ETHICON ENDO-SURGERY GENERATOR G11, MODEL GEN11

K101990 · Ethicon Endo-Sugery, LLC · GEI · Jun 13, 2011 · General, Plastic Surgery

Device Facts

Record IDK101990
Device NameETHICON ENDO-SURGERY GENERATOR G11, MODEL GEN11
ApplicantEthicon Endo-Sugery, LLC
Product CodeGEI · General, Plastic Surgery
Decision DateJun 13, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Generator G11 provides radiofrequency power to drive EnSeal® electrosurgical instruments that are used during open or laparoscopic general and gynecological surgery to cut and seal vessels and to cut, grasp, and dissect tissue. In addition the generator provides power to drive Harmonic ultrasonic surgical instruments that are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. EnSeal® and Harmonic® instruments when used with the Generator G11 have not been shown to be effective for sterilization procedures or tubal coagulation. Do not use these instruments for these procedures.

Device Story

Generator G11 serves as a dual-modality power source for surgical instruments; provides radiofrequency (RF) energy for EnSeal electrosurgical instruments and ultrasonic energy for Harmonic surgical instruments. Used in open or laparoscopic general and gynecological surgery; operated by surgeons. Device features a touchscreen interface for user control, information, and feedback. Accepts both EnSeal and Harmonic instruments via dedicated receptacle ports. By consolidating power delivery for both modalities into one unit, the generator enables surgeons to perform vessel sealing, tissue dissection, and soft tissue incisions with bleeding control. Clinical benefit includes streamlined surgical workflow and precise energy delivery for tissue management.

Clinical Evidence

Bench testing and laboratory evaluations in an animal model were conducted to demonstrate that the Generator G11 performs as intended.

Technological Characteristics

Dual-modality surgical generator providing RF and ultrasonic energy. Features a touchscreen user interface for control and feedback. Supports a wider range of power, voltage, current, and frequency compared to predicate devices. Connectivity via specific receptacle ports for EnSeal and Harmonic instruments.

Indications for Use

Indicated for patients undergoing open or laparoscopic general and gynecological surgery requiring vessel sealing, tissue cutting, grasping, dissection, or soft tissue incisions where bleeding control and minimal thermal injury are desired. Contraindicated for sterilization procedures or tubal coagulation.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary # K101990 - Company Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969 - Donovan May Contact Regulatory Affairs Associate II Ethicon Endo-Surgery, Inc. Telephone: (513) 337-7981 Fax: (513) 337-2793 Email: dmay3@its.jnj.com July 14, 2010 Date Prepared Device Name Trade Name: Ethicon Endo-Surgery Generator G11 Common Name: Electrosurgical & Ultrasonic Surgical Generator # Classification Names - · Electrosurgical, Cutting & Coagulation & Accessories (21 CFR 878.4400, Product Code GEI) - · Electrocautery, Gynecologic and Accessories (21 CFR 884.4120, Product Code HGI) - · Instrument, Ultrasonic Surgical (Unassigned, Product Code LFL) ## Predicate Devices - · EnSeal® RF60 Electrosurgical Generator (cleared under K072177 on September 5, 2007 as part of the EnSeal® Vessel Sealing and Hemostasis System) - · EnSeal® Universal Generator (cleared under K081129 on July 28, 2008 as a system with the EnSeal PowerTip instrument) - · UltraCision Harmonic® Scalpel Generator 300 (cleared under K002906 on December 15, - 2000 as part of the UltraCision Harmonic® Scalpel Generator 300 System) #### Device Description: The Ethicon Endo-Surgery Generator G11 supplies energy to EnSeal® electrosurgical instruments and Harmonic ultrasonic surgical instruments. The generator utilizes a touchscreen display interface and has a receptacle port that accepts either EnSeal® or Harmonic® devices. Connectors (one for Harmonic and one for EnSeal® instruments) are used to enable the generator to power currently cleared surgical instruments. {1}------------------------------------------------ ## Indications for Use: The Generator G11 provides radiofrequency power to drive EnSeal® electrosurgical instruments that are used during open or laparoscopic general and gynecological surgery to cut and seal vessels and to cut, grasp, and dissect tissue. In addition the generator provides power to drive Harmonic ultrasonic surgical instruments that are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. EnSeal® and Harmonic® instruments when used with the Generator G11 have not been shown to be effective for sterilization procedures or tubal coagulation. Do not use these instruments for these procedures. The differences between the intended use of the subject device and those of the predicate devices do not affect the safety and effectiveness of the subject device when used as labeled for the following reasons: - · The Generator G11 provides power for the use of both EnSeal electrosurgical and Harmonic ultrasonic surgical open and laparoscopic surgical instruments. The predicate devices referenced are only able to provide power for one of these surgical instrument : types. - · The predicate devices were each submitted as part of a surgical system thus the Indications for Use statements included in the submissions associated with them did not contain indications specific to the function of the generators except for those in the labeling. These differences do not alter the intended surgical effects of the subject device from that of the predicate devices. Technological Characteristics: The Ethicon Endo-Surgery Generator G11 incorporates the technological characteristics of the predicate devices in a single surgical generator but allows for a greater range of power, voltage, current, and frequency for both the EnSeal electrosurgical and Harmonic ultrasonic surgical modes. A touchscreen user interface has been incorporated to provide the user with information, control, and feedback during use. Performance Data: Bench testing and laboratory evaluations in an animal model were conducted to demonstrate that the Generator G11 performs as intended. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ethicon Endo-Surgery, LLC % Mr. Donovan May Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, OH 45242 JUN 1 3 2011 Re: K101990 Trade Name: Generator G11 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, LFL, HGI Dated: June 1, 2011 Received: June 2, 2011 Dear Mr. May: We have reviewed your Section 510(k) premarket notification of intent to market the device we nevers forrewed your bother is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreativent date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may are see of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny 1 with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Donovan May forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Euis D. Keith for Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use 510(k) Number (if known): K101990 Device Name: Generator G11 Indications for Use: The Generator GI I provides radiofrequency power to drive EnSeal® electrosurgical instruments that are used during open or laparoscopic general and gynecological surgery to cut and seal vessels and to cut, grasp, and dissect tissue. In addition the generator provides power to drive Harmonic ultrasonic surgical instruments that are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. EnSeal® and Harmonic® instruments when used with the Generator G11 have not been shown to be effective for sterilization procedures or tubal coagulation. Do not use these instruments for these procedures. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) *(Division Sign-Off)* Division of Surgical, Orthopedic, and Restorative Devices Page 1 of 1 KIO 990 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...