Published clinical literature/studies on orbital floor reconstruction and pediatric facial fracture management
The sponsor used a literature review of clinical outcomes in patients aged 15-21 to demonstrate that the safety profile and anatomical suitability for the device in this age group are equivalent to the adult population.
Literature review; Pediatric subpopulation; Clinical outcomes; Safety profile
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review of orbital repair in pediatric subpopulation (15-21 years); Literature review of clinical studies; Follow-up/Duration: Over 25 years of clinical use
Patients 15 to 21 years of age; Number of Sites: Not applicable
Not applicable for this study
Complication rates (infection, migration, foreign body reaction), growth effects
Indications for Use
The Stryker Universal Orbital Floor System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit.
Device Story
System consists of pre-bent titanium orbital floor plates, globe retractor, and plate holding forceps; used for reconstruction of orbital floor and/or medial wall. Plates are pre-contoured based on average anatomical model derived from 300 CT scans; surgeon can further trim and contour plates to fit specific patient anatomy. Device used in clinical setting by surgeons; plates implanted permanently via screws. System facilitates restoration of orbital volume; simplifies surgical fitting process. Benefits include effective reconstruction of orbital defects; risks include infection, migration, and foreign body reaction, consistent with established alloplastic implant profiles.
Clinical Evidence
No clinical performance testing performed. Substantial equivalence supported by literature review of analogous devices demonstrating that the 15-21 year old subpopulation is anatomically and physiologically equivalent to adults for orbital reconstruction purposes.
Technological Characteristics
Titanium plates; pre-bent based on average anatomical model; available in two sizes (small/large) and left/right configurations. Includes globe retractor and plate holding forceps. Permanent implant; non-sterile. No software or energy source.
Indications for Use
Indicated for the reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in patients 15 years of age and older.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Journal of Craniofacial Surgery 21(4), 1142-6 (Tabrizi et al., 2010)
Ophthalmic Plastic and Reconstructive Surgery 17(3): 174-9 (Hatton et al., 2001)
CMF Trauma and Reconstruction 2(2): 103-12 (Chao & Losee, 2009)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three faces overlapping to symbolize the department's focus on people. The logo is accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1, 2014
Stryker Leibinger GmbH & Co. KG c/o Ms. Julie Schoell, M.S., RAC Staff Regulatory Affairs Specialist Stryker Craniomaxillofacial 750 Trade Centre Way, Suite 200 Portage, MI 49002
Re: K133461
Trade/Device Name: Stryker Universal Orbital Floor System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: July 3, 2014 Received: July 7, 2014
Dear Ms. Schoell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known): K133461
Device Name: Stryker Universal Orbital Floor System
Indications For Use: Intended Use:
> The Stryker Universal Orbital Floor System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit.
#### Indication for Use:
The Stryker Universal Orbital Floor System is indicated for the reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in patients 15 years of age and older.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Andrew I. Steen -S 2014.08.01 12:27:10-04'00'
Page 1 of 1 -
{3}------------------------------------------------
Revised 510(k) Summary
{4}------------------------------------------------
| Submitter: | Stryker Leibinger GmbH & Co. KG<br>Bötzinger Straße 41<br>79111 Freiburg<br>Germany |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Julie Schoell<br>Staff Regulatory Affairs Specialist<br>Phone: (269) 389-3419<br>Fax: (877) 648-7114<br>Stryker Craniomaxillofacial<br>750 Trade Centre Way, Suite 200<br>Portage, MI 49002, USA |
| Date prepared: | July 28, 2014 |
| Proprietary Name: | Stryker Universal Orbital Floor System |
| Common Name: | Bone plate |
| Regulatory Class: | Class II |
| Regulation Number: | 872.4760 Bone Plate |
| Product Codes: | JEY |
| Predicate Device: | Stryker Universal Orbital Floor System, K123786 |
# 510/k) Summarv
## Introduction:
This Traditional 510(k) is being submitted to the FDA to grant clearance to market the Stryker Universal Orbital Floor System cleared via K123786 with the following modification.
## Proposed Modification:
The Stryker Universal Orbital Floor System was cleared via K123786 for reconstruction of the floor and/or medial wall of the orbit in adult patients. This Traditional 510(k) is being submitted for the same system to expand the currently cleared Indications for Use to allow use of the device in patients 15 years of age and older. The rationale for limiting the indications to those patients 15 years of age and older is that the orbital volume has reached its adult size at that age. Except for this proposed change in indications, the device remains unchanged from its earlier cleared version.
Confidential Proprietary July 2014
{5}------------------------------------------------
## Intended Use:
The Stryker Universal Orbital Floor System is intended to be used in the reconstruction of the floor and/or medial wall of the orbit.
## Indications for Use:
The Stryker Universal Orbital Floor System is indicated for the reconstructive treatment of orbital floor and/or medial wall trauma or bone excision in patients 15 years of age and older.
## Device Description:
The Stryker Universal Orbital Floor System includes pre-bent titanium orbital floor plates, a globe retractor and a plate holding forceps.
The pre-bent titanium plates are available in a small (L=31mm W=34mm, H=12mm) and a large size (L=35mm, W=36mm, H=16mm) along with left and right configurations. The plates can be trimmed along cutting lines and contoured to fit the specific needs of the patient.
The plates are designed based on an average anatomical model of CT-scan data taken from 300 subjects (92% Caucasian). The metadata of the 300 subjects that have been included in the generation of the average anatomical model is listed in Table 1. The selected scans were obtained from healthy subjects without any deformation of the bony orbital structures.
| Age [years] | | Gender | | Ethnic Group | |
|-------------|----|--------|-----|--------------|-----|
| 10-19 | 18 | f | 122 | ca | 276 |
| 20-29 | 19 | m | 178 | me | 3 |
| 30-39 | 20 | | | xx | 21 |
| 40-49 | 28 | | | | |
| 50-59 | 36 | | | | |
| 60-69 | 56 | | | | |
| 70-79 | 72 | | | | |
| 80-89 | 44 | | | | |
| 90-99 | 6 | | | | |
#### Table 1: Meta data of the 300 CT scans
ದ 59 years;
Gender: 122 female (f) and 178 male (m);
Ethnic Group: 276 Caucasians (ca), 3 Middle-East (me), 21 unknown (xx)
The globe retractor is an instrument designed to allow retraction of the orbital contents off the orbital floor and walls during implantation of the above mentioned pre-bent titanium plates.
{6}------------------------------------------------
Lastly, the plate holding forceps is an instrument designed specifically for use with above mentioned pre-bent titanium plates. They facilitate the insertion and positioning of the plate within the orbit.
## Comparison with the Predicate Device:
The Stryker Universal Orbital Floor System is compared to its predicate device for substantial equivalence based on the following criteria:
- A. Intended Use
- B. Principle of Operation
- C. Technological Characteristics
- D. Benefit-Risk Evaluation
## A: Intended Use/Indications for Use
## Table 2: Substantial Equivalence of Indications for Use
| | Subject Device | Predicate Device |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Intended Use:<br>The Stryker Universal Orbital<br>Floor System is intended to be<br>used in the reconstruction of<br>the floor and/or medial wall of<br>the orbit. | Intended Use:<br>The Stryker Universal Orbital<br>Floor System is intended to be<br>used in the reconstruction of<br>the floor and/or medial wall of<br>the orbit |
| | Indications for Use: | Indications for Use: |
| | The Stryker Universal Orbital<br>Floor System is indicated for<br>the reconstructive treatment of<br>orbital floor and/or medial wall<br>trauma or bone excision in<br>patients 15 years of age and<br>older. | The Stryker Universal Orbital<br>Floor System is indicated for<br>the reconstructive treatment of<br>orbital floor and/or medial wall<br>trauma or bone excision in<br>adult patients. |
The Indications for Use statement for the Stryker Universal Orbital Floor system is not identical to the predicate device; however, the differences do not alter the therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Available evidence from the literature demonstrates that the subpopulation of patients age 15 to 21 is anatomically and physiologically identical to the population greater than 22 years of age. Both the Subiect and Predicate Device have the same intended use for reconstruction of the floor and/or medial wall of the orbit.
## B. Principle of Operation
The basic operational principle of the Subject Device, as well as the Predicate Device, is to reconstruct the orbital floor and/or medial wall. The fixation method of both the Subject Device and the Predicate is with screws inserted through
{7}------------------------------------------------
dedicated screw holes. The Subject Device and the Predicate Device are both permanent implants and have the same craniomaxillofacial (CMF)/orbital area of application.
## C. Technological Characteristics
There is no design change; the Stryker Universal Orbital Floor System has the exact same design as the Predicate Device:
- -Same patient contacting surface: orbital floor and wall
- -Same area of contact and contact duration (tissue/bone/permanent implant)
- -Same material: titanium
- -Same design: pre-bent shape based on average anatomical model
- -Same sizes and shapes as the Predicate Device
- -Same mode of plate adjustment: trimming along cutting lines and bending
- -Both the Subject Device and Predicate Device are provided non-sterile
## D. Benefit-Risk Evaluation
The design of the device approximates the shape and dimensions of the average normal orbital floor and medial wall in order to facilitate the operative goal of restoring normal (pre-traumatic) orbital volume as accurately as possible. Additionally, this approximation simplifies the bending and fitting process for the surgeon during an orbital procedure, while still allowing the surgeon to trim and contour the material to fit the specific needs of the patient. A review of the literature shows that titanium mesh and other alloplastic implants have been used for orbital repair in a pediatric subpopulation of 15 to 21 years of age for over 25 years. In published studies on the use of permanent fixation in the craniomaxillofacial skeleton in a pediatric subpopulation of 15 to 21 years of age, only the detrimental effects on growth have been noted, with no other reported complications due to age. Complications associated with the use of alloplastic material in the pediatric population were no different than the risks identified for the Subject Device, which include infection, migration, and foreign body reaction.2 The complication rate in this age group is not shown to be greater than that seen in the 22 and over population. 3 Based on the risk assessment and data reported in the clinical literature, any adverse events are minimized and acceptable when weighed against the benefits of the intended performance.
## Substantial Equivalence Analysis:
The conclusions drawn from the comparison of the intended use, principle of operation, and technological characteristics, including the clinical performance data from the literature and benefit-risk evaluation, demonstrate that the Stryker Universal Orbital Floor System performs substantially equivalent to the Predicate Device. The differences in Indications for Use do not raise new questions of safety and effectiveness.
{8}------------------------------------------------
## Clinical Testing:
There was no clinical performance testing required because the differences in Indications for Use do not raise new questions of safety and effectiveness. The clinical performance data provided in the literature on analogous reference devices supports an indication in patients 15 years of age and older.
## Non-Clinical Testing:
The proposed modification does not have any impact on the performance of the device. Hence, no additional non-clinical testing has been performed. Previously conducted performance tests as described in K123786 included Verification and Validation (V&V) testing of biocompatibility, cleaning, sterilization, corrosion resistance, stability of plate, functionality over lifetime, transportation of the complete set and of packaging, an end product test and a design validation - end user test. The Stryker Universal Orbital Floor System passed all tests.
## Conclusions:
The results of the non-clinical data demonstrate the Stryker Universal Orbital Floor System will perform as intended in the specified use conditions. The intended use, technological characteristics, clinical performance data presented in the literature, and the benefit-risk evaluation supports that known and foreseeable risks, and any adverse events, are minimized and acceptable when weighed against the benefits of the intended performance of the device during normal conditions of use executed per the product labeling. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
## References
1. Tabrizi. R., Ozkan. T.B., Mohammadineiad. C., Minaee. N. (2010). Orbital floor reconstruction. Journal of Craniofacial Surgery. 21(4), 1142-6.
2. Hatton, M.P., Watkins, L.M., Rubin, P.A.D. (2001). Orbital Fractures in Children. Ophthalmic Plastic and Reconstructive Surgery. 17(3): 174-9.
3. Chao, M.T., Losee, J.E. (2009). Complications in Pediatric Facial Fractures. CMF Trauma and Reconstruction. 2(2): 103-12.
Confidential Proprietary July 2014
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.