K072325 · Jawon Medical Co., Ltd. · MNW · Nov 30, 2007 · Cardiovascular
Device Facts
Record ID
K072325
Device Name
BODY COMPOSITION ANALYZER, EASYBODY 202, 203, 205
Applicant
Jawon Medical Co., Ltd.
Product Code
MNW · Cardiovascular
Decision Date
Nov 30, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Body Composition Analyzer Models EasyBody 202 and EasyBody 203 are intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), LBM(Lean Body Mass), BMR(Basic Metabolic Rate), recommended daily calorie intake, daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method. Body Composition Analyzer Model EasyBody 205 is intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), LBM(Lean Body Mass), BMR(Basic Metabolic Rate), Segmental LBM, recommended daily calorie intake. daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method. The EasyBody 202, EasyBody 203 and EasyBody 205 measure the impedance and weight of the user, and are intended for use only in healthy subjects between the age of 5-89.
Device Story
Non-invasive body composition analyzers (EasyBody 202, 203, 205) measure user weight and bio-electrical impedance (BIA) at 50kHz. Devices use 4 electrodes (202/203) or 8 electrodes (205) in a tetra-polar configuration. Inputs (impedance, weight) processed via experimentally derived algorithm to estimate body fat percentage, lean body mass, BMR, and exercise recommendations. Operated by user in upright position; results displayed to user. Provides health-related metrics to assist in fitness and lifestyle management. Powered by AA batteries or AC adapter.
Clinical Evidence
Clinical comparison tests performed against predicate device ZEUS 9.9. Results demonstrate no significant difference between subject devices and predicate. Bench testing confirmed compliance with IEC 60601-1, EN 60601-1-2, and ISO 10993 standards.
Technological Characteristics
BIA-based body composition analyzer. 50kHz operating frequency. Tetra-polar electrode method (4 electrodes for 202/203; 8 electrodes for 205). Powered by 6V DC (AA batteries) or AC adapter. Measures impedance (100-950 ohms) and weight (1-150 kg). Upright patient position.
Indications for Use
Indicated for healthy subjects aged 5-89 for estimation of body composition metrics including PBF, MBF, LBM, BMR, recommended daily calorie intake, daily exercise time, and exercise recommendations via BIA method.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Body Composition Analyzer Model ZEUS 9.9 (K053556)
Submission Summary (Full Text)
{0}------------------------------------------------
K072325
Page 1 of 4
# Section 5
# 510(k) Summary
NOV 3 0 2007
[As Required by 21 CFR 807.92]
| Date Prepared: | August 03, 2007 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Jawon Medical Co., Ltd. |
| | #1208-12, Sinsang-Li, Jinryang-Myun, Kyungsan-city, Kyungsang-Bukdo, |
| | 712-830, Republic of Korea |
| | Tel: 82-53-856-0993 |
| | Establishment Registration Number: 9616164 |
| Contact Person: | MI Consulting Co., Ltd. (Mr. HL Jung) |
| | RM 624, Life Officetel, 61-3, Yoido-dong, Youngdeungpo-gu, Seoul, 150-731, |
| | Republic of Korea |
| | Tel: 82-2-6264-9001 |
| | Fax: 82-2-6269-9001 |
| | Email: jung_halyong@yahoo.com |
| Trade Name: | Body Composition Analyzer |
| | Models EasyBody 202, EasyBody 203 and EasyBody 205 |
| Common Name: | Body Fat Analyzer / Body Fat Monitor / Body Composition Monitor |
| Classification Name: | Body Composition Analyzer (Impedance plethysmograph) / MNW |
| Predicate Device: | Body Composition Analyzer Model ZEUS 9.9 (K053556) |
| Device Description: | The EasyBody 202, EasyBody 203 and EasyBody 205 are non-invasive body<br>composition analyzers intended for use only in healthy subjects between the<br>age of 5-89. The devices employ BIA(Bio-electrical Impedance Analysis)<br>method, 4 electrodes for EasyBody 202 and EasyBody 203, 8 electrodes for<br>EasyBody 205, and then measure body composition using an experimentally<br>derived algorithm. The devices are powered by four AA batteries (DC 6V) or<br>AC adapter. |
| Intended use: | Body Composition Analyzer Models EasyBody 202 and EasyBody 203 are<br>intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), |
| Jawon Medical | EasyBody |
Section 5: Page 1 of 4
{1}------------------------------------------------
LBM(Lean Body Mass), BMR(Basic Metabolic Rate), recommended daily calorie intake, daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method.
Body Composition Analyzer Model EasyBody 205 is intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), LBM(Lean Body Mass), BMR(Basic Metabolic Rate), Segmental LBM, recommended daily calorie intake. daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method.
The EasyBody 202, EasyBody 203 and EasyBody 205 measure the impedance and weight of the user, and are intended for use only in healthy subjects between the age of 5-89.
- Technologic characteristics: The devices EasyBody 202, EasyBody 203 and EasyBody 205 have been compared to the predicate device ZEUS 9.9 such as a comparing table of technology characteristics. The differences in this submission don't raise new questions concerning either safety or effectiveness.
- Non-clinical and clinical tests: The subject devices EasyBody 203 and EasyBody 203 meet the requirements of IEC 60601-1, EN 60601-1-2 and ISO 10993 series. The results of clinical comparison tests with predicate device ZEUS 9.9 demonstrate that there is no significant difference between the subjected devises and the predicate device.
- Conclusions: Based on non-clinical and clinical tests, the subject devices EasyBody 202, EasyBody 203 and EasyBody 205 are as safe, as effective, and perform as well as the predicate device ZEUS 9.9. Accordingly, the subject devices are substantially equivalent to the predicate device.
{2}------------------------------------------------
| ltems | Zeus 9.9 | EasyBody 202 | EasyBody 203 | EasyBody 205 |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K053556 | New | New | New |
| Intended Use | Body composition<br>Analyzer | Same | Same | Same |
| Indications<br>for use | Estimated of:<br>PBF(Percentage of<br>Body Fat), MBF(Mass<br>of Body Fat),<br>LBM(Lean Body<br>Mass), TBW(Total<br>Body Water,<br>BMI(Body Mass<br>Index), BMR(Basic<br>Metabolic Rate),<br>Segmental LBM,<br>ICW(Intra-Cellular<br>Water), ECW(Extra-<br>Cellular Water), and<br>Ratio of ECW/TBW<br>Actual :<br>Impedance,<br>BMI(Body Mass<br>Index), weight,<br>WHR(Waist to Hip<br>Ratio)<br>Healthy subjects of<br>the age of 7-89 | Estimated of:<br>PBF(Percentage of<br>Body Fat),<br>MBF(Mass of Body<br>Fat), LBM(Lean<br>Body Mass),<br>BMR(Basic<br>Metabolic Rate),<br>recommended<br>daily calorie intake,<br>daily exercise time,<br>sort and term of<br>exercise<br>Actual:<br>Impedance, weight<br>Healthy subjects of<br>the age of 5-89 | Estimated of:<br>PBF(Percentage of<br>Body Fat),<br>MBF(Mass of Body<br>Fat), LBM(Lean<br>Body Mass),<br>BMR(Basic<br>Metabolic Rate),<br>recommended<br>daily calorie intake,<br>daily exercise time,<br>sort and term of<br>exercise<br>Actual:<br>Impedance, weight<br>Healthy subjects of<br>the age of 5-89 | Estimated of:<br>PBF(Percentage of<br>Body Fat),<br>MBF(Mass of Body<br>Fat), LBM(Lean Body<br>Mass), BMR(Basic<br>Metabolic Rate),<br>Segmental LBM,<br>recommended<br>daily calorie intake,<br>daily exercise time,<br>sort and term of<br>exercise<br>Actual:<br>Impedance, weight<br>Healthy subjects of<br>the age of 5-89 |
| Analysis<br>Method | Bio-electrical<br>Impedance<br>Analysis (BIA) | Same | Same | Same |
| Operating<br>parameters | Frequency:<br>1,5,50,<br>250,550,1000kHz | Frequency:<br>50kHz | Frequency:<br>50kHz | Frequency:<br>50kHz |
| Electrode<br>type | Tactile | Same | Same | Same |
| Number/<br>Placement of<br>Electrodes | 8 electrodes<br>Placed on Hands and<br>feet(or ankle)<br>(Tetra-polar Electrode<br>method) | 4 electrodes<br>Placed on feet<br>(Tetra-polar<br>Electrode method) | 4 electrodes<br>Placed on feet<br>(Tetra-polar<br>Electrode method) | 8 electrodes<br>Placed on Hands<br>and feet<br>(Tetra-polar<br>Electrode method) |
| Items | Zeus 9.9 | EasyBody 202 | EasyBody 203 | EasyBody 205 |
| Patient<br>Position | Upright | Same | Same | Same |
| Impedance<br>Measuring<br>Site | Whole body,<br>Right Arm, Right Leg,<br>Left Arm, Left Leg<br>Trunk | Foot to Foot | Foot to Foot | Whole body,<br>Right Arm, Right Leg,<br>Left Arm, Left Leg,<br>Trunk |
| Impedance<br>Measurement<br>Range | 100 ~ 950 ohms | Same | Same | Same |
| Measuring<br>height | 110 ~ 200 cm | 100 ~ 200 cm | 100 ~ 200 cm | 100 ~ 200 cm |
| Measuring<br>weight | 10 ~ 250 kg | 1~150 kg | 1~150 kg | 1~150 kg |
| Measuring<br>PBF | 3 ~ 50% | Same | Same | Same |
| Measuring<br>Time | Within 1 minute | About 15sec | About 15sec | About 15sec |
| Electrical<br>current<br>applied<br>during<br>measurement | Maximum current<br>360uA | Maximum current<br>180uA | Maximum current<br>180uA | Maximum current<br>180uA |
| Power<br>consumption | 50VA | 0.36VA | 0.36VA | 0.36VA |
| Power<br>Supply | AC120V 50/60Hz | AA Battery DC6V<br>or AC Adapter | AA Battery DC6V<br>or AC Adapter | AA Battery DC6V<br>or AC Adapter |
| Operating<br>ambient | Temp: 10 ~ 40 ℃<br>RH: 30 ~ 75% | Same | Same | Same |
| Storage<br>ambient | Temp: -20 ~ 60 ℃<br>RH: lower than 95% | Same | Same | Same |
| Dimensions<br>(W x L x H) | 470 x 655 x 1220mm | 310 x 311 x 38.5mm | 300 x 356 x 79mm | 300 x 300 x 369mm |
| Weight | 45Kg | 1.7kg | 2.2kg | 2.7kg |
## < Comparison Table of technical characteristics >
Jawon Medical
EasyBody
Section 5: Page 3 of 4
ે છે. રે
{3}------------------------------------------------
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three stripes representing its wings.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### NOV 3 0 2007
Jawon Medical Co., Ltd. c/o H. L. Jung MI Consulting Co., Ltd. Room 624, Life Officetel, 61-3 Yoido-dong, Youngdeungpo-gu Seoul. 150-731 REPUBLIC OF KOREA
K072325 Re:
> Trade/Device Name: JAWON MEDICAL BODY COMPOSITION ANALYZER MODELS EasyBody 202, EasyBody 203, and EasyBody 205 Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: November 19, 2007 Received: November 21, 2007
Dear Mr. Jung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA' may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Section 4
## Indications for Use
510(k) Number (if known): _ KO72325
Device Name: JAWON MEDICAL BODY COMPOSITION ANALYZER MODELS EasyBody 202, EasyBody 203 and EasyBody 205
Indications For Use:
The EasyBody 202 and EasyBody 203 are intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), LBM(Lean Body Mass), BMR(Basic Metabolic Rate), recommended daily calorie intake, daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method.
The EasyBody 205 is intended to estimate PBF(Percentage of Body Fat), MBF(Mass of Body Fat), LBM(Lean Body Mass), BMR(Basic Metabolic Rate), Segmental LBM, recommended daily calorie intake, daily exercise time, and sort and term of exercise using the BIA(Bio-electrical Impedance Analysis) method.
The EasyBody 202, EasyBody 203 and EasyBody 205 measure the impedance and weight of the user, and are intended for use only in healthy subjects between the age of 5-89.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use
(Per 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IN NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert Lewin
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
EasyBody 202/203/205
Section 4: Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.