K132307 · Convida Healthcare & Systems Corporation · FSY · Nov 5, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132307
Device Name
VIDA LED SURGICAL LIGHTING SYSTEM
Applicant
Convida Healthcare & Systems Corporation
Product Code
FSY · General, Plastic Surgery
Decision Date
Nov 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
VIDA LED Surgical Lighting System - X series provide illumination for surgery or examination at operating room.
Device Story
VIDA LED Surgical Lighting System-X series provides surgical and diagnostic illumination using high-efficacy LEDs. System features modular design, including mounting module, horizontal arm, and suspension device, allowing 360° horizontal rotation and vertical adjustment. Used in operating rooms by surgical staff. Optional accessories include 1080p camera handle module, medical-grade flat monitor, and fiber optic coder/decoder set. Device provides bright, shadow-reduced light to assist clinicians in visualizing surgical sites, potentially improving procedural accuracy and patient outcomes.
Clinical Evidence
No clinical data provided; substantial equivalence established via bench testing and adherence to recognized international standards (IEC 60601-1, IEC 60601-2-41, IEC 60601-1-2).
Technological Characteristics
LED-based surgical luminaire. Features manual focusing, 4200K color temperature, 95 CRI, and 160,000 lux central illumination. Modular suspension system with 360° rotation. Power input 100-240V, 50-60Hz. Complies with IEC 60601-1 (safety), IEC 60601-2-41 (surgical luminaires), and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for illumination of surgical or examination fields in operating rooms.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
ﺮ
VIDA LED Surgical Lighting System-X series
# 510(k) Summary
| 5.1 Type of Submission: | Traditional |
|-------------------------|---------------|
| 5.2 Preparation Date: | July 19, 2013 |
NOV 0 5 2013
| 5.3 Revised Date: | N/A |
|------------------------------------|-------------------------------------------------------------------------|
| 5.4 Submitter: | CONVIDA HEALTHCARE & SYSTEMS CORPORATION |
| Address: | 2F and B2., No.33, Dinghu Rd., Guishan Township, Taoyuan County. Taiwan |
| Phone: | +886-3-318-3777 |
| Fax: | +886-3-318-7799. |
| Contact: | Ryan Hung / Regulatory Engineer |
| Establishment Registration Number: | N/A |
### 5.5 Identification of the Device:
| Proprietary/Trade name: | VIDA LED Surgical Lighting System-X serie |
|-------------------------|-------------------------------------------|
| Common Name: | Surgical lamp |
| Classification Name: | Light. Surgical, Ceiling Mounted |
| Device Classification: | 2 |
| Regulation Number: | 878.4580 |
| Panel: | General & Plastic Surgery |
| Product Code: | FSY |
### Identification of the Predicate Device: 5.6
.
| Predicate Device Name: | Mach LED SC |
|-----------------------------------------|----------------------|
| Manufacturer: | Dr. Mach GmbH& Co.KG |
| 510(k) Number or Clearance Information: | K093009 |
## 5.7 Intended Use and Indications for Use of the subject device.
Intended Use: VIDA LED Surgical Lighting System - X series provide illumination for surgery or examination at operating room.
Indications for Use: VIDA LED Surgical Lighting System-X series provide illumination for surgery or examination at operating room.
{1}------------------------------------------------
VIDA LED Surgical Lighting System-X series
### Device Description 5.8
The VIDA LED Surgical Lighting System-X series provide illumination for surgery or examination at operating room.
The VIDA LED Surgical Lighting System-X series (VIDA X) use high-efficacy light-emitting diodes (LED) for surgical operations and diagnostic purposes. LEDs provide bright and shadow reduced light. Thanks to its modular design. VIDA X also offers many options for configurations and can be installed in operating rooms with or without laminar air flow systems.
A lighthead to the main spindle can be rotated horizontally with 360°. the spring arms can be rotated horizontally with 360° and moved vertically with 45° downwards and 30° upwards. The VIDA LED Surgical Lighting System-X series consists of Mounting Module, Horizontal Arm module, Suspension Device Module, sterile handle cover and Optional Accessories.
Optional accessories for the VIDA LED Surgical Lighting System-X series are as follows:
---- Sterile Handle Cover (for the X50 and X70 lighthead which doesn't mount the Camera, XV, S12. S15 and D24 suspensions.)
---- Camera Handle module (1080p) (which can be installed on X50 or X70 lighthead or XV suspension.)
---- Medical Grade Flat Monitor
---- Fiber Optic Coder Set (including fiber optic transmitter and receptor)
### Performance Summary 5.9.
This device conforms to the standards listed as below:
- IEC 60601-1:2005 Medical Electrical Equipment ~ Part 1: General Requirements for . Basic Safety and Essential Performance
- IEC 60601-2-41:2009 Medical Electrical Equipment Part 2-41: Particular requirements for the basic and essential performance of surgical luminaires and luminaires for diagnosis
- IEC 60601-1-2:2007 Medical Electrical Equipment, Part 1-2: General requirements for basic Safety and essential performance Collateral Standard: Electromagnetic Compatibility Requirements and tests
- IEC 62304:2006 Medical Device Software: Software Life Cycle Process
- 150 14971:2007 Medical Device: Application of Risk Management to Medical Devices
- ISO 15223-1:2012 Medical Device: Symbols to be used with medical device labels, labeling and information to be supplied, part 1: General requirements
- ISO 17664:2004 Sterilization of medical devices: information to be provided by the
20
{2}------------------------------------------------
VIDA LED Surgical Lighting System-X series
manufacturer for the processing of resterilizable medical devices
## 5.10 Safety and Effectiveness
The result of bench testing indicates that the new device is as safe and effective as the predicate device.
## 5.11 Substantial Equivalence Determination
The VIDA LED Surgical Lighting System-X series is substantially equivalent to Mach LED SC. Any difference that exists between the Mach LED SC and the predicate device has no negative effect on safety or effectiveness and actually enhances the usefulness in the chosen application.
| Item | Proposed Device | Predicate Device | | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| | VIDA LED Surgical Lighting<br>System-X series<br>(CONVIDA HEALTHCARE<br>& SYSTEMS<br>CORPORATION) | Mach LED SC<br>(Dr. Mach GmbH& Co.KG.<br>K093009) | | |
| Classification | Class II | Class II | | |
| Code or Federal Regulations | 878.4580 | 878.4580 | | |
| Intended Use | VIDA LED Surgical Lighting<br>System - X series provide<br>illumination for surgery or<br>examination at operating<br>room. | The Mach LED SC lighting<br>system is designed for<br>illuminating an examination<br>area and surgical field at the<br>hospital and doctor's<br>practice. | | |
| Similarity | X70 | | | |
| | X50 | 5sc | 3sc | |
| Over-the-counter Medical Devices | No | No | | |
| Light Source | LED | LED | LED | LED |
| Color Rendering Index | 95 | 95 | 95 | 95 |
| Central illumination (at 1m) (lux) | 160,000 | 130.000 | 160,000 | 130.000 |
| Input Power | 100-240V, single phase,<br>50-60Hz | - | | |
| Difference | | | | |
| Diameter of Light Head | 72cm | 58cm | 72cm | 57cm |
| Number of LEDS | 90 | 60 | 40 | 28 |
| Lighthead Wattage | 150W | 120W | | |
| Power Consumption | 100W | 80W | 65W | 45W |
| Color temperature (Kelvin) | 4200 | 4200 | 4500°K | 4500°K |
| Light field diameter | D10:<br>20-32cm<br>D50/D10:<br>>50% | D10:<br>18-28cm<br>D50/D10:<br>>50% | 20-32cm | 17~28cm |
| Depth of illumination | L1+L2<br>20%: 140cm<br>L1+L2<br>60%: 71cm | L1+L2<br>20%: 145cm.<br>L1+L2<br>60%: 82cm | | |
| Total irradiance at maximum<br>intensity | <656W/m² | <533W/m² | | |
| Average LED Life time(hours) | 50.000 | 50,000 | >40,000 | >40,000 |
| Light focusing mechanism | Manual<br>focusing | Manual<br>focusing | - | - |
| Ultraviolet light filter mechanism | NA | NA | | |
| Available accessories | Standard accessories:<br>-Sterile Handle Cover x2<br>(X50, X70, XV, S12, S15)<br>-Sterile Handle Cover x4<br>(D24)<br>Optional Accessories:<br>-Sterile Handle Cover (for the<br>X50 and X70 lighthead which<br>doesn't mount the Camera,<br>XV, S12, S15 and D24<br>suspensions.)<br>-Camera Handle module<br>(1080p) (which can be<br>installed on X50 or X70<br>lighthead or XV suspension.)<br>-Medical Grade Flat Monitor<br>-Fiber Optic Coder/Decoder | | -Camera module<br>-Remote control for camera<br>module<br>-Remote control with<br>network interface for camera<br>module<br>-Single monitor yoke for flat<br>panel monitors<br>-Double monitor yoke for flat<br>panel monitors<br>-Instrument trays<br>-Trays for CRT monitors<br>-24V DC battery backup<br>support<br>-Low profile wall control<br>unit<br>-Integrated laser pointer | |
{3}------------------------------------------------
---
LED Surgical Lighting System-X series
{4}------------------------------------------------
### VIDA LED Surgical Lighting System-X series
| | Set (including fiber optic transmitter and receptor) | -Sterilizable handle sleeves |
|--|------------------------------------------------------|------------------------------|
| | | |
### 5.12 Non-clinical Performance Testing
Performance testings were conducted to verify that the VIDA LED Surgical Lighting System-X series meet the requirements for Medical Electrical Equipment as defined in IEC 60601-1, IEC 60601-2-42, and IEC 60601-1-2.
# 5.13 Clinical Performance Testing
No clinical testing is required for this device classification submission.
### 5.14 Conclusion
.
The VIDA LED Surgical Lighting System-X series submitted in this 510(k) file is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared Mach LED SC which is the subject of K093009. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness. After analyzing bench tests, safety testing data, it can be concluded that VIDA LED Surgical Lighting System-X series is substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the left side of the eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
November 5, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
CONVIDA HEALTHCARE & SYSTEMS COPRORATION Mr. Ryan Hung Regulatory Engineer 2F and B2., No.33, Dinghu Road., Guishan Township Taoyuan County Taiwan 33378
Re: K132307
Trade/Device Name: VIDA LED Surgical Lighting System-X series Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FSY Dated: September 18, 2013 Received: September 19, 2013
Dear Mr. Hung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Mr. Ryan Hung
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark Night Jierson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use
· 510(k) Number (if known): K132307
Device Name: VIDA LED Surgical Lighting System-X series
### Indications for Use:
VIDA LED Surgical Lighting System - X series provide illumination for surgery or examination at operating room.
Prescription Use -X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use-(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lona 01 3.11.05 14.37.31 -05 00
for MXM
(Division Sign-off) Division of Surgical Devices 510(k) Number: K132307
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.