K140930 · Convida Healthcare & Systems Corporation · FSY · Jun 9, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140930
Device Name
VIDA LED SURGICAL LIGHTING SYSTEM, V SERIES
Applicant
Convida Healthcare & Systems Corporation
Product Code
FSY · General, Plastic Surgery
Decision Date
Jun 9, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room.
Device Story
VIDA LED Surgical Lighting System, V series is a ceiling-mounted surgical lamp providing illumination for operating rooms. Device utilizes high-efficacy LEDs as light source; features prism-based light delivery technology to focus light into 110cm focal cylinder, reducing shadow. System offers 5 color temperature settings (3500K-5000K) and adjustable intensity up to 160,000 lux. Operated by surgeons or OR staff via lighthead handle or wall-mounted control center. Includes ambient lighting modes (central luminance or green light) for endoscopy/diagnostic procedures. Supports optional HD 1080p camera handles and medical-grade LCD monitors via pre-wired suspension arms. Designed for durability, ease of cleaning, and continuous operation during power failure. Provides consistent, glare-free illumination to assist surgical teams in visualizing the operative site.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and compliance with international standards (IEC 60601-1, IEC 60601-2-41, IEC 60601-1-2).
Technological Characteristics
Ceiling-mounted surgical luminaire; LED light source; prism-based optical delivery; 5 color temperature settings (3500K-5000K); CRI 95; 160,000 lux max intensity. Modular design; pre-wired for HD video/camera integration. Materials: tough molded, sealed housing for cleaning. Standards: IEC 60601-1, IEC 60601-2-41, IEC 60601-1-2, ISO 14971, ISO 15223-1, ISO 17664, IEC 62304.
Indications for Use
Indicated for illumination during surgery or examination in an operating room setting.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
VIDA LED Surgical Lighting System-X series (K132307)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for ConVida Healthcare & Systems. The logo features a geometric pattern resembling a flower or starburst on the left. To the right of the pattern is the text "ConVida" in a bold, sans-serif font. Below "ConVida" is the text "Healthcare & Systems" in a smaller font size.
ConVida Healthcare & Systems Corp. 510(K) Notification, K number: N/A
# JUN 0 9 2014
Integrate Visual Clarity at the Highest Precision
VIDA LED Surgical Lighting System, V series
## 510(K) Summary
| 1. | Type of Submission: | Traditional |
|----|----------------------|---------------------------------------------------------------------|
| 2. | Preparation Date: | April 3, 2014 |
| 3. | Revised Date: | N/A |
| 4. | Submitter: | ConVida Healthcare & Systems Corporation |
| | Address: | No.33, Dinghu Rd., Guishan Township, Taoyua<br>County 33378, Taiwan |
| | Phone: | +886-3-318-3777 #332 |
| | Fax: | +886-3-318-7799 |
| | Contact: | Sharon Lee / Regulatory Engineer |
| | Registration Number: | 3010402711 |
#### 5. Identification of the Device:
| Proprietary/Trade name: | VIDA LED Surgical Lighting System, V series |
|-------------------------|---------------------------------------------|
| Common Name: | Surgical lamp |
| Classification Name: | Light, Surgical, Ceiling Mounted |
| Device Classification: | 2 |
| Regulation Number: | 878.4580 |
| Panel: | General & Plastic Surgery |
| Product Code: | FSY |
#### б. Identification of the Predicate Device:
| Predicate Device Name: | VIDA LED Surgical Lighting System-X series |
|------------------------|--------------------------------------------|
| Manufacturer: | ConVida Healthcare & Systems Corporation |
| Product Code: | FSY |
| 510(K) Number: | K132307 |
{1}------------------------------------------------
Integrate Visual Cinrity at the Highest Precision
Image /page/1/Picture/1 description: The image contains the logo for ConVida Healthcare & Systems. The logo consists of a grid-like pattern on the left and the word "ConVida" in bold, sans-serif font on the right. Below "ConVida" are the words "Healthcare & Systems" in a smaller font size.
ConVida Healthcare & Syst 510(K) Notification, K number: N/A VIDA LED Surgical Lighting System, V series
#### 7. Intended Use and Indications for Use of the Subject Device:
#### Intended Use:
The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room.
### Indications for Use:
The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room.
#### 8. Device Description:
The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room.
VIDA LED Surgical Lighting System, V series uses high-efficacy light-emitting-diodes (LED) as light source which provides high brightness and no extra heat; the design of the optical mechanism gives a characteristic of less shadow which is significant to the surgeon during the operation. Given of its modular design, VIDA V provided many options for configurations and can be installed in an operating room with or without a laminar air flow system.
The features of VIDA LED Surgical Lighting System, V series are as followed:
- Prism-Based Light Delivery Technology: ConVida's patented optical system, that . combines an advanced refractive prism lens and innovative high-power LED technology, focuses light into an intense 110cm (43") focal cylinder, eliminating the need for refocusing when adjusting surgical table.
- . Consistent Illumination at Different Color Temperature Settings: VIDA V series lighthead's color temperature can be individually adjusted at any of the 5 settings (3500K, 4000K, 4250K, 4500K, 5000K) with a high CRI of 95, R9 94.
- . Optimal Brightness without Glare: Elegant, sophisticated optical system consists up to 2 types of high-power LEDs and multiple refractive prisms produces up to 160,000 lux of intensity without discomfort flare/glare.
- . VIDA V light gives the Surgeon Control: On the VIDA V, both lighthead focus and light intensity can easily be adjusted via turning the lighthead handle. They are also adjustable from the standard lighthead control or optional VIDA V Control Center on the wall.
- Long Lifetime: Under normal use, VIDA V LEDs enjoy an extremely long lifetime of . 50,000+ hours (70% lumen maintenance at 50,000 hours).
- . Two Types of Ambient Lights (Option): 5% central luminance (GUIDE lighting mode) c be configured through standard lighthead control. Guide lighting mode may ambient lighting for manipulating endoscopy instruments or lighting diagnos procedures. Green ambient light sheds low-level green light from the suspension hub
{2}------------------------------------------------
Integrate Visual Clarity at the Highest Precision
ConVida Healthcare & Systems Corp.
510(K) Notification, K number: N/A
VIDA LED Surgical Lighting System, V series
and is controllable on the optional VIDA V Control Center.
- Camera-Ready Lighthead is Built into the System: All VIDA V lightheads are pre-wired . with HD cabling for taking an optional HD 1080p Camera Handle. The HD 1080p Camera Handle is removable from one VIDA V lighthead to another providing superb flexibility for recording.
- Exceptional Maneuvering Flexibility: The exceptionally durable non-sterile handles allow easy positioning of all our lightheads. Each lighthead rotates a full 360°so that the system can be quickly and easily positioned for any surgical procedure. Individual lightheads can be lower to 810mm above the floor providing excellent low lateral illumination.
- . Sleek Non-Sterile Handle: The lighthead has is molded with a sleek all-around handle that facilitates easy repositioning and adds modern elegance to O.R. setting
- Easy-to-clean Compact Lighthead: Made of tough molded materials, the VIDA V series . lightheads are completely sealed which helps prevent biocontaminants from entering the operative site and lighthead itself, and facilitates cleaning.
- . HD Supported Suspensions: All VIDA V suspensions are prewired with video cables to deliver information users need within and without the surgical field. Images are in crisp and accurate HD format with no signal conversion necessary and no loss of video quality.
- Heavy Duty HD Flat Monitor Suspension: Flat Monitor Suspension Arms support single/dual monitors up to a total of 24 kg (53 lbs), making them ideal for surgical displays up to 32". Large display compatibility allows the surgical team to see more of the procedural area. Optional Medical Grade LCD Displays from SONY also offer uncompressed HD video at 1080p which shows true images from SD- and HD-endoscopes, ultrasound, C-arms and other equipment.
- Optimal HD Connectivity: The built-in fiber optic cable allows the suspension to support monitors featuring Full HD 1920 x 1080 video signals.
- . Low Profile Suspension: Low-profile cardinal lighthead suspension is offered for rooms with ceiling lower than 280mm.
- . Failsafe Design: The unit is protected against an electrical power failure, thus ensuring that all surgical procedures can continue without interruption.
Image /page/2/Picture/14 description: The image shows a collection of squares of varying sizes arranged in a pattern. The squares are outlined in black, and some of them have a gray fill. The squares are arranged in a way that creates a visually interesting composition. The text "page 5-4" is visible in the lower right corner of the image.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the ConVida Healthcare & Systems logo. The logo consists of a grid-like pattern on the left and the text "ConVida Healthcare & Systems" on the right. The text is in a bold, sans-serif font, with "ConVida" appearing larger than "Healthcare & Systems".
ConVida Healthcare & Syste 510(K) Notification, K number: N/A
Integrate Visual Clarity at the Highest Precision
VIDA LED Surgical Lighting System, V series
#### 9. Performance Summary:
This device conforms to the standards listed as below:
- IEC 60601-1:2005 Medical Electrical Equipment Part 1: General Requirements for Basic . Safety and Essential Performance
- . IEC 60601-2-41:2009 Medical Electrical Equipment - Part 2-41: Particular requirements for the basic and essential performance of surgical luminaires and luminaires for diagnosis
- . IEC 60601-1-2:2007 Medical Electrical Equipment, Part 1-2: General requirements for basic Safety and essential performance Collateral Standard: Electromagnetic Compatibility Requirements and tests
- IEC 62304:2006 Medical Device Software: Software Life Cycle Process •
- ISO 14971:2007 Medical Device: Application of Risk Management to Medical Devices .
- ISO 15223-1:2012 Medical Device: Symbols to be used with medical device labels, labeling and information to be supplied, part 1: General requirements
- . ISO 17664:2004 Sterilization of medical devices: information to be provided by the manufacturer for the processing of resterilizable medical devices
### 10. Safety and Effectiveness
The bench testing result shows that the proposed device is complied with IEC standards, and as safe as the predicate device.
### 11. Substantial Equivalence Determination:
The VIDA LED Surgical Lighting System, V series is substantially equivalent to VIDA LED Surgical System-X series. There are no negative effect on safety or effectiveness between VIDA LED Surgical Lighting System, V series and the predicate device. The following is the detail of the compared item.
Image /page/3/Picture/18 description: The image shows a pattern of squares with rounded corners. The squares are arranged in a grid-like fashion, with some squares being larger than others. The squares are all outlined in black, and the background is white. The text "page 5-5" is written in the bottom right corner of the image.
{4}------------------------------------------------
page 5-6
ﻨﺔ ﺍﻟ
Integrate Visual Clarity at the Highest Precision
Image /page/4/Picture/1 description: The image shows the logo for ConVida Healthcare & Systems. The logo consists of a geometric pattern on the left and the text "ConVida" on the right. Below "ConVida" is the text "Healthcare & Systems" in a smaller font. The geometric pattern is made up of small squares arranged in a grid-like pattern.
ConVida Healthcare & Systems Corp. 510(K) Notification, K number: N/A
VIDA LED Surgical Lighting System, V series
| Item | Proposed Device | | Predicate Device | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|----------|
| Manufacturer | ConVida Healthcare & Systems Corp. | | ConVida Healthcare & Systems Corp. | |
| Classification | Class II | | Class II | |
| Regulation Number | 878.4580 | | 878.4580 | |
| Product Code | FSY | | FSY | |
| Intended Use | The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room. | | VIDA LED Surgical Lighting System- X series provide illumination for surgery or examination at operating room. | |
| Lighthead model | V70 | V50 | X70 | X50 |
| Similarity | | | | |
| Over-the-counter Medical Device | No | | No | |
| Light Source | LED | LED | LED | LED |
| Color Rendering Index (CRI) | 95 | 95 | 95 | 95 |
| LED Life time (hrs) | 50,000 | 50,000 | 50,000 | 50,000 |
| Difference | | | | |
| Diameter of Light Head (cm) | 82 | 72 | 72 | 58 |
| Number of LEDs | 55 | 44 | 90 | 60 |
| Central illumination (at 1m) (lux) | 160,000 | 140,000 | 160,000 | 130,000 |
| Color temperature, Kelvins | 3500-5000K<br>(3500K,<br>4000K,<br>4250K,<br>4500K,<br>5000K) | 3500-5000K<br>(3500K,<br>4000K,<br>4250K,<br>4500K,<br>5000K) | 4200 | 4200 |
| Total irradiance (W/m²) | 592 | 518 | < 656 | 533 |
| Irradiance intensity (mW/m²lux) | 3.7 | 3.7 | 4.1 | 4.1 |
| d50/d10 (%) | 56 | ਟੇਰੇ | > 50 | > 50 |
| Light field size, d10 (cm) | 17~30 | 16~27 | 20~32 | 18~28 |
| Depth of illumination | 20%: 110 | 20%: 100 | 20%: 140 | 20%: 145 |
| L1+L2 (cm) | 60%: 45 | 60%: 45 | 60%: 71 | 60%: 82 |
| Power Consumption | 100W@24Vdc | 100@24Vdc | 100W | 80W |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for ConVida Healthcare & Systems. The logo consists of a grid-like pattern on the left and the text "ConVida" in bold font on the right. Below "ConVida" is the text "Healthcare & Systems" in a smaller font.
Integrate Visual Clarity at the Highest Precision
510(K) Notification, K number: N/A
VIDA LED Surgical Lighting System, V series
### 12. Non-clinical Performance Testing:
Performance testing were conducted to verify that the VIDA LED Surgical Lighting System, V series meets the requirements for Medical Electrical Equipment as defined in IEC 60601-1, IEC60601-2-41, and IEC 60601-1-2
### 13. Clinical Performance Testing:
No Clinical testing is required for this device classification submission.
### 14. Conclusion:
The VIDA LED Surgical Lighting System, V series submitted in this 510(K) file is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared VIDA LED Surgical Lighting System-X series which is the subject of K132307. Differences between the devices cited in this section do not raise any new issue of safety or effectiveness. After analyzing bench test, safety testing data, it can be concluded that VIDA LED Surgical Lighting System, V series is substantially equivalent to the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract shapes resembling birds in flight, arranged in a diagonal line.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 9, 2014
ConVida Healthcare & Systems Corporation Sharon Lee, Regulatory Engineer No. 33, Dinghu Road . Guishan Township, Taoyuan County 33378, Taiwan
Re: K140930
Trade/Device Name: VIDA LED Surgical Lighting System, V Series Regulation Number: 21 CFR 878.4580 Regulatory Name: Surgical lamp Regulatory Class: Class II Product Code: FSY Dated: April 09, 2014 Received: April 11, 2014
Dear Ms. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dri has made a and regulations administered by other Federal agencies. You must of any I edital bandler and a courtements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{7}------------------------------------------------
Page 2 - Ms. Sharon Lee
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
- cc: DMC 510(k) Staff Division D.O.
{8}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K140930
#### Device Name
VIDA LED Surgical Lighting System, V series
#### Indications for Use (Describe)
The VIDA LED Surgical Lighting System, V series provide illumination for surgery or examination at operating room.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
NEW YOU WARE OF FOR FDA USE ONLY MASE ONLY CHANGE ON WARRER OF THE WORLD Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
Image /page/8/Picture/12 description: The image shows the name Neil R Ogden -S, followed by the date and time 2014.06.05 15:06:29 -04. There is also a logo to the right of the date and time. The logo is not clear, but it appears to be a seal or emblem.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.