RENOVIS TESERA TRABECULAR TECHNOLOGY (T3) ACETABULAR SHELL SYSTEM
Applicant
Renovis Surgical Technologies, LLC
Product Code
LPH · Orthopedic
Decision Date
Apr 11, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Renovis Tesera Trabecular Technology (T) Acetabular Shell System components are to be used with components of the Renovis A400 Surgical Hip Replacement System, and are indicated for patients suffering from the following: 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis: 2. Rheumatoid arthritis: 3. Correction of functional deformity; 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and 5. Revision procedures where other treatment or devices have failed. The Renovis Tesera Trabecular Technology (T) Acetabular Shell System is intended for cementless applications.
Device Story
The Renovis Tesera Trabecular Technology (T3) Acetabular Shell System is a cementless acetabular component for total hip arthroplasty. It is manufactured via additive manufacturing (powder bed fusion) using titanium alloy (Ti-6Al-4V ELI). The device features a porous structure fabricated concurrently with the solid substrate. It is designed for use with polyethylene liners, femoral components, and instruments from the Renovis A400 Surgical Hip Replacement System. The shell is available in multiple configurations (no hole or cluster hole) and sizes (44-66 mm). Liners attach to the shell via a snap-lock mechanism during impaction. The device is intended for use by orthopedic surgeons in a clinical/surgical setting to replace diseased or damaged hip joints, aiming to restore function and alleviate pain in patients with degenerative joint disease, fractures, or failed prior procedures.
Clinical Evidence
Bench testing only. Performance testing included ASTM F1044-05 (shear), ASTM F1160-05 (fatigue), ASTM F1147-05 (tension), ASTM F1854-09 (stereological evaluation of porous coatings), ASTM F1978-12 (abrasion resistance), and ASTM F1820-97 (axial disassembly force). Results demonstrated compliance with 21 CFR 888.3358 requirements for volume porosity, pore size, and coating thickness.
Technological Characteristics
Material: Titanium-6Aluminum-4Vanadium ELI (ASTM F136-12a, ASTM F3001-13). Manufacturing: Additive manufacturing (powder bed fusion). Design: Porous-coated uncemented acetabular shell with solid substrate. Sizes: 44-66 mm diameter. Configurations: No hole or cluster hole. Attachment: Snap-lock mechanism for polyethylene liners. Energy source: N/A (mechanical implant).
Indications for Use
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, non-union, femoral neck/trochanteric fractures of the proximal femur, or revision procedures requiring cementless hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Renovis A400 Surgical Hip Replacement System (K112897)
Exactech Novation Crown Cup with InteGrip Acetabular Shell (K102975)
Submission Summary (Full Text)
{0}------------------------------------------------
### APR 1 1 2014
## ORENOVIS
K132312 Page 1 of 3
#### Traditional 510(k) Summary as required by 21 CFR 807.92(a) K132312
| A ) Submitted by: | Renovis Surgical Technologies Inc.<br>1901 W. Lugonia Ave, Ste 340<br>Redlands, CA 92374<br>Phone: 909-557-2360<br>Fax: 909-307-8571 |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------|
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------|
Official Contact: Anthony DeBenedictis Vice President of Quality Assurance
Sharyn Orton, Ph.D. Consultant: MEDIcept, Inc. 200 Homer Ave Ashland, MA 01721
Date prepared: April 2, 2014
B) Device Name: Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented
Common Name: Total Hip Arthroplasty - Acetabular Components
Renovis Tesera Trabecular Technology (T3) Acetabular Shell System Proprietary Name:
Device Class: Class II
21 CFR 888.3358 Regulation number:
Regulation name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis
Product code: LPH
Classification panel: Orthopedic
- C) Predicates:
Renovis A400 Surgical Hip Replacement System (acetabular shells) K112897 Exactech Novation Crown Cup with InteGrip Acetabular Shell K102975
I
{1}------------------------------------------------
## ORENOVIS
D) Device Description:
The Renovis Tesera Trabecular Technology (T3) Acetabular Shell System ("T" Acetabular Shell System") acetabular shells ("T" shells") are modifications of the Renovis A400 Surgical Hip Replacement System acetabular shells cleared through premarket notification K112897. This submission proposes the following design change: The proposed T shells will be manufactured from the same titanium alloy using an additive manufacturing process to fabricate the implant and porous structure concurrently, and have a thicker solid substrate and porous layer.
The T shells are to be used with polyethylene liners, femoral components, acetabular shell trials, and other system specific instruments originally cleared in the Renovis A400 Surgical Hip System (K112897). The T shell is available in multiple configurations (no hole shell; cluster hole shell) and sizes (44 - 66 mm diameter). The liners attach to the shells under impaction with a snap lock mechanism.
The T3 shells comply with the following material standards:
- ASTM F136-12(a) Standard Specification for Wrought Titanium-6Aluminum-. 4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications
- ASTM F3001-13 Standard Specification for Additive Manufacturing Titanium-6 . Aluminum-4 Vanadium ELI (Extra Low Interstitial) with Powder Bed Fusion
E) Indications For Use:
The Renovis Tesera Trabecular Technology (T) Acetabular Shell System components are to be used with components of the Renovis A400 Surgical Hip Replacement System, and are indicated for patients suffering from the following:
- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis:
- 2. Rheumatoid arthritis:
- 3. Correction of functional deformity;
- 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and
- 5. Revision procedures where other treatment or devices have failed.
The Renovis Tesera Trabecular Technology (T) Acetabular Shell System is intended for cementless applications.
F) Substantial Equivalence Comparison and Discussion
The T Acetabular Shell System has similar Indications for Use, design, material, sizes and geometries, and compatible liners and femoral components as the K112897 A400 acetabular shells. The T Acetabular Shell System additive manufacturing process is similar to that of the K 102975 Exactech acetabular shells. The T shells have a thicker porous structure and solid substrate than the K112897 A400 acetabular shells. Differences between the Renovis T' shells and the predicate devices do not raise any new issues of safety or effectiveness.
{2}------------------------------------------------
# ORENOVIS
#### G) Biocompatibility
The biocompatibility of Ti-6A1-4V EBM manufactured FDA cleared orthopedic implants has been demonstrated.
H) Performance Testing - Bench
The following performance tests were successfully conducted:
- ASTM F1044-05 (2011) Standard Test Method for Shear Testing of Calcium Phosphate . Coatings and Metallic Coatings
- . ASTM F1160-05 (2011) Standard Test Method for Shear and Bending Fatigue Testing of Calcium Phosphate and Metallic Medical and Composite Calcium/Phosphate Metallic Coatings
- ASTM F1147-05 (2011) Standard Test Method for Tension Testing of Calcium . Phosphate and Metallic Coatings
- . ASTM F1854-09 Standard Test Method for Stereological Evaluation of Porous Coatings on Medical Implants The results of ASTM F1854 testing demonstrate compliance with the requirements in 21 CFR 888.3358 for volume porosity, average pore size, and interconnecting porosity and
- porous coating thickness
- ASTM F1978-12 Standard Test Method for Measuring Abrasion Resistance of Metallic . Thermal Spray Coatings by Using the Taber Abraser
- Acetabular shell fatigue testing compared to a predicate device per Renovis protocol and . ASTM F1820-97 Standard Test Methods for Determining the Axial Disassembly Force of a Modular Acetabular Device
Sterilization and aging validation studies were conducted previously.
#### Conclusion
The similarity in Indications for Use, sizes, geometries and technology and the performance testing and biocompatibility data submitted support that the Renovis T Acetabular Shell System is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human services.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-GKA99 Silver Spring MI) 20993-0002
#### April 11, 2014
Renovis Surgical Technologies Incorporated % Dr. Sharyn Orton Senior Consultant MEDIcept. Incorporated 200 Homer Avenue Ashland, Massachusetts 01721
Re: K132312
Trade/Device Name: Renovis Tesera Trabecular Technology (T3) Acetabular Shell System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II
Product Code: LPH Dated: March 12, 2014 Received: March 13, 2014
Dear Dr. Orton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{4}------------------------------------------------
#### Page 2 - Dr. Sharyn Orton
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Lori A. Wiggins
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Form
510(k) Number (if known): K132312
Device Name: Renovis Tesera Trabecular Technologies (T) Acetabular Shell System
Indications for Use:
The Renovis Tesera Trabecular Technologies (T) Acetabular Shell System components are to be used with components of the Renovis A400 Surgical Hip Replacement System, and are indicated for patients suffering from: 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
2. Rheumatoid arthritis;
3. Correction of functional deformity;
4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and 5. Revision procedures where other treatment or devices have failed.
The Renovis Tesera Trabecular Technologies (T) Acetabular Shell System is intended for cementless applications.
Prescription Use AND/OR (21 CFR 801 Subpart D)
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
### Elizabet情診所rank -S
Division of Orthopedic Devices
Renovis T Acetabular Shell System
Traditional 510(k) Premarket Notification 6-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.