MEDIQA SINGLE PASS REVERSE OSMOSIS SYSTEM WITH HEAT SANITIZATION, MEDIQA DOUBLE PASS REVERSE OSMOSIS SYSTEM WITH HEAT SA
Applicant
Ameriwater
Product Code
FIP · Gastroenterology, Urology
Decision Date
Nov 13, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
The AmeriWater MediQA Reverse Osmosis System is one component of a water treatment system designed to pre-treat and purify potable water using reverse osmosis for making dialysate for hemodialysis applications. The device is intended to be a component in a complete water purification system, and is not a complete water treatment system. It must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well to meet current AAMI and Federal (U.S.) standards. The AmeriWater MediQA is intended for use in water rooms in a hospital, clinic, or dialysis center. The device includes an integrated heat sanitization process.
Device Story
Reverse osmosis (RO) system for hemodialysis water purification; utilizes high-pressure pump to force potable feed water through RO modules; separates water into permeate (purified) and concentrate (rejected contaminants). Available in single-pass or double-pass configurations; double-pass models reprocess concentrate from second stage. Includes integrated heat sanitization (185-203°F) for membranes and piping, activated manually or via timer. System monitored via touch screen displaying temperature, conductivity, and flow data. Used in hospital/clinic water rooms by trained staff. Provides purified water for dialysate preparation; heat sanitization reduces bacterial content, ensuring compliance with AAMI/Federal standards.
Clinical Evidence
Non-clinical bench testing performed. Verified RO performance and efficacy of heat sanitization in reducing bacterial content. Biocompatibility, software validation, and electrical safety testing conducted. Results indicate device produces water meeting AAMI and Federal standards for hemodialysis.
Technological Characteristics
Reverse osmosis water purification system; single or double-pass configurations. Features high-pressure pumps, RO membranes, and integrated heat sanitization (185-203°F). Touch screen interface for monitoring temperature, conductivity, and flow. Power: 208/230V, 60 Hz, 3-phase. Flow rates: 4.5 to 12.0 gpm. System includes 16-gallon working tank.
Indications for Use
Indicated for use in hospitals, clinics, or dialysis centers as a component of a water treatment system to purify potable water for hemodialysis dialysate production. Must be used with pre-treatment and potentially post-treatment systems to meet AAMI/Federal standards.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Gambro Central Water Treatment System CWP 100 - WRO H (K974899)
Submission Summary (Full Text)
{0}------------------------------------------------
AmeriWater .
បានចំបងប្រពន្ធបានប្រជាប្រ ប្រ
1303 STANLEY AVENUE DAYTON, OH 45404 USA TEL 937-461-8833, 800-535-5585 FAX 937-461-1988 www.AMERIWATER.com
Image /page/0/Picture/3 description: The image contains two circular logos that are partially overlapping. The logo on the left has the text "SOZ Environment" written in a stylized font. The logo on the right has the text "Department" written in a similar stylized font, along with a graphic above the text. Both logos appear to be for environmental or governmental organizations.
510(K) SUMMARY
| | November 5, 2013 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number: | K131904 |
| Submitter: | AmeriWater |
| Contact: | Brian R. Bowman, Quality & Regulatory Administrator<br>1303 Stanley Avenue, Dayton, OH 45404<br>Phone: (937)461-8833 Fax: (937)461-1988<br>brianbowman@ameriwater.com |
| Proprietary Name: | AmeriWater MediQA Reverse Osmosis System |
| Common Name: | Reverse Osmosis System |
| Classification Name: | Water purification system for hemodialysis |
| Classification: | Class II Medical Device under §876.5665<br>Panel: Gastroenterology<br>Product Code: FIP |
| Equivalent Device: | K974899, Gambro Central Water Treatment System CWP 100 - WRO H |
NOV 1 3 2013Device Description: Reverse osmosis, the scientific concept used in the AmeriWater MEDIQA Reverse Osmosis System, is the opposite of osmosis. The MEDIQA system uses a pump to apply the pressure required for reverse osmosis. Potable tap water enters the MEDIQA through an inlet solenoid valve filling the feed water tank. A high pressure pump forces water from the feed tank through the RQ modules, each module containing a high performance membrane. The water entering the RO module is split into two flows. The water which passes through the membrane is known as permeate and is purified water. The water rejected by the membrane contains an increased level of dissolved contaminants. It passes out of the RO module as the second flow stream called concentrate, and is sent to drain.
The MEDIQA is available as a single pass or a double pass reverse osmosis (RO) unit. In a double pass (two RO stages) system, permeate from the first stage is pressurized by a second high pressure oump and fed to the second stage RO module set. Permeate from the second stage RO module set is fed via a manifold to the distribution loop. Unused permeate returning from the distribution loop is fed back into the feed tank. The concentrate from the first stage is sent to drain while the concentrate from the second stage is returned to the feed water tank for reprocessing. The feed and permeate water flows are monitored at various points in the process to verify temperature, conductivity, and flow. This data is displayed on a touch screen panel to give instant feedback of water quality and process activity. The MEDIQA system is controlled and operated by a touch screen mounted on the machine.
The AmeriWater MEDIQA Reverse Osmosis System also includes a built in heat sanitization feature to enable a heat sanitization of the reverse osmosis membranes and its pipe work. Heat sanitization can be activated manually using the HEATSAN buttons on the touch screen display or automatically if timer clock settings are implemented. The frequency of heat sanitization will depend upon usage and
{1}------------------------------------------------
application demands. Monitoring of the system for bacterial content should be conducted at regular intervals to determine the optimum frequency for heat sanitization.
Indications for Use: The AmeriWater MediQA Reverse Osmosis System is one component of a water treatment system designed to pre-treat and purify potable water using reverse osmosis for making dialysate for hemodialysis applications. The device is intended to be a component in a complete water purification system, and is not a complete water treatment system. It must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well to meet current AAMI and Federal (U.S.) standards. The AmeriWater MediQA is intended for use in water rooms in a hospital, clinic, or dialysis center. The device includes an integrated heat sanitization process.
The MediQA is available in both single pass and double pass models that supply from 4.5 to 12.0 gallons per minute (gpm) of product water. Model MSP1 is a single membrane RO that produces up to 5.0 gpm of product water. Model MSP2 is a single pass, dual-membrane RO that produces up to 9.4 gpm of product water. Model MSP3 is a single pass, 3-membrane RO that produces up to 12.6 gpm of product water. Model MDP1 is a double pass, 2-membrane RO that produces up to 5.0 pm of product water. Model MDP2 is a double pass, 3-membrane RO that produces up to 7.0 gpm of product water. Model MDP3 is a double pass, 4-membrane RO that produces up to 10.0 gpm of product water. Model MDP4 is a double pass, 5-membrane RO that produces up to 12.0 gpm of product water.
Statement of Substantial Equivalence: The AmeriWater MEDIQA is substantially equivalent to the Gambro Central Water Treatment System CWP 100 - WRO H cleared for market under K974899. The following table compares and contrasts the predicate device and the new device. This table alona with the documentation included in this submission demonstrates that there are no new issues of safety or effectiveness associated with this design change, and that the new device is substantially equivalent to the predicate device.
| | Predicate Device | AmeriWater |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CWP 100 - WRO H (K974899) | MEDIQA |
| Indications for use | The Gambro Central Water Treatment<br>System CWP 100-WRO H is designed to<br>produce water of adequate quality for<br>hemodialysis, both chemically and<br>microbiologically with an adequate flow,<br>provided that the feed water complies with<br>the existing standards for drinking water<br>and has been properly pre-treated. | The AmeriWater MEDIQA Reverse<br>Osmosis System is a water treatment<br>systems intended for use in hemodialysis<br>applications. The MEDIQA is designed to<br>pre-treat and purify potable water for use in<br>making dialysate for hemodialysis and to<br>meet current AAMI and Federal (U.S.)<br>standards. The AmeriWater MEDIQA is<br>intended for use in a hospital, clinic, or<br>dialysis center. The device includes an<br>integrated heat sanitization process. |
| For Use In: | Hospitals | Hospitals, clinics, or dialysis centers |
| Power Requirements | 120/208V, 60 Hz, 3-phase | 208/230V; 60 Hz; 3-phase |
| Permeate Flow Rates | 3.4 - 9.2 gpm | 4.5 - 12.0 gpm |
| RO Disinfection | Chemical | Chemical or Heat |
| Heat Sanitization For | Water Distribution System | MEDIQA System only |
| Heater Power Rating | 7.0 - 9.0 kW | 9.0 kW |
| Heated Temperature | 185°F | 185 - 203°F |
| Working Tank Volume | 68 - 87 gallons | 16 gallons |
{2}------------------------------------------------
Summary of Performance Testing: Non-clinical testing was conducted to verify and validate the performance of the reverse osmosis function and the efficacy of the heat sanitization in the reduction of bacteria. Results of performance testing indicate that the device produces water that meets current AAM and Federal (U.S.) standards. Microbiological testing results show evidence that the heat sanitization function is effective in the reduction of bacteria. Clinical studies show evidence that the device, when used in accordance with the instructions for use, will produce water that meets current AAMI and Federal (U.S.) standards for hemodialysis. Test resuits from biocompatibility testing, software validation, and electrical safety testing indicate that the device is safe and effective for its intended purpose.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three lines representing its wings and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 13, 2013
Ameri Water Brian R. Bowman Quality & Regulatory Administrator 1303 Stanley Avenue Dayton, OH 45404
Re: K131904
> Trade/Device Name: AmeriWater MediQA Reverse Osmosis System Regulation Number: 21 CFR\$ 876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: FIP Dated: August 12, 2013 Received: August 15, 2013
Dear Brian R. Bowman,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
1f your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of
{4}------------------------------------------------
Page 2 - Brian R. Bowman
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.lda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/yem115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/delault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
## Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
AmeriWater
្រប្រចម្ជីដនរបស់ប្រជាជនជន្រ
1303 STANLEY AVENUE DAYTON, OH 45404 USA TEL 937-461-8833.800-535-5585 FAX 937-461-1988 www.ANGERIWATER.com
Image /page/5/Picture/3 description: The image contains two circular logos that are overlapping. The logo on the left has some text that is difficult to read. The logo on the right has a symbol at the top and the text "Degremont" at the bottom.
## Indications for Use
510(k) Number (if known): K131904
AmeriWater MediQA Reverse Osmosis System Device Name:
Indications For Use:
The AmeriWater MediQA Reverse Osmosis System is one component of a water treatment system designed to pre-treat and purify potable water using reverse osmosis for making dialysate for hemodialysis applications. The device is intended to be a component in a complete water purification system, and is not applioditions. The do not no must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well to meet current AAMI and Federal (U.S.) standards. The AmeriWater MediQA is intended for use in water rooms in a hospital, clinic, or dialysis center. The device includes an integrated heat sanitization process.
The MediQA is available in both single pass and double pass models that supply from 4.5 to 12.0 gallons t no moute (gpm) of product water. Model MSP1 is a single membrane RO that produces up to 5.0 gom of product water. Model MSP2 is a single pass, dual-membrane RO that produces up to 9.4 gpm 5.0 gplires product water. Model more pass, 3-membrane RO that produces up to 12.6 gpm of product water. Model MDP1 is a double pass, 2-membrane RO that produces up to 5.0 gpm of product water. Water MDP2 is a double pass, 3-membrane RO that produces up to 7.0 gpm of product water. Model MDP3 is a double pass, 4-membrane RO that produces up to 10.0 gpm of product water. Model MDP4 is a double pass, 5-membrane RO that produces up to 12.0 gpm of product water.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert P. Lerner -S 2013.11.13 15:10:58 -05'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.