K141122 · Ethicon Endo-Surgery, LLC · GEI · Sep 15, 2014 · General, Plastic Surgery
Device Facts
Record ID
K141122
Device Name
HARMONIC SCALLOP BLADE, GENERATOR G11
Applicant
Ethicon Endo-Surgery, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Harmonic Scallop Blade is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, gynecologic, exposure to orthopedic structures (such as spine and joint space), ENT (Ears, Nose, Throat), and thoracic surgery, including mobilization of the Internal Mammary Artery (IMA). The Generator G11 provides radiofrequency power to drive EnSeal electrosurgical instruments that are used during open or laparoscopic general and gynecological surgery to cut and seal vessels and to cut, grasp, and dissect tissues. In addition, the generator provides power to drive Harmonic ultrasonic surgical instruments that are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. EnSeal and Harmonic instruments when used with the Generator G11 have not been shown to be effective for sterilization procedures or tubal coagulation. Do not use these instruments for these procedures.
Device Story
System comprises Generator G11, footswitch, handpiece, and Harmonic Scallop Blade; titanium blade with non-removable gray sheath. Generator supplies ultrasonic energy to blade; converts electrical energy to mechanical vibration for tissue cutting and coagulation. Used in OR by surgeons for soft tissue incision and vessel sealing. Blade features wide concave design for scraping soft tissue from bone; integrated torque wrench for assembly. Generator G11 software (v2013_1) updated to support 60W maximum power output for Scallop Blade. Surgeon operates via footswitch/handpiece; visual feedback via touchscreen. Provides precise cutting with minimal thermal injury; reduces bleeding; serves as alternative to traditional electrosurgery or steel scalpels.
Clinical Evidence
No clinical data. Evidence based on ex-vivo and in-vivo bench and preclinical testing. Studies included acute and 30-day survival models assessing spot coagulation, general procedure capability, hemostasis, tissue healing response, and blade strength. Software verification and validation performed per FDA guidance.
Technological Characteristics
Titanium blade with gray sheath; ultrasonic energy source; 60W max power output. Generator G11 features touchscreen interface and dual-port connectivity for EnSeal/Harmonic instruments. Software version 2013_1. Sterile, single-patient use.
Indications for Use
Indicated for patients undergoing general, plastic, gynecologic, orthopedic (spine/joint space), ENT, or thoracic surgery requiring soft tissue incision, vessel coagulation (up to 2mm), or tissue dissection. Contraindicated for sterilization procedures or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Medtronic Integrated Power Console (IPS®) System (K081475)
Codman® Cobb Spinal Elevator (Class I exempt)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 15, 2014
Ethincon Endo-Surgery, LLC Ms. Emily Kruetzkamp Regulatory Affairs Associate 475 Calle C Guaynabo, Puerto Rico 00969
Re: K141122
Trade/Device Name: Harmonic® Scallop Blade and Generator G11 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, HGI, LFL Dated: August 19, 2014 Received: August 20, 2014
Dear Ms. Kruetzkamp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K141122
Device Name Harmonic Scallop Blade
#### Indications for Use (Describe)
The Harmonic Scallop Blade is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, gynecologic, exposure to orthopedic structures (such as spine and joint space), ENT (Ears, Nose, Throat), and thoracic surgery, including mobilization of the Internal Mammary Artery (IMA).
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141122
Device Name Generator G11
#### Indications for Use (Describe)
The Generator G11 provides radiofrequency power to drive EnSeal electrosurgical instruments that are used during open or laparoscopic general and gynecological surgery to cut and seal vessels and to cut, grasp, and dissect tissues. In addition, the generator provides power to drive Harmonic ultrasonic surgical instruments that are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. EnSeal and Harmonic instruments when used with the Generator G11 have not been shown to be effective for sterilization procedures or tubal coagulation. Do not use these instruments for these procedures.
#### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# Company
Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969
# Contact
Emily Kruetzkamp, Regulatory Affairs Associate Ethicon Endo-Surgery, Inc. Telephone: (513) 337-1546 Fax: (513) 337-2802 Email: ekruetzk@its.jnj.com
# Date Prepared: April 25, 2014
| Trade Name: | Harmonic® Scallop Blade |
|-----------------------------|-----------------------------------------------------------|
| Common Name: | Instrument, ultrasonic surgical |
| Classification Name: | Instrument, ultrasonic surgical |
| Device Class: | Unclassified |
| Classification Regulation: | Unclassified |
| Panel: | 79, General and Plastic Surgery |
| Classification Code: | LFL |
| Trade Name: | Generator G11 |
| Common Name: | Electrosurgical & Ultrasonic Surgical Generator |
| Classification Names: | Electrosurgical, Cutting & Coagulation & Accessories; |
| | Electrocautery, Gynecologic (and Accessories); |
| | Instrument, Ultrasonic Surgical |
| Device Class: | Class II |
| Classification Regulations: | 21 CFR 878.4400, 21 CFR 884.4120, |
| | and Unclassified (LFL) |
| Panel: | 79, General and Plastic Surgery |
| Classification Codes: | GEI, HGI, LFL |
| Predicate Devices | Harmonic® 10cm Combination Hook Blade, K072203 |
| | Generator G11, K101990 |
| | Medtronic Integrated Power Console (IPS®) System, K081475 |
| | Codman® Cobb Spinal Elevator, Class I exempt per |
| | CFR 878.4800 |
## Device Description:
The Harmonic® Scallop Blade is a sterile, single patient use device, consisting of a titanium blade with a non-removable gray sheath. The Harmonic® Scallop Blade instrument allows for the cutting of soft tissue and coagulation of vessels up to and including 2 mm in diameter. A soft grip pad on the handle housing facilitates grasping. The instrument is equipped with an integrated internal torque wrench, which is used for assembly to the Harmonic® Hand Piece.
{5}------------------------------------------------
The device system has four essential parts: the generator, the Footswitch, the Hand Piece, and the instrument. The Generator G11 supplies energy to Enseal® electrosurgical instruments and Harmonic® ultrasonic surgical instruments. The generator utilizes a touchscreen display interface and has a receptacle port that accepts either Enseal® or Harmonic® devices. Connectors (one for Harmonic® and one for Enseal® instruments) are used to enable the generator to power currently cleared surgical instruments. The Generator G11 hardware has not changed from the Predicate. The Harmonic® Scallop Blade instrument is designed for use exclusively with the Generator G11 software version 2013 1 or later, which has been updated to increase the maximum Power output to 60W required to power the Harmonic Scallop Blade.
# Indications for Use:
## Harmonic Scallop Blade
The Harmonic Scallop Blade is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, gynecologic, exposure to orthopedic structures (such as spine and joint space), ENT (Ears, Nose, Throat), and thoracic surgery, including mobilization of the Internal Mammary Artery (IMA).
## Generator G11
The Generator G11 provides radiofrequency power to drive EnSeal electrosurgical instruments that are used during open or laparoscopic general and gynecological surgery to cut and seal vessels and to cut, grasp, and dissect tissues. In addition, the generator provides power to drive Harmonic ultrasonic surgical instruments that are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. EnSeal and Harmonic instruments when used with the Generator G11 have not been shown to be effective for sterilization procedures or tubal coagulation. Do not use these instruments for these procedures.
## Technological Characteristics:
The Harmonic® Scallop Blade is similar in technological characteristics to the Predicate Device, the Harmonic® Combination Hook Blade HK105. The Harmonic Scallop Blade end effector design is different in that it incorporates a wide concave blade design.
The Subject Device Harmonic® Scallop Blade is similar to the Predicate Device Cobb Spinal Elevator in its ability to scrape soft tissue from orthopedic structures. The Subject Device Harmonic Scallop Blade is similar to the Predicate Device Medtronic IPC® System with Midas Rex Legend instruments in that the devices have the same durability when touching orthopedic structures in removing soft tissue from bone.
The Generator G11 supplies power to both the Enseal® electrosurgical and Harmonic® ultrasonic surgical devices. The Subject Device Generator G11 software version 2013_1 enables the maximum ultrasonic power output to 60W.
## Performance Data:
Ex-vivo and in-vivo tests were performed to verify that the performance of the Harmonic Scallop Blade instrument meets the definition of substantial equivalence to the Predicate Devices. Harmonic® Blade HK105, Codman Cobb, and Medtronic Integrated Power Console (IPC®) System, Midas Rex® Legend® instruments. Device performance was assessed against design requirements, including bench and preclinical testing. Applicable software verification and validation testing was completed per FDA Guidance for industry and staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "General Principles of Software Validation; Final Guidance for Industry and FDA Staff." Bench testing includes device durability, acoustics, and reliability testing. Preclinical studies include acute and 30-day survival studies, and demonstrate device performance on spot coagulation, general
{6}------------------------------------------------
procedure capability, hemostasis following muscle transection, tissue effects and healing response, and blade strength to access orthopedic procedures. These testing criteria demonstrate that the Harmonic® Scallop Blade performs as intended and is substantially equivalent to the Predicate Devices.
This submission does not include data from Clinical Studies.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.