The Stryker Universal Orbital Floor System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit.
Device Story
System comprises pre-bent titanium orbital floor implant, globe retractor, and plate-holding forceps. Implant design derived from average anatomical model of 300 healthy CT scans; intended to restore normal orbital volume. Used in clinical setting by surgeons for reconstruction of orbital floor/medial wall following trauma or bone excision. Implant features medial wall extension (removable via cutting lines) and fixation arms for infra-orbital rim attachment. Surgeon uses retractor to expose fracture site, positions plate, and secures with screws. Provides structural support for orbital contents; facilitates anatomical restoration.
Clinical Evidence
No clinical testing was performed to support this submission.
Technological Characteristics
Material: Commercially Pure Titanium (CP Ti Grade 2). Design: 3D pre-bent plate, 0.4mm thickness, based on average anatomical model of 300 CT scans. Fixation: Screw holes for 1.2mm, 1.4mm, and 1.7mm screws. Delivery: Non-sterile. Mechanical modification: Cutting/trimming lines, pliable mesh for bending.
Indications for Use
Indicated for reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in adult patients.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
Synthes Craniofacial Plate and Screw System (K080331)
Submission Summary (Full Text)
{0}------------------------------------------------
K123726
# stryker®
| 510(k) Summary | | APR 0 9 2013 |
|----------------------------|--------------------------------------------------------------------------------------|--------------|
| Date prepared: | December 6, 2012 | |
| Submitter: | Stryker Leibinger GmbH & Co. KG<br>Bötzinger Straße 41<br>79111 Freiburg.<br>Germany | |
| Contact: | Jamshed Badarpura<br>Phone: (269) 389-4260<br>jamshed.badarpura@stryker.com | |
| | Stryker Craniomaxillofacial<br>750 Trade Center Way<br>Portage, MI 49002<br>USA | |
| Proprietary Name: | Stryker Universal Orbital Floor System | |
| Common Name: | Universal Orbital Floor | |
| Proposed Regulatory Class: | Class II | |
| Product Codes: | JEY - Bone Plate | |
| Predicate Device: | Synthes Craniofacial Plate and Screw System -<br>K080331 | |
# Intended Use:
The Stryker Universal Orbital Floor System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit.
#### Indication for Use:
The Stryker Universal Orbital Floor System is indicated for the reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in adult patients.
# Device Description:
The Stryker Universal Orbital Floor System comprises a pre-bent titanium orbital floor implant that approximates the shape and dimensions of the average normal orbital floor and medial wall (in order to facilitate the operative goal of restoring normal (pretraumatic) orbital volume as accurately as possible), a globe retractor for retraction of
Traditional 510(k)
Page 2 of 6
{1}------------------------------------------------
**stryker**®
the orbital contents off the orbital floor and walls (in order to provide exposure of the fracture site(s)), and a plate-holding forceps (that facilitates the insertion and positioning of the plate within the orbit).
The plates are designed based on an average anatomical model of CT-scan data taken from 300 subjects (92% Caucasian). The metadata of the 300 subjects that have been included in the generation of the average anatomical model is listed in table 1. The selected scans were obtained from healthy subjects without any deformation of the bony orbital structures.
# Table 1: Meta data of the 300 CT scans
Image /page/1/Figure/5 description: This image shows a table of data organized by age, gender, and ethnic group. The age ranges are in increments of 10 years, from 10-19 to 90-99. For the age range of 10-19, there are 18 people, 122 females, and 276 people in the 'ca' ethnic group. The average age is 59 years.
Gender: 122 female (f) and 178 male (m);
Ethnic Group: 276 Caucasians (ca), 3 Middle-East (me), 21 unknown (xx)
The implant is provided in pre-bent left and right and in small and large sizes. A medial wall extension is attached to the floor section of the implant in order to provide coverage for a co-extensive medial wall fracture. If not needed, this medial wall extension may be removed by cutting through the designated cutting lines on the implant. The implant is fixed to the infra-orbital rim via either the holes on the provided fixation arms or through the anterior screw holes on the plate itself.
# Clinical Testing:
No clinical testing was performed to support this submission.
# Non-Clinical Testing:
The Stryker Universal Orbital Floor System was subjected to Verification and Validation testing of biocompatibility, cleaning, sterilization, corrosion resistance, stability of plate.
Traditional 510(k)
{2}------------------------------------------------
**stryker**®
functionality over lifetime, transportation of the complete set and of packaging, an end product test and a design validation - end user test. The System passed all tests.
#### Comparison to Predicate Device
The Stryker Universal Orbital Floor System ("Subject Device") is compared to its Predicate Device for Substantial Equivalence. Side-by-side comparison tables summarizing the following criteria are provided below: .
- Intended Use A:
- B: Principle of Operation
- Technological Characteristics ்.
- D: Packaging
#### A: Indications for Use
#### Table 1: Substantial Equivalence of Indications for Use
| | Subject Device | Predicate Device |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Intended Use:<br>The Stryker Universal Orbital Floor<br>System is intended to be used in the<br>reconstruction of the floor and/or<br>medial wall of the orbit. | Intended Use / Indications for Use:<br>The Synthes Craniofacial Plate and Screw<br>System is intended for use in selective trauma<br>of the midface and craniofacial skeleton,<br>craniofacial surgery, reconstructive procedures<br>and selective orthognathic surgery of the<br>maxilla and chin. |
| Intended Use | Indications for Use:<br>The Stryker Universal Orbital Floor<br>System is indicated for the<br>reconstructive treatment of orbital<br>floor and/or medial wall trauma or<br>bone excision in adult patients. | Indications<br>Synthes MatrixMIDFACE Preformed Orbital<br>Plates are intended for trauma repair and<br>reconstruction of the craniofacial skeleton.<br>- Orbital floor fractures<br>- Medial orbital wall fractures<br>– Combined orbital floor and medial wall<br>fractures |
The intended use of the Subject Device is included in the intended use of the Predicate Device. The Predicate has the same indications for use as the Subject Device, though without age specification.
{3}------------------------------------------------
#### B: Principle of Operation
#### Table 2: Substantial Equivalence of Principle of Operation
| | Subject Device | Predicate Device |
|------------------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|
| Application<br>Area | CMF/Orbital | CMF/Orbital |
| Operational<br>principle | Reconstruct the orbital floor and/or<br>medial wall | Reconstruct the orbital floor and/or medial wall |
| Mode of<br>fixation | Plate fixation with screws inserted<br>through dedicated screw holes. | Plate fixation with screws inserted through<br>dedicated screw holes. |
| Duration of<br>implantation<br>within the body | Permanent implant | Permanent implant |
The basic operational principle of the Subject Device, as well as the Predicate Device, is to reconstruct the orbital floor and/or medial wall. The fixation method of both the Subject Device and the Predicate is with screws inserted through dedicated screw holes. The Subject Device and the Predicate Device are both permanent implants.
# C: Technological Characteristics
#### Table 3: Substantial Equivalence of Technological Characteristics
| | Subject Device | Predicate Device |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | Commercially Pure Titanium | Pure Titanium |
| Design | 3D pre-bent plate based on average<br>anatomical model; 0.4mm thickness;<br>Based on average anatomical model of<br>300 CT-scans and comprising 276<br>Caucasian subjects (92%) | 3D pre-bent plate based on average<br>anatomical model; 0.4mm thickness<br>Based on average anatomical model of 279<br>Caucasian CT-scans |
| Mode of<br>Modification | Cutting / trimming along dedicated<br>lines; pliable mesh structure for<br>bending | Cutting / trimming along dedicated lines;<br>pliable mesh structure for bending |
| Sizes | 2 sizes: large and small for left and<br>right each | 2 sizes: large and small for left and right each |
| Dimensions | Large Plate:<br>L=35mm, W=36mm, H=16mm<br>Small Plate:<br>L=31mm W=34mm, H=12mm | Large Plate:<br>L=35mm, W=38mm, H=18mm<br>Small Plate:<br>L=31mm, W=34mm, H=10mm |
| Screw Holes | · For 1.2 mm, 1.4 mm and 1.7 mm<br>screws<br>· Number of screw holes along rim: 4<br>· Two screw fixation arms with a<br>number of 2 screw holes each<br>· Position of screw fixation arms at<br>1st and 3rd screw hole from medial | · For 1.5 mm and 1.8 mm screws<br>· Number of screw holes along rim: 4<br>· Two screw fixation arms with a number of 2<br>screw holes each<br>· Position of screw fixation arms at 1st and 2nd<br>screw hole from medial |
#### Traditional 510(k)
{4}------------------------------------------------
The Subject Device is made of biocompatible commercially pure titanium (CP Ti Grade 2); the Predicate is made of pure titanium.
The Subject Device is pre-bent in a 3D shape and size similar to the Predicate Device.
# D: Packaging
#### Table 4: Technological Characteristics of Packaging
| | IN AN ANNA EN A FRIEND A BE LEAR A SECURITIE E LE PROVECE IN LE FRIENDE E LE FRUITE LE FREE LE LE LEVEL<br>Ipvice<br>Iniant ' | <br>Tedicate<br>Device | |
|----------------|-------------------------------------------------------------------------------------------------------------------------------|------------------------|--|
| Packaging<br>2 | Non-sterile | sterile<br>· STATIC · | |
The Subject Device is delivered non-sterile. The Predicate Device is delivered nonsterile and sterile.
# Substantial Equivalence Conclusion:
The Stryker Universal Orbital Floor System has similar material and technological characteristics, and the same principal of operation as the Predicate Device. The intended use and indications for use are nearly identical to the Predicate Device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three wave-like lines representing the flow of health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the caduceus, indicating the department's name and national affiliation.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### April 9, 2013
Mr. Jamshed Badarpura Regulatory Compliance Analyst Stryker 750 Trade Center Way PORTAGE MI 49002
Re: K123786
Trade/Device Name: Stryker Universal Orbital Floor System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: March 4, 2013 Received: March 7, 2013
Dear Mr. Badarpura:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Badarpura
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/6/Figure/6 description: The image shows the name "Kwame O. Ulmer" in bold, black font at the top. Below the name is a faded logo that appears to be for the FDA. To the right of the logo, the word "for" is written in a smaller, plain font.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
# K123786
Device Name:
Indications For Use:
Stryker Universal Orbital Floor System
The Stryker Universal Orbital Floor System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit.
The Stryker Universal Orbital Floor System is indicated for the reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in adult patients.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary S. Runner -S ﺎ 2013:04.05 And DA ﺍﻟﻤﺴﺎﻋﺪﺓ 11 12:24:33 -04'00'
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Division Control, Dental Devices
123786
Page 1 of _ 1
510(k) Number;_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.