K140045 · Jawon Medical Co., Ltd. · MNW · Feb 18, 2015 · Cardiovascular
Device Facts
Record ID
K140045
Device Name
BODY COMPOSITION ANALYZER
Applicant
Jawon Medical Co., Ltd.
Product Code
MNW · Cardiovascular
Decision Date
Feb 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
These devices calculates and estimates body composition parameters by using measured weight and impedance and entered height, age and gender. It shows body composition of Weight, Standard weight, Lean Body Mass, Total Body Water, Intra Cellular Water, Extra Cellular Water, B.M.I. (Body Mass Index), Mass of Body Fat, Percent of Body Fat, Basal Metabolic Rate, Impedance, Body Type, Target to Control, Goal setter. In case of whole body, it shows more items including Waist to Hip Ratio, Ratio of E.C.W. /T.B.W., Segmental analysis (lean body mass of arms, legs, and trunk). These devices can only estimate the body composition parameters with the exception of Weight, BMI, and impedance. These devices are only for use with generally healthy individuals of a specified age range.
Device Story
Jawon Body Composition Analyzers (X-CONTACT 350, 350M, 350S) are bioelectrical impedance analysis (BIA) devices. They measure weight and electrical impedance via electrodes while users input height, age, and gender. The device processes these inputs to estimate body composition metrics, including lean body mass, total body water, intracellular/extracellular water, BMI, body fat mass/percentage, and basal metabolic rate. Whole-body models provide additional segmental analysis (arms, legs, trunk) and waist-to-hip ratios. Used in clinical or home settings by generally healthy individuals. Output is displayed to the user to assist in monitoring body composition and setting health goals. The device provides estimates only; weight, BMI, and raw impedance are direct measurements.
Clinical Evidence
Performance testing was conducted to compare the subject devices against predicate devices. Results demonstrated no significant difference in measurements. The devices meet IEC 60601-1 and IEC 60601-1-2 safety and electromagnetic compatibility standards.
Technological Characteristics
Bioelectrical impedance analysis (BIA) device. Measures weight and electrical impedance. Provides estimates for body composition parameters. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for generally healthy individuals to estimate body composition parameters including weight, lean body mass, body water, BMI, body fat, and basal metabolic rate using impedance and weight measurements.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 18, 2015
Jawon Medical Co., Ltd. Won Hee Park President 1208-12, Sinsang-ri, Jinryang-eup Kyungsan-city Kyungsang-Bukdo 712-830 Republic of Korea
Re: K140045
> Trade/Device Name: Jawon Body Composition Analyzers, Model X-CONTACT 350; X-CONTACT 350M; X-CONTACT 350S Regulation Number: 21 CFR 870.2770 Regulation Name: Analyzer, body composition Regulatory Class: II Product Code: MNW Dated: November 7, 2014 Received: November 20, 2014
Dear Won Hee Park,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known): K140045
## Device Name: Jawon Body Composition Analyzers Model X-CONTACT 350; X-CONTACT 350M; X-CONTACT 350S
Indications For Use:
These devices calculates and estimates body composition parameters by using measured weight and impedance and entered height, age and gender. It shows body composition of Weight, Standard weight, Lean Body Mass, Total Body Water, Intra Cellular Water, Extra Cellular Water, B.M.I. (Body Mass Index), Mass of Body Fat, Percent of Body Fat, Basal Metabolic Rate, Impedance, Body Type, Target to Control, Goal setter.
In case of whole body, it shows more items including Waist to Hip Ratio, Ratio of E.C.W. /T.B.W., Segmental analysis (lean body mass of arms, legs, and trunk).
These devices can only estimate the body composition parameters with the exception of Weight, BMI, and impedance. These devices are only for use with generally healthy individuals of a specified age range.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
{4}------------------------------------------------
#1208-12 Sinsang-ri Jinryang-eup Kyungsan-City. Kyungsan-Bukdo712-830 Republic of Korea Tel-82-53-856—0993 Fax 82-53-856-0995
Doc. No. 4 510(K) Summary [K140045]
[As required by 21 CFR 807.92]
Date Prepared: 02/18/15
| Submitter: | Jawon Medical Co. LTD.<br>#1208-12 Sinsang-ri Jinryang-eup Kyungsan-City.<br>Kyungsan-Bukdo712-830 Republic of Korea<br>Tel-82-53-856-0993 Fax 82-53-856-0995<br>Establishment Registration Number 9616164 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | MEDMONTS Co. Ltd. (W.S. Park)<br>Life-officetel 320, 40 63 ro, Youngdeungpo-gu, Seoul 150-731, |
| Trade Name: | Body Composition Analyzer<br>Model X-Contact 350, X-Contact 350M, X-Contact 350S |
| Common Name: | Body Fat Analyzer/Body Fat Monitor/Body Composition Monitor |
| Classification Name: | Body Composition Analyzer (Impedance plethysmograph) /MNW |
| Predicate Device: | Body Composition Analyzer Models ioi 353{K092431) and EasyBody<br>202(K072325) |
X-Contact 350, X-Contact 350M, X-Contact 350S A list of 510(k) Devices To Be Cleared
| No. | Model Name | Indication for Use |
|-----|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | X-Contact 350 | Whole Body Composition Analyzer<br>These devices calculates and estimates body composition parameters by using<br>measured weight and impedance and entered height, age and gender. It shows body<br>composition of Weight, Standard weight, Lean Body Mass, Total Body Water, Intra<br>Cellular Water, Extra Cellular Water, B.M.I. (Body Mass Index), Mass of Body Fat,<br>Percent of Body Fat, Basal Metabolic Rate, Impedance, Body Type, Target to Control,<br>Goal setter.<br><br>In case of whole body, it shows more items including Waist to Hip Ratio, Ratio of<br>E.C.W. /T.B.W., Segmental analysis (lean body mass of arms, legs, and trunk).<br><br>These devices can only estimate the body composition parameters with the exception<br>of Weight, BMI, and impedance. These devices are only for use with generally healthy<br>individuals of a specified age range. |
| 2 | X-Contact 350M | Upper Body Composition Analyzer<br>These devices calculates and estimates body composition parameters by using |
{5}------------------------------------------------
| 3 | X-Contact 350S | Lower Body Composition Analyzer<br>These devices calculates and estimates body composition parameters by using<br>measured weight and impedance and entered height, age and gender. It shows body<br>composition of Weight, Standard weight, Lean Body Mass, Total Body Water, Intra<br>Cellular Water, Extra Cellular Water, B.M.I. (Body Mass Index), Mass of Body Fat,<br>Percent of Body Fat, Basal Metabolic Rate, Impedance, Body Type, Target to Control,<br>Goal setter.<br>In case of whole body, it shows more items including Waist to Hip Ratio, Ratio of<br>E.C.W. /T.B.W., Segmental analysis (lean body mass of arms, legs, and trunk).<br>These devices can only estimate the body composition parameters with the exception<br>of Weight, BMI, and impedance. These devices are only for use with generally healthy<br>individuals of a specified age range. |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technologic<br>Characteristics | These subject devices X-CONTACT 350, X-CONTACT 350M, X-CONTACT<br>350S have same intended use and technology characteristics as predicate<br>devices. See enclosed document number 9 (Substantial Equivalence<br>Discussion). The differences in this submission do not raise new questions<br>concerning either their safety or effectiveness. | |
| Non-Clinical and<br>Clinical Tests | These subject devices X-CONTACT 350, X-CONTACT 350M, and X-CONTACT<br>350S meets the requirements of IEC 60601-1, IEC 60601-1-2 and in-house test<br>criteria. The results of performance tests using X-Contact 350 demonstrated that<br>there is no significant difference in measurements. See the enclosed document<br>number 15 (Performance Test-Clinical) | |
| Conclusion: | Based on non-clinical and clinical tests, the modified devices X-CONTACT 350,<br>X-CONTACT 350M, and X-CONTACT 350S, are as safe, as effective, and<br>perform as predicate devices perform. (See document number 9-Substantial<br>Equivalents Discussion). Accordingly the subject devices X-CONTACT 350, X-<br>CONTACT 350M, and X-CONTACT 350S, are substantially equivalent to the<br>predicate devices | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.