K101329 · Health Line International Corporation · LJS · Aug 19, 2010 · General Hospital
Device Facts
Record ID
K101329
Device Name
SYNERGY CT PICC
Applicant
Health Line International Corporation
Product Code
LJS · General Hospital
Decision Date
Aug 19, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
The SYNERGY™ CT PICC is intended to be used by medical professionals for short-term or long-term peripheral access to the central venous system for infusion, intravenous therapy, blood sampling and for power injection of contrast media. The intended use has not changed from that of the predicate devices. All SYNERGY™ CT PICC products have a maximum recommended infusion rating of 5 ml/sec. The maximum pressure or pounds per square inch (psi) of the power injector utilized should not exceed 300 psi.
Device Story
SYNERGY™ CT PICC is a peripherally inserted central venous catheter (PICC) for infusion, IV therapy, blood sampling, and power injection of contrast media. Device consists of radiopaque polyurethane tubing with a kink-resistant, reverse-tapered design and an injection-molded polyurethane hub with Luer lock fittings. Available in 4 Fr/5 Fr single lumen, 5 Fr/6 Fr dual lumen, and 6 Fr triple lumen configurations, approximately 60 cm in length. Used by medical professionals in clinical settings for short- or long-term central venous access. Supports power injection up to 5 ml/sec at a maximum pressure of 300 psi. Provides reliable venous access, reducing need for repeated venipuncture and facilitating contrast-enhanced imaging procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Verification testing performed according to FDA guidance for short-term and long-term intravascular catheters and relevant ISO standards (ISO 11135:2007, ISO 10993). Testing confirmed device meets performance criteria for power injection (5 ml/sec at 300 psi) and biocompatibility.
Indicated for short or long term (less than or greater than 30 days) peripheral access to the central venous system for infusion, intravenous therapy, blood sampling, and power injection of contrast media in patients requiring central venous access.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
K101329
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (21 CFR 807.92) for SYNERGY™ CT PICC
# SUBMITTER:
Health Line International Corporation 803 N. 1250 W. - STE 1 Centerville, Utah 84014
AUG 1 9 2010
# ESTABLISHMENT REGISTRATION NUMBER: 3006097687
CONTACT:
Nola L. Benstog QA/RA Director Telephone: 801-773-7798 Fax: 801-820-8007 Email: nbenstog@hlic.net
# DATE PREPARED:
Proprietary Name:
Regulation Name:
Common/Usual Name:
May 11, 2010
# NAME OF MEDICAL DEVICE:
# SYNERGY™ CT PICC
Percutaneous, implanted, long-term intravascular catheter Peripherally Inserted Central Catheter (PICC), single, double and triple lumen
# DEVICE CLASSIFICATION:
Classification Panel: Regulatory Class: Product Code: Regulation Number: General Hospital Class II ાર 21 CFR 880.5970
#### PREDICATE DEVICES:
Proprietary Name: Regulation Name: Common/Usual Name: Classification Panel: Regulatory Class: Product Code: Regulation Number:
### PICC (CT Rated and Non-Rated) (K083873)
Percutaneous, Implanted, Long-term Intravascular Catheter Catheter, Intravascular, Therapeutic, Long Term General Hospital Class II ।।
21 CFR 880.5970
CUNFIDENTIAL
{1}------------------------------------------------
| Proprietary Name: | OmniPICC P.I. (K051102) |
|-----------------------|---------------------------------------------------------------------------------------------|
| Regulation Name: | Percutaneous, implanted, long-term intravascular catheter |
| Common/Usual Name: | Peripherally Inserted Central Catheter (PICC), single and<br>double lumen |
| Classification Panel: | General Hospital |
| Regulatory Class: | Class II |
| Product Code: | LJS |
| Regulation Number: | 21 CFR 880.5970 |
| Proprietary Name: | OmniPICC P.I. (K062579) |
| Regulation Name: | Percutaneous, implanted, long-term intravascular catheter |
| Common/Usual Name: | Peripherally Inserted Central Catheter (PICC), 4 French single<br>and 5 French double lumen |
| Classification Panel: | General Hospital |
| Regulatory Class: | Class II |
| Product Code: | LJS |
| Regulation Number: | 21 CFR 880.5970 |
| Proprietary Name: | PRO-LINE CT Pressure Injectable CVC (K093309) |
| Regulation Name: | Percutaneous, Implanted, Long-Term intravascular Catheter |
| Common/Usual Name: | Percutaneous, implanted, long-term intravascular catheter |
| Classification Panel: | General Hospital |
| Regulatory Class: | Class II |
| Product Code: | LJS |
| Regulation Number: | 21 CFR 880.5970 |
#### DEVICE DESCRIPTION:
The SYNERGY™ CT PICC is a family of peripherally inserted central venous catheters designed to perform infusion, intravenous therapy, blood sampling and also power injection of contrast media studies. The catheters, made of radiopaque polyurethane tubing, are inserted peripherally. Each SYNERGY™ CT PICC has a kink resistant, reverse tapered catheter design. The SYNERGY™ CT PICC kit includes a catheter and introduction components. The catheter is supplied sterile and non-pyrogenic in a variety of kit configurations.
The SYNERGY™ CT PICC is indicated for dwell times shorter or greater than 30 days. The SYNERGY™ CT PICC catheter assemblies have been tested to withstand power injection of worst-case viscosity injection media at 5 ml/sec with a maximum power injector pressure of 300 psi.
The SYNERGY™ CT PICC product line has catheters in 4 Fr and 5 Fr single lumen, 5 Fr and 6 Fr dual lumen and 6 Fr triple lumen. All catheters are approximately 60 cm long. The catheters are attached to an injection-molded polyurethane hub that has extension legs with Luer
CONFIDENTIAL
{2}------------------------------------------------
lock fittings for access attachment. All SYNERGY™ CT PICC products have a maximum recommended infusion rating is 5 ml/sec. The maximum pressure or pounds per square inch (psi) of the power injector utilized should not exceed 300 psi.
#### CONFIDENTIAL
{3}------------------------------------------------
# INTENDED USE:
The SYNERGY™ CT PICC is intended to be used by medical professionals for short-term or long-term peripheral access to the central venous system for infusion, intravenous therapy, blood sampling and for power injection of contrast media. The intended use has not changed from that of the predicate devices. All SYNERGY™ CT PICC products have a maximum recommended infusion rating of 5 ml/sec. The maximum pressure or pounds per square inch (psi) of the power injector utilized should not exceed 300 psi.
# INDICATIONS FOR USE:
The SYNERGY™ CT PICC is indicated for short or long term (less than or greater than 30 days) peripheral access to the central venous system for infusion, intravenous therapy, blood sampling and power injection of contrast media. All SYNERGY™ CT PICC products have a maximum recommended infusion rating of 5 ml/sec.
# TECHNOLOGICAL COMPARISON TO PREDICATE DEVICES:
New device is compared to Marketed Device? Yes. It is compared to legally marketed predicates.
Does the new device have the same indication statements? Yes.
Do the differences alter the intended therapeutic/diagnostic/etc. effect (i.e. deciding may consider impact on safety and effectiveness)? No, the differences do not alter the intended use of the device.
Does the new device have the same technological characteristics, e.g. design, material, etc.? Yes. The PICC device of this submission, the SYNERGY™ CT PICC, is identical to that of (K083873), PICC (CT Rated and Non-Rated). In fact, the SYNERGY™ CT PICC and the PICC (CT Rated and Non-Rated) are manufactured by the same company, Health Line International Corporation, the submitter of this submission. In addition, the SYNERGY™ CT PICC is substantially equivalent to predicate devices, OmniPICC P.I. (K051102) and (K062579) and PRO-LINE CT Pressure Injectable CVC (K093309). The basic fundamental scientific technology of the device has not changed. There may be minor variations in the contents of the introduction kit components.
Could the new characteristics affect safety or effectiveness? No.
Do the new characteristics raise new types of safety and effectiveness questions? No. There are no new types of safety and effectiveness questions.
Do accepted scientific methods exist for assessing effects of the new characteristics? Yes.
CONFIDENTIAL
{4}------------------------------------------------
The FDA's Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, dated 3/16/95 was used to determine the appropriate methods for evaluating the device's performance.
Sterilization requirements of ISO 11135:2007, Sterilization of Health Care Products -Requirements for Validation and Routine Control -- Ethylene Oxide Sterilization.
Biocompatibility requirements according to of ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing. Test profiles for externally communication of Irences contacting, long-term devices will be met.
These and other standards were used to determine the appropriate methods for evaluating the device's performance.
Are performance data available to assess effects of new characteristics? Yes. Verification testing was performed according to protocols based on the above-referenced vicince document recommendations and additional standards.
Do performance data demonstrate equivalence? Yes. Performance data gathered in design verification testing demonstrated that the SYNERGY™ CT PICC is substantially equivalent to the noted predicate devices.
#### CONCLUSION
The SYNERGY™ CT PICC met all established acceptance criteria for performance testing and design verification testing. This testing demonstrated that the SYNERGY™ CT PICC is safe and effective for its intended use, and based on FDA's decision tree is substantially equivalent to the following predicate devices: PICC (CT Rated and Non-Rated)(K083873), OmniPICC P.I. (K051102) and (K062579) and PRO-LINE CT Pressure Injectable CVC (K093309).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Nola Benstog Ouality Assurance/ Regulatory Affairs Director Health Line International Corporation 803 N. 1250 W street, Suite 1 Centerville, Utah 84014
AUG 1 9 2010
Re: K101329
Trade/Device Name: SYNERGY™ CT PICC Regulation Number: 21 CFR 880.5970 Regulation Name: Pertaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: July 26, 2010 Received: July 27, 2010
Dear Ms. Benstog:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2- Ms. Benstog
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Susan Lumore
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K101329
Page 14 of 169
### Indications For Use
510(k) Number (if known):
Device Name: SYNERGY™ CT PICC
Indications For Use:
The SYNERGY™ CT PICC is indicated for short or long term (less than or greater than 30 days) peripheral access to the central venous system for infusion, intravenous therapy, blood sampling and power injection of contrast media. All SYNERGY™ CT PICC products have a maximum recommended infusion rating of 5 ml/sec.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Uid C. Ch 8/19/10
Division Sign-Off) (Division Sign=ON)
Division of Anesthesiology, General Hospital
Division of Anest Dantal Devices infection Control, Dental Devices
510(k) Number: k101329
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.