K140020 · Bsp-Biiological Signal Processing, Ltd. · DPS · Oct 28, 2014 · Cardiovascular
Device Facts
Record ID
K140020
Device Name
HYPERQ ACS-100 SYSTEM
Applicant
Bsp-Biiological Signal Processing, Ltd.
Product Code
DPS · Cardiovascular
Decision Date
Oct 28, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K140020 · Oct 28, 2014
HYPERQ ACS-100 SYSTEM
Bsp-Biiological Signal Processing, Ltd.
Clinical database (ECG recordings, demographic data, and gold-standard diagnostics)
Retrospective analysis of existing clinical database records to validate the diagnostic accuracy of the HyperQ Rest system in identifying Reduced Amplitude Zones (RAZ) in patients with acute chest pain.
Patients presenting with acute chest pain to an Emergency Department, aged 18 years and above
Gold standard diagnosis from the database
Identification of Reduced Amplitude Zones (RAZ) in ECG signals
Indications for Use
The BSP HyperQ™ Rest System is intended to be used during rest ECG tests, using the analysis of high frequency components present within the central portion of the QRS complex (HF-QRS). The device outputs a reduced amplitude zone (RAZ) value for the physician to interpret (based on reported clinical study results). The device is indicated for use in the ECG interpretation of patients with acute chest pain presenting to the Emergency Department, aged 18 years and above, as an additional adjunctive aid during initial patient presentation. The significance of the HF-QRS changes and interpretation of the HyperQTM Rest analysis must be determined by a physician, based on the reported clinical study results presented in the device labeling.
Device Story
System measures, processes, and displays high-frequency (1000Hz) ECG signals; analyzes high-frequency components of the QRS complex (HF-QRS). Device identifies Reduced Amplitude Zones (RAZ) and extracts High Frequency Morphological Indexes (HFMI) across 12 leads. Used in Emergency Departments by physicians to aid in clinical evaluation of acute chest pain. Output provides quantitative RAZ values and visual HFQRS envelope; physician interprets findings alongside clinical context to assist in diagnostic decision-making. Benefits include providing additional objective data during initial patient presentation to support clinical assessment.
Clinical Evidence
Clinical study analyzed high-frequency ECG signals to validate diagnostic accuracy in identifying RAZ in patients aged 18+ with acute chest pain. Analysis compared device output to gold-standard diagnostics from a database. Results demonstrated the device serves as an effective aid in the diagnostic process.
Technological Characteristics
Analyzes HF-QRS components of ECG signals via 1000Hz sampling rate. Extracts HFMI and identifies RAZ. Hardware identical to HyperQ AD-100; complies with IEC 60601-1 and IEC 60601-1-2. No patient contact materials.
Indications for Use
Indicated for patients aged 18+ presenting to the Emergency Department with acute chest pain; used as an adjunctive aid for ECG interpretation.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
Predicate Devices
GE Medical Systems Information Technologies MAC 800 Resting ECG Analysis System (K090212)
{0}------------------------------------------------
Public Health Service
WO66-G609
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -
Silver Spring, MD 20993-0002
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
October 28, 2014
Biological Signal Processing, Ltd. % Ahava Stein A. Stein Regulatory Affairs Consulting Ltd. 20 Hata' As Str., Suite 102 Kfar Saba, 44425 Israel
Re: K140020 Trade/Device Name: HyperQ Rest System Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph; Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) Regulatory Class: Class II Product Code: DPS, MWI Dated: August 3, 2014 Received: August 8, 2014
Dear Ahava Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140020
Device Name HyperQ Rest
#### Indications for Use (Describe)
The BSP HyperO™ Rest System is intended to be used during rest ECG tests, using the analysis of high frequency components present within the central portion of the QRS complex (HF-QRS). The device outputs a reduced amplitude zone (RAZ) value for the physician to interpret (based on reported clinical study results). The device is indicated for use in the ECG interpretation of patients with acute chest pain presenting to the Emergency Department, aged 18 years and above, as an additional adjunctive aid during initial patient presentation. The significance of the HF-QRS changes and interpretation of the HyperQTM Rest analysis must be determined by a physician, based on the reported clinical study results presented in the device labeling.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY BSP HYPERQ™ ACS-100 SYSTEM
## 510(k) Number K140020
### Applicant Name:
| Company Name: | Biological Signal Processing Ltd. |
|---------------|-----------------------------------|
| Address: | 22a Raoul Wallenberg Street |
| | Tel Aviv 69719 |
| | Israel |
| | Tel: +972-3-647-4840 |
| | Fax: +972-3-647-1498 |
| | E-mail: yair@bsp.co.il |
### Contact Person:
| Official Correspondent: | Ahava Stein |
|-------------------------|---------------------------------------------------------|
| Name: | A. Stein - Regulatory Affairs Consulting Ltd. |
| Address: | 20 Hata'as Str., Suite 102<br>Kfar Saba 44425<br>Israel |
| | Tel: +972-9-7670002 |
| | Fax: +972-9-7668534 |
| | E-mail: ahava@asteinrac.com |
| Date Prepared: | October 12, 2014 |
| Trade Name: | BSP HyperQTM Rest System |
| Classification Name: | CFR Classification section 870.2340; (Product code DPS) |
| Classification: | Class II Medical Device |
## Predicate Device:
The BSP HyperQ™ Rest System is substantially equivalent to the following predicate devices.
| Manufacturer | Device | 510(k) No. |
|------------------------------------------------|-------------------------------------|------------|
| GE Medical Systems<br>Information Technologies | MAC 800 Resting ECG Analysis System | K090212 |
| BSP Ltd. | HyperQ™ AD-100 System | K102579 |
{4}------------------------------------------------
### Device Description:
The HyperQTM Rest System measures, processes, stores, and displays information derived from an electrocardiogram (ECG) with a high sampling rate (1000Hz). The device analyzes and records the High Frequency components of the QRS complex of standard ECG (HFQRS). The device displays the HFQRS to the user and also summarizes the analysis to generate an indication whether observed changes are within normal HyperO range, or constitute significant HyperO changes (Reduced Amplitude Zones, a.k.a RAZ); or represent a non-diagnostic test (unreadable). These findings are used as an aid in the clinical evaluation of acute chest pain on initial presentation to the Emergency Department.
### Intended Use/Indication for Use:
The BSP HyperQTM Rest System is intended to be used during rest ECG tests, using the analysis of high frequency components present within the central portion of the QRS complex (HF-QRS). The device outputs a reduced amplitude zone (RAZ) value for the physician to interpret (based on reported clinical study results). The device is indicated for use in the ECG interpretation of patients with acute chest pain presenting to the Emergency Department, aged 18 years and above, as an additional adjunctive aid during initial patient presentation. The significance of the HF-QRS changes and interpretation of the HyperQTM Rest analysis must be determined by a physician, based on the reported clinical study results presented in the device labeling.
#### Performance Standards:
The BSP HyperQTM Rest System is identical in hardware to the BSP HyperQTM AD-100, which was cleared under K102579. The Rest therefore maintains compliance with the recognized consensus standards previously described for the AD-100 device.
#### Non-Clinical (Bench) Performance Data:
Not applicable.
# Pre-Clinical (Animal Study) Performance Data:
Not Applicable
{5}------------------------------------------------
#### Clinical Performance Data:
The effectiveness of the BSP HyperQTM Rest system was demonstrated in a clinical study. High frequency ECG signals were analyzed to obtain an automatic interpretation indicating significant HyperQ changes or changes within normal range for all tests that were readable. The analysis was based solely on the digital data from the ECG device and was compared to a gold standard diagnosis from the database.
Demographic data, ECG recordings and gold-standard diagnostics were used as a source for validating the diagnostic accuracy of the system. Results demonstrated that the device can be used as an aid in the identification of Reduced Amplitude Zones (RAZ) values in patients presenting with acute chest pain to an Emergency Department, aged 18 years and above.
### Technological Characteristics:
The technology of the BSP devices involves analysis of High Frequency QRS (HFQRS) components of the ECG signal (HFQRS), made possible by the high frequency sampling rate. The high frequency range is represented by the HyperQ™ signal. The morphology of the HyperO envelope is examined for existence of reduced amplitude zones (RAZ) and High Frequency Morphological Indexes (HFMI) are extracted. The level of RAZ of all 12 leads is quantified by the HFMI, providing the user with a quantitative value.
## Substantial Equivalence:
The indications for use and technological characteristics of the BSP HyperQ™ Rest device are substantially equivalent to the indications for use and technological characteristics of the Predicate GE MAC 800 Resting ECG Analysis System and the HyperQ AD-100 System.
The primary difference between the HyperQ AD-100 and the HyperQ Rest is in the indication for use for rest ECG analysis from the acquired ECG. The predicate device GE MAC 800 Resting ECG Analysis System (K090212) is specifically indicated for
#### K140020
{6}------------------------------------------------
rest ECG Testing. Based upon the validation results, BSP believes this change does not raise additional safety or efficacy concerns.
The design and components in the BSP HyperQ™ Rest system, including the hardware and software, are similar to the design and components found in the predicate HyperQ™ AD-100 System. The algorithm was modified to allow for analysis of rest ECG signals. The minor differences in the software user interface are insignificant and do not influence the safety or efficacy of the device. The performance specifications (including display of High Frequency QRS signals, and morphological assessment of them) of the BSP HyperQ™ Rest system are substantially equivalent to those in the HyperQ™ AD-100 System. There are no patient contact materials. Any minor differences in the technological characteristics do not raise new safety or effectiveness concerns. Furthermore, the new BSP HyperQTM Rest System is identical in its hardware to the HyperQ™ AD-100 System. The AD-100 system underwent performance testing, including electrical and mechanical safety testing according to IEC 60601-1 and electromagnetic compatibility testing according to IEC 60601-1-2. These were provided in the original 510(k) submission of the HyperQ™ AD-100 System, and are not repeated in this submission. As the software algorithm was modified, software validation testing was performed. The results of the software validation tests demonstrated that the device meets its specifications, which are similar to those of the predicate device. These performance tests demonstrated that the minor differences in the device specifications meet the system requirements and do not raise new safety or effectiveness concerns. Furthermore, the results of the clinical study performed demonstrate that the device provides an effective tool as an aid in the diagnostic process of acute chest pain in patients aged 18 years and above, for the identification of RAZ in the ECG signal.
#### Conclusions:
Based on the performance testing and comparison to predicate devices, the BSP HyperQTM Rest System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.