POWERPORT POLYMETRIC PORT WITH 8 FR S/L CHRONOFLEX CATHETER
K063377 · C.R. Bard, Inc. · LJT · Jan 25, 2007 · General Hospital
Device Facts
Record ID
K063377
Device Name
POWERPORT POLYMETRIC PORT WITH 8 FR S/L CHRONOFLEX CATHETER
Applicant
C.R. Bard, Inc.
Product Code
LJT · General Hospital
Decision Date
Jan 25, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerPort" Polymeric Port is a totally implantable vascular access device designed to provide long-term. repeated access to the vascular system long-term, repeated access to the vascular system.
Device Story
Totally implantable vascular access device; consists of plastic port and 8 Fr open-ended single lumen ChronoFlex® catheter. Triangular port design features three palpation bumps for identification; purple colorants added to port and catheter materials to signify power injectability. Used in clinical settings by physicians for long-term vascular access. Device facilitates infusion of medications, fluids, nutrition, blood products, and blood sampling. When paired with PowerLoc Safety Infusion Set, enables power injection of contrast media. Provides patient benefit through reliable, repeated vascular access while minimizing trauma associated with frequent venipuncture.
Clinical Evidence
Bench testing only. Verification testing performed according to protocols based on FDA guidance for implanted infusion ports and intravascular catheters. Standards utilized include BS/EN/ISO 10555-1, BS/EN/ISO 10555-3, AAMI/ANSI/ISO 11135, and AAMI/ANSI/ISO 10993-1. Performance data demonstrated equivalence to the predicate device.
Technological Characteristics
Totally implantable vascular access port made of polymer; 8 Fr open-ended single lumen ChronoFlex® catheter. Triangular form factor with palpation bumps. Sterilization via ethylene oxide (AAMI/ANSI/ISO 11135). Complies with ISO 10555-1/3 and NF S 94-370 standards. Biocompatibility evaluated per ISO 10993-1.
Indications for Use
Indicated for patients requiring repeated vascular access for infusion of medications, I.V. fluids, parenteral nutrition, blood products, and blood withdrawal. When used with PowerLoc Safety Infusion Set, indicated for power injection of contrast media at max rate 5 ml/s.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
PowerPort™ Implanted Titanium Port with 8 Fr ChronoFlex® Catheter (K060812)
Submission Summary (Full Text)
{0}------------------------------------------------
K06-337
Carpe 1 of 3
PowerPort™ Polymeric Port Traditional 510(k)
## Section 5
## PowerPort™ Implanted Polymeric Port with 8 Fr ChronoFlex® Catheter 510(k) Summary of Safety and Effectiveness 21 CFR 807.92(a)
JAN 2 5 2007
### 5.1 General Information
| Submitter Name: | Bard Access Systems, Inc. (BAS)<br>[Wholly owned Subsidiary of C.R. Bard, Inc.] |
|-----------------------------------------------|---------------------------------------------------------------------------------|
| Address: | 5425 W. Amelia Earhart Drive<br>Salt Lake City, UT 84116 |
| Telephone Number: | (801) 595-0700 ext. 5541 |
| Fax Number: | (801) 595-5425 |
| Contact Person: | Michaela Rivkowich |
| Date of Preparation: | November 6, 2006 |
| Registration Number: | 1720496 |
| Additional Registration Numbers:<br>C.R. Bard | 2212754 |
### 5.2 Subject Device Information
| Device Name: | PowerPort <sup>TM</sup> Implanted Polymeric Port with 8 Fr ChronoFlex <sup>®</sup> Catheter |
|-----------------------|---------------------------------------------------------------------------------------------|
| Trade Name: | PowerPort <sup>TM</sup> , ChronoFlex <sup>®</sup> |
| Common/Usual Name: | Implanted Infusion Port and Catheter |
| Classification Name: | Class II, 80 LJT |
| Classification Panel: | Subcutaneous, Implanted, Intravascular Infusion Port and Catheter<br>General Hospital |
### 5.3 Predicate Device Information
| Device Name: | PowerPort™ Implanted Titanium Port with 8 Fr ChronoFlex® Catheter |
|-----------------------|---------------------------------------------------------------------------------------|
| Trade Name: | PowerPort™, ChronoFlex® |
| Common/Usual Name: | Implanted Infusion Port and Catheter |
| Classification Name: | Class II, 80 LJT |
| Classification Panel: | Subcutaneous, Implanted, Intravascular Infusion Port and Catheter<br>General Hospital |
### 5.4 Intended Use
The intended use is the same as for the predicate device.
The PowerPort" Polymeric Port is a totally implantable vascular access device designed to provide
long-term. repeated access to the vascular system long-term, repeated access to the vascular system.
{1}------------------------------------------------
### Indications for Use ર્ડ
The indications for use are the same as those for the predicate device.
The PowerPort™ Implanted Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
When used with the PowerLoc" Safety Infusion Set, the PowerPort™ device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 m}/s.
### Device Description ર્સ્ડર્
The PowerPort™ device consists of a plastic port and an 8 Fr open-ended single lumen ChronoFlex® The rower of - device ocables of a pith a cathlock. The port is triangular in shape and has three palpation bumps to distinguish it as power injectable. Purple colorants have been added to the port and parpater materials to provide the device with an appearance that allows the placing physician to identify the PowerPort device as power injectable.
### Technological Comparison to Predicate Device 5.7
The technological characteristics of the PowerPort™ Polymeric Port and 8 Fr ChronoFlex® Catheter are The teoliners of the se of the predicate PowerPort™ Titanium Port and 8 Fr ChronoFlex® Catheter in terms of intended use, indications for use, application, user population, basic design, performance and labeling.
### 510(k) Substantial Equivalence Decision Tree 5.8
## New device is compared to Marketed Device?
- Pres. PowerPort™ Polymeric Port device is compared to the predicate PowerPort™ Titanium A Port device.
## Does the new device have the same indication statement and the same intended use as the predicate device?
- Yes. The intended use and the indications for use are the same as those for the predicate A device.
- Does the new device have the same technological characteristics, e.g. design, material, etc?
predicate PowerPort™ Titanium Port; however, the fundamental scientific technology of the device has not changed.
### Could the new characteristics affect safety or effectiveness?
- Yes. The design changes may affect safety or effectiveness of the device. A
## Do the new characteristics raise new types of safety and effectiveness questions?
- > No. There are no new types of safety and effectiveness questions.
## Do accepted scientific methods exist for assessing effects of the new characteristics?
- Yes. Testing was based on FDA guidance documents and recognized standards to evaluate A the device's performance:
- Guidance on 510(k) Submissions for Implanted Infusion Ports, dated October 1990 "
- Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term I Intravascular Catheters, dated 3/16/95
{2}------------------------------------------------
- BS/EN/ISO 10555-1: 1997, Sterile, single-use intravascular catheters, Part 1. General . requirements.
- ISO 10555-1: 1995, Sterile, single-use intravascular catheters, Part 1. General ' requirements, Amendment 1: 1999
- BS/EN/ISO 10555-3: 1997, Sterile, single-use intravascular catheters, Part 3. Central 트 venous catheters
- AAMI/ANSI/ISO 11135: 1994, Medical Devices -- Validation and Routine Control of 1 Ethylene Oxide Sterilization
- AAMI/ANSI/ISO 10993-1: 2003, Biological Evaluation of medical Devices Part 1: l Evaluation and Testing, and the FDA Modified ISO 10993 Test Profile
- NF S 94-370, French Standard, Surgical Implants, implantable catheter chambers, . intravenous, intraarterial, intraperitoneal, intrathecal and epidural use (April 1999)
### Are performance data available to assess effects of new characteristics?
- Yes. Verification testing was performed according to protocols based on the above-A referenced guidance documents recommendations and additional standards.
### Do performance data demonstrate equivalence?
- Yes. Performance data gathered in design verification testing demonstrated that the subject A PowerPort" Polymeric Port device is substantially equivalent to the predicate device, the PowerPort™ Titanium Port, K060812, concurrence date July 14, 2006.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo with a stylized eagle or bird symbol on the right side. To the left of the bird symbol, there is text arranged vertically, which reads 'DEPARTMENT OF HEALTH AND HUMAN SERVICES'. The logo appears to be a representation of a government agency or organization related to health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
C.R. Bard, Incorporated Mr. Michaela Rivkowich Associate Regulatory Affairs Manager Bard Access Systems, Incorporated 5425 West Amelia Earhart Drive Salt Lake City, Utah 84116
JAN 2 5 2007
Re: K063377
Trade/Device Name: PowerPort™ Implanted Polymeric Port Regulation Number: 880.5965 Regulation Name: Subcutaneous Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT Dated: November 6, 2006 Received: November 8, 2006
Dear Mr. Rivkowich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent(for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations offecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Rivkowich
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing proctice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
2/11-1-77-4 PowerPort™ Polymeric Port Traditional 510(k)
### Section 4
## Indications for Use
510(k) Number (if known): ______________
Device Name: PowerPort™ Implanted Polymeric Port
## Indications for Use:
The PowerPort" Implanted Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples,
When used with the PowerLoc " Safety Infusion Set, the PowerPort" device is indicated for power injection of contrast media. For power injection of contrast infusion rate is 5 ml/s.
Prescription Use ਮ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF
NEED OF AND AFFERED AFF AFFEDED) NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
::
Anthony D. Avistar
长463377
000016
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.