Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not for mammography
Device Story
MobileDiagnost wDR 2.0 is a motorized mobile radiographic system; consists of mobile base unit, user interface (computer, keyboard, display, mouse), and wireless portable SkyPlate X-ray detectors. Operator generates, processes, and handles digital X-ray images at point-of-care. System uses X-ray tube, generator, and collimator to capture images via flat solid-state detector (thin film transistor, scintillator). Wireless technology transmits images to archive systems; integrates with RIS for patient/exam data. Healthcare providers use output for diagnostic assessment; supports clinical decision-making through high-quality digital imaging. Benefits include improved mobility and efficient workflow for bedside radiography.
Clinical Evidence
No clinical trials required for substantial equivalence. Bench testing performed for DQE, MTF, aliasing, signal levels, and latent image decay. Design validation included a single-blinded clinical image concurrence study comparing SkyPlate detectors to the predicate WPD FD-W17 detector, confirming equivalent diagnostic capability.
Indicated for adult and pediatric patients requiring diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, and extremities. Positioning includes sitting, standing, or lying (prone/supine). Contraindicated for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Philips Eleva Workspot for DigitalDiagnost (K141736)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine and health. The caduceus is composed of three abstract human figures, representing the department's focus on people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 18, 2014
Sedecal SA % Ms. Jennifer Cartledge Consultant REU Associates Inc. 409 Woodridge Drive SENECA SC 29672
Re: K141895
Trade/Device Name: Mobile Diagnost wDR 2.0 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL, MQB Dated: August 18, 2014 Received: August 19, 2014
Dear Ms. Cartledge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Sm.h.7)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K141895
Device Name MobileDiagnost wDR 2.0
#### Indications for Use (Describe)
Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not for mammography
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 7 510(k) Summary
510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Manufacturer: | SEDECAL SA<br>C/ Pelaya, 9 – 13<br>Pol. Ind. Río de Janeiro<br>28110 Algete, Madrid<br>Spain (España)<br>Establishment registration number:9617251 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jennifer Cartledge<br>REU Associates Inc.<br>409 Woodridge Drive<br>Seneca, SC 29672<br>Tel.: (864) 500-0097<br>Email: jcartledge@reuassociates.com |
| Device Name: | <i>MobileDiagnost wDR 2.0</i> |
| Classification: | (Primary) Product code IZL<br>(Primary) Classification Name: Mobile X-Ray System<br>(Primary) Classification Regulation: 21CFR 892.1720<br>Classification Panel: Radiology<br>Device Class: Class II |
| Predicate Device: | Trade Name: <i>MobileDiagnost wDR</i><br>Manufacturer: Sedecal SA<br>510(k) Clearance: K111725 - July 19, 2011<br>(Primary) Product code IZL<br>(Primary) Classification Name: Mobile X-Ray System<br>(Primary) Classification Regulation: 21CFR 892.1720<br>(Secondary) Product code: MQB<br>(Secondary) Classification Name: Stationary x-ray system<br>(Secondary) Classification Regulation: 21CFR 892.1680<br>Classification Panel: Radiology<br>Device Class: Class II |
| Device description: | The <i>MobileDiagnost wDR 2.0</i> system is a motorized mobile radiographic<br>system consisting of a mobile base unit, and a user interface (computer, key-<br>board, display, mouse), combined with a flat solid state X-ray detector. It is<br>used by the operator to generate, process and handle digital X-ray images.<br>The <i>MobileDiagnost wDR 2.0</i> integrates a new generation of wireless porta-<br>ble x-ray detectors ( <i>SkyPlate</i> ) to replace the former detector WPD FD-W17<br>cleared under the predicate submission (K111725). |
| Indications for | The Indication for Use of the <i>MobileDiagnost wDR 2.0</i> is identical to that of |
{4}------------------------------------------------
the currently marketed and predicate device, MobileDiagnos wDR, K111725 Use: -July 19, 2011, and is as follows: Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not for mammography.
Fundamental The MobileDiagnos wDR 2.0 employs the same basic construction and fun-Scientific damental scientific technology as provided with the currently marketed and Technology: predicate device, MobileDiagnos wDR, K111725 -July 19, 2011, with regards to the functionality of the following: image receptor type, image processor, automatic image processing, manual image processing, advanced image processing, image export (interfaces), and the use of standard monitors. The new SkyPlate detectors employ the same basic construction and fundamental scientific technology as provided with the detector integrated into the currently marketed predicate device. The proposed MobileDiagnost wDR 2.0 employs the same basic construction and fundamental scientific technology as provided with the currently marketed and predicate device, MobileDiagnost wDR (K111725). With the exception of the detector of the proposed MobileDiagnost wDR 2.0, the mobile base unit is equivalent in comparison with the currently marketed and predicate, MobileDiagnost wDR. As stated in the premarket notification, the integration of the new SkyPlate detector did not necessitate or cause any modification in
- the x-ray tube. .
- . the x-ray generator,
- the x-ray control, or
- . the collimator of the MobileDiagnost wDR 2.0.
Additionally, the software modifications for this submission are only made to ensure proper integration of the new detector in accordance with CDRH's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
The design features of the major components of the currently marketed and predicate MobileDiagnost wDR detector (K111725), which include X-ray absorber, thin film transistor, scintillator and image processor are equivalent to the new SkyPlate wireless portable detectors of the proposed MobileDiagnost wDR 2.0. Additionally, the wireless technology utilized in the currently marketed and predicate MobileDiagnost wDR detector (K111725) is the same as the wireless technology provided with the new SkyPlate detectors of the proposed MobileDiagnost wDR 2.0. The currently marketed and predicate MobileDiagnost wDR device (K111725) offers a detector size of 489 x 466 x 25 mm, while the SkyPlate detectors of the proposed MobileDiagnost wDR 2.0 is provided in both Large: 384 x 460 x 16 mm and Small: 268 x 328 x 16 mm detector sizes. The image size of currently marketed and predicate MobileDiagnost wDR device (K111725)is 3000 x 2400 Pixels, while the SkyPlate detectors of the proposed MobileDiagnost wDR 2.0 is 2330 x 2846 Pixels for the large and 1500 x 1920 Pixels for the small. Finally, the X-ray field of the currently marketed and predicate MobileDiagnost wDR device (K111725) is 432 x 341.1 mm as compared to X-ray fields of 344.8 x 421.2 mm and 222.0 x 284.1 mm provided with the Large and Small Sky
{5}------------------------------------------------
Plate detectors, respectively, of the proposed MobileDiagnost wDR 2.0. The slightly modified specifications of the additional design features, such as detector size, detector weight, image, Pixel size, etc. of the new SkyPlate wireless portable detectors do not affect the safety or effectiveness of the device since it does not affect the image quality. Regarding the major components of x-ray systems as per 21 CFR 1020.30(a), only the collimator was replaced with an equivalent type from the same manufacturer with almost identical specifications. The following attributes of the equivalent collimator (R221 A DHHS) provided with the proposed MobileDiagnost wDR 2.0 are identical to the collimator, (R108F DHHS), provided with the currently marketed and predicate MobileDiagnost wDR device (K111725): the Field Type, Additional Variable Filtration, Light Source, SID Measurement, Mounting of DAP Chamber, Minimum Inherent Filtration (Al Equivalent), cCSAus Certification, DHHS Certification, Proximity Sensor, X-Ray Rating, Power Supply, and Multilayer Square specification . The equivalent collimator does provide an additional Centering Light option for the Image Receptor Alignment and an Automatic Centering with Metal Flange option for the Alignment to Focus and eliminates the option for a Plastic -120° Flange. Apart from these minor differences, the specification of the replacement collimator is identical to that of the predicate collimator and therefore, does not raise any new issues of safety or effectiveness. Additionally, the alignment and collimation procedures of the proposed MobileDiagnos wDR 2.0 remain identical to that of the currently marketed and predicate MobileDiagnos wDR (K111725).
Note, both the proposed MobileDiagnost wDR 2.0 and the currently marketed and predicate MobileDiagnost wDR device (K111725) use wireless technology for image acquisition and the transmission of images to archive systems and provide patient and exam information from radiology information systems (RIS), but will still be able to utilize CR or traditional film cassettes as a back-up scenario.
Based on the information provided above, the MobileDiagnos wDR 2.0 is considered substantially equivalent to the currently marketed and predicate device, MobileDiagnos wDR, K111725 -July 19, 2011 in terms of fundamental scientific technology.
The non-clinical performance testing constrains that the main physical values for comparison of X-ray devices like DOE and MTF are basically equal or better than the predicate device ranging from 61% to 14% for MTF (predicate 60% to 15%) and from 66% to 24% for DQE (predicate 66% to 22%).
The non-clinical performance data demonstrate substantial equivalence. They include (but are not limited to):
- > Detective Quantum Efficiency (DQE)
- A Modulation Transfer Function (MTF)
- > Aliasing
- > Output signal level
- A Lag, Memory Effects and Ghost Images
- A Allowable types and quantity of defects
- A Change in detection sensitivity
Summary of Non-Clinical Performance Data:
{6}------------------------------------------------
- Latent image decay characteristic A
- A Recovery time for radiographic devices
- A Dose requirements and reciprocity changes
- A Stability of device characteristics
- A Uniformity of device characteristics
- A Reuse rate for radiographic devices
- A Test pattern image tests
The MobileDiagnos wDR 2.0 complies with the following international and FDA-recognized consensus standards:
- IEC 62304 Medical device software – Software life cycle processes
- . AAMI/ANSI IEC 62366 Application of usability engineering to medical devices
- ISO 14971 Application of risk management to medical devices
- ANSI/AAMI ES60601-1 Medical Electrical Equipment -- Part 1: General Requirements For Basic Safety And Essential Performance
- . IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
- . IEC 60601-1-3 Medical Electrical Equipment - Part 1-3: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Radiation Protection In Diagnostic X-Ray Equipment
- . IEC 60601-2-54 Medical Electrical Equipment - Part 2-54: Particular Requirements For The Basic Safety And Essential Performance Of X-Ray Equipment For Radiography And Radioscopy
- . IEC 62220-1 ME equipment - Part 1-2: Determination of the detective quantum efficiency
- . NEMA PS 3.1 - 3.20 Digital Imaging And Communications In Medicine (DICOM) Set
- ISO 10993-1 Biological Evaluation Of Medical Devices - Part 1: Evaluation And Testing Within A Risk Management Process
Non-clinical software verification and validation tests have been performed with regards to the intended use, technical claims, requirements specifications and risk management results.
The non-clinical software verification and validation test results demonstrate that the MobileDiagnos wDR 2.0 complies with international and FDArecognized consensus standards and meets the acceptance criteria and is adequate for its intended use. Therefore, the MobileDiagnos wDR 2.0 is substantially equivalent to the currently marketed device, MobileDiagnos wDR, K111725 -July 19, 2011 in terms of safety and effectiveness.
Summary of No clinical data is necessary to evaluate safety or effectiveness for purposes Clinical Data: of determining substantial equivalence of the proposed modification. Bench testing is sufficient to assess the device safety and effectiveness, including demonstrating equivalent image quality. However, as part of design validation, clinical images were collected and analyzed, to ensure that images constructed by the modified system meet user needs, as per 21 CFR 820.30(g) "Design Validation". This study consisted of a single blinded, concurrence study, according to CDRH's Guidance for the Submission of 510(k)'s for
{7}------------------------------------------------
Solid State X-ray Imaging Devices. The study confirmed that the new x-ray detectors, SkyPlate, meet user needs and provide images of equivalent diagnostic capability to the predicate device, the WPD FD-W17 detector cleared under K111725 with the MobileDiagnos wDR –July 19, 2011, meeting all requirements regarding substantial equivalence pertaining to user needs. Substantial The MobileDiagnos wDR 2.0 is substantially equivalent to the currently Equivalence marketed and predicate device MobileDiagnos wDR, K111725 -July 19, Conclusion: 2011 in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Furthermore. The new portable x-ray detectors SkyPlate have already received a 510(k) market clearance under K133259 on January 24, 2014 (integrated into the Siemens Ysio Max x-ray system), under K141381 on June 12, 2014 (integrated into the Philips DuraDiagnost radiography system), and under K141736 on July 25, 2014 (integrated into the Philips ProGrade and Philips Eleva Workspot for DigitalDiagnost).Additionally, substantial equivalence was demonstrated with non-clinical performance (verification and validation) tests, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC 62304, IEC 62366, ISO 14971, ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-54, IEC 62220-1, and NEMA PS 3.1 - 3.20 and ISO 10993-1. The results of these tests demonstrate that MobileDiagnost wDR 2.0 met the acceptance criteria and are adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, software validation, and clinical image concurrence data demonstrates that the device is as safe, as effective, and performs as well or better than the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.