OSTEOMED LOW PROFILE ORBITAL FLOOR PLATE SYSTEM

K140978 · Osteomed · JEY · Dec 18, 2014 · Dental

Device Facts

Record IDK140978
Device NameOSTEOMED LOW PROFILE ORBITAL FLOOR PLATE SYSTEM
ApplicantOsteomed
Product CodeJEY · Dental
Decision DateDec 18, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4760
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The OsteoMed Low Profile Plate System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit.

Device Story

Permanent implant system for orbital floor and medial wall reconstruction. Comprised of commercially pure titanium (ASTM F 67) plates in various shapes (preformed or non-preformed) and thicknesses (0.2mm–0.5mm). Fixed to infraorbital rim via screw holes using 1.2mm/1.6mm standard/auto-drive screws or 1.9mm safety screws. Used by surgeons in clinical settings to restore orbital anatomy following trauma or bone excision. System utilizes existing CFX system instrumentation (stainless steel, anodized aluminum, medical-grade plastic). Provided non-sterile; requires sterilization before use.

Clinical Evidence

Bench testing only. Mechanical testing demonstrated the system meets the performance of the predicate Synthes MatrixMIDFACE System.

Technological Characteristics

Material: Commercially pure Titanium (ASTM F 67). Fixation: Bone screws (1.2mm, 1.6mm, 1.9mm). Plate thickness: 0.2mm–0.5mm. Form factor: Preformed and non-preformed plates. Sterilization: Non-sterile, requires user sterilization. No software or electronic components.

Indications for Use

Indicated for reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in adult patients.

Regulatory Classification

Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping to create a sense of depth and connection. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 18,2014 OsteoMed c/o Mrs. Piedad Pena, M.S. Manager, Regulatory Affairs 3885 Arapaho Road Addison, Texas 75001 Re: K140978 Trade/Device Name: OsteoMed Low Profile Orbital Plate System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: November 17, 2014 Received: November 18, 2014 Dear Mrs. Piedad Pena: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {1}------------------------------------------------ comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Susan Runno DDS, MA Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Statement K140978 510(k) Number (if known): OsteoMed Low Profile Orbital Floor Plate System Device Name: Indications for Use: The OsteoMed Low Profile Plate System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit. The OsteoMed Low Profile Plate System is indicated for the reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in adult patients. Prescription Use: _X (Part 21 CFR 801 Subpart D) And/Or Over-The-Counter (Part 21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Osteomed. The logo is in blue and white, and it is set against a white background. The word "OSTEOMED" is written in large, bold letters. The logo is simple and modern. ## 510(k) SUMMARY ## OsteoMed Low Profile Orbital Floor Plate System Summary Date: November 17, 2014 Submitter Data: OsteoMed 3885 Arapaho Road Addison, TX 75001 972 677 4600 972 677 4601 (fax) Primary Contact: Mrs. Piedad Pena, M.S. Device Trade Name: OsteoMed Low Profile Orbital Plate System Common Name: Orbital Floor Plate System Classification Name: Plate, bone Product Code: JEY- Bone Plate Device Classification: 21 CFR Part §872.4760 - Bone plate ## Legally Marketed Predicate Devices: Synthes MatrixMIDFACE System (K031761) Stryker Universal Orbital Floor Plate System (K123786) ## Device Description: The OsteoMED Low Profile Orbital Floor Plate System is a permanent implant that is provided non sterile and is manufactured from commercially pure Titanium (ASTM F 67). The OSTEOMED Low Profile Orbital Floor Plates are offered with hole sizes which are able to accept 1.2mm and 1.6mm standard and auto drive bone screws and 1.9mm and 1.9mm safety screws and are offered in thicknesses ranging from .2mm to .5mm. The implant is also provided in a variety of shapes including both nonpreformed shapes and pre formed left and right profiles. The implant is fixed to the infra orbital rim via either the holes on the provided fixation arms or through the anterior screw holes on the plate itself to reconstruct the orbital floor and/or medial wall in adult patients. The OsteoMed CFX system which is already available in the marketplace includes the screws (ASTM F136) and instrumentation to facilitate implantation of the Low Profile Orbital Floor Plate System. The instrumentation is made from various grades of stainless steel, anodized aluminum, and/or medical grade Image /page/3/Picture/18 description: The image shows a logo with the word "Colson" in a stylized font. To the left of the word is a circular emblem with a vertical line extending upwards. To the right of the word "Colson" are the words "A COLSON A". {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Osteomed. The logo is in a stylized font, with the letters in a dark blue color. The logo is surrounded by a light blue swoosh shape. There is a registered trademark symbol in the upper right corner of the logo. plastic. All instruments included in the CFX system are Class I, standard, manual, surgical instruments. ## Intended Use/ Indications for Use: The OsteoMed Low Profile Plate System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit. The OsteoMed Low Profile Plate System is indicated for the reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in adult patients. ### Performance Testing: Mechanical testing was performed demonstrating the new OsteoMed Low Profile Plate System meets that of the predicate device, the Synthes MatrixMIDFACE System. Performance equivalence was shown through the verification comparison to the predicate Synthes MatrixMIDFACE System (K031761). ### Technological Characteristics: The OsteoMed Orbital Floor Plate System is considered to be substantially equivalent in design, intended use and material to the predicate devices. There are not any design differences between the systems that would affect safety or effectiveness or raise new questions regarding safety and effectiveness. ## Sterilization: The OsteoMed Low Profile Orbital Floor Plate System components are single use, supplied non sterile and require sterilization prior to use. #### Substantial Equivalence: The basis of substantial equivalence for this device is based on similarities in intended use, indications for use, material, function, technology, performance, and operational principles to the predicate device; Synthes MatrixMIDFACE System which was cleared under K031761 on September 3, 2003 and the Stryker Universal Orbital Floor Plate System which was cleared under K123786 on April 9, 2013. Image /page/4/Picture/15 description: The image shows the Colson logo. The logo consists of the word "Colson" in a stylized font, with the "C" being larger than the other letters. The word is enclosed in a black circle. To the right of the logo is the letter "A". {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Osteomed. The logo is in blue and purple, and the word "OSTEOMED" is in bold, sans-serif letters. The logo is surrounded by an oval shape. The top half of the oval is blue, and the bottom half is purple. # Device Comparison Chart: | Features | OsteoMed Low Profile<br>Orbital Floor Plate<br>System (NEW) | Synthes MatrixMIDFACE<br>System (K031761) | Stryker Universal<br>Orbital Floor System<br>(K123786) | |------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use /<br>Indications for<br>Use | Intended Use:<br>The OsteoMed Low Profile<br>Orbital Floor Plate System is<br>intended to be used in the<br>reconstruction of the floor<br>and/or medial wall of the<br>orbit.<br><br>Indications for Use:<br>The OsteoMed Low Profile<br>Orbital Floor Plate System is<br>indicated for the<br>reconstructive treatment of<br>the orbital floor and/or<br>medial wall trauma or bone<br>excision in adult patients. | Intended<br>Use/Indications for Use:<br>The Synthes Craniofacial<br>Plate and Screw System is<br>intended for use in selective<br>trauma of the midface and<br>craniofacial skeleton,<br>craniofacial surgery,<br>reconstructive surgery of<br>the maxilla and chin.<br><br>Indications: Synthes<br>MatrixMIDFACE Preformed<br>Orbital Plates are intended<br>for trauma repair and<br>reconstruction of the<br>craniofacial skeleton.<br>-Orbital floor fractures<br>-Medial orbital wall<br>fractures<br>-Combined orbital floor and<br>medial wall fractures. | Intended Use:<br>The Stryker Universal<br>Orbital Floor System is<br>intended to be used in<br>the reconstruction of the<br>floor and/or medial wall<br>of the orbit.<br><br>Indications for Use:<br>The Stryker Universal<br>Orbital Floor System is<br>indicated for the<br>reconstructive treatment<br>of the orbital floor<br>and/or medial wall<br>trauma or bone excision<br>in adult patients. | | Material &<br>Biocompatibility | Plates are manufactured<br>from CP (Pure) Titanium (as<br>per ASTM F67). | Plates are manufactured<br>from CP (Pure) Titanium. | Plates are manufactured<br>from CP (Pure) Titanium,<br>Grade 2. | | Method of<br>Fixation | Plate fixation with screws<br>inserted through dedicated<br>screw holes. | Plate fixation with screws<br>inserted through dedicated<br>screw holes. | Plate fixation with screws<br>inserted through<br>dedicated screw holes. | | Sterility | Provided as single use, non<br>sterile | Non sterile/sterile | Provided non sterile | | Profile<br>(Thickness)<br>Range | .2mm-.5mm | .2mm-.5mm | .3mm-.6mm | | Screw Diameters<br>Accepted | 1.2mm and 1.6mm screws<br>1.5mm and 1.9mm safety<br>screws | 1.5mm screws 1.85 mm<br>safety screws | 1.2mm and 1.4mm<br>screws 1.7mm safety<br>screws | | Application Area | CMF/Orbital | CMF/Orbital | CMF/Orbital | | Target<br>Population | Adults | Unknown | Adults | | Technology | Plates & Screw Fixation | Plates & Screw Fixation | Plates & Screw Fixation | | Operational<br>Principle | Reconstruct the orbital floor<br>and/or medial wall. | Reconstruct the orbital floor<br>and/or medial wall. | Reconstruct the orbital<br>floor and/or medial wall. | Image /page/5/Picture/4 description: The image shows the logo for Colson Associates. The logo consists of the word "Colson" in a bold, sans-serif font, with a vertical line extending above the "l". To the right of the logo is the text "A COLSON ASSOC", also in a sans-serif font, but smaller and less bold than the main logo. a colson associate
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...