, and procedures such as the removal of septal spurs, antrostomy, frontal sinus trephination and irrigation, frontal sinus drill-out, endoscopic DCR and trans-spehnoidal procedures. Nasopharyngeal/laryngeal procedures would comprise adenoidectorny, laryngeal lesion de-bulking, laryngeal polypectomy, tracheal procedures, and tonsillectomy. Head & neck procedures would encompass soft tissue shaving, rhinoplasty (narrowing of the bony vault and revision of the bony pyramid), removal of fatty (adipose) tissue (lipodebridement) in the maxillary and mandibular regions of the face, and acoustic neuroma removal. Otology procedures would include mastoidectomy and mastoidotomy.
The Diego Powered Dissector and Drill system that is described in this notification has the same technological characteristics, power modality and mode of operation as the predicate device. The intended uses are substantially equivalent to the described predicate Turbo 7000 System. The Diego Powered Dissector and Drill system is designed to meet UL 2601-1 including Australian deviations, CSA 22.2 No. 601, IEC 601-1-1 (EN 60601-1), IEC 601-1-2 (EN 60601-1-2), IEC 601-1-4 (EN 60601-1-2), IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-4, IEC 61000-4-5, IEC 529, ISO 10993-1, EN 55011. Class B.
The Diego Powered Dissector and Drill System is substantially equivalent to the Turbo 7000 System and difference of the processors that control the units should not affect the safety or effectiveness of the device.
{1}------------------------------------------------
Comparison Chart
**vs**
## vs.
Turbo 7000 Shaver/Drill Syster
# Console, Footswitch, and Irrigation Pump
| | Diego Powered Dissector and Drill System | Turbo 7000 Shaver Drill System |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | General ENT, Head and Neck, and Otoneurologic Procedures | General ENT, Head and Neck, and Otoneurologic Procedures |
| Driver Configuration<br>Operating Modes | Console with separate footswitch<br>Forward, Reverse, Oscillate | Console with separate footswitch<br>Forward, Reverse, Oscillate |
| Operating Speeds | Up to 18,000 rpm for sinus surgery and<br>up to 44,000 rpm for otologic,<br>otoneurologic, and acoustic neuroma<br>procedures | Up to 7,000 rpm for sinus surgery and<br>up to 44,000 rpm for otologic,<br>otoneurologic, and acoustic neuroma<br>procedures |
| Speed Indication | Digital | Digital |
| Footswitch | Electric | Electric |
| Irrigation Pump | Yes, integrated into console | Yes, integrated into console |
| Number of Handpieces supported | 4 | 5 |
| | Diego, Stapes, and Viper Straight and<br>Angled | Turbo 7000, Enhanced ESSential<br>Shaver, Stapes, and Viper Straight and<br>Angled |
| Processor Control | 8-bit Microprocessor, Re-<br>programmable (Factory Only) | FPGA (Field Programmable Gate<br>Array) chip. Not re-programmable |
| | Diego Powered Dissector and Drill System | Turbo 7000 Shaver Drill System |
| Intended Use | General ENT, Head and Neck, and Otoneurologic Procedures | General ENT, Head and Neck, and Otoneurologic Procedures<br>Enhanced ESSential Shaver and Stapes |
| Material of Housing | Diego - Stainless Steel Nosecone and Aluminum Housing<br>Viper Handpiece - Stainless Steel Nosecone and Aluminum Housing | - Stainless Steel<br>Viper Handpiece - Stainless Steel Nosecone and Aluminum Housing<br>Turbo 7000 – Aluminum with Stainless Steel Sleeve |
| Suction Capability | Yes for Shaver Handpiece<br>No for Drill Handpiece | Yes for Shaver Handpiece<br>No for Drill Handpiece |
| Irrigation Capability | Yes, with the exception of the Stapes Drill Handpiece | Yes, with the exception of the Stapes Drill Handpiece |
| Chucking Mechanism | Diego is Spring Loaded and able to chuck in multiple positions. The Viper handpieces utilize locking rings and the Stapes uses a collet to secure the burrs | Spring Loaded; ESSential Shaver and Turbo 7000 Shaver Blades/Burrs are able to chuck in multiple positions.<br>The Viper handpieces utilize locking rings and the Stapes uses a collet to secure the burrs |
| Sterilization Method | Autoclave or EO | Autoclave or EO |
{2}------------------------------------------------
Handpieces
: 上一
.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design of three horizontal lines that curve downwards, resembling a human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 8 2002
Gyrus ENT LLC c/o Gregory Sredin Sr. Regulatory Affairs Specialist 2925 Appling Road Bartlett, TN 38133
Re: K020594
Trade/Device Name: Diego™ Powered Dissector and Drill System Regulation Number: 21 CFR 874.4250 Regulation Name: ENT Pneumatic or Electric Drill Regulatory Class: Class II Product Code: ERL Dated: February 21, 2002 Received: February 22, 2002
Dear:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address . http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Food and Drug Administration 510(k) Notification -- Diego™ Powered Dissector and Drill System February 21, 2002
KU20594
510(k) Number: Device Name:
Diego™ Powered Dissector and Drill System
### Indications for Use:
The Diego Powered Dissector and Drill System's intended use is the cutting and The Dlego Powered Dissociol and Bhil System. Beath and otoneurologic procedures.
Sinus applications would embody:
- ethmoidectory/sphenoethmoidectomy .
- polypectomy .
- septoplasty and .
- procedures such as .
- antrostomy ●
- endoscopic DCR .
- frontal sinus drill-out .
- frontal sinus trephination and irrigation ●
- septal spurs removal .
- trans-spehnoidal procedures .
Nasopharyngeal/laryngeal procedures would comprise:
- adenoidectomy .
- laryngeal lesion de-bulking .
- laryngeal polypectomy .
- tracheal procedures ●
- tonsillectomy .
Head & Neck procedures would encompass:
- soft tissue shaving .
- rhinoplasty (narrowing of the bony vault and revision of the bony pyramid) .
- minopial of fatty (adipose) tissue (lipodebridement) in the maxillary and mandibular . regions of the face
- acoustic neuroma removal .
Otology procedures would include:
- mastoidectomy ●
- mastoidotomy. .
{6}------------------------------------------------
### Food and Drug Administration 510(k) Notification - Diego™ Powered Dissector and Drill System February 21, 2002
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | ✓ |
|----------------------|---|
| Counter | |
| (Per 21 CFR 801.109) | |
OR
Over-the-
(Optional Format 1-2-96)
are
(Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises
510(k) Number
Page 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.