K141331 · Biomet, Inc. · JDI · Oct 27, 2014 · Orthopedic
Device Facts
Record ID
K141331
Device Name
BIOMET ORTHOPAEDIC SALVAGE SYSTEM (OSS)
Applicant
Biomet, Inc.
Product Code
JDI · Orthopedic
Decision Date
Oct 27, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Proximal and Distal Femoral Reconstruction, Tibial Reconstruction.
Device Story
The Biomet Orthopaedic Salvage System (OSS) is a modular orthopedic implant system for limb salvage reconstruction. It consists of various components including distal/proximal femoral components, proximal tibial sleeves, diaphyseal segments, splined stems, and augments. These components are used by orthopedic surgeons in clinical settings to replace bone and joint structures compromised by tumor, trauma, or failed previous arthroplasty. The system is typically used with bone cement, though certain OsseoTi (titanium alloy) components allow for uncemented bone integration. The device provides structural support and joint function, aiming to restore mobility and alleviate pain in patients with significant bone loss or deformity. The system is modular, allowing surgeons to assemble specific configurations based on the patient's anatomical needs. The output is a reconstructed skeletal segment that restores joint mechanics and limb stability.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence was determined based on non-clinical bench testing, including device fatigue testing, static disassociation testing, tensile properties testing, and material characterization for OsseoTi.
Technological Characteristics
Modular orthopedic implant system. Materials: CoCrMo (for strength) and OsseoTi (porous titanium alloy for bone integration). Features: External taper posts, augment tapers, and suture holes for secondary tissue fixation. Connectivity: None. Energy: None. Sterilization: Not specified.
Indications for Use
Indicated for patients with painful/disabled joints due to avascular necrosis, osteoarthritis, rheumatoid arthritis, or traumatic arthritis; correction of varus, valgus, or posttraumatic deformity; correction/revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement; ligament deficiencies; tumor resections; treatment of non-unions, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement; revision of previously failed total joint arthroplasty; and trauma. Reduced size components indicated for small adults and adolescents (12-21 years) requiring proximal femoral, distal femoral, total femur, or proximal tibial replacement, or resurfacing of proximal tibia/distal femur.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Zimmer Segmental System Trabecular Metal Proximal Tibial Component, Trabecular Metal Proximal Femoral Component, and Additional Segment with Male/Female Taper Components (K110940)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus-like symbol with three human profiles facing right, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2014
Biomet, Incorporated Tracy Bickel Johnson Regulatory Global Project Manager 56 East Bell Drive Post Office Box 587 Warsaw, Indiana 46581
Re: K141331
Trade/Device Name: Biomet Orthopaedic Salvage System (OSS) Line Extension Regulation Number: 21 CFR 888.3350 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JDI, LPH, KRO Dated: September 24, 2014 Received: September 26, 2014
Dear Ms. Tracy Bickel Johnson,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141331
#### Device Name
Biomet Orthopedic Salvage System (OSS) Line Extension
#### Indications for Use (Describe)
#### OSS INDICATIONS
- 1. Painful and disabled joint resulting from avascular necrosis, osteoarthritis, or traumatic arthritis.
- 2. Correction of varus, valgus, or posttraumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
- 4. Ligament deficiencies.
- 5. Tumor resections.
- 6. Treatment of non-unions, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. *
- 7. Revision of previously failed total joint arthroplasty.
#### 8. Trauma.
These devices are to be used with bone cement unless composed of OsseoTi (titanium allov, not licensed in Canada) or a proximal femur is indicated for use (USA).
Biomet OSS Reduced size (RS) components offers a variety of component options for treatment in small adults and adolescents (12-21 years) that require proximal femoral, total femur, or proximal tibial replacement as well as, resurfacing components for the proximal tibia and distal femur (USA).
* Not applicable to Regenerex Ultra Porous Construct titanium knee augment usage (not licensed in Canada), or any other knee component.
#### COMPRESS INDICATIONS
The Compress Segmental Femoral Replacement System is indicated for:
- 1. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
- 2. Tumor resections.
- 3. Revision of previously failed total joint arthroplasty.
- 4. Trauma.
The Compress Segmental Femoral Replacement System components are intended for uncemented use.
When components of the Orthopaedic Salvage System are used with Biomet's Compress Segmental Femoral Replacement System, the user should refer to the package insert contained with the Compress components for full prescription information.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <svg class="bi bi-check-square-fill" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm10.03 4.97a.75.75 0 0 0-1.08.022L7.477 9.417 5.384 7.323a.75.75 0 0 0-1.06 1.06L6.97 11.03a.75.75 0 0 0 1.079-.02l3.992-4.99a.75.75 0 0 0-.01-1.05z" fill-rule="evenodd"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;"> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "BIOMET" in a stylized, sans-serif font. The letters are all capitalized and appear to be outlined or embossed, giving them a three-dimensional effect. A small registered trademark symbol is visible to the upper right of the letter "T". The overall design is clean and modern.
# 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
| Submitter Information | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Biomet Manufacturing, LLC |
| Address | 56 East Bell Drive |
| | PO Box 587 |
| | Warsaw, IN 46581-0857 |
| Phone number | (574) 372-1761 |
| Fax number | (574) 372-1683 |
| Establishment<br>Registration Number | 1825034 |
| Name of contact person | Tracy Bickel Johnson, RAC |
| Date prepared | 22 August 2014 |
| Name of device | |
| Trade or proprietary<br>name | Biomet Orthopaedic Salvage System (OSS) |
| Common or usual<br>name | Knee/Hip Implants |
| Classification name /<br>Regulation | Prosthesis, hip, semi-constrained, metal/polymer, porous<br>uncemented (21 CFR § 888.3350); |
| | Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented (21<br>CFR § 888.3350); |
| | Prosthesis, Knee, Femorotibial, Constrained, Cemented,<br>Metal/Polymer (21 CFR § 888.3510) |
| Classification panel | Orthopedic |
| Product Code(s) | LPH; JDI; KRO |
| Legally marketed device(s)<br>to which equivalence is | Biomet Orthopaedic Salvage System (K002757, K052685,<br>K123501); |
| claimed | Biomet RS (Reduced Size) OSS Additional Components<br>(K051479/K021260); |
| | Zimmer Segmental System Trabecular Metal Proximal Tibial<br>Component, Trabecular Metal Proximal Femoral Component, and<br>Additional Segment with Male/Female Taper Components<br>(K110940) |
| | Reference Item: Biomet Reconstructive Wedges (K122770)<br>Reference Item: Regenerex Sleeve Augments (K072336) |
| Reason for 510(k)<br>submission | New device |
| Device description | The new devices included in this submission are additional<br>components to Biomet's Orthopaedic Salvage System (OSS) that<br>offer surgeons additional prostheses options to be used in limb<br>salvage reconstruction. The new devices include distal femoral<br>component, proximal femoral component, proximal tibial sleeve,<br>diaphyseal segment, splined stem, diaphyseal augment and<br>metaphyseal augment. |
| Intended use of the device | Proximal and Distal Femoral Reconstruction, Tibial<br>Reconstruction. |
| Indications for use | OSS INDICATIONS<br>1. Painful and disabled joint resulting from avascular necrosis,<br>osteoarthritis, rheumatoid arthritis, or traumatic arthritis.<br>2. Correction of varus, valgus, or posttraumatic deformity.<br>3. Correction or revision of unsuccessful osteotomy, arthrodesis,<br>or previous joint replacement.<br>4. Ligament deficiencies.<br>5. Tumor resections.<br>6. Treatment of non-unions, femoral neck fracture, and<br>trochanteric fractures of the proximal femur with head<br>involvement, unmanageable using other techniques. *<br>7. Revision of previously failed total joint arthroplasty.<br>8. Trauma.<br><br>These devices are to be used with bone cement unless composed<br>of OsseoTi (titanium alloy, not licensed in Canada) or a proximal<br>femur is indicated for use (USA).<br><br>Biomet OSS Reduced size (RS) components offers a variety of<br>component options for treatment in small adults and adolescents<br>(12-21 years) that require proximal femoral, distal femoral, total<br>femur, or proximal tibial replacement as well as, resurfacing<br>components for the proximal tibia and distal femur (USA).<br><br>* Not applicable to Regenerex Ultra Porous Construct titanium<br>knee augment usage (not licensed in Canada), or any other knee<br>component.<br><br>COMPRESS INDICATIONS<br>The Compress Segmental Femoral Replacement System is<br>indicated for:<br>1. Correction of revision of unsuccessful osteotomy,<br>arthrodesis, or previous joint replacement.<br>2. Tumor resections.<br>3. Revision of previously failed total joint arthroplasty.<br>4. Trauma. |
| Summary of the Technologies | The Compress Segmental Femoral Replacement System components are intended for uncemented use.<br><br>When components of the Orthopaedic Salvage System are used with Biomet's Compress Segmental Femoral Replacement System, the user should refer to the package insert contained with the Compress components for full prescription information. |
Mailing Address:
P.O. Box 587
Warsaw, IN 46581-0587 Toll Free: 800.348.9500
Office: 574.267.6639
Main Fax: 574.267.8137 www.biomet.com
Shipping Address:
56 East Bell Drive
Warsaw, IN 46582
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "BIOMET" in a stylized, sans-serif font. The letters are bold and appear to be outlined, giving them a three-dimensional effect. A registered trademark symbol is present to the right of the word.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Biomet. The logo is in a sans-serif font and is all uppercase. The letters are all connected and have a thick, blocky appearance. There is a registered trademark symbol in the upper right corner of the logo.
The new devices are modifications to the predicate devices (K002757, K052685,
K051479/K021260 and K110940) in the following ways:
- Distal femoral components (with optional augments) are modified with aesthetic profile update and an external taper post feature;
- Splined stems are modified with material change to CoCrMo;
- . Diaphyseal segments are modified with an external augment taper feature and material change to CoCrMo;
- . Proximal tibial sleeves are modified with material change to OsseoTi® and additional suture holes for optional tissue attachment method utilizing optional spiked washers and bolts;
- . Metaphyseal augments are modified with augment taper feature and material change to OsseoTi":
- . Diaphyseal augments are modified with material change to OsseoTi™ and geometry updates to be compatible with OSS devices
The external taper post feature and external augment taper feature are incorporated so that new devices are compatible with other OSS components (K002757, K052685, and K051479). Material change to CoCrMo for its material strength, and to OsseoTi™ to enhance the potential for bone integration. The OsseoTi™ material and process for the devices included in this submission is identical as which cleared in Biomet Reconstructive Wedges K122770. Cleaning of OsseoTi is identical to Regenerex materials which were previously cleared in K072336.
New proximal sleeves are updated with additional suture holes for optional use with spiked washers and bolts as optional secondary tissue fixation method. Primary tissue fixation method remains the same as predicate device.
# PERFORMANCE DATA
# SUMMARY OF NON-CLINICAL TESTS
## Performance Test Summary-New Device
Device Fatique Testing Device Static Disassociation Testing Tensile Properties Testing Material Characterization for OsseoTi Cleaning
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "BIOMET" in a stylized, sans-serif font. The letters are thick and blocky, with a slightly rounded appearance. A registered trademark symbol is present to the right of the letter "T". The word is in a gray color, and the background is white.
## SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
Clinical Performance Data/Information: N/A
## CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
No clinical testing was necessary for a determination of substantial equivalence.
The results of device fatigue testing, static disassociation testing and material characterization studies indicated the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.