K133646 · Siemens Medical Systems, Inc. · JAK · Jun 20, 2014 · Radiology
Device Facts
Record ID
K133646
Device Name
ADMIRE
Applicant
Siemens Medical Systems, Inc.
Product Code
JAK · Radiology
Decision Date
Jun 20, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
ADMIRE is a CT reconstruction software. The end user can choose to apply either ADMIRE or the weighted filter back-projection (WFBP) to the acquired raw data. Depending on the clinical task, patient size, anatomical location, and clinical practice, the use of ADMIRE can help to reduce radiation dose while maintaining pixel noise, low contrast detectability and high contrast resolution. Phantom measurements showed that high contrast resolution and pixel noise are equivalent between full dose WFBP images and reduced dose ADMIRE images. Additionally, ADMIRE can reduce spiral artifacts by using iterations going back and forth between image space and raw data space. Images reconstructed with ADMIRE are not intended to be evaluated with syngo Osteo CT or syngo Calcium Scoring.
Device Story
ADMIRE is a software-based iterative reconstruction option for CT systems. It processes raw CT data to generate reconstructed images with improved quality—specifically higher pixel noise reduction, improved high-contrast edge resolution, and noise texture similar to filtered back projection (FBP). The device operates by performing iterative cycles between image space and raw data space, which also serves to reduce spiral artifacts. Clinicians select ADMIRE as an alternative to standard weighted filtered back projection (WFBP) during the reconstruction workflow. By improving image quality, the device enables clinicians to acquire scans at lower radiation doses without compromising diagnostic utility. It is intended for use in clinical settings where CT imaging is performed. The output is a reconstructed CT image used by radiologists and physicians for diagnostic decision-making.
Clinical Evidence
Bench testing only. Performance was validated using phantom measurements to compare high-contrast resolution and pixel noise between full-dose WFBP and reduced-dose ADMIRE images. Results demonstrated equivalence in image quality metrics. Sample clinical images were provided to support performance claims. No human clinical trials were conducted.
Technological Characteristics
Software-based iterative reconstruction algorithm for CT systems. Operates on raw data to produce reconstructed images. Complies with IEC 62304 (software life cycle), ISO 14971 (risk management), and IEC 60601-1-4/1-6 (safety). Supports DICOM standards. Moderate level of concern software.
Indications for Use
Indicated for use as a CT reconstruction software for patients undergoing CT imaging. Used to assist in radiation dose reduction while maintaining image quality (pixel noise, low contrast detectability, high contrast resolution). Not for use with syngo Osteo CT or syngo Calcium Scoring applications.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
JUN 2 0 2014
K133646
page 1 of 5
## 510(K) SUMMARY FOR ADMIRE
Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355
Date Prepared: May 5, 2014
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
- 1. General Information
SIEMENS
Importer/Distributor: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355
Establishment Registration Number: 2240869
Manufacturing Site: SIEMENS AG, MEDICAL SOLUTIONS Siemensstrasse 1 91301 Forchheim, GERMANY
Establishment Registration Number: 3004977335
#### 2. Contact Person:
Mrs. Kimberly Mangum Technical Specialist, Regulatory Affairs Submissions Siemens Medical Solutions, Inc. USA 51 Valley Stream Parkway D02 Malvern, PA 19355-1406 Fax: (610) 448-1778 Phone: (610) 448-1772 Email: kimberly.mangum@siemens.com
- Device Name and Classification 3.
| Product Name: | ADMIRE |
|-----------------------|----------------------------------|
| Propriety Trade Name: | ADMIRE |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
{1}------------------------------------------------
Device Class: Class II Product Code: 90JAK
Legally Marketed Predicate Devices Trade Name: SAFIRE 510(k)#: K103424 Clearance Date: Clearance Date: Classification Panel: Radiology Classification Regulation:21 CFR § 892.1750 Device Class: . Product Code: 90 JAK
#### 4. Substantial Equivalence:
Siemens ADMIRE is substantially equivalent to the following medical devices in commercial distribution:
| Predicate Device Name | FDA Clearance Number | FDA Clearance Date |
|-----------------------|----------------------|--------------------|
| SAFIRE | K103424 | November 22, 2011 |
#### 5. Device Description:
Siemens ADMIRE is an extension of the previously cleared Sinogram Affirmed Iterative Reconstruction (SAFIRE) reconstruction algorithm. ADMIRE is a software option for CT operating systems that provides an improved image quality or reciprocally can allow the physician to acquire scans with reduced radiation dose without reduction of image quality compared to today's standard.
ADMIRE is designed to improved reconstructed image quality through the integration of additional processing steps in image reconstruction. These additional steps result in the following improvements in image quality:
- . Higher pixel noise reduction
- A noise texture closer to filtered back projection (FBP) .
- . Improved resolution for high contrast edges
#### Summary of Technical Characteristics of the Subject Device as 6. Compared with the Predicate Device:
ADMIRE reconstruction software do not have significant changes in the indications for use, materials, energy source, or technological characteristics when compared to the predicate device.
The additional/extended algorithm processing steps result in the following improvements in reconstructed image quality for ADMIRE in comparison to the predicate device SAFIRE:
{2}------------------------------------------------
- Higher pixel noise reduction in thicker slices (for example 3mm and . 5mm)
- Less outliers in the noise texture resulting in a noise texture closer to . that of filtered back projection (FBP)
- Improved Resolution at High Contrast Edges compared to weighted . filtered back projection (WFBP)
The intended use and fundamental scientific technology are similar to the predicate device; therefore Siemens believes that they are substantially equivalent to the predicate device.
#### 7. Nonclinical Testing:
ADMIRE is designed to fulfill the requirements of following safety and performance standards:
| Recognition<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and<br>Date | Publication<br>Date | Standards<br>Development<br>Organization |
|-----------------------|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|---------------------|------------------------------------------|
| 5-40 | General | Medical devices - Application of<br>risk management to medical<br>devices | 14971 Second<br>Edition 2007-<br>03-01 | 08/20/2012 | ISO |
| 13-8 | Software | Medical device software -<br>Software life cycle processes | 62304 First<br>edition 2006-05 | 08/20/2012 | IEC |
| 5-41 | General | Medical electrical equipment -<br>Part 1-4: General requirements<br>for safety - Collateral standard:<br>Programmable electrical medical<br>systems, edition 1.1 | 60601-1-4:<br>2000<br>Consol.Ed 1.1 | 09/08/2009 | IEC |
| | General | Medical electrical equipment -<br>Part 1-6: General requirements<br>for basic safety and performance | 60601-1-6 | 2006 | IEC |
| 12-218 | Radiology | Digital Imaging and<br>Communications in Medicine<br>(DICOM) | PS 3.1 - 3.18<br>(2008) | 2008 | NEMA |
{3}------------------------------------------------
This submission contains performance data to demonstrate continued conformance with special controls for medical devices containing software. Non clinical tests (integration and functional) and phantom testing were conducted for ADMIRE during product development. The modifications described in this Premarket Notification were supported with verification/validation testing.
The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission.
EMC/electrical safety was evaluated according to the IEC Standards. Siemens certify conformance to Voluntary Standards covering Electrical and Mechanical Safety. In conclusion, the identified risk of electrical hazards was mitigated and is substantially equivalent to the predicate device in terms of safety and effectiveness. All testing and validation has been completed.
### Summary of Additional Testing
In addition to successful completion of verification and validation testing, additional bench testing was performed to substantiate performance claims and demonstrate substantial equivalence to the predicate device SAFIRE. Sample clinical images were also provided within the submission.
### 8. Indications for Use:
ADMIRE is a CT reconstruction software. The end user can choose to apply either ADMIRE or the weighted filter back-projection (WFBP) to the acquired raw data. Depending on the clinical task, patient size, anatomical location, and clinical practice, the use of ADMIRE can help to reduce radiation dose while maintaining pixel noise, low contrast detectability and high contrast resolution. Phantom measurements showed that high contrast resolution and pixel noise are equivalent between full dose WFBP images and reduced dose ADMIRE images. Additionally, ADMIRE can reduce spiral artifacts by using iterations going back and forth between image space and raw data space.
ADMIRE at highest noise reduction strength for thin (0.6 mm) reconstruction slices in simulated body and head phantoms for low contrast objects with different contrasts.
Images reconstructed with ADMIRE are not intended to be evaluated with syngo Osteo CT or syngo Calcium Scoring.
{4}------------------------------------------------
K133646
Page 5 of 5
## 9. General Safety and Effectiveness Concerns:
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards. Siemens adheres to recognized and established industry practice and standards.
### 10. Conclusion as to Substantial Equivalence
In summary, Siemens is of the opinion that the ADMIRE does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate devices.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" arranged in a circular pattern around the symbol. The logo is black and white.
Public Flealth Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
June 20, 2014
Siemens Medical Systems, Inc. % Ms. Kimberlv Mangum Regulatory Affairs Specialist 51 Valley Stream Parkway MALVERN PA 19301
K133646 Rc:
> Trade/Device Name: ADMIRE Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: H Product Code: JAK Dated: May 8, 2014 Received: May 13, 2014
Dear Ms. Mangum:
We have reviewed your Section 510(k) premarket.notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Ms. Kimberly Mangum
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Smh.7)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K133646
Device Name ADMIRE
Indications for Use (Describe)
ADMIRE is a CT reconstruction software. The end user can choose to apply either ADMIRE or the weighted filter backprojection (WFBP) to the acquired raw data. Depending on the clinical task, patient size, anatomical location, and clinical practice, the use of ADMIRE can help to reduce radiation dose while maintaining pixel noise, low contrast detectability and high contrast resolution. Phantom measurements showed that high contrast resolution and pixel noise are equivalent between full dose WFBP images and reduced dose ADMIRE images. Additionally, ADMIRE can reduce spiral artifacts by using iterations going back and forth between image space and raw data space.
Images reconstructed with ADMIRE are not intended to be evaluated with syngo Osteo CT or syngo Calcium Scoring.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Sm.h.7)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.