BIOMET SPORTS MEDICINE JUGGERKNOT SOFT ANCHOR

K110145 · Biomet, Inc. · MBI · Mar 4, 2011 · Orthopedic

Device Facts

Record IDK110145
Device NameBIOMET SPORTS MEDICINE JUGGERKNOT SOFT ANCHOR
ApplicantBiomet, Inc.
Product CodeMBI · Orthopedic
Decision DateMar 4, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The JuggerKnot™ Soft Anchor is intended to be used for soft tissue to bone fixation with indications for use in: Shoulder Bankart lesion repair, SLAP lesion repair, Acromio-clavicular repair, Capsular shift/capsulolabral reconstruction, Deltoid repair, Rotator cuff tear repair, Biceps tenodesis Foot and Ankle Medial/lateral repair and reconstruction, Mid- and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair Elbow Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment Knee Extra-capsular repair: MCL, LCL, and posterior oblique ligament Iliotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure Hand and Wrist Collateral ligament repair, Scapholunate ligament reconstruction, Tendon transfers in phalanx, Volar plate reconstruction Hip Acetabular labral repair

Device Story

JuggerKnot™ Soft Anchors consist of coreless sleeve structure and suture; designed for soft tissue-to-bone fixation. Deployed by bunching against bone. Used by surgeons in clinical settings for orthopedic repairs including shoulder, foot/ankle, elbow, knee, hand/wrist, and hip. Device provides mechanical fixation strength for soft tissue attachment; benefits patient by enabling secure tissue-to-bone reattachment during surgical procedures.

Clinical Evidence

Bench testing only. Mechanical pullout testing performed to compare fixation strength of JuggerKnot™ Soft Anchors against predicate device. Results indicate equivalent fixation strength.

Technological Characteristics

Coreless sleeve structure with suture. Materials, design, and sizing are similar or identical to predicate devices. Non-degradable soft tissue fixation fastener (21 CFR 888.3040).

Indications for Use

Indicated for soft tissue to bone fixation in shoulder, foot, ankle, elbow, knee, hand, wrist, and hip procedures, including ligament, tendon, and capsule repairs/reconstructions.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows handwritten text that appears to be a combination of letters and numbers. The text starts with "K110145" on the first line, followed by "p1/2" on the second line. The handwriting is somewhat uneven, and the characters are spaced irregularly. Image /page/0/Picture/1 description: The image shows the logo for Biomet Sports Medicine. The logo is in black and white and features the word "BIOMET" in a stylized font. Below the word "BIOMET" is the phrase "SPORTS MEDICINE" in a smaller, sans-serif font. The logo is simple and professional, and it is likely used to represent the company's brand. MAR - 4 2011 # 510(k) Summary | Preparation Date: | January 17, 2011 | |----------------------|---------------------------------------------------------------------------------------------------------------------| | Applicant/Sponsor: | Biomet Sports Medicine | | Contact Person: | Elizabeth Wray / Regulatory Project Manager<br>Victor Rodgers / Director of Quality, Clinical, & Regulatory Affairs | | Proprietary Name: | JuggerKnot™ Soft Anchors | | Common Name: | Soft Tissue Fixation Device | | Classification Name: | Fastener, fixation, nondegradable, soft tissue<br>(21CFR §888.3040) MBI | # Legally Marketed Devices To Which Substantial Equivalence Is Claimed: K071704 Sleeve and Button Soft Tissue Fixation Devices / Biomet Sports Medicine ## Device Description: The JuggerKnot™ Soft Anchors consist of a coreless sleeve structure and suture. The anchors are intended for use in soft tissue fixation by bunching against bone when deployed. ## Intended Use / Indications for Use: The JuggerKnot™ Soft Anchor is intended to be used for soft tissue to bone fixation with indications for use in: ## Shoulder Bankart lesion repair, SLAP lesion repair, Acromio-clavicular repair, Capsular shift/capsulolabral reconstruction, Deltoid repair, Rotator cuff tear repair, Biceps tenodesis ## Foot and Ankle Medial/lateral repair and reconstruction, Mid- and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair ## Elbow Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment # Knee Extra-capsular repair: MCL, LCL, and posterior oblique ligament Iliotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure Mailing Address: P.O. Box 587 Warsaw, IN 46581-0587 Toll Free: 800.348.9500 Office: 574.267.6639 Main Fax: 574.267.8137 www.biomet.com Shipping Address: 56 East Bell Drive Warsaw, IN 46582 {1}------------------------------------------------ K110145 pH2 # Hand and Wrist Collateral ligament repair, Scapholunate ligament reconstruction, Tendon transfers in phalanx, Volar plate reconstruction # Hip Acetabular labral repair # Summary of Technologies: The technological characteristics (materials, design, sizing and indications) of the JuggerKnot™ Soft Anchors are similar or identical to the predicate devices or other previously cleared devices. # Non-Clinical Testing: Non-clinical laboratory testing was performed to verify the fixation strength of the JuggerKnot™ Soft Anchors in mechanical pullout testing as compared to the predicate devices for specific indications for use. The efficacy of the JuggerKnot™ Soft Anchors was compared to that of the Biomet Sports Medicine Sleeve and Button Soft Tissue Fixation Devices. The test results indicate that the Biomet Sports Medicine JuggerKnot™ Soft Anchors provide equivalent fixation strength to the predicate devices and would be functional within their intended use. # Clinical Testing: None provided as a basis for substantial equivalence. All trademarks are the property of Biomet, Inc. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that form the wings and body of the bird. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Biomet, Inc. % Ms. Elizabeth Wray Regulatory Project Manager 56 East Bell Drive - P.O. Box 587 Warsaw, Indiana 46581-0587 MAR - 4 2011 Re: K110145 Trade/Device Name: JuggerKnot" Soft Anchors Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: February 11, 2011 Received: February 14, 2011 Dear Ms. Wray: We have reviewed your Section 510(k) premarket notification of-intent to-market the device referenced above ând have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ Page 2 - Ms. Elizabeth Wray comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance, You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. /www.fda.gov/MedicalDevices/ResourcesforYouIndustry/ucm.htm Sincerely yours, Aliz B. nute for Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use KI10145 510(k) Number (if known): Device Name: JuggerKnot™ Soft Anchors Indications For Use: The JuggerKnot™ Soft Anchors are intended for soft tissue to bone fixation for the following indications: ## Shoulder Bankart lesion repair, SLAP lesion repair, Acromio-clavicular repair, Capsular shift/capsulolabral reconstruction, Deltoid repair, Rotator cuff tear repair, Biceps tenodesis ## Foot and Ankle Medial/lateral repair and reconstruction, Mid- and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair #### Elbow Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment #### Knee Extra-capsular repair: MCL, LCL, and posterior oblique ligament Iliotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure ## Hand and Wrist Collateral ligament repair, Scapholunate ligament reconstruction, Tendon transfers in phalanx, Volar plate reconstruction ## Hip Acetabular labral repair Prescription Use_ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ NO (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) for M. Melkerson Page 1 of 1 (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices K110145 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...