INFILL 41-TLIF CONVEX OBLIQUE DEVICE, INFILL 43-TLIF CONTOUR OBLIQUE, INFILL 44-TLIF CONTOUR OBLIQUE AND THE INFILL 60
K133721 · Pinnacle Spine Group, LLC · MAX · Mar 31, 2014 · Orthopedic
Device Facts
Record ID
K133721
Device Name
INFILL 41-TLIF CONVEX OBLIQUE DEVICE, INFILL 43-TLIF CONTOUR OBLIQUE, INFILL 44-TLIF CONTOUR OBLIQUE AND THE INFILL 60
Applicant
Pinnacle Spine Group, LLC
Product Code
MAX · Orthopedic
Decision Date
Mar 31, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
InFill® interbody fusion device is indicated for intervertebral body fusion of the spine in skeletally mature patients. InFill® interbody fusion device is designed for use with autogenous bone graft to facilitate fusion. InFill® interbody fusion device is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar devices are to be used in patients who have had at least six months of non-operative treatment. InFill® interbody fusion device is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine. These implants may be implanted via a variety of open or minimally invasive approaches. These approaches include anterior, lateral and transforaminal.
Device Story
Radiolucent spinal implants (InFill 41-TLIF, 43-TLIF, 44-TLIF, 60-ALIF) designed for intervertebral body fusion; manufactured from PEEK with tantalum markers. Implants feature central fenestration for bone graft support and surface teeth for fixation. Used by surgeons in clinical settings via anterior, lateral, or transforaminal approaches. Implants provide structural support to the disc space to facilitate fusion in patients with degenerative disc disease. Used in conjunction with supplemental internal spinal fixation systems. Provided sterile for single use.
Clinical Evidence
Bench testing only. Performance data includes static and dynamic compression (ASTM F2077), static subsidence (ASTM F2267), and expulsion testing to verify mechanical integrity and equivalence to predicate devices.
Technological Characteristics
Materials: PEEK OPTIMA, tantalum markers. Design: Box, oval, or banana-shaped with bullet nose; central fenestration for bone graft; surface teeth for bone-to-implant interface. Standards: ASTM F2077 (compression), ASTM F2267 (subsidence). Sterilization: Sterile, single-use.
Indications for Use
Indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis at one or two contiguous lumbar levels (L2-S1). Requires at least six months of failed non-operative treatment. Must be used with autogenous bone graft and supplemental internal spinal fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
## MAR 3 1 2014
## 6. 510(k) Summary
This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Pinnacle Spine Group, LLC |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| DATE PREPARED: | March 22, 2014 |
| CONTACT PERSON: | Rebecca K Pine<br>1601 Elm Street, Suite 1930<br>Dallas, TX 75201<br>Phone: 760.809.5178<br>Fax: 760.290.3216 |
| TRADE NAME: | InFill® 41-TLIF Convex Oblique device, InFill® 43-TLIF Contour Oblique, InFill® 44-TLIF Contour Oblique and the<br>InFill® 60-ALIF |
| COMMON NAME: | Spinal Implant |
| CLASSIFICATION NAME: | Intervertebral Body Fusion Device |
| DEVICE CLASSIFICATION: | Class II |
| REGULATION NUMBER | 888.3080 (product code: MAX) |
| PREDICATE DEVICES: | InFill Interbody Fusion Device (K103729, K121733, K124012)<br>Novel TL Spinal Spacer System, (K080699)<br>Calix-A Lumbar Spinal Implant System (K131350) |
## Substantially Equivalent To:
The InFill® 41-TLIF Convex Oblique device, InFill® 43-TLIF Contour Oblique, InFill® 44-TLIF Contour Oblique and the InFill® 60-ALIF devices are substantially equivalent in intended use, principal of operation and technological characteristics to the InFill Interbody Fusion Device, Novel Spinal Spacer System and the Calix-A Lumbar Spinal Implant System.
## Description of the Device Subject to Premarket Notification:
The InFill® 41-TLIF Convex Oblique device, InFill® 43-TLIF Contour Oblique, InFill® 44-TLIF Contour Oblique and the InFill® 60-ALIF are radiolucent implantable devices manufactured from PEEK and tantalum (marker material). The implants are available in various sizes to suit the individual pathology and anatomical conditions of the patient.
. The InFill® 41-TLIF Convex Oblique device, InFill® 43-TLIF Contour Oblique, InFill® 44-TLIF Contour Oblique and the InFill® 60-ALIF are provided sterile, for single use
{1}------------------------------------------------
only.
## Indication for Use:
InFill® interbody fusion device is indicated for intervertebral body fusion of the spine in skeletally mature patients. InFill® interbody fusion device is designed for use with autogenous bone graft to facilitate fusion. InFill® interbody fusion device is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar devices are to be used in patients who have had at least six months of non-operative treatment. InFill® interbody fusion device is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine. These implants may be implanted via a variety of open or minimally invasive approaches. These approaches include anterior, lateral and transforaminal.
#### Technical Characteristics:
The InFill® 41-TLIF Convex Oblique device, InFill® 43-TLIF Contour Oblique, InFill® 44-TLIF Contour Oblique and the InFill® 60-ALIF have similar physical and technical characteristics to the predicate devices, as illustrated in the table below.
| Technical<br>Characteristics | In Fill®<br>intervertebral<br>body fusion<br>devices, (41-<br>TLIF Convex<br>Oblique, 43-<br>/44-TLIF<br>Contour<br>Oblique, 60-<br>ALIF) | In Fill®<br>intervertebral<br>body fusion<br>device<br>(K103729,<br>K121733,<br>K124012) | Novel TL<br>Spinal Spacer<br>System,<br>(K080699) | Calix-A<br>Lumbar<br>Spinal<br>Implant<br>System<br>(K131350) | Biomet<br>Enclave<br>(K081636) |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|---------------------------------------------------|---------------------------------------------------------------|--------------------------------|
| Shape | Box-shaped,<br>bullet nose<br>Oval shape,<br>bullet nose<br>Banana-shaped,<br>bullet nosed | Box-shaped,<br>bullet nose | Banana-shaped,<br>bullet nosed | Oval shape,<br>bullet nose | Oval<br>shape,<br>bullet nose |
| Bone to implant<br>surface | Surface teeth | SAME | SAME | SAME | SAME |
| Bone graft<br>support feature | Central<br>fenestration | SAME | SAME | SAME | SAME |
| Primary<br>implant<br>material | PEEK OPTIMA<br>® | SAME | SAME | SAME | SAME |
| Surgical<br>Approach | Transforaminal<br>Anterior | Lateral<br>Transforaminal | Transforaminal | Anterior | Anterior |
Pinnacle Spine Group, LLC InFill® Product Family
{2}------------------------------------------------
## Performance Data:
All necessary performance testing, has been completed for the InFill® 41-TLIF Convex Oblique device, InFill® 43-TLIF Contour Oblique, InFill® 44-TLIF Contour Oblique and the InFill® 60-ALIF devices including static/dynamic compression (ASTM F2077), static subsidence (ASTM F2267) and expulsion to assure substantial equivalence to the predicate devices.
## Basis for Determination of Substantial Equivalence:
Upon reviewing the performance data provided in this submission and comparing intended use, design, materials, principle of operation and overall technological characteristics, the InFill® 41-TLIF Convex Oblique device, InFill® 43-TLIF Contour Oblique, InFill® 44-TLIF Contour Oblique and the InFill® 60-ALIF devices are determined by Pinnacle Spine Group, LLC, to be substantially equivalent to existing legally marketed devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 31, 2014
Pinnacle Spine Group, LLC % Ms. Rebecca K. Pine Consultant 1601 Elm Street, Suite 1930 Dallas, Texas 75201
Re: K133721
Trade/Device Name: Infill® Interbody Fusion Devices (Infill® 41-TLIF Convex Oblique, Infill® 43-TLIF Contour Oblique, Infill® 44-TLIF Contour Oblique, and the Infill® 60-ALIF)
Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: January 9, 2014 Received: January 10, 2014
Dear Ms. Pine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
#### Page 2 - Ms. Rebecca K. Pine
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known) K133721
Device Name InFill® Interbody Fusion Devices
#### Indications for Use (Describe)
InFill® interbody fusion device is indicated for intervertebral body fusion of the spine in skeletally mature patients. InFill® interbody fusion device is designed for use with autogenous bone graft to facilitate fusion. InFill® interbody fusion device is intended for use at either one levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbardevices are to be used in patients who have had at least six months of non-operative treatment. InFill® interbody fusion device is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine. These implants may be implanted via a variety of open or minimally invasive approaches. These approaches include anterior, lateral and transforaminal.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
球的方法: 大利亚 FOR FDA USE ONLY ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
## Anton E. Dmitriev, PhD
Division of Orthopedic Devices
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.