AMERIWATER CENTURION 1500+ REVERSE OSMOSIS SYSTEM WITH HEAT SANITIZATION
Applicant
Ameriwater
Product Code
FIP · Gastroenterology, Urology
Decision Date
Apr 16, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
The AmeriWater Centurion 1500+ Reverse Osmosis Systems are water treatment systems intended for use in hemodialysis applications. They are designed to pre-treat and purify potable water for use in making dialysate for hemodialysis and to meet current AAM and Federal (U.S.) standards. The device is intended to be a component in a complete water purification system, and is not a complete water treatment system. It must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well to meet current AAM and Federal (U.S.) standards. The Centurion 1500+ Reverse Osmosis System is intended for use in a hospital, clinic, dialysis center, or for home care for single patient use. The device includes an integrated heat sanitization process.
Device Story
The AmeriWater Centurion 1500+ is a reverse osmosis (RO) water treatment system for hemodialysis. It processes potable water by applying pressure across a semi-permeable polyamide membrane to remove organic and inorganic contaminants, producing purified permeate for dialysate preparation. The system includes an inlet solenoid valve, feed water tank, and high-pressure pump. It features an integrated, manually activated heat sanitization function to manage bacterial content. Used in hospitals, clinics, dialysis centers, and home care environments, the device is operated by healthcare staff or patients. It functions as a component within a larger water purification system, requiring external pre-treatment and potentially post-treatment stages to ensure compliance with AAMI and federal standards. The system provides purified water, helping ensure the safety and quality of dialysate used in patient treatment.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including performance verification of the RO function, microbiological efficacy testing of the heat sanitization process, biocompatibility testing, software validation, and electrical safety testing (IEC 60601-1).
Technological Characteristics
Materials: Nylon, Polypropylene, Stainless Steel, EPDM, Polyamide (TFC membrane). Principle: Reverse osmosis via semi-permeable membrane. Energy: 115V, 60 Hz, 20 Amp. Form factor: Water treatment system component. Connectivity: Not specified. Sanitization: Integrated thermal/heat sanitization. Software: Embedded controller for system operation and heat sanitization management.
Indications for Use
Indicated for hemodialysis patients in hospitals, clinics, dialysis centers, or home care settings. Used to pre-treat and purify potable water for dialysate preparation. Must be used as part of a larger water purification system with pre- and post-treatment components.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
AmeriWater MediQA Reverse Osmosis System (K131904)
AmeriWater MRO Portable Reverse Osmosis System (K111740)
Submission Summary (Full Text)
{0}------------------------------------------------
000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
1303 STANLEY AVENUE DAYTON, OH 45404 USA TEL 937-461-8833, 800-535-5585 FAX 937-461-1988 www.AMERIWATER.com
February 27, 2014
## 510(K) SUMMARY
| 510(K) Number: | K133783 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | AmeriWater |
| Contact: | Brian R. Bowman, Quality & Regulatory Administrator<br>1303 Stanley Avenue, Dayton, OH 45404<br>Phone: (937)461-8833 Fax: (937)461-1988<br>brianbowman@ameriwater.com |
| Proprietary Name: | AmeriWater Centurion 1500+ Reverse Osmosis System |
| Common Name: | Reverse Osmosis System |
| Classification Name: | Water purification system for hemodialysis |
| Classification: | Class II Medical Device under §876.5665<br>Panel: Gastroenterology<br>Product Code: FIP |
| Equivalent Devices: | K131904, AmeriWater MediQA Reverse Osmosis System |
K111740, AmeriWater MRO Portable Reverse Osmosis System MROS Device Description: The AmeriWater Centurion 1500+ Reverse Osmosis System is a water treatment system intended for use in hemodialysis applications. The system is designed to remove organic and inorganic substances and contaminants from potable water. The purified (or treated) water will then be used to prepare and dilute dialysate concentrate to form dialysate. The AmeriWater Centurion 1500+ Reverse Osmosis System is intended to be used in hospitals, clinics, dialysis centers, and for home care use. There is no direct contact between the patient and any part of the device invasive. Materials that contact that contact the product water include: Nylon, Polypropylene, Stainless Steel, EPDM, and Thin Film Composite Membrane (Polyamide). The Centurion 1500+ Reverse Osmosis System purifies water by applying pressure (greater than the osmotic
pressure difference) to the feed water to reverse the water flow through a semi-permeable membrane so that the water moves from a more concentrated solution to a less concentrated solution resulting in purified permeate flow. The feed water enters the Centurion through an inlet solenoid valve filling the feed water tank. A high pressure pump forces water from the feed tank through the RO membrane where it is split into permeate which has passed through the membrane and the concentrate which passes over the membrane and carries the contaminants to drain. The AmeriWater Centurion 1500+ Reverse Osmosis System produces water that meets current AAMI and Federal (U.S.) standards for water used in hemodialysis applications.
The AmeriWater Centurion 1500+ Reverse Osmosis System also includes a built in heat sanitization feature. Heat sanitization can be activated manually using the HEAT SANITIZATION function from the controller's main menu. The frequency of thermal sanitization will depend upon usage and application demands. Monitoring of the system for bacterial content should be conducted at regular intervals to determine the optimum frequency for thermal sanitization.
Indications for Use: The AmeriWater Centurion 1500+ Reverse Osmosis Systems are water treatment systems intended for use in hemodialysis applications. They are designed to pre-treat and purify potable water for use in making dialysate for hemodialysis and to meet current AAMI and Federal (U.S.) standards. The device is intended to be a component in a complete water purification system, and is not a complete water treatment system. It must be preceded by pre-treatment devices, and may need to be followed by post-
## K133783
Page 1 of 3
{1}------------------------------------------------
treatment devices as well to meet current AAM and Federal (U.S.) standards. The Centurion 1500+ Reverse Osmosis System is intended for use in a hospital, clinic, dialysis center, or for home care for single patient use. The device includes an integrated heat sanitization process.
Statement of Substantial Equivalence: The AmeriWater Centurion is substantially equivalent to the AmeriWater MRO Portable Reverse Osmosis System (MROS) cleared for market under K111740, and the AmeriWater MediQA Reverse Osmosis System cleared for market under K131904. The following table compares and contrasts the predicate devices and the new device. This table along with the documentation included in this submission demonstrates that there are no new issues of safety or effectiveness associated with this design change, and that the new device is substantially equivalent to the predicate device.
| | Predicate Device<br>AmeriWater MEDIQA (K131904) | Predicate Device<br>AmeriWater MROS<br>(K111740) | AmeriWater<br>Centurion 1500+ |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The AmeriWater MediQA Reverse<br>Osmosis System is one<br>component of a water treatment<br>system designed to pre-treat and<br>purify potable water using reverse<br>osmosis for making dialysate for<br>hemodialysis applications. The<br>device is intended to be a<br>component in a complete water<br>purification system, and is not a<br>complete water treatment system.<br>It must be preceded by pre-<br>treatment devices, and may need<br>to be followed by post-treatment<br>devices as well to meet current<br>AAMI and Federal (U.S.)<br>standards. The AmeriWater<br>MediQA is intended for use in<br>water rooms in a hospital, clinic, or<br>dialysis center. The device<br>includes an integrated heat<br>sanitization process. | The AmeriWater MRO Portable<br>Reverse Osmosis Systems are<br>water treatment systems intended<br>for use in hemodialysis<br>applications. They are designed to<br>pre-treat and purify potable water<br>for use in making dialysate for<br>hemodialysis and to meet current<br>AAMI and Federal (U.S.)<br>standards. The AmeriWater<br>Portable MROS model is intended<br>for use in a hospital, clinic, dialysis<br>center, or for home care for single<br>patient use. The AmeriWater<br>Portable MRO1 model is for<br>treatment of up to two patients in a<br>hospital, clinic, or dialysis centers. | The AmeriWater Centurion 1500+<br>Reverse Osmosis Systems are<br>water treatment systems intended<br>for use in hemodialysis<br>applications. They are designed to<br>pre-treat and purify potable water<br>for use in making dialysate for<br>hemodialysis and to meet current<br>AAMI and Federal (U.S.)<br>standards. The device is intended<br>to be a component in a complete<br>water purification system, and is<br>not a complete water treatment<br>system. It must be preceded by<br>pre-treatment devices, and may<br>need to be followed by post-<br>treatment devices as well to meet<br>current AAMI and Federal (U.S.)<br>standards. The Centurion 1500+<br>Reverse Osmosis System is<br>intended for use in a hospital,<br>clinic, dialysis center, or for home<br>care for single patient use. The<br>device includes an integrated heat<br>sanitization process. |
| For Use In: | Hospitals, clinics, or dialysis<br>centers | Hospitals, clinics, dialysis centers,<br>home care | Hospitals, clinics, dialysis centers,<br>home care |
| Power Requirements | 208/230V; 60 Hz; 3-phase | 115V 60 Hz 20 Amp | 115V 60 Hz 20 Amp |
| Purification process | Reverse Osmosis | Reverse Osmosis | Reverse Osmosis |
| RO Membrane | Polyamide, thin film composite,<br>spiral wound | Polyamide, thin film composite,<br>spiral wound | Polyamide, thin film composite,<br>spiral wound |
| Rejection rates | Total dissolved salts > 96% | Total dissolved salts > 94% | Total dissolved salts > 98% |
| Permeate Flow Rates | 19 to 48 l/min at 10°C | 1.8 l/min at 10°C | 1.5 l/min at 10°C |
| Drain Requirements | Max 96 l/min | Max 3.6 l/min | Max 3.0 l/min |
| RO Disinfection | Chemical or Heat | Chemical | Chemical or Heat |
| Water Contacting<br>Materials | Polyamide, Polypropylene, CPVC,<br>Nitrile, EPDM, Stainless steel,<br>Polyacetal, PTFE, Viton(FKM),<br>Fiberglass, PA66 (Nylon), PET<br>Noryl | ABS, Acrylic, Carbon, Nylon, PVC,<br>Polyester, Polyethylene,<br>Polypropylene, Stainless Steel,<br>Tygon, Buna N | Nylon, Polypropylene, Stainless<br>Steel, EPDM, and TFC Membrane<br>(Polyamide) |
{2}------------------------------------------------
Summary of Performance Testing: Non-clinical testing was conducted to verify and validate the performance of the reverse osmosis function and the efficacy of the heat sanitization of bacteria. Results of performance testing indicate that the device produces water that meets current AAM and Federal (U.S.) standards. Microbiological testing results show evidence that the heat sanitization function is effective in the reduction of bacteria. Clinical studies show evidence that the device, when used in accordance with the instructions for use, will produce water that meets current AAMI and Federal (U.S.) standards for hemodialysis. Test results from biocompatibility testing, software validation, and electrical safety testing (in accordance with IEC 60601-1) indicate that the device is safe and effective for its intended purpose.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, and three horizontal bars above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
### April 16, 2014
AmeriWater, Inc. Brian R. Bowman Quality and Regulatory Administrator 1303 Stanley Avenue Dayton, OH 45404
Re: K133783
> Trade/Device Name: AmeriWater Centurion 1500 + Reverse Osmosis System Regulation Number: 21 CFR§ 876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: FIP Dated: March 12, 2014 Received: March 20, 2014
Dear Brian R. Bowman,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Brian R. Bowman
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Ameriliorer
ច្បាទចេចចេចចេនចេននេះជាជាវេរ
1303 STANLEY AVENUE DAYTON, OH 45404 USA TEL 937-461-8833, 800-535-5585 FAX 937-461-1988 WWW.AMERIWATER.com
Image /page/5/Picture/3 description: The image contains two overlapping circles with text inside. The larger circle on the left has the text "Suez environnement". The smaller circle on the right has the text "Degremont" and a logo above it. The image appears to be a logo or emblem for a company or organization.
## Indications for Use
510(k) Number (if known): K133783
Device Name: AmeriWater Centurion 1500+ Reverse Osmosis System
Indications For Use:
The AmeriWater Centurion 1500+ Reverse Osmosis Systems are water treatment systems intended for use in hemodialysis applications. They are designed to pre-treat and purify potable water for use in making dialysate for hemodialysis and to meet current AAM and Federal (U.S.) standards. The device is intended to be a component in a complete water purification system, and is not a complete water treatment system. It must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well to meet current AAM and Federal (U.S.) standards. The Centurion 1500+ Reverse Osmosis System is intended for use in a hospital, clinic, dialysis center, or for home care for single patient use. The device includes an integrated heat sanitization process.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert P. Leerne 2014.04.16 14:29:32
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.