DIGITAL DENTAL INTRA ORAL SENSOR, HDI 2000, HDI 2000A
Applicant
Rayence Co., Ltd.
Product Code
MQB · Radiology
Decision Date
May 13, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
HDI 2000 / HDI 2000A, an Intra-oral Imaging System, is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists.
Device Story
Intra-oral imaging system; acquires digital dental X-ray images using solid-state sensor. Sensor connects via wire to USB box; USB box connects to computer. System captures, stores, views, and manipulates images for diagnostic use by dentists. Operates via viewer software (EasyDent or ProraView). ProraView includes consulting simulation tool. Used in dental clinics. Provides instantaneous digital images following exposure; facilitates patient information management and diagnostic reporting.
Clinical Evidence
Bench testing only. Electrical, mechanical, and environmental safety testing performed per IEC 60601-1:2005 and EMC testing per IEC 60601-1-2:2007. Non-clinical and clinical considerations addressed per FDA guidance for solid-state X-ray imaging devices. All test results satisfactory.
Technological Characteristics
Solid-state intra-oral sensor; USB 2.0 connectivity; analog-to-digital conversion. Dimensions: 38.7x29.2x4.95mm or 42.8x31.5x4.95mm. Software-based image processing and management. Standards: IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for use by dentists to collect dental x-ray photons and convert them into electronic impulses for storage, viewing, and manipulation for diagnostic purposes.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K140060
Page 1 of 4
510(k) Submission – HDI 2000 / HDI 2000A
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirement s of 21 CFR Part 807.92.
### Date 510K summary prepared: January 3, 2014
Submitter's Name, address, telephone number, a contact person:
| Submitter's Name : | Rayence Co., Ltd. |
|-------------------------|---------------------------------------------------------|
| Submitter's Address: | 14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea |
| Submitter's Telephone: | +82-31-8015-6459 |
| Contact person: | Mr. Kee Dock Kim / Manager / +82-31-8015-6459 |
| Official Correspondent: | Dave Kim (davekim@mtech-inc.net) |
| (U.S. Designated agent) | |
| Address: | 8310 Buffalo Speedway, Houston, TX 77025 |
| Telephone: | +713-467-2607 |
| Fax: | +713-583-8988 |
Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:
| Trade/proprietary name: | HDI 2000, HDI 2000A |
|-------------------------|-----------------------------------------|
| Common Name: | Medical Image Processing Unit |
| Classification Name : | 21CFR892.1680 / Stationary x-ray system |
| Product Code: | MQB |
### Predicate Device :
| Manufacturer | : Rayence Co., Ltd. |
|---------------|-------------------------------------------|
| Device | : EzSensor P |
| 510(k) Number | : K121132 (Decision Date - MAY. 10. 2012) |
2
{1}------------------------------------------------
#### Device Description :
The HDI 2000 / HDI 2000A Intra-oral imaging system is a device which acquires digital image. Direct digital systems acquire images with a solid-state sensor that is connected to a computer to produce an image almost instantaneously following exposure. The primary advantage of direct sensor systems is the speed with which images are acquired. The sensor is connected by a wire to an analog-to-digital USB box, which is connected to the computer. Images are produced within seconds of sensor exposure. The wire length from a direct sensor is about 3 meters and less. The USB box connects to the USB 2.0 port of the computer.
#### Indication for use :
HDI 2000 / HDI 2000A, an Intra-oral Imaging System, is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists.
### Summary of the technological characteristics of the device compared to the predicate device
The The indications for use, material, form factor, performance, and safety characteristics of HDI 2000 / HDI 2000A described in this 510(k) are the same as that of the predicate device, EzSensor P of Rayence Co., Ltd.
The primary difference is the size of new sensors and availability of optional image viewing software, Prora view, which is alreay cleared by FDA (K131594). As demonstrated by the technical comparison between the subject and predicate device, the difference in the physical dimension of the sensor does not present any new concerns in terms of safety and effectiveness.
{2}------------------------------------------------
#### 1. HDI 2000 / HDI 2000A Summary
#### 1.1 Hardware
The hardware of HDI 2000 and HDI 2000A are identical.
The shape of the top cover and case of EzSensor P ( K121132) are modified without affecting the sensor performance.
| Change | HDI 2000 / HDI 2000A | EzSensor P |
|-------------------------------------------------|-------------------------------------------|---------------------------------|
| Image: Sensor ASS'Y | Image: HDI 2000 / HDI 2000A exploded view | Image: EzSensor P exploded view |
| Sensor_ASS'Y<br>③ : Upper Case<br>⑤ : Top Cover | | |
#### 1.2 Viewer Software
This device is operated using viewer software. With the identical hardware configuration, the model name is distinguished by the type of image viewing software available. The software functions include the patient information management, image capture, and viewer for captured images.
| Model | Sensor Type | Size (mm) | Viewer Software | Viewer Software<br>Manufacturer |
|-----------|-------------|--------------------|---------------------|---------------------------------|
| HDI 2000 | 1.5 | 38.7 x 29.2 x 4.95 | Easydent | Rayence Co., Ltd. |
| HDI 2000A | 1.5 | 38.7 x 29.2 x 4.95 | ProraView (K131594) | EWOO Soft |
| HDI 2000 | 2.0 | 42.8 x 31.5 x 4.95 | Easydent | Rayence Co., Ltd. |
| HDI 2000A | 2.0 | 42.8 x 31.5 x 4.95 | ProraView (K131594) | EWOO Soft |
Easydent is the same software used for the predicate device whereas ProraView image viewing software is a stand alone imaging viewer program with 510(k) clearance of its own. Both viewing programs have the similar functionality and performance. The main difference is the design of the user interface (UJ) and new consulting simulation tool for Prora View. Prora View requires a fee based license registration for the right to use whereas EasyDent is provided to users free of charge.
The design control risk analysis for ProraView describes hazards associated with respect to SRS and SDS. All risks are mitigated and any residual risks are determined acceptable.
{3}------------------------------------------------
K140060
Page 4 of 4
- 2.1) Difference for Easydent and ProraView
| | Easydent | Prora View |
|----------|--------------|--------------|
| Function | Patient | Patient |
| | View | View |
| | Acquisition | Acquisition |
| | Consultation | Consultation |
| | Report | Report |
| | | License |
### Summary for any testing in the submission
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1: 2005 + CORR.I(2006) + CORR(2007) (Medical electrical equipment Part 1:General requirements for basic safety and essential performance) was performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2:2007.
Non-clinical & Clinical considerations according to FDA Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed.
All test results were satisfactory.
### Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Rayence Co., Ltd. concludes that HDI 2000 / HDI 2000A is safe and effective and substantially equivalent to predicate device as described herein.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular text element and a symbol. The text element encircles the symbol and reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". The symbol is a stylized representation of a human figure, with three overlapping figures in a dynamic, forward-leaning pose.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 13, 2014
Rayence Co. Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025
Re: K140060
Trade/Device Name: Digital Dental Intra Oral Sensor (HDI 2000, HDI 2000A) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB and MUH Dated: April 15, 2014 Received: April 18, 2014
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2—Mr. Kim
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fdagov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Smh.7)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
#### 510(k) Number (if known) K140060
#### Device Name
HDI 2000, HDI 2000A digital dental image processing system
#### Indications for Use (Describe)
HDI 2000 / HDI 2000A, an Intra-Oral Intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
(\$m, 3)
FORM FDA 3881 (6/13)
PSL" Pakkatung Services { 101 \ 44 \-47 \u
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.