EzDent-i is dental imaging software that is intended to provide diagnostic tools for maxillofacial radiographic imaging. These tools are available to view and interpret a series of DICOM compliant dental radiology images and are meant to be used by trained medical professionals such as radiologist and dentist. EzDent-i is intended for use as software to acquire, view and save 2D image files, load DICOM project files from panorama, cephalometric, and intra-oral imaging equipment.
Device Story
EzDent-i is dental imaging software for storage, processing, and viewing of 2D radiographic images. Inputs include digital X-ray images from VATECH equipment, intraoral photographs, and imported DICOM files. Software functions as a central database for patient information and images, supporting multi-workstation network environments. Clinicians use tools for image enhancement (brightness, contrast, sharpness, filters), measurement (linear, angle), and annotation (text, drawing, implant libraries). Output is displayed on standard PC/Mac monitors for diagnostic review by dentists or radiologists. Device facilitates clinical decision-making by providing diagnostic visualization and measurement tools for maxillofacial imaging. Benefits include centralized image management and improved diagnostic utility of radiographic data.
Clinical Evidence
Bench testing only. Software validation testing evaluated all input/output functions and operational modes against predetermined acceptance criteria. No clinical data provided.
Technological Characteristics
Software-only device; no patient contact. Compatible with IBM-compatible PC or Mac. Supports DICOM image format and SQL databases. Features include linear/angle measurement, image enhancement filters, and generic implant libraries. Complies with IEC 62304 (software life-cycle) and ISO 14971 (risk management).
Indications for Use
Indicated for trained medical professionals (radiologists, dentists) to view, interpret, acquire, and save 2D maxillofacial radiographic images (panoramic, cephalometric, intra-oral) in DICOM format.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
KB1594
Page 1 of 6
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
## Date
May 29, 2013
#### Manufacturer
Ewoo Soft Co., Ltd S-402, 231 Pangyoyeok-ro, Bundang-Gu Seongnam-Si, Gyeonggi-do, 463-440, Republic of Korea Tel: +82 31 8015 6171 Fax: +82 31 6966 7420
Contact person: Mr. Young Seok Kim Email: ys.kim@ewoosoft.com
#### United States Sales Representative (U.S. Designated agent)
Mtech Group 12946 Kimberley Ln Houston, TX 770779 Tel: +713-467-2607 Fax: +713-464-8880 Contact person: Mr. Dave Kim
## Trade/Proprietary Name:
EzDent-i /E2 / ProraView
#### Common Name:
Dental Imaging Software
#### Classification Name:
System, image processing, radiological (21CFR 892.2050, Product code LLZ, Class2)
AUG 2 9 2013
{1}------------------------------------------------
#### Description :
EzDent-i is a dental imaging software solution that stores, analyzes and diagnoses patient images that have been acquired through VATECH dental equipment.
EzDent-i is equipped with everything you need for digital panoramic and cephalometric image storage, processing and viewing. EzDent-i functions as a central storage point for digital images and associated patient data. Images can be acquired directly from equipment that EzDent-i currently supports. In addition, images can be imported from other digital sources.
#### The Main Functions of EzDent-i
With EzDent-i you can perform the following operations assuming that all the other equipment is ready to use.
- Create and store new patient information in a database 1.
- Capture and store digital X-ray images with exposure values from the device. 2.
- 3. Capture and store intraoral photographs.
- Export and import digital images 4.
- Process images to enhance their diagnostic value with dental-specific tools 5.
- Analyze the image with application-specific measurement tools 6.
- Build an environment with multiple workstations using a database shared over a network. 7.
- 8. Printing images and image related information.
EzDent-i can be used in a network environment. If EzDent-i is installed in several computers, the patient and image database can be shared among them and used from different workstations
{2}------------------------------------------------
#### Indication for use:
EzDent-i is dental imaging software that is intended to provide diagnostic tools for maxillofacial radiographic imaging. These tools are available to view and interpret a series of DICOM compliant dental radiology images and are meant to be used by trained medical professionals such as radiologist and dentist.
EzDent-i is intended for use as software to acquire, view and save 2D image files, load DICOM project files from panorama, cephalometric, and intra-oral imaging equipment.
#### Predicate Device:
| Manufacturer | : Gendex Dental System |
|---------------|--------------------------------------------|
| Device | : VixWin Pro |
| 510(k) Number | : K060178 (Decision Date - March 10, 2006) |
## Comparison of features and specifications of the device and SE device:
| Characteristic | EzDent-i | Vix Win Pro |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K number | K131594 | K060178 |
| Manufacturer | EWOO Soft | Gendex Dental System |
| Indications for use | EzDent-i is dental imaging<br>software that is intended to<br>provide diagnostic tools for<br>maxillofacial radiographic<br>imaging. These tools are<br>available to view and interpret<br>a series of DICOM compliant<br>dental radiology images and<br>are meant to be used by<br>trained medical professionals<br>such as radiologist and dentist.<br>EzDent-i is intended for use as<br>software to acquire, view and<br>save 2D image files, load<br>DICOM project files from<br>panorama, cephalometric, and<br>intra-oral imaging equipment. | Vix Win Pro controls capture,<br>display, treatment, analysis<br>and saving of X-ray digital<br>images from<br>DenOptix®,Visualix®/GX-S,<br>Orthoalix 9200 DPI and DDE<br>digital imaging systems<br>produced by Gendex. It can<br>also handle other types of<br>digital images, e.g. color<br>images from an intraoral or<br>extraoral dental camera, such<br>as the Gendex Concept IV<br>series, or images acquired by<br>digitizing film with a flat bed<br>scanner. |
| Platform | IBM-compatible PC or Mac | PC |
| Operating System | Microsoft Window 7, Window<br>8, Mac(Leopard, Snow<br>Leopard, Lion) | Windows 98, 2000 and XP |
{3}------------------------------------------------
| User Interface | | |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Image Input<br>Sources | Images can be scanned, loaded<br>from digital cameras or card<br>readers, or imported from a<br>radiographic imaging device | Images can be scanned,<br>loaded from digital cameras<br>or card readers, or imported<br>from a radiographic imaging<br>device |
| 32 bit / 64 bit | 32 / 64 bit | 32 / 64 bit |
| Image format | DICOM | DICOM |
| Patient Database<br>Compatibility | SQL | SQL |
| Includes Image<br>Measurement<br>tools | Linear distance, angle | Angle, length measurement |
| Image viewing | Full, side by side, gallery,<br>thumbnail | Full, side by side, gallery,<br>thumbnail |
| Image<br>manipulation | Brightness, contrast,<br>sharpness, inverse, film view,<br>rotate, zooming, whitening,<br>nerve canal tracing, memo | Sharpen, smooth, area<br>measurement, rotate, flip,<br>mirror, reverse, grayscale<br>enhance, emboss, brightness,<br>contrast, equalize, contrast,<br>emboss, despeckle, optimizer,<br>colorize, magnify, spotlight,<br>annotation, soft tissue filter-<br>for cephalometric only, |
| Implant module | Generic implant libraries | Include implant libraries from<br>Nobel Biocare, Bicon, 3i, and<br>Straumann, and generic |
| 3D imaging<br>capability | Includes interface to 3D<br>imaging software, Ez3D-i.<br>EzDent-i imaging software<br>does not view, transfer or<br>process 3D radiographs. | Includes interface to 3D<br>imaging software provided<br>with Gendex GXDP-700<br>series, GXCB-500series.<br>Vix Win imaging software<br>does not view, transfer or<br>process 3D radiographs. |
| Image annotation | Text, paint, ellipse, pointer,<br>select, draw, magnify, line,<br>rectangle, polygon, ruler,<br>protractor, smile library,<br>smudge, brush, redeye<br>reduction, select region, copy /<br>paste | Select Mark, Hollow<br>Rectangle, Straight Line.<br>Attach a Note, Freehand Line.<br>Hollow Ellipse, Hollow<br>Polygon, Polyline, Length<br>Measurement, Filled<br>rectangle, Highlighter, text,<br>text Stamp, Arrow, Filled<br>Ellipse, Filled Polygon,<br>Hide/Show Marks, Angle |
# Substantial Equivalence:
The subject device and predicate device are substantially equivalent in terms of 2D image
{4}------------------------------------------------
diagnostic analysis, having the similar indications for use and functionalities like functions to sort and save DICOM files and 2D image files from different modality [Panorama, cephalometric, Both are compatible with similar operation software and offer similar image viewing, intraoral]. annotation and simulation features. Both EzDent-i, the proposed device, and VixWin Pro, the predicate device are categorized in product code LLZ; equivalence between these models is evident.
EzDent-i radiographic imaging viewer is similar to the predicate device and the proposed device does not raise any new or potential safety risks to the user or patient and questions of safety or The proposed device is equivalent in performance to existing legally marketed effectiveness. devices.
#### Technological Characteristics:
EzDent-i is a software device that does not contact the patient, nor does it control any life sustaining devices. Results produced by the software's diagnostic, treatment planning and simulation tools are dependent on the interpretation of trained and licensed radiologists, clinicians and referring physicians as an adjunctive to standard radiology practices for diagnosis. A physician, providing ample opportunity for competent human interprets images and information being presented.
#### Nonclinical Testing:
The complete system configuration has been assessed and tested by the manufacturer and passed all in-house testing criteria. The software validation test was designed to evaluate all input functions, output functions, and actions performed by EzDent-i. Each operational mode and the process followed are documented in the Software Validation Report.
The validation testing verified and validated the risk analysis and individual performance results were within the predetermined acceptance criteria.
#### Safety and Performance Data:
- IEC 62304 Medical device software - Software life-cycle processes : 2006
- ISO 14971 Medical Devices -- Application of risk management to medical device : 2007
#### Conclusion:
{5}------------------------------------------------
The premarket notification for Ez.Dent-i contains adequate information and data to determine substantial equivalence to the predicate device and predicate device are substantially equivalent in the areas of technical characteristics, general function, application, The new device does not introduce a fundamentally new scientific and intended use. technology and the nonclinical tests demonstrate that the device is safe and effective. Therefore, it is our opinion that the EzDent-i described in this submission is substantially equivalent to the predicate device.
END
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2013
Ewoo Soft Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 12946 Kimberley Lane HOUSTON TX 77079
Re: K131594
Trade/Device Name: EzDent-i /E2/ProraView Regulation Number: 21 CFR 892,2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 09, 2013 Received: August 13, 2013
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must of any reath all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2-Mr. Kim
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131594
EzDent-i / E2 / Prora View Device Name:
Indications for Use:
EzDent-i is dental imaging software that is intended to provide diagnostic tools for maxillofacial radiographic imaging. These tools are available to view and interpret a series of DICOM compliant dental radiology images and are meant to be used by trained medical professionals such as radiologist and dentist.
EzDent-i is intended for use as software to acquire, view and save 2D image files, load DICOM project files from panorama, cephalometric, and intra-oral imaging equipment.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
\$\left(y_{m} \right)\$
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
K131594 510(k)
Page 1 of 1 __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.