K133335 · C.R. Bard, Inc. · LJT · Feb 14, 2014 · General Hospital
Device Facts
Record ID
K133335
Device Name
POWERPORT IMPLANTABLE PORT
Applicant
C.R. Bard, Inc.
Product Code
LJT · General Hospital
Decision Date
Feb 14, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K133335 · Feb 14, 2014
POWERPORT IMPLANTABLE PORT
C.R. Bard, Inc.
Medical records from five major oncology centers
Clinical data were collected from patients at five oncology centers to evaluate the safety of extending the maintenance flushing interval for Bard ports with Groshong catheters from four weeks to 90 days.
Retrospective analysis of clinical data; Follow-up/Duration: 28,452 patient-days
Post-infusional adult males or females ≥ 21 years of age with a Bard port implanted; Sample Size: 465 saline-only intervals; 28,452 patient-days of follow-up; Number of Sites: 5
Not applicable for this study
Rate of adverse events
Indications for Use
Power and Non Power-Injectable Implantable Ports are intended to be an implanted vascular access device designed to provide long-term, repeated access to the vascular system.
Device Story
Implantable vascular access system consisting of rigid housing, self-sealing septum, and distally-valved Groshong® catheter; implanted by physician. Provides repeated vascular access for infusion/withdrawal without daily external catheter care. PowerPort® models support power injection of contrast media when used with specific PowerLoc® safety infusion sets. Device remains implanted long-term; requires periodic maintenance flushing. Clinical study supported extending maintenance flushing interval from 4 weeks to 90 days using saline-only locks. Benefits include reduced frequency of maintenance procedures for patients requiring long-term therapy.
Clinical Evidence
Prospective clinical study at 5 US oncology centers involving adult patients (≥ 21 years) with implanted Bard ports. Evaluated safety of extending maintenance flushing interval to 90 days. Primary endpoint: adverse event rate. Results: 465 saline-only intervals, 28,452 patient-days, 1 adverse event recorded. Incidence rate 0.35/10,000 patient-days (95% CI: 0.01, 1.96).
Technological Characteristics
Implantable port with rigid housing and self-sealing septum; distally-valved Groshong® catheter. No changes to materials or design. Maintenance protocol extended to 90-day saline flush interval.
Indications for Use
Indicated for adult patients (≥ 21 years) requiring repeated vascular access for infusion of medications, I.V. fluids, parenteral nutrition, blood products, and blood withdrawal. PowerPort® models, when used with PowerLoc® Safety Infusion Set, are indicated for power injection of contrast media at rates up to 5 mL/s.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
PowerPort® Implantable Port with Groshong® Catheter (K081311)
PowerPort® Polymeric Port with 8 F Single Lumen ChronoFlex® Catheter (K063377)
{0}------------------------------------------------
(33335
Bard Access Systems, Inc. Ports
with Groshong® Catheter – Extended Maintenance Protocol
with Groshong® Catheter – Extended Maintenance Protocol
Traditional 510(k) Premarket Notification
# ACCESS SYS
## FEB 1 4 2014
Section 5 - 510(k) Summary
### 510(k) Summary 21 CFR 807.92(a)
| | Submitter Name: | Bard Access Systems, Inc. |
|-------------------------|--------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| General<br>Provisions | Address: | 605 North 5600 West<br>Salt Lake City, UT 84116 |
| | Contact Person: | Amy Honey<br>Regulatory Affairs Specialist |
| | Telephone Number: | (801) 522-5671 |
| | Fax Number: | (801) 522-5425 |
| | Date of Preparation: | October 15, 2013 |
| Subject<br>Device | Trade Name: | Power and Non Power-Injectable Implantable Ports with<br>Groshong® Catheter |
| | Common Name: | Implanted Infusion Port & Catheter |
| | Classification Name: | LJT - Subcutaneous, Implanted, Intravascular Infusion Port<br>& Catheter |
| | Product Code: | LJT |
| | Regulation: | 21 CFR §880.5965 |
| Predicate<br>Devices | Trade Name: | PowerPort® Implantable Port with Groshong® Catheter |
| | Common Name: | Implanted Infusion Port & Catheter |
| | Classification Name: | LJT - Subcutaneous, Implanted, Intravascular Infusion Port<br>& Catheter |
| | Product Code: | LJT |
| | Regulation: | 21 CFR §880.5965 |
| | Premarket Notification: | K081311 |
| | Trade Name: | PowerPort® Polymeric Port with 8 F Single Lumen<br>ChronoFlex® Catheter |
| | Common Name: | Implanted Infusion Port & Catheter |
| | Classification Name: | LJT - Subcutaneous, Implanted, Intravascular Infusion Port<br>& Catheter |
| | Product Code: | LJT |
| | Regulation: | 21 CFR §880.5965 |
| | Premarket Notification: | K063377 |
| Trade Name: | BardPort® Titanium Ports | |
| Common Name: | Implanted Infusion Port & Catheter | |
| Classification Name: | LJT - Subcutaneous, Implanted, Intravascular Infusion Port & Catheter | |
| Product Code: | LJT | |
| Regulation: | 21 CFR §880.5965 | |
| Premarket Notification: | K050310 | |
| Trade Name: | X-Port® Duo Port | |
| Common Name: | Implanted Infusion Port & Catheter | |
| Classification Name: | LJT - Subcutaneous, Implanted, Intravascular Infusion Port & Catheter | |
| Product Code: | LJT | |
| Regulation: | 21 CFR §880.5965 | |
| Premarket Notification: | K032044 | |
| Trade Name: | BardPort® X-Port® isp Port | |
| Common Name: | Implanted Infusion Port & Catheter | |
| Classification Name: | LJT - Subcutaneous, Implanted, Intravascular Infusion Port & Catheter | |
| Product Code: | LJT | |
| Regulation: | 21 CFR §880.5965 | |
| Premarket Notification: | K022983 | |
| Trade Name: | Plastic Low-Profile Subcutaneous Port | |
| Common Name: | Implanted Infusion Port & Catheter | |
| Classification Name: | LJT - Subcutaneous, Implanted, Intravascular Infusion Port & Catheter | |
| Product Code: | LJT | |
| Regulation: | 21 CFR §880.5965 | |
| Premarket Notification: | K924250 | |
| Trade Name: | Plastic Attachable Dual Port | |
| Common Name: | Implanted Infusion Port & Catheter | |
| Classification Name: | LJT - Subcutaneous, Implanted, Intravascular Infusion Port & Catheter | |
| Product Code: | LJT | |
| Regulation: | 21 CFR §880.5965 | |
| Premarket Notification: | K912702 | |
| Trade Name: | Cath-Tech Port Implantable Vascular Access System | |
| Common Name: | Implanted Infusion Port & Catheter | |
| Classification Name: | LJT - Subcutaneous, Implanted, Intravascular Infusion Port<br>& Catheter | |
| Product Code: | LJT | |
| Regulation: | 21 CFR §880.5965 | |
| Premarket Notification: | K880571 | |
| Trade Name: | Hickman Plastic Subcutaneous Port | |
| Common Name: | Implanted Infusion Port & Catheter | |
| Classification Name: | LJT - Subcutaneous, Implanted, Intravascular Infusion Port<br>& Catheter | |
| Product Code: | LJT | |
| Regulation: | 21 CFR §880.5965 | |
| Premarket Notification: | K873213 | |
{1}------------------------------------------------
Section 5 - 510(k) Summary
{2}------------------------------------------------
#### PowerPort®
PowerPort® Implantable Ports are designed to provide repeated access to the vascular system without the need for repeated venipuncture or daily care of an external catheter. Long-Term Implantable Ports consist of a rigid housing and a self-sealing septum. The catheters used with infusion ports are essentially the same design as externalized, stand-alone intravascular catheters. Groshong® catheters are attached to the port by the physician during implantation.
PowerPort® Implantable Ports can be used for routine vascular access using a noncoring access needle. However, for power injection procedures, PowerPort® ports must be accessed with a Bard PowerLoc® Safety Infusion Set (SIS) to create a powerinjectable system.
#### Device Description
#### Non Power-Injectable Ports
Long-Term Implantable Ports are designed to provide repeated access to the vascular system without the need for repeated venipuncture or daily care of an external catheter. Long-Term Implantable Ports consist of a rigid housing and a self-sealing septum. The catheters used with infusion ports are essentially the same design as externalized, stand-alone intravascular catheters. Groshong® catheters are attached to the port by the physician during implantation.
Long-Term Implantable Ports can be used for routine vascular access using a non-coring access needle.
{3}------------------------------------------------
、
| Intended Use | Power and Non Power-Injectable Implantable Ports are intended to be an implanted<br>vascular access device designed to provide long-term, repeated access to the vascular<br>system. | |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Indications For<br>Use | PowerPort®<br>The PowerPort® Implantable Port is indicated for patient therapies requiring repeated<br>access to the vascular system. The port system can be used for infusion of medications,<br>I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of<br>blood samples. | |
| | When used with a PowerLoc® Brand Safety Infusion Set, the PowerPort® device is<br>indicated for power injection of contrast media. For power injection of contrast media,<br>the maximum recommended infusion rate is 5 mL/s. | |
| | Non Power-Injectable Ports<br>The BardPort®, SlimPort®, and X-Port® implantable ports are indicated for patient<br>therapies requiring repeated access to the vascular system. The port system can be<br>used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood<br>products, and for the withdrawal of blood samples. | |
| Technological<br>Characteristics | There have been no changes to the technological characteristics or design of the Power<br>and Non Power-Injectable Implantable Ports with Groshong® Catheter; therefore, they<br>are substantially equivalent with respect to basic design and function of the of the of the<br>predicate devices. Extending the maintenance protocol from a four-week flushing<br>period to a 90-day flushing period does not impact the intended use, and has been<br>shown through clinical data not to raise any new questions regarding safety or efficacy. | |
| Safety &<br>Performance<br>Tests: Non-<br>Clinical | Verification and validation testing were not required for this submission. Safety and<br>efficacy of the change was supported through clinical data; no physical testing was<br>required. | |
| | User requirements and use-related risks have been evaluated in and mitigated through<br>a clinical study, as well as in the Instructions for Use. | |
{4}------------------------------------------------
.
| Safety &<br>Performance<br>Tests: Clinical | Bard currently recommends that Power and Non Power-Injectable Implantable Ports<br>with Groshong® Catheter be filled with normal saline after each use, and if the port<br>remains unused for long periods of time, the saline lock should be changed by flushing<br>at least every four weeks. Under Institutional Review Board oversight, clinical data were<br>independently collected from patients at five major oncology centers to provide clinical<br>evidence in support of extending the maximum recommended maintenance flushing<br>interval to at least every 90 days for Bard ports with distally-valved (Groshong®)<br>catheters.<br>The study was conducted at five sites in the United States. Study subjects consisted of<br>post-infusional adult males or females ≥ 21 years of age that had a Bard port implanted.<br>The primary endpoint was to compare the rate of adverse events at three time points in<br>Group A versus extended accession intervals in Group B and Group C.<br>Considering solely the saline-only intervals, the study demonstrated that approximately<br>57% of the collected patient-day data represented flushing with saline only. Overall,<br>there were 465 saline-only intervals recorded with 28,452 patient-days of follow-up.<br>Only one adverse event was recorded in the saline-only group, yielding an incidence<br>rate of 0.35 adverse events per 10,000 patient-days, with a 95% Poisson confidence<br>interval (0.01, 1.96).<br>Maintenance flushing of Groshong® port systems has been shown to be safe for a wide<br>variety of time intervals based on the very low adverse event rates noted in the clinical<br>data. An extension of the maintenance flushing schedule from its current label of every<br>four weeks to every 90 days with saline only is recommended. |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, technological characteristics, and safety and<br>performance testing, the subject Power and Non Power-Injectable Implantable Ports<br>with Groshong® Catheter met the minimum requirements for its intended use and is<br>substantially equivalent in design, materials, sterilization, principles of operation, and<br>indications for use to the predicate devices cited. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 14, 2014
Bard Access Systems, Incorporated Ms. Amy Honey Regulatory Affairs Specialist 605 North 5600 West Salt Lake City, UT 84116
Re: K133335
Trade/Device Name: PowerPort®, Non Power-Injectable Ports Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port And Catheter Regulatory Class: II Product Code: LJT Dated: October 18, 2013 Received: October 29, 2014
Dear Ms. Honey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Honey
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fdagov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/6/Picture/8 description: The image shows the name "Erin Freith-S" in a bold, sans-serif font. The name is written in black ink on a white background. The letters are evenly spaced and the name is easy to read. The text appears to be a logo or a brand name.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Section 4 - Indications for Use Statement
#### Indications for Use
510(k) Number (if known): K133335
Device Name:
PowerPort®
Indications for Use:
The PowerPort® Implantable Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
When used with a PowerLoc® safety infusion set, the PowerPort® device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 mL/s.
Device Name:
Non Power-Injectable Ports
Indications for Use:
The BardPort®, SlimPort®, and X-Port® implantable ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR §801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR §801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Richard C. 2014.02.27 10:55:21 -05'00'
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.