K141355 · Besmed Health Business Corporation · BWF · Sep 3, 2014 · Anesthesiology
Device Facts
Record ID
K141355
Device Name
VOLUMETRIC INCENTIVE SPIROMETER
Applicant
Besmed Health Business Corporation
Product Code
BWF · Anesthesiology
Decision Date
Sep 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Besmed Volumetric Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser. Intended for single-patient, multi-use in a hospital or home care setting.
Device Story
Besmed Volumetric Incentive Spirometer is a mechanical, non-sterile, single-patient, multi-use device for hospital or home care. It consists of a housing, flexible tubing, and a mouthpiece. Patient inhales through the mouthpiece, creating a vacuum that raises an internal piston and ball. The piston provides a visual measurement of inspiratory volume (2500 cc or 5000 cc models), while the ball acts as an indicator for maintaining optimal inspiratory flow rate. The device serves as a 'lung exerciser' to provide visual feedback, encouraging patients to improve respiratory capacity through repeated use. The mouthpiece is removable for cleaning, but the main unit is disposable and not intended for cleaning. It functions as a passive mechanical tool to incentivize deep breathing.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included volume accuracy/repeatability, comparative performance against the predicate, environmental stress testing (high/low temperature and humidity), and drop testing. Biocompatibility testing (cytotoxicity, sensitization, intracutaneous/irritation) was performed based on external communicating/surface contact classification.
Technological Characteristics
Mechanical, non-sterile, single-patient, multi-use device. Components: housing, 1 ball/piston, flexible tubing, mouthpiece. Available in 2500 cc and 5000 cc volumes. Sensing principle: vacuum-actuated mechanical piston and ball indicator. No energy source. Materials identical to K133873. Biocompatibility: External communicating (indirect gas pathway) and surface contact (< 24 hours).
Indications for Use
Indicated for patients requiring inspiratory exercise to improve respiratory capacity and function.
Regulatory Classification
Identification
An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three overlapping faces suggesting community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure, emphasizing the department's role and national scope. The logo is simple, using a single color, and is designed to be easily recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 3, 2014
Besmed Health Business Corporation C/O Mr. Paul Dryden President ProMedic, Inc. No. 5. Lane 116, Wu-Kong 2nd Rd. Wu-Ku District, New Taipei City TAIWAN
Re: K141355
Trade/Device Name: Besmed Volumetric Incentive Spirometer Regulation Number: 21 CFR 868.5690 Regulation Name: Spirometer, therapeutic (incentive) Regulatory Class: Class II Product Code: BWF Dated: August 4, 2014 Received: August 5, 2014
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name "Tejashri Purohit-Sheth, M.D." in a serif font. The text is black and centered on a white background. The letters are bolded and slightly spaced apart.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K141355
Device Name
Besmed Volumetric Incentive Spirometer
#### Indications for Use (Describe)
The Besmed Volumetric Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser. Intended for single-patient, multi-use in a hospital or home care setting.
Type of Use (Select one or both, as applicable)
|× Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/2/Picture/15 description: The image shows the logo for the Food and Drug Administration (FDA). The logo is in a light gray color and features the letters "FDA" in a stylized font. The logo is simple and recognizable, and it is often used on products that have been approved by the FDA.
Anya C. Harry -S 2014.08.29 07:11:29 -04'00'
{3}------------------------------------------------
#### 510(k) Summary Page 1 of 4
| Date Prepared: | 20-Aug-2014 |
|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Besmed Health Business Corp.<br>No. 5, Lane 116, Wu-Kong 2nd Rd,<br>Wu-Ku District, New Taipei City, Taiwan | Tel - 011-886-2-2290-3959<br>Fax - 011-886-2-2299-9076 |
| Official Contact: | Winnie Chung, Regulatory Affairs |
| Proprietary or Trade Name: | Volumetric Incentive Spirometer |
| Common/Usual Name: | Incentive spirometer |
| Classification Name: | 21CFR 868.5690, Class 2<br>BWF - spirometer, therapeutic (incentive) |
| Predicate Devices: | K801199 – Hudson RCI Voldyne |
#### Device Description:
The Besmed Volumetric Incentive Spirometer has a flexible tube and mouthpiece which the patient inhales through. This tube connects to the inspiration port of the unit. The chamber has a ball and piston which upon the patient inhaling, creates a vacuum, which causes the ball and piston to rise, the piston reflects the inspiratory volume while the ball is an indicator for the user to maintain the optimum inspiratory flow rate. Two (2) models are offered - 2500 cc and 5000 cc. It is a single patient, multi-use, disposable, non-sterile device. The mouthpiece may be removed and cleaned by the user as needed but the main device is not to be cleaned.
As a patient improves their respiratory capacity or gets stronger they improve their inspiratory volume raising the piston. The principle of the ball indicator is to provide the user a visual indicator of their inspiratory flow rate and help them to improve the respiratory function with repeated uses. Incentive spirometers are commonly referred to as "lung exercisers".
The Besmed Volumetric ICS achieves the therapeutic intent by helping the patient to improve their inspiratory volume. It is a relative improvement device that as described offers the patient the "incentive" to improve. This is the identical therapeutic intent of the predicate volume based incentive spirometer.
#### Indications for Use:
The Besmed Volumetric Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser. Intended for single-patient, multi-use in a hospital or home care setting.
## Patient Population:
Patients requiring inspiratory exercise.
## Environments of use:
Hospital and home care settings
{4}------------------------------------------------
# 510(k) Summary
## Comparison to Predicates
| Attribute | Hudson RCI<br>K801199 | Proposed<br>Besmed Volumetric ICS |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Intended as an inspiratory deep breathing positive exerciser.<br><br>Intended for single-patient, multi-use in a hospital or home care setting. | The Besmed Volumetric Incentive Spirometer is intended as an<br>inspiratory deep breathing positive exerciser.<br><br>Intended for single-patient, multi-use in a hospital or home care setting. |
| Environments of use | Home care settings and hospitals | Home care settings and hospitals |
| Prescriptive | Yes | Yes |
| Patient population | Patients requiring inspiratory exercise | Patients requiring inspiratory exercise |
| Single patient, multi-use | Yes | Yes |
| Patient interface | Mouthpiece (maybe removed and cleaned) | Mouthpiece (maybe removed and cleaned) |
| Basic components | Housing<br>1 ball / piston<br>Tubing<br>Mouthpiece | Housing<br>1 ball / piston<br>Tubing<br>Mouthpiece |
| Volume range | 2500 and 5000 cc | 2500 and 5000 cc |
| Accuracy | Not reported but ~ 21% when comparative testing performed | +/- 15% |
| Performance testing | Volume accuracy<br>2500 cc – 14 – 21%<br>5000 cc – 16 – 21% | Volume accuracy<br>2500 cc – 4 – 15%<br>5000 cc – 3 – 12%<br>Age Testing<br>Pre and post- exposure<br>Environmental Testing<br>High / Low and Humidity conditions<br>Drop test |
{5}------------------------------------------------
## 510(k) Summary Page 3 of 4 20-Aug-2014
#### Substantial Equivalence Discussion
Table 1 compares the key features of the proposed Besmed Volumetric incentive spirometer with the identified predicate and demonstrates that the device can be found to be substantially equivalent.
In summary one can conclude that substantial equivalence is met based upon the following:
#### Indications for Use -
The indications for use are identical for the proposed device when compared to the predicate -K801199 - Hudson RCI Voldyne incentive spirometer.
Discussion - Each device is indicated for use volumetric inspiratory deep breathing positive exerciser.
#### Technology and construction -
The design, components, shape, size, etc., are equivalent to the predicate - K801199 - Hudson RCI Voldyne incentive spirometer.
Discussion - The design is a 1 ball / piston volume based system that upon the patient inhaling raises the piston to indicate the volume of inspired air while there is a ball which indicates the inspiratory flow rate that the patient should try and maintain.
## Environment of Use -
The environments of use are identical to predicate - K801199 - Hudson RCI Voldyne incentive spirometer.
Discussion - The environments of use are identical to the predicate K801199 - Hudson RCI Voldyne incentive spirometer.
## Patient Population -
The patient population of patients requiring inspiratory exercise is equivalent to the predicate -K801199 – Hudson RCI Voldyne incentive spirometer.
Discussion - The patient populations are equivalent to the predicate - K801199 - Hudson RCI Voldyne incentive spirometer.
## Non-Clinical Testing Summary -
#### Materials:
The materials are identical to our reference Besmed TriBall ICS cleared under K133873. We consider the device as External Communicating (Indirect gas pathway) and Surface Contact (direct skin) with limited duration of use (< 24 hours). The following tests were performed.
- Cytotoxicity
- Sensitization ●
- Intracutaneous / Irritation ●
{6}------------------------------------------------
#### 510(k) Summary Page 4 of 4 20-Aug-2014
#### Environmental:
The proposed device was exposed to various environmental conditions of high and low temperatures over time and the performance evaluated and compared before and after these tests to confirm that the proposed device met it performance specifications.
#### Performance Testing including Comparative:
We performed volume accuracy / repeatability and comparative volume accuracy and the results demonstrated equivalent (or better) performance demonstrating the proposed device is equivalent to the - K801199 - Hudson RCI Voldyne incentive spirometer.
Multiple samples of each device were tested multiple times and then evaluated for consistency of performance.
## Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.