K141851 · Od-Os GmbH · GEX · Jan 2, 2015 · General, Plastic Surgery
Device Facts
Record ID
K141851
Device Name
NAVILAS LASER SYSTEM
Applicant
Od-Os GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Navilas Laser System/Navilas Laser System 532+/ Navilas Laser System 577+ are indicated for use: · In Retinal Photocoagulation for the treatment of Clinically Significant Diabetic Macular Edema (Focal or Grid Laser), Proliferative Diabetic Retinopathy (Panretinal Photocoagulation), Sub-retinal (Choroidal) Neovascularization (Focal Laser), Central and Branch Retinal Vein Occlusion (Scatter Laser Photocoagulation, Focal or Grid Laser), Latice Degeneration, Retinal Tears and Detachments (Laser Retinopexy). · For the imaging (capture, display, storage and manipulation) of the eve, including via color. fluorescein angiography and infrared imaging; and for aiding in the diagnosis and treatment of ocular pathology in the posterior segment of the eye. · In Laser Trabeculoplasty for Primary Open Angle Glaucoma, as well as Iridotomy and Iridoplasty for Closed Angle Glaucoma.
Device Story
Navilas Laser System integrates ophthalmic laser photocoagulation with digital fundus imaging (color, infrared, fluorescein angiography). System captures, displays, stores, and manipulates retinal images for treatment planning. Physician operates system to deliver laser energy via motorized mirrors through ophthalmoscope optics. Laser modes include single shot, repeat, and scanned patterns. Imaging uses slit illumination panned at 25 Hz. Device aids clinicians in visualizing ocular pathology and precisely targeting retinal tissue for photocoagulation or glaucoma procedures. Benefits include improved treatment planning and documentation through integrated imaging and automated laser delivery.
Clinical Evidence
Bench testing only. Performance verification and validation testing, including software and hardware testing and system-level bench testing, confirmed the device meets all performance specifications and requirements.
Technological Characteristics
Laser sources: frequency-doubled Nd:YVO4 (532nm) or optically-pumped semiconductor (532nm or 577nm). Delivery: motorized mirrors through ophthalmoscope optics. Imaging: digital camera, slit illumination (25 Hz). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 62366, IEC 60825-1, ISO 15004-2, ISO 14971, 21 CFR 1040.
Indications for Use
Indicated for patients requiring retinal photocoagulation for diabetic macular edema, proliferative diabetic retinopathy, choroidal neovascularization, retinal vein occlusion, lattice degeneration, retinal tears, or detachments; and for patients requiring laser trabeculoplasty, iridotomy, or iridoplasty for glaucoma. Includes imaging for ocular pathology diagnosis and treatment.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines above and below them. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 2, 2015
OD-OS GMBH c/o Judy F. Gordon ClinReg Consulting Services, Inc. 733 Bolsana Drive Laguna Beach, CA 92651
Re: K141851
Trade/Device Name: Navilas Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: HGF, GEX, HKI, NFF, NFG Dated: November 29, 2014 Received: December 1, 2014
Dear Dr. Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA).
You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Deborah L. Falls -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141851
Device Name Navilas Laser System
### Indications for Use (Describe)
The Navilas Laser System/Navilas Laser System 532+/ Navilas Laser System 577+ are indicated for use:
· In Retinal Photocoagulation for the treatment of Clinically Significant Diabetic Macular Edema (Focal or Grid Laser), Proliferative Diabetic Retinopathy (Panretinal Photocoagulation), Sub-retinal (Choroidal) Neovascularization (Focal Laser), Central and Branch Retinal Vein Occlusion (Scatter Laser Photocoagulation, Focal or Grid Laser), Latice Degeneration, Retinal Tears and Detachments (Laser Retinopexy).
· For the imaging (capture, display, storage and manipulation) of the eve, including via color. fluorescein angiography and infrared imaging; and for aiding in the diagnosis and treatment of ocular pathology in the posterior segment of the eye.
· In Laser Trabeculoplasty for Primary Open Angle Glaucoma, as well as Iridotomy and Iridoplasty for Closed Angle Glaucoma.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 4.0 510(K) SUMMARY
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).
### 4.1 SUBMITTER'S NAME, ADDRESS, TELEPHONE NUMBER, CONTACT PERSON, AND DATE SUMMARY PREPARED
| APPLICANT: | OD-OS GmbH<br>Warthestr. 21<br>14513 Teltow<br>Germany |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Judy Gordon, D.V.M.<br>ClinReg Consulting Services, Inc.<br>733 Bolsana Drive<br>Laguna Beach, CA 92651<br>judy@clinregconsulting.com<br>Tel: (949) 715-0609<br>Fax: (949) 715-0610 |
DATE SUMMARY PREPARED: December 22, 2014
### 4.2 NAME OF DEVICE, INCLUDING TRADE NAME AND CLASSIFICATION NAME
| COMMON NAME: | Photocoagulator with a Digital Camera |
|-------------------------|---------------------------------------------------------------------------|
| DEVICE CLASSIFICATION: | Laser Instrument, Surgical, Powered<br>(Class II, per 21 CFR §878.4810 |
| | Camera, Ophthalmic<br>Class II, per 21 CFR §886.1120 |
| | Device, Storage, Images, Ophthalmic<br>Class I per 21 CFR §892.2010 |
| | Device, Communication, Images, Ophthalmic<br>Class I per 21 CFR §892.2020 |
| PRODUCT CODES: | GEX; HKI, NFF, NFG |
| TRADE/PROPRIETARY NAME: | Navilas® Laser System |
{4}------------------------------------------------
### K141851
### 4.3 PREDICATE DEVICES
| 510(K) # | TRADE NAME | MANUFACTURER |
|----------|-------------------------------------------------------|------------------------|
| K091064 | Navilas Laser System | OD-OS |
| K123542 | PASCAL® Synthesis Ophthalmic<br>Scanning Laser System | Topcon Medical Systems |
| K071687 | Iridex IQ Family | Iridex |
## PREDICATE DEVICES FOR NAVILAS LASER SYSTEM
## PREDICATE DEVICES FOR NAVILAS LASER SYSTEM 532+
| 510(K) # | TRADE NAME | MANUFACTURER |
|----------|-------------------------------------------------------|------------------------|
| K091064 | Navilas Laser System | OD-OS |
| K123542 | PASCAL® Synthesis Ophthalmic<br>Scanning Laser System | Topcon Medical Systems |
| K071687 | Iridex IQ Family | Iridex |
## PREDICATE DEVICES FOR NAVILAS LASER SYSTEM 577+
| 510(K) # | TRADE NAME | MANUFACTURER |
|----------|-------------------------------------------------------|------------------------|
| K091064 | Navilas Laser System | OD-OS |
| K123542 | PASCAL® Synthesis Ophthalmic<br>Scanning Laser System | Topcon Medical Systems |
| K071687 | Iridex IQ Family | Iridex |
| K091581 | Quantel Supra 577 | Quantel |
#### 4.4 DEVICE DESCRIPTION
The NAVILAS Laser System combines imaging technologies (fundus live imaging, infra-red imaging and fluorescein angiography) with established retinal laser photocoagulation treatment methods, providing the doctor with a system for image capture, display, storage and manipulation for treatment planning and documentation.
The primary components of the Navilas Laser System include:
- . One of three optional ophthalmic laser sources:
- a frequency doubled ND:YVO4 laser source that operates at 532nm, or -
- an optically-pumped semiconductor laser source that also operates at the same -532nm, or
- an optically-pumped semiconductor laser source that operates at 577nm -
{5}------------------------------------------------
- An integrated delivery system that directs the laser beam through ophthalmoscope optics using motorized mirrors,
- A digital camera and computer hardware that provides continuous real-time imaging ● using slit illumination that is projected through the ophthalmoscope optics and panned automatically at a rapid rate of 25 Hz across the subject area using the motorized mirrors. Imaging can be in color (using white light illumination) or in monochrome (using infrared illumination or blue light illuminations).
Laser photocoagulation with the NAVILAS can be performed using single shot (Single Spot Mode), repeated shots (Repeat Mode), and scanned patterns (Pattern Mode).
### 4.5 STATEMENT OF INTENDED USE
The Navilas Laser System/Navilas Laser System 532+/ Navilas Laser System 577+ are indicated for use:
- In Retinal Photocoagulation for the treatment of Clinically Significant Diabetic Macular Edema (Focal or Grid Laser), Proliferative Diabetic Retinopathy (Panretinal Photocoagulation), Sub-retinal (Choroidal) Neovascularization (Focal Laser), Central and Branch Retinal Vein Occlusion (Scatter Laser Photocoagulation, Focal or Grid Laser), Lattice Degeneration, Retinal Tears and Detachments (Laser Retinopexy).
- . For the imaging (capture, display, storage and manipulation) of the retina of the eye, including via color, fluorescein angiography and infrared imaging; and for aiding in the diagnosis and treatment of ocular pathology in the posterior segment of the eye.
- . In Laser Trabeculoplasty for Primary Open Angle Glaucoma, as well as Iridotomy and Iridoplasty for Closed Angle Glaucoma.
### TECHNOLOGICAL CHARACTERISTICS COMPARISON 4.6
The Navilas Laser System has the same indications for use, similar design and functional features and is therefore substantially equivalent to the predicate devices.
#### 4.7 PERFORMANCE STANDARDS
The Navilas is designed, manufactured and tested in accordance with both mandatory and voluntary standards including:
- . IEC 60601-1 Medical Electric Equipment, Part 1: General requirements for safety Medical Electrical Equipment - Part 1-2: General requirements ● IEC 60601-1-2 for safety (Collateral standard: electromagnetic compatibility requirements and tests)
{6}------------------------------------------------
| • IEC 60601-2-22 | Medical Electrical Equipment, Part 2: Particular requirements for<br>the safety and diagnostic and therapeutic laser equipment |
|------------------|--------------------------------------------------------------------------------------------------------------------------------|
| • IEC 62366 | Medical Devices – Application of usability engineering to medical<br>devices |
| • IEC 60825-1 | Safety of Laser Products, Part 1: Equipment classification and<br>requirements |
| • ISO 15004-2 | Ophthalmic Instruments – Fundamental requirements and test<br>methods – Part 2: Light hazard protection |
| • ISO 14971 | Application of risk management to medical devices |
This device also complies with European Medical Device Directive 93/42/EEC and U.S. Federal Performance Standards per 21 CFR Part 1040 for light emitting products.
### SUMMARY OF PERFORMANCE TEST RESULTS 4.8
Performance verification and validation testing was completed to demonstrate that the device performance complies with specifications and requirements identified for the Navilas Laser System. This was accomplished by software and hardware verification & validation testing, along with system level bench testing of the Navilas Laser System. All criteria for this testing were met and results demonstrate that the Navilas Laser System meets all performance specifications and requirements.
#### 4.9 CONCLUSIONS
As described in this 510(k) Summary, all testing deemed necessary was conducted on the Navilas Laser System to ensure that the device is safe and effective for its intended use and is substantially equivalent to legally marketed devices intended for laser treatment and imaging of the eye.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.