K141676 · Renovis Surgical Technologies, Inc. · LPH · Nov 4, 2014 · Orthopedic
Device Facts
Record ID
K141676
Device Name
RENOVIS SURGICAL POROUS ACETABULAR CUP SYSTEM
Applicant
Renovis Surgical Technologies, Inc.
Product Code
LPH · Orthopedic
Decision Date
Nov 4, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Renovis Surgical Hip Replacement System is indicated for patients suffering from: 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 2. Rheumatoid arthritis; 3. Correction of functional deformity; 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques; and 5. Revision procedures where other treatment or devices have failed.
Device Story
Renovis Surgical Porous Acetabular Cup System consists of titanium, hemispherical acetabular shells; varying sizes (44-66 mm); options for no-hole or cluster-hole configurations. Used as part of total hip arthroplasty system; implanted by orthopedic surgeons in clinical/surgical settings. Device provides stable acetabular component for hip joint replacement; facilitates bone ingrowth via porous coating. Output is physical structural support for hip joint; aids in restoring joint function and mobility. Benefits include pain relief and improved joint mechanics for patients with degenerative or traumatic hip conditions.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included stereological evaluation of porous coatings (ASTM F1854-09), shear testing (ASTM F1044-05), tension testing (ASTM F1147-05), and disassembly force testing (ASTM F1820-97). Sterilization, shelf life, packaging, and shipping validation were performed per ISO 11137-2, ASTM F1980-07, ISO 11607-2, and ASTM D4169-09.
Technological Characteristics
Hemispherical titanium alloy (ASTM F136-13, F620-11, F67-13) acetabular cups. Features porous coating for bone apposition. Dimensions: 44-66 mm outer diameter. Gamma sterilized. Mechanical fixation via press-fit/screws. Complies with ASTM F1854, F1044, F1147, F1820, and F983.
Indications for Use
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, proximal femur fractures (non-union, femoral neck, trochanteric) unmanageable by other techniques, and revision procedures.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Renovis A400 Surgical Hip Joint Replacement System (acetabular shells) (K112897)
Renovis Tesera Trabecular Technology (T3) Acetabular Shell System (K132312)
Encore Orthopedics Foundation Porous Coated Hemispherical Acetabular Cup (K974093)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure, possibly representing the department's focus on health and human well-being. The figure is composed of three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 4, 2014
Renovis Surgical Technologies, Incorporated % Sharyn Orton, Ph.D. Senior Consultant MEDIcept, Incorporated 200 Homer Avenue Ashland, Massachusetts 01721
Re: K141676
Trade/Device Name: Renovis Surgical Porous Acetabular Cup System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH Dated: September 30, 2014 Received: October 3, 2014
Dear Dr. Orton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K141676
Device Name
Renovis Surgical Porous Acetabular Cup System
Indications for Use (Describe)
The Renovis Surgical Hip Replacement System is indicated for patients suffering from:
1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
2. Rheumatoid arthritis;
3. Correction of functional deformity;
4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques; and
5. Revision procedures where other treatment or devices have failed.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Traditional 510(k) Summary as required by 21 CFR 807.92(a) K141676
- A ) Submitted by: Renovis Surgical Technologies Inc. 1901 W. Lugonia Ave, Ste 340 Redlands, CA 92374 Phone: 909-557-2360 Fax: 909-307-8571
- Official Contact: Anthony DeBenedictis Vice President of Quality Assurance
- Consultant: Sharyn Orton, Ph.D. MEDIcept, Inc. 200 Homer Ave Ashland, MA 01721
- Date prepared: September 25, 2014
- B) Device Name: Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented
- Common Name: Total Hip Arthroplasty - Acetabular Components
- Proprietary Name: Renovis Surgical Porous Acetabular Cup System
- Device Class: Class II
- Regulation number: 21 CFR 888.3358
- Regulation name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis
- Product code: LPH
- Classification panel: Orthopedic
- C) Predicates:
| K112897 | Renovis A400 Surgical Hip Joint Replacement System (acetabular shells) |
|---------|-----------------------------------------------------------------------------|
| K132312 | Renovis Tesera Trabecular Technology (T3) Acetabular Shell System |
| K974093 | Encore Orthopedics Foundation Porous Coated Hemispherical<br>Acetabular Cup |
{4}------------------------------------------------
#### D) Device Description:
The Renovis Porous Acetabular Cup System ("Porous shell") are titanium, hemispherical cups consisting of varying sizes (outer diameter of 44 - 66 mm) and screw hole options (no hole shell and cluster hole shells) to accommodate patient anatomy and/or surgical technique. Instrumentation is provided with the system.
The Renovis Porous Acetabular Cup System is part of the Renovis Surgical Hip Replacement System and is to be used with Renovis Surgical Hip Replacement System acetabular and femoral components.
The Renovis Surgical Porous Cup System complies with the following material standards:
- o ASTM F136-13 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) or
- ASTM F620-11 Standard Specification for Titanium Alloy Forgings for Surgical ● Implants in the Alpha Plus Beta Condition
- ASTM F67-13 Standard Specification for Unalloyed Titanium, for Surgical Implant ● Applications (UNS R50250, UNS R50400, UNS R50550 and UNS R50700)
- ASTM A564-13 Standard Specification for Hot-Rolled and Cold-Finished Age-● Hardening Stainless Steel Bars and Shapes
- E) Indications For Use:
The Renovis Surgical Hip Replacement System is indicated for patients suffering from:
- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- 2. Rheumatoid arthritis;
- 3. Correction of functional deformity;
- 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and
- 5. Revision procedures where other treatment or devices have failed.
- F) Substantial Equivalence Comparison and Discussion
The Renovis Surgical Porous Acetabular Cup System has similar Indications for Use, operating principle and basic design, sizes and geometries, biocompatible materials, and is offered gamma sterilized, like the predicate devices. The manufacturing processes used for the outer coating differ between the Renovis Surgical Porous Acetabular Cup System and two of the predicate devices. The Renovis Surgical Porous Acetabular Cup System meets the coating characterization criteria in 21 CFR 888.3358, and expected performance testing. The Renovis Surgical Porous Acetabular Cup System is substantially equivalent to the predicate devices.
{5}------------------------------------------------
- G) Performance Testing Bench
Performance testing was successfully completed.
The Renovis Surgical Porous Acetabular Cup System complies with:
- 21 CFR 888.3358 ●
- 0 FDA "Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement", April 8, 1994
- ASTM F1854-09 Standard Test Method for Stereological Evaluation of Porous Coatings ● on Medical Implants
- o ASTM F1472-08e1 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium Alloy for Surgical Implant Applications (UNS R56400)
- ASTM F1044-05 (2011) Standard Test Method for Shear Testing of Calcium Phosphate o Coatings and Metallic Coatings
- ASTM F1147-05 (2011) Standard Test Method for Tension Testing of Calcium Phosphate and Metallic Coatings
- ASTM F1820-97:2009 Standard Test Method for Determining the Forces for ● Disassembly of Modular Acetabular Devices
Sterilization, shelf life, packaging and shipping validation studies have been conducted. The Renovis Surgical Porous Acetabular Cup System complies with:
- o ISO 11137-2:2013 Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose
- ASTM F1980-07 (reapproved 2011) Standard Guide for Accelerated Aging of Sterile ● Barrier Systems for Medical Devices
- ISO 11607-2:2009: Packaging for terminally sterilized medical devices Part 2: ● Validation requirements for forming, sealing and assembly processes
- ISTA 2A Partial Simulation Performance Tests; Packaged-Products weighing 150 lb (68 kg) or Less
- ASTM D4169-09: Standard Practice for Performance Testing of Shipping Containers and Systems
## Conclusion
Renovis Surgical Porous Acetabular Cup System is substantially equivalent to the predicate devices, and does not raise new issues of safety or effectiveness.
- H) Other Standards
- o ASTM F983-86 (Reapproved 2013) Standard Practice for Permanent Marking of Orthopaedic Implant Components
- ASTM F565-04 (Reapproved 2013) Standard Practice for Care and Handling of ● Orthopedic Implants and Instruments
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.