The DuraDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctor or technician. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications.
Device Story
DuraDiagnost is a digital radiography (DR) system for diagnostic imaging of human anatomy. System components include an X-ray tube assembly, tube column, wall stand, patient table with floating top, high-voltage generator, and an acquisition/reviewing workstation. The device captures X-ray signals via fixed or wireless digital detectors (scintillator-based). The workstation performs image post-processing, storage, and viewing; images are transferable via DICOM network. Operated by physicians or technicians in clinical settings. The system supports various patient positions (sitting, standing, lying). Output provides radiographic images for clinical diagnostic review, aiding healthcare providers in patient assessment and diagnosis.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including verification and validation of design features, technical specifications, and risk management. Testing demonstrated compliance with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, and IEC 60601-2-54. Wireless technology performance verified per FDA guidance.
Indicated for generating radiographic images of human anatomy in patients sitting, standing, or lying (prone/supine). Not for mammography. Intended for use by qualified doctors or technicians.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Brivo XR385, Digital Diagnostic Radiographic System (K103448)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "PHILIPS" in large, bold, black letters. Above and to the right of the word, there is some handwritten text that reads "K14381 Page 1 of 12". The handwritten text is smaller and less distinct than the word "PHILIPS".
# 510(k) Summary of Safety and Effectiveness
.
JUN 1 2 2014
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92. .
| Date Prepared: | March 21, 2014 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd.<br>No. 258, Zhong Yuan Road, Suzhou Industrial Park, 215024 Suzhou,<br>Jiangsu Province, PEOPLE'S REPUBLIC OF CHINA<br>Establishment Registration Number: Not registered yet |
| Contact Person: | Gordon Shu<br>Regulatory Affairs Manager<br>Phone: +86-512-67336804<br>Fax: +86-512-68018677<br>E-mail: Gordon.Shu@philips.com |
| Trade Name: | DuraDiagnost |
| Common Name: | Digital Diagnostic Radiographic System |
| Classification: | Classification Name: Stationary X-Ray System<br>Classification Regulation: 21CFR §892.1680<br>Classification Panel: Radiology<br>Device Class: Class II<br>Classification Product Code: KPR (System, X-Ray, Stationary)<br>Subsequent Product Code: MQB (solid state x-ray imager (flat panel/digital imager)) |
| Predicate Device 1: | Trade Name: Brivo XR385, Digital Diagnostic Radiographic System<br>Manufacturer: GE MEDICAL SYSTEM, LCC<br>510(k) Clearance: K103448 (August 12, 2011)<br>Classification Regulation: 21 CFR, Part 892.1680<br>Classification Name: Stationary X-Ray System<br>Classification Panel: Radiology<br>Device Class: Class II<br>Product Code: KPR |
{1}------------------------------------------------
Device Trade Name: Philips BuckyVision Predicate Manufacturer: Philips Medical Systems, Inc. 510(k) Clearance: K982795 (November 24,1998) Classification 21CFR §892.1680 Regulation: Classification Name: Stationary X-Ray System Classification Panel: Radiology Device Class: Class II KPR & MQB Product Code: ddRVersa™ Motion Trade Name: Predicate Device
> Manufacturer: Swissray Medical AG 510(k) Clearance: K123005 (December 7, 2012) 21CFR \$892.1680 Classification Regulation: Classification Name: Stationary X-Ray System
Classification Panel: Radiology
Device Class:
Class II
Device Code: KPR & MQB
Device description:
2:
3:
The Philips DuraDiagnost Digital Diagnostic Radiographic System (DuraDiagnost) is a flexible digital radiography (DR) system that is designed to provide fast and smooth radiography examinations of sitting, standing or lying patients. The Philips DuraDiagnost consist of the following components: Tube column with X-ray assembly, wall stand with detector carrier, patient table with detector carrier and floating table top, high voltage generator, and acquisition and reviewing workstation for post-processing, storage and viewing of images. Images may be transferred via a DICOM network for printing, storage and detailed review.
{2}------------------------------------------------
Indications Use:
for
The DuraDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctor or technician. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications.
Summarv of the Technological Characteristics & Fundamental Scientific Technology:
The components of the Philips DuraDiagnost employ similar basic construction and fundamental scientific technology as provided with the currently marketed and predicate GE Brivo XR385 and Philips BuckyVision with regards to the functionality of the following components: Integrated tube column, patient table with a floating table top, high-voltage generator, dual-focus rotation anode X-Ray tube, manual beam limiting device, digital detector, wall stand and workstation for images post-processing, storage and viewing (See the comparison Table provided below, comparing the Philips DuraDiagnost to the currently marketed predicate devices). The outcome of this comparison demonstrates that the minor differences in the technological characteristics do not affect the safety or effectiveness of the Philips DuraDiagnost when compared to the currently marketed and predicate devices.
The wireless portable detector (Model No. 3543EZ) of the Philips DuraDiagnost is identical to the wireless portable detector (Model No. 3543EZ) of the currently marketed and predicate ddRVersaTM Motion (K123005 - December 7, 2012) - Swissray Medical AG and is manufactured by Trixell. Therefore, both the wireless portable detector of the Philips DuraDiagnost and the wireless portable detector of the currently marketed and predicate ddRVersa™ Motion employ identical fundamental scientific technology.
Based on the information provided above, the Philips DuraDiagnost is considered substantially equivalent to the currently marketed and predicate device, GE Brivo XR385 (K103448 - August 12, 2011) and Philips Bucky Vision (K982795 - November 24, 1998) in terms of fundamental scientific technology. With regards to wireless portable detector, the Philips DuraDiagnost is considered substantially equivalent to the currently marketed and predicate ddRVersa™ Motion (K123005 - December 7, 2012) - Swissray Medical AG in terms of fundamental scientific technology.
{3}------------------------------------------------
. . .
| Feature | Currently<br>marketed and<br>Predicate GE Brivo<br>XR385 (K103448) | Currently marketed<br>and Predicate Philips<br>BuckyVision<br>(K982795) | Currently marketed<br>and Predicate Swissray<br>ddRVersa™ Motion<br>(K123005) | Philips<br>DuraDiagnost | Comparison Results |
|-----------------------|--------------------------------------------------------------------|-------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design characteristic | Toshiba E7843X | RO 1750 & SRO 33100 | Note: The currently<br>marketed and predicate<br>Swissray ddRVersa™<br>Motion (K123005 –<br>December 7, 2012) is only<br>used to demonstrate<br>substantial equivalence to<br>the wireless portable<br>detector (model 3543EZ)<br>of the Philips<br>DuraDiagnost, since it is<br>manufactured by the same<br>manufacturer Trixell. | RO 1750 & SRO 33100 | Identical. The Philips<br>DuraDiagnost and the currently<br>marketed and predicate Philips<br>BuckyVision both use the<br>identical X-ray tube. Thus,<br>demonstrating SE. |
| Max Tube<br>Voltage | 150kV | 150kV | | 150kV | Identical. Thus, demonstrating<br>SE. |
| Focal spot<br>size | 0.6mm/1.2mm | 0.6mm/1.2mm | | 0.6mm/1.2mm | Identical. Thus, demonstrating<br>SE. |
| Tube Max<br>power | 50K W | 50KW/100KW | | 50KW/100KW | Identical. The Philips<br>DuraDiagnost and the currently<br>marketed and predicate Philips<br>BuckyVision both use the<br>identical X-ray tube. Thus,<br>demonstrating SE. |
| Anode Type | Rotation | Rotation | | Rotation | Identical. Thus, demonstrating<br>SE. |
Philips DuraDiagnost 510 (K)
510(K) Summary
.
・
Page 4 of 12
:
{4}------------------------------------------------
:
| Generator | Max Power | KV range | Milli ampere<br>sec (mAs)<br>product | Collimator<br>Operation<br>Mode | | Shape of<br>Beam | Detector<br>Type | X-ray<br>Scintillator<br>Material | Image Area | | | |
|-------------------------------------------------------------------------------------------|----------------|----------|------------------------------------------|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|-----------------------------|-------------------------------------------|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Unknown | 50KW | 40-150 | 0.02 mAs-512<br>mAs(with AEC<br>control) | Manual collimation | | Rectangular | Digital Detector (Fixed) | Gadolinium<br>Oxysulphide (GdOS) | 40.4cm x 40.4 cm | | | Identical. Thus, demonstrating<br>SE. |
| EMD Technologies,<br>M-CABINET CXA 50kW,<br>M-CABINET CXA 65kW,<br>M-CABINET CXA 80kW | 50KW/65KW/80KW | 40-150 | 0.5 mAs-600 mAs (with<br>AEC control) | Motorized automatic<br>collimation or manual<br>collimation | Note: The currently<br>marketed and predicate<br>Swissray ddRVersaTM<br>Motion (K123005 –<br>December 7, 2012) is only<br>used to demonstrate<br>substantial equivalence to | Rectangular | Digital Detector (Fixed) | Cesium Iodide | 42.5 cm x 42 cm | the wireless portable<br>detector (model 3543EZ)<br>of the Philips<br>DuraDiagnost, since it is<br>manufactured by the same<br>manufacturer Trixell. | | |
| EMD Technologies,<br>M-CABINET CXA<br>50kW, M-CABINET<br>CXA 65kW, M-<br>CABINET CXA 80kW | 50KW/65KW/80KW | 40-150 | 0.5 mAs-600 mAs (with<br>AEC control) | Manual collimation | Identical. The Philips<br>DuraDiagnost and the currently<br>marketed and predicate Philips<br>BuckyVision both use the<br>identical Generator. Thus,<br>demonstrating SE. | Rectangular | Digital Detector (Wireless) | Cesium Iodide | 42.4cm x 34.8cm<br>(wireless) | | Digital Detector (Fixed<br>& Wireless) | Identical. The Fixed and<br>Wireless Portable Detector of the<br>Philips DuraDiagnost is<br>identical to the currently<br>marketed and predicate GE's<br>Brivo XR385, Philips Bucky<br>Vision and Swissray's<br>ddRVersa™ Motion in terms of<br>the type of the detector. Thus,<br>demonstrating SE. |
| | | | | | Identical. The Philips<br>DuraDiagnost and the currently<br>marketed and predicate Philips<br>BuckyVision both use the<br>identical power. Thus,<br>demonstrating SE. | | | GdOS(Fixed)<br>Cesium<br>Iodide(Wireless) | 43cm x 43cm(Fixed)<br>42.4cm x 34.8cm<br>(wireless) | | | Identical. The Fixed and<br>Wireless Portable Detector of the<br>Philips DuraDiagnost and the<br>currently marketed and predicate<br>GE's Brivo XR385 and<br>Swissray's ddRVersa™ Motion<br>utilize identical scintillator<br>material fabricated from GdOS<br>and Cesium Iodide, respectively.<br>Thus demonstrating SE. |
| | | | | | Identical. Thus, demonstrating<br>SE. | | | | | | | Identical for Wireless Portable<br>Detector. |
| | | | | | Identical. The Philips<br>DuraDiagnost and the currently<br>marketed and predicate Philips<br>BuckyVision both use the<br>identical mAs. Thus,<br>demonstrating SE. | | | | | | | The Wireless Portable Detector<br>of the Philips DuraDiagnost and |
| | | | | | Same. The Philips<br>DuraDiagnost and the currently<br>marketed and predicate GE's<br>Brivo XR385 both use the<br>manual collimation mode. Thus,<br>demonstrating SE. | | | | | | | |
# Philips DuraDiagnost 510 (K)
.
Page 5 of 12
.
{5}------------------------------------------------
Philips DuraDiagnost 510 (K)
Page 6 of 12
"
{6}------------------------------------------------
| Image<br>Matrix Size | 2048 x 2048 | the currently marketed and<br>predicate Swissray's<br>ddRVersa™ Motion utilize<br>identical image area. Thus<br>demonstrating SE.<br><br>Similar for Fixed Detector. |
|----------------------|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 2874 x 2840 | The slight differences between<br>the image area of the Philips<br>DuraDiagnost and the currently<br>marketed and predicate predicate<br>Philips BuckyVision and GE's<br>Brivo XR385 are not anticipated<br>to significantly alter the<br>diagnostic image quality. Thus,<br>demonstrating SE. |
| | 2866 x 2350 (wireless) | Identical for Wireless Portable<br>Detector. |
| | 2874 x 2869 (Fixed) | |
| | 2866 x 2350 (wireless) | The Wireless Portable Detector<br>of the Philips DuraDiagnost and<br>the currently marketed and<br>predicate Swissray's<br>ddRVersa™ Motion utilize<br>identical image matrix size. Thus<br>demonstrating SE.<br><br>Similar for Fixed Detector.<br><br>The slight differences between<br>the image matrix size of the<br>Philips DuraDiagnost and the<br>currently marketed and predicate |
(
PHILIPS
.
Philips DuraDiagnost 510 (K)
Summary
Page 7 of 12
{7}------------------------------------------------
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>- | |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | | |
| Class Concession | Comments of Children Comments of Children | |
| I | | |
| - | | |
| | l | |
| | l | |
| Analog / Digital (A/D) conversion | 14 bits | 16 bits | 16 bits | 16 bits |
|-----------------------------------|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Identical. The Philips DuraDiagnost and the currently marketed and predicate Philips BuckyVision and Swissray's ddRVersa™ Motion all use the same A/D conversion. Thus, demonstrating SE. | Note: The currently marketed and predicate Swissray ddRVersa™ Motion (K123005 – December 7, 2012) is only used to demonstrate substantial equivalence to the wireless portable detector (model 3543EZ) of the Philips DuraDiagnost, since it is manufactured by the same manufacturer Trixell. | predicate Philips BuckyVision and GE's Brivo XR385 are not anticipated to significantly alter the diagnostic image resolution. Thus, demonstrating SE. |
| Source to Image Distance (SID) | Table: 50-110cm;<br>Wallstand: 100-180cm | SID depends on different configurations, because BuckyVision is ceiling suspension X-ray system. | Table: 40-115cm;<br>Wallstand: 110-245cm | The slight differences between the SID of the proposed DuraDiagnost and the currently marketed and predicate GE's Brivo XR385 are not anticipated to significantly alter the application usage. Thus, demonstrating SE. |
Philips DuraDiagnost 510 (K)
Summary S10(K) Summary
ﺮ
Page 8 of 12
{8}------------------------------------------------
| | External connectivity | | | |
|-------------------|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DICOM | DICOM 3.0 compatible | DICOM 3.0 compatible | DICOM 3.0 compatible | Same. Thus, demonstrating SE. |
| Software platform | Unknown | Eleva workspot<br>(K063781-January 05, 2007) | Eleva workspot | Same. The Philips<br>DuraDiagnost and the currently<br>marketed predicate Philips<br>BuckyVision both use the same<br>software platform. Thus,<br>demonstrating SE. |
| | | Note: The currently<br>marketed and predicate<br>Swissray ddRVersa™<br>Motion (K123005 –<br>December 7, 2012) is only<br>used to demonstrate<br>substantial equivalence to<br>the wireless portable<br>detector (model 3543EZ)<br>of the Philips<br>DuraDiagnost, since it is<br>manufactured by the same<br>manufacturer Trixell. | Note: The currently<br>marketed and predicate<br>Swissray ddRVersa™<br>Motion (K123005 –<br>December 7, 2012) is only<br>used to demonstrate<br>substantial equivalence to<br>the wireless portable<br>detector (model 3543EZ)<br>of the Philips<br>DuraDiagnost, since it is<br>manufactured by the same<br>manufacturer Trixell. | |
hilips DuraDiagnost 510 (K)
.
. ............................................................................................................................................................................
{9}------------------------------------------------
Summary of Nonclinical Performance Data:
Non-clinical performance testing performed on the Philips DuraDiagnost demonstrates compliance with the following International and FDA-recognized consensus standards and FDA Guidance documents:
- AAMI / ANSI ES60601-1: 2005/(R)2012 and C1:2009/(R)2012 . and, a2:2010/(r)2012 (consolidated text) Medical electrical equipment -Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2007, Medical electrical equipment Part 1-2: . General requirements for basic safety and essential performance Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC 60601-1-3 Edition 2.0 2008-01. Medical electrical . equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment.
- . IEC 60601-2-28 Edition 2.0 2010-03, Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis.
- . IEC 60601-2-54 Edition 1.0 2009-06, Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy.
- IEC 62304 First edition 2006-05, Medical device software -. software life cycle processes.
- ISO 14971 Second edition 2007-03-01, Medical devices -. Application of risk management to medical devices.
- IEC 62366 Edition 1.0 2007-10. Medical devices Application . of usability engineering to medical devices.
- CFR 1020.30 Diagnostic x-ray systems and their major . components.
- CFR 1020.31 Radiographic equipment. ●
{10}------------------------------------------------
- Device specific guidance document, entitled "Guidance for the . Submission of 510(k)'s for Solid State X-ray Imaging Devices - August 6, 1999."
- FDA's Guidance document entitled, "Guidance for the . Content of Premarket Submissions for Software Contained in Medical Devices - May 11, 2005."
- FDA guidance document entitled, "Radio Frequency Wireless . Technology in Medical Devices - Guidance for Industry and Food and Drug Administration Staff - August 13, 2013."
- FDA draft guidance document entitled, "Pediatric Information . for X-ray Imaging Device Premarket Notifications issued on May 10, 2012."
Additionally, verification / validation tests have been performed to intended use, the technical claims, requirement address specifications and the risk management results. The test results demonstrate that Philips DuraDiagnost:
- Complies with the aforementioned international and FDA-. recognized consensus standards and FDA guidance documents.
- Meets the acceptance criteria and is adequate for its . intended use.
Therefore, the Philips DuraDiagnost is substantially equivalent to the currently marketed and predicate device, GE Brivo XR385 (K103448, Aug. 12, 2011) Philips BuckyVision (K982795-Nov.24, 1998) and ddRVersa™ Motion (K123005 - December 7, 2005) - Swissray Medical AG in terms of safety and effectiveness.
Clinical study was not warranted to support this 510(k) submission, since substantial equivalence to the currently marketed and predicate devices was demonstrated with the following attributes:
- . Design features;
Clinical Performance
Data:
- . Indication for use;
- Fundamental scientific technology; and .
- Safety and effectiveness. .
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The word is black and stands out against a white background. The font is simple and modern.
Substantial Equivalence Conclusion:
The Philips DuraDiagnost is substantially equivalent to the currently marketed and predicate device GE Brivo XR385 (K103448. August 12, 2011) and Philips BuckyVision (K982795, November 24,1998) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. With regards to the wireless portable detector, the Philips DuraDiagnost is considered substantially equivalent in terms of design and fundamental scientific technology to the currently marketed and predicate ddRVersa™ Motion (K123005 -December 7, 2005) - Swissray Medical AG.
Additionally, substantially equivalence was demonstrated with the following:
- . Non-clinical performance (verification / validation) tests, which complied with the requirements specified in the international and FDA recognized consensus standards;
- Tests identified in the FDA device specific guidance . document entitled, "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices - August 6, 1999." The results of these tests demonstrate that the Philips DuraDiagnost met the acceptance criteria, provides similar diagnostic image quality when compared with the predicate devices and is adequate for its intended use.
- Tests identified in the FDA guidance document entitled, . "Radio Frequency Wireless Technology in Medical Devices - Guidance for Industry and Food and Drug Administration Staff - August 13, 2013." The results of these tests demonstrate that the Philips DuraDiagnost met the acceptance criteria with regards to wireless technology used in the wireless portable Detector and is adequate for its intended use.
{12}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird-like figure with three curved lines representing its wings or body.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 12, 2014
Philips Healthcare (Suzhou) Co., Ltd. % Mr. Gordon Shu Regulatory Affairs Manager No. 258, Zhong Yuan Road Suzhou Industrial Park Suzhou, Jiangsu 215024 CHINA
Re: K141381 Trade/Device Name: Duradiagnost Regulation Number: 21 CFR 892.1680 Regulation Name: DuraDiagnost radiographic system Regulatory Class: II Product Code: KPR, MQB Dated: May 20, 2014 Received: May 28, 2014
Dear Mr. Shu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{13}------------------------------------------------
Page 2—Mr. Gordon Shu
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
ាប់
Sincerely yours.
Smh.7)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{14}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K141381
Device Name
DuraDiagnost
Indications for Use (Describe)
The DuraDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctor or technician. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017 See PRA Statement below.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.