CALIX LUMBAR SPINAL IMPLANT SYSTEM

K131350 · X-Spine Systems, Inc. · MAX · Oct 23, 2013 · Orthopedic

Device Facts

Record IDK131350
Device NameCALIX LUMBAR SPINAL IMPLANT SYSTEM
ApplicantX-Spine Systems, Inc.
Product CodeMAX · Orthopedic
Decision DateOct 23, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The X-spine Calix Lumbar Spinal Implant System is intended for spinal fusion procedures at one or two contiguous levels (L2 - S1 inclusive) in skeletally mature patients with degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the lumbosacral spine. DDD patients may also have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved level(s). These implants are to be packed with autogenous bone graft and implanted via an anterior, posterior, and/or transforaminal approach. Patients should receive at least six (6) months of non-operative treatment prior to treatment with a lumbosacral intervertebral fusion device. This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems).

Device Story

Calix Lumbar Spinal Implant System is a box/oval-shaped intervertebral body fusion device; manufactured from PEEK-Optima LT1; features superior/inferior surface teeth for fixation and internal cavity for autogenous bone graft packing; includes tantalum radiographic markers. Implanted by surgeons via anterior, posterior, or transforaminal approaches to facilitate spinal fusion in lumbosacral DDD patients. Used in conjunction with cleared supplemental spinal fixation systems (pedicle screw/rod or anterior plate systems). Provides structural support to the intervertebral space to promote fusion; benefits patients by stabilizing the spinal segment and reducing discogenic back pain.

Clinical Evidence

Bench testing only. No clinical data provided. Performance established via standardized mechanical testing: static/dynamic axial compression, static/dynamic axial compression-shear (ASTM F2077), load-induced subsidence (ASTM F2267), and expulsion testing per FDA guidance.

Technological Characteristics

Material: Invibio PEEK-Optima LT1 (ASTM F2026). Radiographic markers: Tantalum (ASTM F560). Geometry: Box/oval shape with surface teeth and hollow center. Dimensions: Variable heights/widths. Energy source: None (mechanical). Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) of the lumbosacral spine (L2-S1) requiring spinal fusion at one or two contiguous levels. Includes patients with Grade 1 spondylolisthesis or retrolisthesis and those with prior non-fusion spinal surgery. Requires at least 6 months of failed non-operative treatment. Contraindicated in skeletally immature patients.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) Summary ### ADMINISTRATIVE INFORMATION | Manufacturer Name: | X-spine Systems, Inc.<br>452 Alexandersville Road<br>Miamisburg, OH 45342<br><br>Telephone (937) 847-8400<br>Fax (937) 847-8410 | | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------| | Official Contact: | David Kirschman, M.D.<br>Chief Medical Officer | OCT 2 3 2013 | | Date Prepared: | May 9, 2013 | | | DEVICE NAME | | | | Trade/Proprietary Name: | Calix ™ Lumbar Spinal Implant System | | | Common Name: | Intervertebral Body Fusion Device | | | Classification Name(s): | Intervertebral Fusion Device with Bone Graft, Lumbosacral | | | Device Class: | Class II | | | Product Code(s): | MAX | | | Regulation Number(s): | §888.3080 | | ## ESTABLISHMENT REGISTRATION NUMBER The X-spine Systems, Inc. establishment registration number is 3005031160. The owner/operator number for X-spine Systems, Inc. is 9063903. #### INTENDED USE The X-spine Calix Lumbar Spinal Implant System is intended for spinal fusion procedures at one . or two contiguous levels (L2 - S1 inclusive) in skeletally mature patients with degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the lumbosacral spine. DDD patients may also have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved level(s). These implants are to be packed with autogenous bone graft and implanted via an anterior, posterior, and/or transforaminal approach. Patients should receive at least six (6) months of non-operative treatment prior to treatment with a lumbosacral intervertebral fusion device. This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). 1 {1}------------------------------------------------ ## DEVICE DESCRIPTION The X-spine Calix Lumbar Spinal Implant System is a generally box or oval shaped device manufactured from Invibio PEEK-Optima LT1 per ASTM F2026 with an array of holes located throughout its geometry as well as teeth on the superior and inferior surfaces. The device is supplied in several widths and heights to accommodate variations in patient anatomy. The devices contain radiographic markers made from tantalum per ASTM F560. The hollow center of the implant allows the device to be packed with bone graft. ## EQUIVALENCE TO MARKETED PRODUCT X-spine Systems, Inc. has submitted information to demonstrate that, for the purposes of FDA's requlation of medical devices, the Calix Lumbar System is substantially equivalent in indications and design principles to predicate devices. # PREDICATE DEVICE(S) - . Medtronic Sofamor Danek - CAPSTONE Spinal System (K073291) - . Globus Medical, Inc. - Patriot Spacer System (K072970) - Globus Medical, Inc. Sustain & Sustain Radiolucent Spacers (K130478) . - K2M, Inc. Aleutian Spinal System (K082698) . - Depuy Spine, Inc. CONCORDE System (K081917) . # PERFORMANCE DATA To establish device performance, the following standardized tests were conducted on full device constructs: ASTM F2077 - Test Methods for Intervertebral Body Fusion Devices - Static and dynamic axial compression ● - . Static and dynamic axial compression-shear ASTM F2267 - Standard Test Methods for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device under Static Axial Compression Expulsion testing as suggested by FDA Guidance - Class II Special Controls Guidance Document: Intervertebral Body Fusion Device ## BASIS OF SUBSTANTIAL EQUIVALENCE In summary, biomechanical testing results indicate that the Calix Lumbar Spinal Implant System is substantially equivalent to predicate device performance and is as effective, and performs at least as safely and effectively as the cited predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized representation of a human figure embracing a globe, symbolizing the department's mission to protect the health of all Americans and provide essential human services. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 23, 2013 X-spine Systems. Incorporated David Kirschman, M.D. Chief Medical Officer 452 Alexandersville Road Miamisburg, Ohio 45342 Re: K131350 > Trade/Device Name: CalixTM Lumbar Spinal Implant System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: October 17, 2013 Received: October 18, 2013 Dear Dr. Kirschman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - David Kirschman, M.D. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Erin同际Keith for Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use 510(k) Number (if known): _K131350 Device Name: Calix™ Lumbar Spinal Implant System Indications for Use: The X-spine Calix Lumbar Spinal Implant System is intended for spinal fusion procedures at one or two contiguous levels (L2 - S1 inclusive) in skeletally mature patients with degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the lumbosacral spine. DDD patients may also have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved level(s). These implants are to be packed with autogenous bone graft and implanted via an anterior, posterior, and/or transforaminal approach. Patients should receive at least six (6) months of non-operative treatment prior to treatment with a lumbosacral intervertebral fusion device. This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anton E. Dmitriev, PhD Division of Orthopedic Devices 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...