K982795 · Philips Medical Systems, Inc. · MQB · Nov 24, 1998 · Radiology
Device Facts
Record ID
K982795
Device Name
PHILIPS BUCKY VISION
Applicant
Philips Medical Systems, Inc.
Product Code
MQB · Radiology
Decision Date
Nov 24, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The Philips BuckyVision is intended for use in general radiographic examinations and applications wherever conventional screen-film systems may be used (excluding fluoroscopy, angiography, and mammography).
Device Story
Philips BuckyVision is a general-purpose digital X-ray system. It utilizes a TRIXELL PIXIUM-4600 Solid State X-ray Imager (SSXI) to convert X-rays directly into electrical signals. These signals undergo processing at a workstation for image management, export, and printing. The system integrates with standard Philips bucky Diagnost components, including X-ray tubes, generators, and collimators. Operated by radiology staff in clinical settings, the device provides digital images on a CRT monitor or laser-imager. This digital workflow replaces traditional screen-film radiography, offering immediate image availability and processing capabilities to support clinical diagnostic decision-making.
Clinical Evidence
Bench testing only.
Technological Characteristics
System utilizes TRIXELL PIXIUM-4600 Solid State X-ray Imager (SSXI) for direct X-ray to electrical signal conversion. Includes X-ray tube, generator, collimator, and workstation. Connectivity provided via workstation for image processing, worklist management, and printing. Operates as a stationary diagnostic X-ray system.
Indications for Use
Indicated for general radiographic examinations in patients requiring diagnostic imaging, excluding fluoroscopy, angiography, and mammography. Intended for use wherever conventional screen-film systems are utilized.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Philips bucky Diagnost System with Radiographic Screen Film Systems (K945287)
Sterling Diagnostic Imaging Direct Radiography™ device (K973206)
{0}------------------------------------------------
NOV 24 1998
Image /page/0/Picture/1 description: The image shows the Philips logo and brand name. The logo is a shield-like shape with a circle inside, containing wavy lines and stars. Next to the logo, the word "PHILIPS" is written in bold, sans-serif font. Below the brand name, there is a handwritten alphanumeric code "K982795".
# Philips Medical Systems
# 510(k) Summary of Safety and Effectiveness
| Company Name: | Philips Medical Systems North America Company |
|----------------------|----------------------------------------------------------------------------------------|
| Address: | 710 Bridegeport Avenue<br>Shelton, CT 06484 |
| Contact Person: | Peter Altman |
| Telephone Number: | 203-926-7031 |
| Prepared (date): | June 15, 1998 |
| Device Name: | Philips BuckyVision |
| Classification Name: | Stationary Diagnostic X-ray System using Digital Solid<br>State X-ray Imaging Detector |
| Common/Usual Name: | General Purpose X-ray System |
| Predicate Device: | Philips bucky Diagnost / Screen Film System |
Philips Medical Systems
710 Bridgestics Archives of Company
Pro Bridges Bronners Arcompany
Pro Box Bell Bromers Court 06484-0917
Fax: (203) 929-6099
{1}------------------------------------------------
### System Description:
The Philips BuckyVision System uses the new x-ray imaging technology ( Solid State X-ray Imager, SSXI ), that converts x-rays directly into electrical signals which can, after appropriate processing, be displayed on a crt-monitor for medical diagnosis or printed out on laser-imager.
The basis of the system is a standard Philips bucky Diagnost System with additional TRIXELL PIXIUM-4600 SSXI connected to a Philips Thoravision-like workstation for worklist-managment, image-processing, image-exporting, and printing. The complete system includes additional subsystems and components, like X-raytube(s), X-ray-generator, collimator, and optionally, a wall bucky with SSXI.
#### Intended Use:
The Philips BuckyVision is intended for use in general radiographic examinations and applications wherever conventional screen-film systems may be used (excluding fluoroscopy, angiography, and mammography).
### Substantial Equivalence Information:
The Philips Bucky Vision is of comparable type and substantially equivalent to:
| the Philips bucky Diagnost System<br>with Radiographic Screen Film Systems, | 510(k) Number : | K945287. |
|-----------------------------------------------------------------------------|-----------------|----------|
| the Sterling Diagnostic Imaging Direct Radiography™ device: | 510(k) Number : | K973206 |
| and the Philips Thoravision | 510(k) Number : | K931071. |
Therefore the Philips Bucky Vision system complies to the same or equivalent standards and has the same intended use as the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three overlapping profiles suggesting a sense of community and support. The profiles are rendered in a simple, flowing line style. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" is arranged in a circular pattern around the central figure, indicating the department's name and national affiliation.
## NOV 2 4 1998
Peter Altman Director of Regulatory Affairs Philips Medical Systems 710 Bridgeport Avenue P. O. Box 860 Shelton, Connecticut 06484-0917 Re:
Food and Drug Administration 9200 Corporate Boulevard K982795 K982795
Philips Bucky Vision Family Comprising: bucky Diagnost TH (+SSXI) with bucky Diagnost CS2: bucky Diagnost TH (+SSXI) with bucky Diagnost CS4: bucky Diagnost VT (+SSXI) with bucky Diagnost CS2: and bucky Diagnost VT (+SSXI) with bucky Diagnost CS4. Dated: November 12, 1998 Received: November 13, 1998 Regulatory class: II 21 CFR 892.1680/Procode: 90 MOB 21 CFR 892,1680/Prodode: 90 KPR
#### Dear Mr. Altman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enacument date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591 for Radiology devices, or 594-4613 for Ear. Nose and Throat devices. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours.
William Yin
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Unknown
Philips BuckyVision Device Name :
Indications For Use :
The Philips BuckyVision is intended for use in general radiographic examinations and applications wherever conventional screen-film systems may be used (excluding fluoroscopy angiography, and mammography).
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDI
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K982795/51
Prescription Use
( Per 21 CFR 801.109
OR
Over-The-Counter Use **__**
(Optional Format 1-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.