K140270 · Health Line International Corporation · FOZ · Aug 20, 2014 · General Hospital
Device Facts
Record ID
K140270
Device Name
NEXUS MIDLINE CT CATHETER
Applicant
Health Line International Corporation
Product Code
FOZ · General Hospital
Decision Date
Aug 20, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NEXUS™ MIDLINE CT Catheter is intended to be used by medical professionals for shortterm percutaneous access to the venous system for infusion, intravenous therapy and for blood sampling.
Device Story
Nexus Midline CT Catheter is a peripherally inserted catheter for short-term (less than 30 days) vascular access. Device consists of a biocompatible polyurethane catheter with an injection-molded hub and Luer lock fittings. Available in 3Fr, 4Fr, and 5Fr sizes (single or dual lumen) and lengths of 10 or 20 cm. Used by medical professionals in clinical settings for infusion, intravenous therapy, and blood sampling. Compatible with power injectors for contrast media delivery. Device is introduced via percutaneous microintroducer using Seldinger or Modified Seldinger technique. Performance verified through bench testing, including tensile strength, flow rate, and power injection pressure testing. Benefits include reliable peripheral venous access for therapeutic and diagnostic procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via tensile strength, power injection, dynamic/static failure, static burst, flow rate, air leakage, and liquid leakage tests. All tests met predefined acceptance criteria.
Technological Characteristics
Biocompatible medical-grade polyurethane catheter; injection-molded hub with Luer lock fittings. Sizes: 3Fr, 4Fr, 5Fr (single/dual lumen). Lengths: 10cm, 20cm. Sterilization: Ethylene Oxide (ISO 11135:2007). Biocompatibility: ISO 10993. Power injection compatible.
Indications for Use
Indicated for short term (less than 30 days) peripheral access to the peripheral venous system for infusion, intravenous therapy, and blood sampling. Suitable for use with power injectors.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 9, 2020
Health Line International Corporation John Lincoln Principal Consultant Po Box 2786 St. George, Utah 84771-2786
Re: K140270
Trade/Device Name: Nexus Midline Ct Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: Class II Product Code: FOZ Dated: January 30, 2014 Received: February 3, 2014
Dear John Lincoln:
This letter corrects our substantially equivalent letter of August 20, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sapana Patel -S
for Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140270
Device Name Nexus™ Midline CT Catheter
#### Indications for Use (Describe)
The NEXUS™ MIDLINE CT CATHETER is indicated for short term (less than 30 days) peripheral access to the peripheral venous system for infusion, intravenous therapy and blood sampling. The NEXUS™ MIDLINE CT CATHETER is suitable for use with power injectors. For maximum flow rate and maximum pressure that can be used during power injection, please refer to the following:
| Catheter Size | Priming Volume<br>(ml) | Max Labeled Flow<br>Rate (ml/sec)* | Max Internal Catheter<br>Pressure at Max Flow Rate<br>(psi) | Rated Burst<br>Pressure<br>(psi)** |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|------------------------------------|-------------------------------------------------------------|------------------------------------|
| 3Fr Single Lumen | 0.29 | 1 | 167 | 251 |
| 4Fr Single Lumen | 0.37 | 5 | 164 | 299 |
| 4Fr Dual Lumen | 0.35/0.35 | 3 | 173/173 | 299 |
| 5Fr Single Lumen | 0.38 | 5 | 112 | 300 |
| 5 Fr Dual Lumen | 0.37/0.37 | 5 | 179/179 | 299 |
| * Pressurized flow rates are determined with pump safety cut-off at 300 psi using viscous fluid at 11.8<br>centipoise (cP) and represents approximate flow capability of power injection of contrast media. | | | | |
| ** Max Burst Pressure is the static burst pressure failure point of the catheter when the lumen is<br>completely occluded. WARNING: Power injector machine pressure limiting feature may not prevent<br>over pressurization of an occluded catheter. | | | | |
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (21 CFR 807.92) for NEXUS™ MIDLINE CT CATHETER
#### SUBMITTER:
Health Line International Corporation 803 N. 1250 W. - STE 1 Centerville, Utah 84014
#### ESTABLISHMENT REGISTRATION NUMBER:
3006097687
### CONTACT:
Nola L. Benstog QA/RA Director Telephone: 801-773-7798 Fax: 855-228-1336 Email: nbenstog@hlic.net
#### DATE PREPARED:
January 30, 2014
#### NAME OF MEDICAL DEVICE:
| Proprietary Name: | NEXUS™ MIDLINE CT CATHETER |
|--------------------|-------------------------------------------|
| Regulation Name: | Intravascular Catheter |
| Common/Usual Name: | Midline Catheter, single and double lumen |
#### DEVICE CLASSIFICATION:
| Classification Panel: | General Hospital |
|-----------------------|------------------|
| Regulatory Class: | Class II |
| Product Code: | FOZ |
| Regulation Number: | 21 CFR 880.5200 |
#### PREDICATE DEVICES:
| Proprietary Name: | SYNERGY™ CT PICC (K101329) |
|-----------------------|-----------------------------------------------------------------------------------|
| Regulation Name: | Percutaneous, Implanted, Long-term Intravascular Catheter |
| Common/Usual Name: | Peripherally Inserted Central Catheter (PICC), single, double<br>and triple lumen |
| Classification Panel: | General Hospital |
| Regulatory Class: | Class II |
| Product Code: | L J S |
| Regulation Number: | 21 CFR 880.5970 |
{4}------------------------------------------------
| Proprietary Name: | UNI-PICCC™ (K130034) |
|-----------------------|-----------------------------------------------------------------------------------|
| Regulation Name: | Percutaneous, Implanted, Long-term Intravascular Catheter |
| Common/Usual Name: | Peripherally Inserted Central Catheter (PICC), single, double<br>and triple lumen |
| Classification Panel: | General Hospital |
| Regulatory Class: | Class II |
| Product Code: | L J S |
| Regulation Number: | 21 CFR 880.5970 |
| Proprietary Name: | Vascu-PICC® and Midline Catheters, Single, Double and<br>Triple Lumen (K091586) |
| Regulation Name: | Percutaneous, Implanted, Long-term Intravascular Catheter |
| Common/Usual Name: | Catheter, Intravascular, Therapeutic, Long-Term |
| Classification Panel: | General Hospital |
| Regulatory Class: | Class II |
| Product Code: | L J S |
| Regulation Number: | 21 CFR 880.5970 |
| Proprietary Name: | PowerGlide Midline Catheter (K121073) |
| Regulation Name: | Intravascular Catheter |
| Common/Usual Name: | Intravascular Catheter |
| Classification Panel: | General Hospital |
| Regulatory Class: | Class II |
| Product Code: | FOZ |
| Regulation Number: | 21 CFR 880.5200 |
### DEVICE DESCRIPTION:
The NEXUS™ MIDLINE CT Catheter is a family of peripherally inserted central catheters made from specially formulated biocompatible medical grade materials. Catheters are packaged in a tray with accessories necessary for a percutaneous microintroducer introduction (Modified Seldinger or Seldinger technique). The device is intended for short term (less than 30days) vascular access.
The NEXUS™ MIDLINE CT Catheter is indicated for dwell times shorter than 30 days. The NEXUS™ MIDLINE CT Catheter product line will have catheters in 3Fr, 4 Fr and 5 Fr single lumen, and 4 Fr and 5 Fr dual lumen. The NEXUS™ MIDLINE CT Catheters are supplied in two lengths, approximately 10 and 20 cm long. All catheters are attached to an injectionmolded polyurethane hub that has extension legs with Luer lock fittings for access attachment.
{5}------------------------------------------------
## INTENDED USE:
The NEXUS™ MIDLINE CT Catheter is intended to be used by medical professionals for shortterm percutaneous access to the venous system for infusion, intravenous therapy and for blood sampling.
The intended use for the NEXUS™ MIDLINE CT Catheter has not changed from that of the SYNERGY™ CT PICC (K101329) and UNI-PICC™ (K130034) predicate devices.
## INDICATIONS FOR USE:
The NEXUS™ MIDLINE CT CATHETER is indicated for short term (less than 30 days) peripheral access to the peripheral venous system for infusion, intravenous therapy and blood sampling. The NEXUS™ MIDLINE CT CATHETER is suitable for use with power injectors. For maximum flow rate and maximum pressure that can be used during power injection, please refer to the following:
| Catheter Size | Priming Volume<br>(ml) | Max Labeled Flow<br>Rate (ml/sec)* | Max Internal Catheter<br>Pressure at Max Flow Rate<br>(psi) | Rated Burst<br>Pressure<br>(psi)** |
|------------------|------------------------|------------------------------------|-------------------------------------------------------------|------------------------------------|
| 3Fr Single Lumen | 0.29 | 1 | 167 | 251 |
| 4Fr Single Lumen | 0.37 | 5 | 164 | 299 |
| 4Fr Dual Lumen | 0.35/0.35 | 3 | 173/173 | 299 |
| 5Fr Single Lumen | 0.38 | 5 | 112 | 300 |
| 5 Fr Dual Lumen | 0.37/0.37 | 5 | 179/179 | 299 |
Pressurized flow rates are determined with pump safety cut-off at 300 psi using viscous fluid at 11.8 centipoise (cP) and represents approximate flow capability of power injection of contrast media.
* * Max Burst Pressure is the static burst pressure failure point of the catheter when the lumen is completely occluded. WARNING: Power injector machine pressure limiting feature may not prevent over pressurization of an occluded catheter.
## TECHNOLOGICAL COMPARISON TO PREDICATE DEVICES:
New device is compared to Marketed Device? Yes. It is compared to four legally marketed predicate devices.
Does the new device have the same indication statements? Yes the indications for use statement is basically the same as the SYNERGY™ CT PICC (K101329) and the UNI-PICC™ (K130034) predicate devices.
{6}------------------------------------------------
Do the differences alter the intended therapeutic/diagnostic/etc. effect (i.e., deciding may consider impact on safety and effectiveness)? No, the differences do not alter the intended effect of the device.
Does the new device have the same technological characteristics, e.g. design, material, etc.? Yes. The device of this submission, the NEXUS™ MIDLINE CT CATHETER, is identical to that of the predicate devices. In fact, the SYNERGY™ CT PICC and the UNI-PICC™ are manufactured by the same company, Health Line International Corporation, the submitter of this submission. The basic fundamental scientific technology of the device has not changed. There may be minor variations in the contents of the introduction kit components.
Could the new characteristics affect safety or effectiveness? No.
Do the new characteristics raise new types of safety and effectiveness questions? No. There are no new types of safety and effectiveness questions.
Do accepted scientific methods exist for assessing effects of the new characteristics? Yes.
The FDA's Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, dated 3/16/95 was used to determine the appropriate methods for evaluating the device's performance.
Sterilization requirements of ISO 11135:2007, Sterilization of Health Care Products -Requirements for Validation and Routine Control -- Ethylene Oxide Sterilization.
Biocompatibility requirements according to of ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing. Test profiles for externally communicating, bloodcontacting, long-term devices will be met.
These and other standards were used to determine the appropriate methods for evaluating the device's performance.
Are performance data available to assess effects of new characteristics? Yes. Verification testing was performed according to protocols based on the above-referenced guidance document recommendations and additional standards.
Nonclinical tests submitted included appropriate biocompatibility tests, as well as performance bench testing. Bench test data including Tensile Strength Test, Power Injection Performance Test, Dynamic Failure Test, Static Pressure Test, Static Burst Test, Life Cycle Power Injection Test, Catheter Flow Rate Test, Air Leakage Test and Liquid Leakage Test all met predefined acceptance criteria, and supported the substantial equivalence determination.
{7}------------------------------------------------
Do performance data demonstrate equivalence? Yes. Performance data gathered in design verification testing demonstrated that the NEXUS™ MIDLINE CT Catheter is substantially equivalent to the noted predicate devices.
## CONCLUSION
The NEXUS™ MIDLINE CT Catheter met all established acceptance criteria for performance testing and design verification testing. This testing demonstrated that the NEXUS™ MIDLINE CT Catheter is safe and effective for its intended use, and based on FDA's decision tree is substantially equivalent to the SYNERGY™ CT PICC (K101329), UNI-PICC™ (K130034), Vascu-PICC® and Midline Catheters, Single, Double and Triple Lumen (K091586) and PowerGlide Midline Catheter (K121073).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.