MYSIGN S PULSE OXIMETER, MYSIGN PC SOFTWARE (OPTIONAL), SOFTTIP LARGE R-3211-31, FINGER CLIP F-3211-31
Applicant
Envitec-Wismar GmbH
Product Code
DQA · Cardiovascular
Decision Date
May 2, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
MySign® S is a handheld pulse oximeter with accessory sensors indicated for continuous noninvasive monitoring of the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult and pediatric (excluding neonatal and infant) patients in hospitals, hospital-type facilities, and mobile transport units. For professional use only.
Device Story
Handheld pulse oximeter for continuous non-invasive monitoring of SpO2 and pulse rate; utilizes reusable/disposable sensors; illuminates arterially perfused tissue with two LED wavelengths; measures time-varying absorbance via photodiode; provides numeric SpO2/pulse rate displays, pulse bar/waveform, and audible/visual alarms; used in hospitals, clinical facilities, and mobile transport by healthcare professionals; includes PC software for retrospective data management, trend charting, and reporting; aids clinical decision-making through real-time saturation monitoring and alarm notification.
Clinical Evidence
Clinical testing performed under controlled hypoxia vs. co-oximetry reference. Results validate accuracy claims for SpO2 range of 70% to 100%. Testing conducted under institutionally approved protocol with informed consent.
Technological Characteristics
Handheld pulse oximeter; two-wavelength LED illumination and photodiode sensing; SpO2 range 1-100%; pulse rate range 0-300 BPM; audible/visual alarms (55-75 db(A)); biocompatible patient contact materials; software-based data management; compliant with IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, and IEC 60601-1-8.
Indications for Use
Indicated for continuous non-invasive monitoring of functional oxygen saturation (SpO2) and pulse rate in adult and pediatric patients (excluding neonates and infants) in hospitals, hospital-type facilities, and mobile transport units. For professional use only.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
K133064
MAY 0 2 2014
Page 1 of 5
### 510(k) Summary in accordance with 21 CFR 807.92
Submitted by: (1) (a)
EnviteC-Wismar GmbH by Honeywell Alter Holzhafen 18 Wismar 23966 Germany Tel.: +49-38 41 360 221 Fax: +49-38 41 360 222 Marko.Sproessel@Honeywell.com
Contact Person:
Position/Title:
R&D Manager
Marko Sproessel
Date of Preparation: May 2, 2014
- (2) Trade Name: EnviteC MySign® S Pulse Oximeter, Accessory SpO2 Sensors, and MySign® PC Software
Common/Classification Name: OXIMETER; EAR OXIMETER
Product Code(s):
Class:
Class II
(3) Predicate Device(s): Substantial Equivalence to:
DQA; 21 CFR §870.2700
DPZ; 21 CFR §870.2710
| K Number | Model | Manufacturer |
|----------|-----------------------------------------------------------------------------------------|---------------------------------------------|
| K070193 | Envitec OxiPen | EnviteC-Wismar GmbH, a<br>Honeywell Company |
| K991823 | Nellcor N-395 | Philips Medical Systems<br>(formerly HP) |
| K122290 | Envitec MySign PC Software<br>(listed accessory of MySign O<br>Oxygen measuring device) | EnviteC-Wismar GmbH, a<br>Honeywell Company |
Reason for Submission: New Device(s)
ನ
- Description of Device: (4)
The EnviteC MySign® S Pulse Oximeter is a handheld pulse oximeter monitor validated for use in portable applications, including out of hospital transport.
{1}------------------------------------------------
The MySign® S Pulse Oximeter features large readable numeric displays for SpO2 and pulse rate and a pulse bar indicator or waveform for visual assessment of pulsation. The device has settable alarms and an audible pulse tone which varies with saturation level.
The MySign® S is offered with a family of Reusable and Disposable SpO2 Sensors for adult and pediatric applications.
MySign® PC Software may be used to obtain retrospective monitoring data from MySign® series monitors and to perform utility functions such as setting the institution identification.
#### (5) Intended use:
The measurement of functional oxygen saturation of arterial hemoglobin (SpO2) has been a standard of care in the USA for 20 years. Applications for oximetry include monitoring in the anesthesia, recovery, and critical care environments, as well as transport monitoring and home care.
### Indications for Use:
MySign® S is a handheld pulse oximeter with accessory sensors indicated for continuous non-invasive monitoring of the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult and pediatric (excluding neonatal and infant) patients in hospitals, hospital-type facilities, and mobile transport units. For professional use only.
#### (6) Technological Characteristics:
The EnviteC MySign® S Pulse Oximeter and Accessory SpO2 Sensors employ the same technological characteristics as the listed predicate devices to determine arterial functional oxygen saturation: arterially perfused tissue is illuminated sequentially by two wavelengths of light emitted by light emitting diodes (LED's), and the time varying absorbance of the tissue is measured from a photodiode light sensor. This method is characteristic of all pulse oximeter monitors and sensors which are the subject of this submission as well as the predicate devices.
{2}------------------------------------------------
# Comparison of Technological Features to Predicate Device(s):
1
| | Envitec MySign® S<br>Pulse Oximeter Monitor | Envitec OxiPen Pulse<br>Oximeter | Nellcor N395 Pulse<br>Oximeter |
|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product/Feature | Envitec MySign® S<br>Pulse Oximeter Monitor | Envitec OxiPen Pulse<br>Oximeter | Nellcor N395 Pulse<br>Oximeter |
| Manufacturer | EnviteC-Wismar GmbH<br>by Honeywell | EnviteC-Wismar GmbH<br>by Honeywell | Nellcor, Inc<br>(now Covidien) |
| 510(k) Number | (pending this<br>submission) | K070193 | K991823 |
| Patient<br>Population | Adult through pediatric<br>(not infant or neonatal)<br>patients | Adult through pediatric<br>(not infant or neonatal)<br>patients | Adult through neonatal<br>patients |
| Indications: | MySign® S is a handheld<br>pulse oximeter with<br>accessory sensors<br>indicated for continuous<br>non-invasive monitoring<br>of the functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and<br>pulse rate for adult and<br>pediatric (excluding<br>neonatal and infant)<br>patients in hospitals,<br>hospital-type facilities,<br>mobile units, and home<br>environments. | The EnviteC OxiPen<br>Pulse Oximeter is<br>intended for noninvasive<br>spot-check measurement<br>of functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2), and<br>pulse rate (measured by<br>SpO2 sensor<br>accessories).<br>The monitor is intended<br>for use on adult and<br>pediatric patients in<br>hospitals, hospital-type<br>facilities, mobile, and<br>home environments. | The intended use of the<br>N-395 Pulse Oximeter<br>is the continuous, non-<br>invasive monitoring of<br>functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and<br>pulse rate. For use with<br>neonatal, pediatric and<br>adult patients, in<br>hospitals, hospital-type<br>facilities and intra-<br>hospital transport<br>environments. For<br>prescription use only. |
| Use | Continuous monitoring<br>of functional oxygen<br>saturation with alarms | Spot monitoring of<br>functional oxygen<br>saturation (no alarms) | Continuous monitoring<br>of functional oxygen<br>saturation with alarms |
| Measurement<br>principle | Two wavelength pulse<br>measurement to obtain<br>functional oxygen<br>saturation | Two wavelength pulse<br>measurement to obtain<br>functional oxygen<br>saturation | Two wavelength pulse<br>measurement to obtain<br>functional oxygen<br>saturation |
| SpO2 range | 1-100% | 0-100% | 1-100% |
| SpO2 accuracy | Monitor:<br>70-100%: ARMS =<br>$\pm$ 2 % (with SoftTip<br>R3211-12 MySign)<br>Accuracy claim varies<br>slightly by sensor type | Monitor:<br>70-100%: ARMS = 2.0<br>Accuracy claim varies<br>slightly by sensor type | Monitor:<br>70-100%: $\pm$ 2 digits<br>Neonatal: $\pm$ 3 digits<br>Saturation (%SpO2 $\pm$ 1<br>SD) |
| Pulse rate range | 0-300 BPM | 20-300 BPM | 20-250 BPM |
| Pulse rate<br>accuracy | $\pm$ 3 BPM (30-250 BPM) | $\pm$ 3 BPM (20-300 BPM) | $\pm$ 3 digits (20-250) |
| Product/Feature | Envitec MySign® S<br>Pulse Oximeter Monitor | Envitec OxiPen Pulse<br>Oximeter | Nellcor N395 Pulse<br>Oximeter |
| Available<br>Oximetry<br>Sensors<br>supported by.<br>IFU's | Finger Clip; SoftTip<br>Finger Sensor(s);<br>SoftTip+; Ear Clip;<br>Disposable<br>Adult/Pediatric | Finger Clip; SoftTip<br>Finger Sensor | Full range of Nellcor<br>reusable/disposable<br>sensors including,<br>Finger Clip, Ear Clip, Y<br>Sensor, Disposable<br>Adult-Neonatal |
| Visible &<br>Auditory Alarms | Audible Alarms/Alerts<br>55-75 db(A) | Audible alerts for sensor<br>off, disconnect, power<br>off | Audible Alarms/Alerts,<br>outputs vary by priority<br>(high, medium, low) |
| IFU structure | Comprehensive user<br>manual, separate<br>accessory instructions | Comprehensive user<br>manual, separate<br>accessory instructions | Comprehensive user<br>manual, separate<br>accessory instructions |
{3}------------------------------------------------
## Page 4 of 5
| Product/Feature | EnviteC MySign® PC Software (new<br>version to support MySign S) | EnviteC MySign® PC Software<br>(cleared version for MySign O) |
|--------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Manufacturer | EnviteC-Wismar GmbH by Honeywell | EnviteC-Wismar GmbH by Honeywell |
| 510(k) Number | (pending - this submission) | K122290 |
| Devices<br>Supported | MySign S (new)<br>MySign O (existing) | MySign O |
| PC Operating<br>System | Microsoft Windows XP or later,<br>Microsoft .NET 4.0 Framework | Microsoft Windows XP or later,<br>Microsoft .NET 3.5 or 4.0 Framework |
| Device<br>connection | USB Serial Port | USB Serial Port |
| Basic menu<br>functions | Dataset, PC Data, Patient Data,<br>Export to Excel, Save, Print (to PDF),<br>Delete, System Settings | Dataset, PC Data, Patient Data,<br>Export to Excel, Save, Print (to PDF),<br>Delete, System Settings |
| Parameters<br>Displayed | %SpO2, Pulse Rate, % O2, alarm<br>limits at time of measurement<br>(tabular or graphical view) | % O2, alarm limits at time of<br>measurement<br>(tabular or graphical view) |
| Data present-<br>ation options | Tabular (data listing)<br>Trend Chart | Tabular (data listing)<br>Trend Chart |
#### (b) (1) Non-Clinical Tests Submitted:
The EnviteC MySign® S Pulse Oximeter and Accessories were tested in accordance with current applicable standards for medical device electrical safety and electromagnetic compatibility and particular standards for pulse oximeter monitors, including the following recognized standards:
{4}------------------------------------------------
### Page 5 of 5
- IEC 60601-1, Medical Electrical Equipment Part 1, General Requirements . for Safety
- IEC 60601-1-2, Medical Electrical Equipment Collateral Standard. . Electromagnetic Compatibility Requirements & Tests
- . ISO 80601-2-61, Medical Electrical Equipment - Particular Requirements for the Basic Safety and Essential Performance of Pulse Oximeters.
- . IEC 60601-1- 8. Medical Electrical Equipment - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
Risk management and software validation was performed on the device, Including MySign® PC, Software, in accordance with established development processes per the following:
- . ANSI/AAMI/ISO 14971. Medical devices-Risk management-Application of risk management to medical devices
- . IEC 62304, Medical Device Software, Software Life Cycle Processes
- . FDA/ODE Guidance for the Content
Sensor patient contact materials meet applicable standards for biocompatibility.
#### (2) Clinical Tests Submitted:
Clinical testing was performed to validate the performance and accuracy of the MySign® S Pulse Oximeter with SpO2 Sensors under controlled hvpoxia versus arterial oxygen saturation as determined by co-oximetry. All testing was performed under an institutionally approved protocol with subject informed consent. Clinical test results support the stated accuracy claims for the specified range of 70% to 100% SaO2.
#### Conclusions from Tests: (3)
As described in (b)(1) and (b)(2) above, EnviteC MySign® S Pulse Oximeter and Accessories are equivalent to predicate devices as substantiated by parameter, bench, and clinical testing. Device safety is substantiated by testing to applicable standards and by biocompatibility of patient contact materials.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the caduceus symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W066-G60 Silver Spring, MD 20993-0002
### May 2, 2014
EnviteC-Wismar GmbH c/o Mr. Stephen Gorski Imagenix Incorporated S65 W35739 Piper Road Eagle, WI 53119
Re: K133064
Trade/Device Name: EnviteC MySign® S Pulse Oximeter, Accessory SpO2 Sensors, and MySign® PC Software Regulation Number: 21 CFR 870.2700 Regulation Name: OXIMETER Regulatory Class: Class II Product Code: DQA, DPZ Dated: March 28, 2014 Received: April 2, 2014
### Dear Mr. Gorski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Gorski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mary S. Runner -
S FDA
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
### 510(k) Number (if known): K133064
### Device Name:
MySign® S Pulse Oximeter and Accessories
### Indications for use:
MySign® S is a handheld pulse oximeter with accessory sensors indicated for continuous noninvasive monitoring of the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adult and pediatric (excluding neonatal and infant) patients in hospitals, hospital-type facilities, and mobile transport units. For professional use only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND / OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neel J. Patel-S
2014.05.02 15:14:52 '04'00'
Page 1 of __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.