ORIGIO® Sequential Fert™ is for the fertilization of oocytes in vitro. ORIGIO® Sequential Cleav™ is for the culture of embryos until the 2-8 cell stage. ORIGIO® Sequential Cleav™ can also be used for embryo transfer at day 2 or 3.
Device Story
ORIGIO® Sequential Fert™ and Cleav™ are ready-to-use, aseptically filtered, non-viscous culture media for assisted reproductive technology (ART). Fert™ supports in vitro fertilization of oocytes; Cleav™ supports embryo culture to the 2-8 cell stage and subsequent transfer. Used in clinical laboratories by embryologists/technicians. Media contain physiological salts, amino acids, energy substrates (glucose, pyruvate, lactate), vitamins (folic acid, calcium pantothenate), and protein (HSA). Cleav™ also contains sodium hyaluronate. These components provide necessary nutrients for specific embryo development stages. Output is a chemical environment for gamete/embryo maintenance. Benefits include optimized conditions for fertilization and embryo development prior to implantation.
Clinical Evidence
Bench testing only. Evidence includes product specification comparisons (pH, osmolality, endotoxin, MEA), stability studies (36-week shelf life), and biocompatibility testing (ISO 10993-1:2009) for Cleav™. No clinical data presented.
Technological Characteristics
Aseptically filtered, non-viscous liquid media. Components: physiological salts, HSA, gentamicin sulphate, vitamins (folic acid, calcium pantothenate), amino acids, energy substrates (glucose, lactate, pyruvate), sodium bicarbonate, citrate, and sodium hyaluronate (Cleav™ only). Packaged in 10 mL or 60 mL PETG bottles with HDPE closures. Non-electronic, chemical-based device.
Indications for Use
Indicated for in vitro fertilization of human oocytes (Fert™) and culture of human embryos until the 2-8 cell stage, including embryo transfer at day 2 or 3 (Cleav™).
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
Cook IVF Fertilization Medium and Cook IVF Cleavage Medium (K002385)
{0}------------------------------------------------
# a CooperSurgical Company
K133912 Page 1 of 5
# K133912
# MAY 1 4 2014
# 510(K) SUMMARY
# Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination
| Submitted by: | ORIGIO a/s<br>Knardrupvej 2,<br>2760 Måløv,<br>Denmark |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Tove Kjær<br>Director Corporate Regulatory Affairs<br>ORIGIO a/s |
| | Phone: +45 4679 0220<br>Fax: +45 4679 0300 |
| | Date Submitted: May 12, 2014 |
| Device Identification<br>Trade name: | ORIGIO® Sequential FertTM<br>ORIGIO® Sequential FertTM with phenol red<br>ORIGIO® Sequential CleavTM<br>ORIGIO® Sequential CleavTM with phenol red |
| Common name: | ORIGIO® Sequential FertTM<br>ORIGIO® Sequential CleavTM |
| Classification name: | Reproductive media and supplements (21 CFR 884.6180, Product<br>Code MQL) |
| Predicate devices:<br>Cook Medical: | Cook IVF Fertilization Medium and Cook IVF Cleavage Medium<br>(K002385). |
# Description
ORIGIO® Sequential Fert™ is intended for in vitro fertilization of human oocytes.
Two versions of ORIGIO® Sequential Fert™ are available:
- Catalogue no. 8301: ORIGIO® Sequential Fert™ .
- Catalogue no. 8302: ORIGIO® Sequential Fert™ with phenol red .
ORIGIO® Sequential Cleav™ is intended for in vitro culture of human embryos until the 2-8 cell stage. The medium can also be used for transfer.
Two versions of ORIGIO® Sequential Cleav™ are available:
{1}------------------------------------------------
- Catalogue no. 8303: ORIGIO® Sequential Cleav™
- Catalogue no. 8304: ORIGIO® Sequential Cleav™ with phenol red
Both ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ are aseptically filtered, non viscous solutions, light pink or colorless solutions, which are ready to use by professionals within assisted reproduction.
ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ are contained in 10 mL or 60 mL transparent polyethylene terephthalate glycol (PETG) bottles with high density polyethylene (HDPE) closures, available in card board boxes of 1 x 10 mL and 1 x 60 mL bottles. The bottles and boxes are individually labeled. The boxes also contain instruction for use provided as package insert.
# Indication for use
ORIGIO® Sequential Fert™ is for the fertilization of oocytes in vitro.
ORIGIO® Sequential Cleav™ is for the culture of embryos until the 2-8 cell stage. ORIGIO® Sequential Cleav™ can also be used for embryo transfer at day 2 or 3.
# Technological characteristics
The design of ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ as well as the predicates listed in this submission is based on each medium containing the appropriate nutrients for the embryo development stage it is intended for. Table 1 compares the technological characteristics of ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ to the predicates Cook IVF Fertilization Medium and Cook IVF Cleavage Medium. Both similarities and differences are illustrated.
ORIGIO® Sequential Fert™ is for the fertilization of oocvtes in vitro. Cook IVF Fertilization Medium is intended for use during in vitro fertilization procedures for insemination and incubation of ocytes. The in vitro fertilization procedure includes insemination and incubation of the oocytes. Thus, the indication for use for ORIGIO® Sequential Fert™ is considered comparable to the predicate and the differences are not considered to represent a new intended use nor do they pose any safety or effectiveness issues.
ORIGIO® Sequential Cleav™ is for culture and transfer of embryos as the predicate Cook IVF Cleavage Medium. Thus, the intended use of ORIGIO® Sequential Cleav™ is considered identical to the predicate.
| Product | ORIGIO®<br>Sequential Fert™ | Cook IVF<br>Fertilization<br>Medium | ORIGIO®<br>Sequential Cleav™ | Cook IVF Cleavage<br>Medium |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for use | ORIGIO® Sequential<br>Fert™ is for the<br>fertilization of<br>oocytes in vitro. | Cook IVF<br>Fertilization Medium<br>is intended for use<br>during in vitro<br>fertilization<br>procedures for<br>insemination and<br>incubation of<br>oocytes. | ORIGIO® Sequential<br>Cleav™ is for the<br>culture of embryos<br>until the 2-8 cell<br>stage.<br>ORIGIO® Sequential<br>Cleav™ can also be<br>used for embryo<br>transfer at day 2 or<br>3. | Cook IVF Cleavage<br>Medium is intended<br>for use during in<br>vitro fertilization<br>procedures for<br>culture and transfer<br>of embryos. |
| Product<br>specification | | | | |
| Product | ORIGIO®<br>Sequential Fert™ | Cook IVF<br>Fertilization<br>Medium | ORIGIO®<br>Sequential Cleav™ | Cook IVF Cleavage<br>Medium |
| pH | 7.3-7.5 | 7.5-7.8 (in air)<br>7.3-7.5 (in 6% CO₂) | 7.2-7.4 | 7.5-7.8 (in air)<br>7.3-7.5 (in 6% CO₂) |
| Osmolality<br>(mOsm/kg) | 277-293 | 285-295 | 272-288 | 285-295 |
| Endotoxin (EU/mL) | <0.15 | <0.4 | <0.15 | <0.4 |
| Aseptically filtered | X | X | X | X |
| 1-cell MEA | ≥80% | ≥80% | ≥80% | ≥80% |
| Formulation | | | | |
| Physiological salts | Sodium chloride<br>Potassium chloride<br>Sodium dihydrogen<br>phosphate<br>Magnesium sulphate<br>Calcium chloride | X | Sodium chloride<br>Potassium chloride<br>Sodium dihydrogen<br>phosphate<br>Magnesium sulphate | X |
| Protein source | | | | |
| HSA Protein<br>Supplementation | 5 mg/mL | X | 5 mg/mL | X |
| Drugs | | | | |
| Gentamicin<br>sulphate | 0.01 mg/mL | X | 0.01 mg/mL | X |
| Vitamins | | | | |
| Calcium<br>pantothenate | X | | X | |
| Folic acid | X | | X | |
| Amino acids | | | | |
| Non essential<br>amino acids | X | X | X | X |
| Essential amino<br>acids | | | X | |
| Energy substrates | | | | |
| D-(+)-Glucose | X | X | X | |
| Calcium lactate | X | X | X | X |
| Sodium pyruvate | X | X | X | X |
| Buffering system | | | | |
| Sodium<br>bicarbonate | X | X | X | X |
| Other | | | | |
| Tri sodium citrate<br>dihydrate | X | | X | |
| Sodium<br>hyaluronate | | | X | |
# Table 1, Comparison of ORIGIO® Sequential Fert™ and ORIGIO® Sequential Fert™ with the predicates
{2}------------------------------------------------
The technological characteristics of ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ are comparable to those of the predicate device. The main differences are:
- Vitamins: ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ contain the . vitamins folic acid and calcium pantothenate in similar concentration ranges as added in other cleared ART media e.g. EmbryoAssist™ (K080473) and EmbryoGen® (K120136),
{3}------------------------------------------------
which both have fertilization of oocytes, culture of embryos until the 2-8 cell stage and embryo transfer as the indication for use. Thus, folic acid and calcium pantothenate have a history of use in other ART media and do not represent a new technology.
- Essential amino acids: ORIGIO® Sequential Cleav™ contains both essential and non . essential amino acids, whereas the predicate for this product. Cook IVF Cleavage Medium, only contains non essential amino acids. Essential amino acids are well known components of cleared ART media with indication for use for fertilization, culture of embryos until the 2-8 cell stage and embryo transfer e.g. ISM1™ (K030490) and SAGE 1-Step™ (K133707) and thus, do not represent a new technology.
- Glucose: ORIGIO® Sequential Fert™ contains glucose which is also present in the predicate Cook IVF Fertilization Medium. ORIGIO® Sequential Cleav™ contains a low level of glucose which is not present in the predicate but is found in a comparable concentration in other cleared ART media with the same indication for use. culture of embryos until the 2-8 cell stage and embryo transfer, e.q. EmbryoAssist™ (K080473). Thus, the addition of glucose has a history of use in cleavage stage culture media and does not represent a new technology.
- Citrate: ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ contain citrate which is a well known component in ART media and has a history of use in media intended for fertilization, culture of embryos until the 2-8 cell stage and embryo transfer e.g. Universal IVF Medium (K991279) and EmbryoAssist™ (K080473). Thus, the addition of citrate does not represent a new technology.
- Sodium hyaluronate: ORIGIO® Sequential Cleav™ contains hyaluronate in a concentration . range similar to that added in other cleared ART media products e.g. SAGE 1-Step™ (K133707) which is indicated for use at the same developmental stage and for embryo transfer. Thus, the addition of sodium hyaluronate does not represent a new technology.
The differences in composition do not impact the substantial equivalence and do not raise any new types of safety or effectiveness concern.
### Performance data
The product specifications for ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ and the predicates are similar regarding sterility, pH, and Mouse Embryo Assay (MEA) test. The endotoxin level for ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ is lower (<0.15 EU/mL) than for the predicates (<0.4 EU/mL) and thus do not raise any safety concerns.
Regarding the osmolality, the specification limit is a bit wider for ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ (277-293 and 272-288 mOsm/kg, respectively) than for the predicates (285-295 mOsm/kq). However, the specification for ORIGIO® Sequential Fert™ is identical to the osmolality specification for Universal IVF Medium (K991279) which is also intended for fertilization of oocvtes, whereas the specification for ORIGIO® Sequential Cleav™ is the same as the one for EmbryoAssist™ (K080473) with the indication for use for culture of embryos until the 2-8 cell stage. Thus, the osmolality specifications for ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ do not raise any safety concerns.
The shelf life of ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ has been validated in stability studies to 36 weeks. The parameters which have been tested in the stability studies through shelf life includes pH, osmolality, endotoxin, HSA concentration, MEA, and sterility.
In general, ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ are subject to the same control methods and, to a significant degree, contain the same components as the predicate devices. ORIGIO® Sequential Fert™ and ORIGIO® Sequential Cleav™ have similar handling procedures and storage conditions. Therefore, ORIGIO® Sequential Fert™ and ORIGIO®
{4}------------------------------------------------
Sequential Cleav™ are considered substantially equivalent to the predicate devices Cook IVF Fertilization Medium and Cook IVF Cleavage Medium (K002385).
# Biocompatibility
ORIGIO® Sequential Fert™ will not come in contact with the body during use and therefore does not require biocompatibility testing to support its intended use.
ORIGIO® Sequential Cleav™ may be used for transferring embryos to the patient's uterus following culture. Biocompatibility testing per ISO 10993-1:2009 for a device with short duration contact with mucosal tissues (i.e., cytotoxicity, sensitization, and irritation) was conducted and . showed device materials to be biocompatible, supporting safety for use in embryo transfer procedures.
#### Conclusion
The conclusion from the performance and safety data, intended use comparison, product formulation comparison and test specification comparison, demonstrates that ORIGIO® Sequential Fert™ (with and without phenol red) and ORIGIO® Sequential Cleav™ (with and without phenol red) are suitable for their intended use, and meet the criteria in the comparison to the predicate devices (Cook IVF Fertilization Medium and Cook IVF Cleavage Medium, K002385) in which substantial equivalence has been demonstrated.
{5}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 14, 2014
ORIGIO a/s Tove Kjaer Director Corporate Regulatory Affairs Knardrupvej2 Måløv 2760 Denmark
Re: K133912
> Trade/Device Name: ORIGIO® Sequential Fert™ ORIGIO® Sequential Fert™ with phenol red ORIGIO® Sequential CleayTM ORIGIO® Sequential Cleav™ with phenol red Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: MQL Dated: April 8, 2014 Received: April 11, 2014
Dear Tove Kjacr,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Tove Kjaer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/6/Picture/7 description: The image shows the text "Benjamin Asher-S". The text is written in a bold, sans-serif font. The letters "FDA" are stylized with a pattern of lines and shapes. The text is arranged horizontally, with "Benjamin" on the left, "FDA" in the middle, and "Asher-S" on the right.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K133912
Device Name
ORIGIO® Sequential Fert™ and ORIGIO® Sequential Feri™ with phenol red
Indications for Use (Describe)
ORIGIO® Sequential Fert™ is for the fertilization of oocytes in vitro.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
ーズ : 【 : 【 : 【 : 】 : 】 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (9/13)
156 Productions Services (101) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of Information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (9/13)
{9}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K133912
### Device Name
ORIGIO& Sequential Cleavi™ and ORIGIO® Sequential Cleav™ with phenol red
# Indications for Use (Describe)
ORIGIO® Sequential Cleav™ is for the culture of embryos until the 2-8 cell stage. ORIGIOR Sequential Cleav™ can also be used for embryo transfer at day 2 or 3.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FOR FOR FOR USE ONLY 2 2 2 Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (9/13)
"SC" Mublisheng Scrivace (303) +43-6740
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{10}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, Including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
*An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.