K133873 · Besmed Health Business Corp · BWF · Apr 4, 2014 · Anesthesiology
Device Facts
Record ID
K133873
Device Name
TRIBALL INCENTIVE SPIROMETER
Applicant
Besmed Health Business Corp
Product Code
BWF · Anesthesiology
Decision Date
Apr 4, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Besmed Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser. Intended for single-patient, multi-use in a hospital or home care setting.
Device Story
TriBall Incentive Spirometer is a mechanical, non-sterile, single-patient, multi-use device used for respiratory therapy. It consists of a housing, three balls, flexible tubing, and a mouthpiece. The patient inhales through the mouthpiece, creating a vacuum that causes the balls to rise within the chamber. The height of the balls provides a visual indicator of the patient's inspiratory flow rate (600, 900, and 1200 cc/sec). Used in hospitals or home care settings, the device serves as a 'lung exerciser' to help patients improve respiratory capacity through repeated use. The visual feedback acts as an incentive for the patient to increase their inspiratory effort, aiding in recovery or strengthening of respiratory function.
Clinical Evidence
Bench testing only. Comparative performance testing evaluated inspiratory rate/volume accuracy (600, 900, 1200 cc/sec) against the predicate, demonstrating accuracy within +/- 5%. Biocompatibility testing (ISO 10993) for cytotoxicity, sensitization, and intracutaneous irritation was performed. Environmental testing included high/low temperature, humidity, and drop testing to confirm performance specifications.
Technological Characteristics
Mechanical 3-ball system; flow-based measurement; includes housing, balls, tubing, mouthpiece, and gross particulate filter. Materials tested per ISO 10993 for external communicating (indirect gas pathway) and surface contact. No energy source required. Single-patient, multi-use, disposable.
Indications for Use
Indicated for patients requiring inspiratory exercise in hospital or home care settings.
Regulatory Classification
Identification
An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.
{0}------------------------------------------------
# APR 0 4 2014
## 510(k) Summary Page 1 of 4
| Date Prepared: | 03-Apr-2014 | |
|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|--|
| Besmed Health Business Corp.<br>No. 5, Lane 116, Wu-Kong 2nd Rd,<br>Wu-Ku District, New Taipei City, Taiwan | Tel - 011-886-2-2290-3959<br>Fax - 011-886-2-2299-9076 | |
| Official Contact: | Winnie Chung<br>Regulatory Affairs Associate | |
| Proprietary or Trade Name: | TriBall Incentive Spirometer | |
| Common/Usual Name: | Incentive spirometer | |
| Classification Name: | 21CFR 868.5690<br>BWF - spirometer, therapeutic (incentive)<br>Class II | |
| Predicate Devices: | K781831 - Hudson RCI TriFlo | |
## Device Description:
The Besmed TriBall Incentive Spirometer has a flexible tube and mouthpiece which the patient inhales through. This tube connects to the inspiration port of the unit. The chamber has a series of balls which upon the patient inhaling, creates a vacuum, which causes the balls to rise, they reflect the inspiratory flow rate in cc/sec (600, 900 and 1200).
It is a single patient, multi-use, disposable, non-sterile device.
As a patient improves their respiratory capacity or gets stronger their inspiratory flow rate raises the balls. The principle is to provide the user a visual indicator of their inspiratory flow rate and help them to improve the respiratory function with repeated uses. Incentive spirometers are commonly referred to as "lung exercisers".
The TriBall achieves the therapeutic intent by helping the patient to improve their inspiratory flow rate. It is a relative improvement device that as described offers the "incentive" to improve. This is the identical therapeutic intent of all predicate flow / volume based incentive spirometers.
#### Indications for Use:
The Besmed Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser.
Intended for single-patient, multi-use in a hospital or home care setting.
### Patient Population:
Patients requiring inspiratory exercise.
## Environments of use:
Hospital and home care settings
{1}------------------------------------------------
510(k) Summary
Comparison to Predicate
| Attribute | Hudson RCI<br>K781831 | Proposed<br>Besmed TriBall |
|---------------------------|------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Intended as an inspiratory deep breathing positive exerciser. | The Besmed Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser. |
| Environments of use | Intended for single-patient, multi-use in a hospital or home care setting.<br>Home care settings and hospitals | Intended for single-patient, multi-use in a hospital or home care setting.<br>Home care settings and hospitals |
| Prescriptive | Yes | Yes |
| Patient population | Patients requiring inspiratory exercise | Patients requiring inspiratory exercise |
| Single patient, multi-use | Yes | Yes |
| Patient interface | Mouthpiece | Mouthpiece |
| Basic components | Housing<br>3 balls<br>Tubing<br>Mouthpiece<br>Gross particulate filter | Housing<br>3 balls<br>Tubing<br>Mouthpiece<br>Gross particulate filter |
| Flow / Volume range | 600 / 900 / 1200 cc/sec | 600 / 900 / 1200 cc/sec |
| Performance testing | Inspiratory Rate / Volume accuracy<br>600 - 16.7%<br>900 - 1.6%<br>1200 - 11.3%<br>Accuracy specification - not stated | Inspiratory Rate / Volume accuracy<br>600 - 0.5%<br>900 - 0.3%<br>1200 - 0.1%<br>Accuracy specification - +/- 5%<br>Age Testing<br>Pre and post- exposure<br>Environmental Testing<br>High/Low and Humidity conditions<br>Drop test |
{2}------------------------------------------------
## 510(k) Summary Page 3 of 4 03-Apr-2014
#### Substantial Equivalence Discussion
Table 1 compares the key features of the proposed Besmed TriBall incentive spirometer with the identified predicate and demonstrates that the device can be found to be substantially equivalent.
In summary one can conclude that substantial equivalence is met based upon the following:
#### Indications for Use -
The indications for use are identical for the proposed device when compared to the predicate -K781831 - Hudson RCI incentive spirometer.
Discussion - Each device is indicated for use volumetric inspiratory deep breathing positive exerciser.
#### Technology and construction -
The design, components, shape, size, etc. are equivalent to the predicate - K 781831 - Hudson RCI incentive spirometer.
Discussion - The design is a 3 ball system that upon the patient inhaling raises the balls to indicate the volume of inspired air.
#### Environment of Use -
The environments of use are identical to predicate - K781831 - Hudson RCI incentive spirometer.
Discussion - The environments of use are identical to the predicate K781831 - Hudson RCI incentive spirometer.
#### Patient Population -
The patient population of patients requiring inspiratory exercise is equivalent to the predicate -K781831 - Hudson RCI incentive spirometer.
Discussion - The patient populations are equivalent to the predicate - K781831 - Hudson RCI incentive spirometer.
#### Non-Clinical Testing Summary -
#### Materials:
We have performed ISO 10993 testing on the component materials of the TriBall Incentive spirometer which is considered as External Communicating (Indirect gas pathway) and Surface Contact (direct skin) with the patient which means the following tests were performed.
- Cytotoxicity .
- Sensitization .
- Intracutaneous / Irritation .
{3}------------------------------------------------
## 510(k) Summary Page 4 of 4 03-Apr-2014
## Environmental:
The proposed device was exposed to various environmental conditions of high and low temperatures over time and the performance evaluated and compared before and after these tests to confirm that the proposed device met it performance specifications.
## Performance Testing including Comparative:
We performed comparative volume accuracy to the predicate and the results demonstrated equivalent (or better) performance demonstrating the proposed device is equivalent to the -K781831 - Hudson RCI incentive spirometer.
Multiple samples of each device were tested multiple times and then evaluated for consistency of performance.
## Substantial Equivalence Conclusion -
The proposed device has been found to be substantially equivalent to the predicate. Differences between the proposed device and the predicate do not raise new questions of safety or efficacy.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to promote health, well-being, and human services. The eagle is positioned to the right of the department's name, which is written in a circular pattern around the logo.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 4, 2014
Besmed Health Business Corp C/O Mr. Paul Dryden, President ProMedic, Inc. 24301 Woodsage Drive Bonita Springs, FL 34134
Re: K133873
Trade/Device Name: TriBall Incentive Spirometer Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive spirometer Class: II · Product Code: BWF Dated: March 5, 2014 Received: March 6, 2014
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract tiability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Ms. Chung
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note
the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Clinical Deputy Director
DAGRID
FOR
FOR
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known)
K33873
Device Name Besmed TriBall Incentive Spirometer
Indications for Use (Describe)
The Besmed Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser.
Intended for single-patient, multi-use in a hospital or home care setting.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY | |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | |
| K133873 | Anya C. Harry -S |
| | Digitally signed by Anya C. Harry -S<br>DN: c=US, o=U.S. Government,<br>ou=HHS, ou=FDA, ou=People,<br>cn=Anya C. Harry -S,<br>0.9.2342.19200300.100.1.1=001131559<br>Date: 2014.04.03 11:45:44 -04'00' |
FORM FDA 3881 (9/13) PDF Page 41 of 266 Page 4.2 EF
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.