JUGGERLOC BONE TO BONE IMPLANTS/ BONE INSTRUMENTS/ BONE IMPLANT AND INSTRUMENT KITS

K141219 · Biomet Manufacturing Corp · HTN · Jul 18, 2014 · Orthopedic

Device Facts

Record IDK141219
Device NameJUGGERLOC BONE TO BONE IMPLANTS/ BONE INSTRUMENTS/ BONE IMPLANT AND INSTRUMENT KITS
ApplicantBiomet Manufacturing Corp
Product CodeHTN · Orthopedic
Decision DateJul 18, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Biomet JuggerLoc™ Bone system is intended for repair in the foot and ankle including indications for: Midfoot repair including but not limited to Lisfranc repair, ankle syndesmosis fixation (syndesmosis disruptions), and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures.

Device Story

JuggerLoc Bone to Bone System is a suture-based construct for bone fracture stabilization. Device utilizes a JuggerKnot soft anchor combined with a metal button and ZipLoop technology. System is used by orthopedic surgeons in clinical settings to repair foot and ankle injuries, including Lisfranc repairs and syndesmosis disruptions. By providing a secure fixation construct, the device assists in stabilizing bone segments during the healing process. It serves as an adjunct to traditional trauma hardware.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics.

Technological Characteristics

Suture-based bone fixation construct. Components include a soft anchor and a metal button. Utilizes ZipLoop technology. Materials and design are consistent with predicate orthopedic fixation devices. Sterilization and specific material standards are not detailed in the provided text.

Indications for Use

Indicated for patients requiring foot and ankle repair, specifically midfoot repair (e.g., Lisfranc), ankle syndesmosis fixation, and as an adjunct for Weber B and C ankle fracture trauma hardware.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JUL 1 8 2014 ### Section 5 - 510(k) Summary A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92 | Submitter Information | | |------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name | Biomet Manufacturing Corp. | | Address | 56 East Bell Drive<br>P.O. Box 587<br>Warsaw, IN 46581 | | Phone number | (574) 372-1941 | | Fax number | (574) 371-1027 | | Establishment | 1825034 | | Registration Number | | | Name of contact person | Jared Cooper<br>Regulatory Affairs Specialist | | Date prepared | 29 April 2014 | | Name of device | | | Trade or proprietary name | JuggerLoc Bone to Bone System | | Common or usual name | Bone Fixation Anchor | | Classification Name | Single/multiple component metallic bone fixation appliances and accessories. | | Classification panel | Orthopedics | | Regulation | 21 CFR § 888.3030 | | Product Code(s) | HTN - washer, bolt nut | | | | | Legally marketed device(s) to which equivalence is claimed | K130033 - Biomet ToggleLoc System<br>K110145 - Biomet JuggerKnot Soft Anchors<br>K133275 - Arthrex Mini-Tightrope | | Reason for 510(k) submission | New Device Construct | | Device description | The new JuggerLoc Bone to Bone system is a suture based construct intended to stabilize fractures. The construct consists of a JuggerKnot Soft Anchor with ZipLoop Technology, featuring combinations of suture and metal button fixation. | | Indications for use | The JuggerLoc Bone to Bone System is intended for repair in the foot and ankle including indications for:<br>Midfoot repair including but not limited to Lisfranc repair, ankle syndesmosis fixation (syndesmosis disruptions), and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures. | | Summary of the Technologies | | The technological characteristics of the JuggerLoc Bone to Bone System are similar to the predicate devices including design, dimensions, function, and materials. The JuggerLoc design differs from the predicate devices by having one soft anchor and one metal button to achieve fixation rather than two metal buttons. {1}------------------------------------------------ #### PERFORMANCE DATA ## SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION Performance Test Summary - New Device The JuggerLoc Bone to Bone System components have the same basic technological and material characteristics as the predicate devices except for slight modifications to the general design as described in this 510(k) notification. Preclinical performance tests are provided to address the subject construct's strength. A biocompatibility rationale was also provided. Results indicate that the subject construct is substantially equivalent to legally marketed devices. #### SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION Clinical Performance Data/Information: None provided as a basis for substantial equivalence. #### CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA The results of preclinical tests and the similarities with legally marketed devices indicate the JuggerLoc Bone to Bone System is substantially equivalent to currently marketed devices. No new concerns of safety or efficacy have been raised. {2}------------------------------------------------ Public Health Service Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 18, 2014 Biomet Manufacturing Corp. Dr. Jared Cooper Regulatory Affairs Specialist 56 East Bell Drive. PO Box 587 Warsaw, Indiana 46581 Re: K141219 Trade/Device Name: JuggerLoc TM Bone to Bone System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HTN Dated: May 23, 2014 Received: May 27, 2014 Dear Dr. Cooper: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {3}------------------------------------------------ Page 2 - Dr. Jared Cooper (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.hun for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Lori A. Wiaqins for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Section 4 - Indications for Use Statement K141219 510(k) Number: JuggerLoc™ Bone to Bone System Device Name: Indications For Use: The Biomet JuggerLoc™ Bone system is intended for repair in the foot and ankle including indications for: Midfoot repair including but not limited to Lisfranc repair, ankle syndesmosis fixation (syndesmosis disruptions), and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures. × Prescription Use_ (Per 21 CFR 801 Subpart D) AND/OR Over-the-Counter_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Elizabeth 臨歐ank -S Division of Orthopedic Devices
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...