K123501 · Biomet, Inc. · JDI · Feb 27, 2013 · Orthopedic
Device Facts
Record ID
K123501
Device Name
OSS MODULAR STEMS, OSS DIAPHYSEAL SEGMENTS
Applicant
Biomet, Inc.
Product Code
JDI · Orthopedic
Decision Date
Feb 27, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Orthopedic Salvage System components are intended for cemented use only unless used in a proximal femoral replacement application. Biomet's Segmental Femoral Replacement components are intended for use in total knee, total hip, or total femoral replacement procedures. Specific indications are: 1. Painful and disabled joint resulting from avascular necrosis, osteoarthritis, rheumatoid arthritis, or traumatic arthritis. 2. Correction of varus, valgus or post-traumatic deformity. 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. 4. Ligament deficiencies. 5. Tumor resections. 6. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques. 7. Revision of previously failed total joint arthroplasty. 8. Trauma. These devices are to be used with bone cement unless a proximal femur is indicated for use (USA). When components of the Orthopedic Salvage System are used with Biomet's Compress Segmental Femoral Replacement System, they are intended for uncemented application and indicated for: 1. Correction or revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement. 2. Tumor resections. 3. Revision of previously failed total joint arthroplasty. 4. Trauma.
Device Story
Orthopedic Salvage System (OSS) is a modular component set for complex revision surgeries and limb salvage. Components replace portions of or total femur, knee, and proximal tibia. Modification involves transitioning male tapers from machined to roller-hardened and machined surfaces. Used in clinical settings by orthopedic surgeons. Device provides structural replacement for diseased or damaged bone/joint tissue; improves patient mobility and function. Bench testing confirms roller-hardened tapers exhibit superior or equivalent fatigue strength and pull-off resistance compared to predicate.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including cantilever fatigue, compressive fatigue, and static axial separation testing.
Technological Characteristics
Modular orthopedic prosthesis components. Modification: roller-hardened and machined male tapers. Materials: metal/polymer. Energy source: N/A (mechanical). Connectivity: N/A. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring total knee, hip, or femoral replacement due to avascular necrosis, osteoarthritis, rheumatoid arthritis, traumatic arthritis, deformity (varus/valgus/post-traumatic), failed previous procedures (osteotomy/arthrodesis/arthroplasty), ligament deficiency, tumor resection, or trauma. Includes proximal femoral fractures unmanageable by other techniques.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for BIOMET MANUFACTURING CORP. The word "BIOMET" is in a stylized font with a thick outline. Below the logo, the words "MANUFACTURING CORP." are printed in a simple, sans-serif font. The text is black and the background is white.
K12350 (pg 1/3)
# FEB 2 7 2013
#### 510(k) Summary
Preparation Date:
23 October, 2012
Applicant/Sponsor: Biomet Manufacturing Corp. 56 East Bell Drive Warsaw, IN 46582 FDA Registration #: 1825034
Contact Person:
Gary Baker Sr. Regulatory Specialist 56 East Bell Drive Warsaw, IN 46582 574-267-6639 Ext. 1568 574-371-1027 gary.baker@biomet.com
Proprietary Name:
· Orthopedic Salvage System (OSS)
Constrained knee Common Name:
Classification Code(s)/Name(s): JDI, KRO
888.3350 (JDI) - Hip joint metal/polymer semi-constrained cemented prosthesis.
888.3510 (KRO) - Knee joint femorotibial metal/polymer constrained cemented prosthesis
# Legally Marketed Devices To Which Substantial Equivalence Is Claimed:
K002757 - Oncology Salvage System (Biomet)
### Device Description:
The Orthopedic Salvage System is a system of modular components that are intended to be used for difficult revision surgeries or limb salvage procedures. The components can replace any portion of, or the total femur and/or the knee and proximal tibia.
# Intended Use:
The Orthopedic Salvage System components are intended for cemented use only unless used in a proximal femoral replacement application.
Mailing Address:
P.O. Box 587
Warsaw, IN 46581-0587
Toll Free: 800.348.9500
Office: 574.267.6639
Main Fax: 574.267.8137
www.biomet.com
{1}------------------------------------------------
# Indications For Use:
Biomet's Segmental Femoral Replacement components are intended for use in total knee, total hip, or total femoral replacement procedures. Specific indications are:
- 1. Painful and disabled joint resulting from avascular necrosis, osteoarthritis, rheumatoid arthritis, or traumatic arthritis.
- 2. Correction of varus, valgus or post-traumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
- 4. Ligament deficiencies.
- 5. Tumor resections.
- 6. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques. *
- 7. Revision of previously failed total joint arthroplasty.
- 8. Trauma.
These devices are to be used with bone cement unless a proximal femur is indicated for use (USA).
*Not applicable to Regenerex Ultra Porous Construct titanium knee augment usage (not licensed in Canada), or any other knee component.
When components of the Orthopedic Salvage System are used with Biomet's Compress Segmental Femoral Replacement System, they are intended for uncemented application and indicated for:
- 1. Correction or revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
- 2. Tumor resections.
- 3. Revision of previously failed total joint arthroplasty.
- 4. Trauma.
# Summarv of Technologies:
The modifications proposed within this submission include expanding the product line and modifying the male tapers from a machined taper to a roller hardened and machined taper.
# Non-Clinical Testing:
Cantilever Fatigue testing was conducted to evaluate the taper cantilever fatigue strength and compare the results with the cantilever fatigue strength of the predicate tapers. Results determined that the roller hardened tapers were stronger in cantilever fatigue than the predicate tapers. The acceptance criteria were met.
Compressive Fatigue testing was conducted to evaluate the taper compressive fatigue strength and compare the results with the compressive fatigue strength of the predicate tapers. Results determined that the compressive fatigue strength of the roller hardened tapers was equivalent to or greater than the compressive fatigue strength of the predicate tapers. The acceptance criteria were met.
{2}------------------------------------------------
K123501 (pg 3/3)
Static Axial Separation testing was conducted to evaluate the pull-off strength of the taper junction and compare the results with the pull-off strength of the taper junction of the predicate tapers. Results determined that the pull-off strength of the roller hardened tapers was equivalent to or greater than the pull-off strength of the predicate tapers. The acceptance criteria were met.
# Clinical Testing:
Clinical testing was not required to demonstrate substantial equivalence of the design modifications proposed to the cleared predicate devices.
All trademarks are property of Biomet, Inc.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular fashion around the symbol. The text is in all caps and is in a sans-serif font. The logo is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 27, 2013
Biomet Manufacturing Corp. % Mr. Gary Baker Senior Regulatory Specialist 56 East Bell Drive Warsaw, Indiana 46582
Re: K123501
Trade/Device Name: Orthopedic Salvage System Regulation Number: 21 CFR 888.3350 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JDI, KRO Dated: January 11, 2013 Received: January 28, 2013
Dear Mr. Baker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
# Page 2 - Mr. Gary Baker
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
ErinI.D.Keith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
510(k) Number (if known): _______K123501
Device Name: Orthopedic Salvage System
#### Indications for Use:
Biomet's Segmental Femoral Replacement components are intended for use in total knee, total hip, or total femoral replacement procedures. Specific indications are:
- 1. Painful and disabled joint resulting from avascular necrosis, osteoarthritis, rheumatoid arthritis, or traumatic arthritis.
- 2. Correction of varus, valgus or post-traumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
- 4. Ligament deficiencies.
- 5. Tumor resections.
- 6. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femirr with head involvement, unmanageable by other techniques. *
- 7. Revision of previously failed total joint arthroplasty.
- 8. Trauma.
These devices are to be used with bone cement unless a proximal femur is indicated for use (USA).
*Not applicable to Regenerex Ultra Porous Construct titanium knee augment usage (not licensed in Canada), or any other knee component.
When components of the Orthopedie Salvage System are used with Biomet's Compress Segmental Femoral Replacement System, there intended for uncemented application and indicated for:
1. Correction or revision of unsucessful osteotomy, arthrodesis, or previous joint replacement.
- 2. Tumor resections.
- 3. Revision of previously fuiled total joint arthroplasty.
- 4. Traumu
Prescription Use _ YES __ AND/OR Over-The-Counter Use NO NO ____(Part (21 CFR 807 Subpart C) 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabethi i Mrank -S
Division of Orthopedic Devices
10
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.