The measurement of functional oxygen saturation of arterial hemoglobin (SpO2) has been a standard of care in the USA for 20 years. Applications for oximetry include monitoring in the anesthesia, recovery, and critical care environments, as well as transport monitoring and home care. The OxiPen device has no alarms or settable alarm limits and is intended for spot checking.
Device Story
Compact, battery-powered handheld pulse oximeter for spot-check monitoring. Inputs: optical signals from SpO2 sensors (two-wavelength LED illumination, photodiode detection). Operation: tissue absorbance measured; signal amplified, digitized, and processed to calculate SpO2 and pulse rate. Output: numeric SpO2 and pulse rate values on LCD, pulse bar graph, and audible pulse tone. Used in hospitals, mobile, and home settings by clinicians or patients. No alarms or settable alarm limits. Provides visual indicators for signal quality, sensor disconnection, and battery status. Benefits: portable, noninvasive assessment of oxygenation status.
Clinical Evidence
Clinical testing performed under an institutionally approved protocol compared OxiPen measurements against co-oximetry (gold standard) during controlled hypoxia. Results validated accuracy claims for the 70% to 100% SaO2 range. Bench testing included electrical safety, EMC, environmental, and pulse rate simulation using a listed simulator. Biocompatibility of patient-contact materials confirmed.
Technological Characteristics
Two-wavelength LED illumination (red/infrared) and photodiode sensor. Two-layer architecture: dedicated ChipOx subsystem for oximetry, separate controller for power/display/UI. LCD display with backlight. Battery-powered. Biocompatible patient-contact materials. Complies with electrical safety, EMC, and environmental standards.
Indications for Use
Indicated for noninvasive spot-check measurement of functional oxygen saturation (SpO2) and pulse rate in adult and pediatric patients in hospital, mobile, and home environments. Prescription use only.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
K070193
Page 1 of 3
# 510(k) Summary of Safety and Effectiveness in accordance with 21 CFR 807.92
| (a) | (1) | Submitted by: | EnviteC-Wismar GmbH<br>Alter Holzhafen 18<br>Wismar 23966<br>Germany<br>Tel.: +49-38 41 360 221<br>Fax: +49-38 41 360 222<br>E-mail: k.huebner@envitec.com | | | MAY - 2 2007 | |
|-----|-----|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------|--------------|--|
| | | Contact Person: | Mr. Klaus Huebner, Ph.D. | | | | |
| | | Position/Title: | Manager of R & D | | | | |
| | | Date of Preparation: | April 4, 2007 | | | | |
| | (2) | Trade Name: | EnviteC OxiPen Pulse Oximeter and Accessories | | | | |
| | | Common/Classification Name: | OXIMETER | | | | |
| | | Product Code(s): | DQA; 21 CFR §870.2700 | | | | |
| | | Class: | Class II | | | | |
| | (3) | Predicate Device(s): | Substantial Equivalence to: | | | | |
| | | K Number | Model | | Manufacturer | | |
| | | K963707 | NBP 40 Handheld Pulse<br>Oximeter | | Nellcor Puritan Bennett<br>(TYCO) | | |
| | | Reason for Submission: | New device | | | | |
- (4) Description of Device:
The OxiPen Pulse Oximeter is a compact battery-powered handheld monitor designed for simplicity and portability for spot checking applications. It does not provide alarms or settable alarm limits.
The OxiPen features large readable numeric displays for SpO2 and pulse rate and a pulse bar indicator for visual assessment of pulsation. Additional indicators include a signal quality LED and heart symbol which show pulsation, a sensor symbol for disconnection alert, and a battery capacity symbol.
{1}------------------------------------------------
## Page 2 of 3
Like the referenced predicate device, the OxiPen provides an audible pulse tone which varies with saturation level, as well as audible and visual alerts for conditions like sensor disconnect and power-off. The LCD display may be backlit by briefly pressing any key.
The OxiPen is provided with sensors specifically designed for use with the device, and features a compact connector with EnviteC proprietary pinout.
#### (5) Intended use:
The measurement of functional oxygen saturation of arterial hemoglobin (SpO2) has been a standard of care in the USA for 20 years. Applications for oximetry include monitoring in the anesthesia, recovery, and critical care environments, as well as transport monitoring and home care. The OxiPen device has no alarms or settable alarm limits and is intended for spot checking.
## Indications for Use:
The EnviteC OxiPen Pulse Oximeter is intended for noninvasive spotcheck measurement of functional oxygen saturation of arterial hemoglobin (SpO2), and pulse rate (measured by SpO2 sensor accessories).
The monitor is intended for use on adult and pediatric patients in hospitals, hospital-type facilities, mobile, and home environments.
Prescription device.
#### (6) Technological Characteristics:
The EnviteC OxiPen Pulse Oximeter and accessory sensors employ the same technological characteristics as the predicate device and other oximeters to determine functional arterial oxygen saturation - arterially perfused tissue is illuminated sequentially by two wavelengths of light emitted by light emitting diodes (LED's), and the time varying absorbance of the tissue is measured from a photodiode light sensor. The resultant electrical signal is amplified, digitized, and analyzed to determine the functional oxygen saturation and pulse rate. The information is displayed on a liquid crystal display with pulse bar graph and audible pulse tone.
The OxiPen utilizes a two-layer architecture in which the oximeter-specific functions are managed by a dedicated subsystem (ChipOx) and the monitor functions including power management, display, and user interface are controlled separately.
{2}------------------------------------------------
#### (b) (1) Non-Clinical Tests Submitted:
The monitor and accessory sensors were tested in accordance with applicable standards for medical device electrical safety, electromagnetic compatibility, environmental operation and storage conditions, resistance to moisture ingress, and shock and vibration. The device passed all of the tests.
The monitor and accessory sensors were tested for pulse rate with a listed simulator. The device passed all of the tests.
Embedded software in the device was verified to requirements and validated to meet intended use by software and system level performance testing.
Sensor patient contact materials meet applicable standards for biocompatibility.
### (2) Clinical Tests Submitted:
Clinical testing was performed to validate the performance and accuracy of the EnviteC OxiPen Pulse Oximeter and accessory sensors under controlled hypoxia versus arterial oxygen saturation as determined by cooximetry. All testing was performed under an institutionally approved protocol with subject confidentiality and informed consent. Clinical test results support the stated accuracy claims for the specified range of 70% to 100% SaO2.
#### (3) Conclusions from Tests:
As described in (b)(1) and (b)(2) above, the EnviteC OxiPen Pulse Oximeter and accessory sensors perform in a manner equivalent to the predicate device as substantiated by parameter, bench, and clinical testing. Device safety is substantiated by compliance testing to applicable standards and by biocompatibility of patient contact materials.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three bars representing its wings, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The logo is black and white and appears to be a simple, vectorized design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
EnviteC-Wismar GmbH C/O Mr. Stephen H. Gorski Imagenix, Incorporated S65 W35739 Piper Road Eagle, Wisconsin 53119
MAY - 2 2007
Re: K070193
Trade/Device Name: EnviteC OxiPen® Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: April 6, 2007 Received: April 10, 2007
Dear Mr. Gorski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Gorski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
### Device Name:
EnviteC OxiPen® Pulse Oximeter
### Indications for use:
The EnviteC OxiPen Pulse Oximeter is intended for noninvasive spot-check measurement of functional oxygen saturation of arterial hemoglobin (SpO2), and pulse rate (measured by SpO2 sensor accessories).
The monitor is intended for use on adult and pediatric patients in hospitals, hospital-type facilities, mobile, and home environments.
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Prescription Use _ × (Part 21 CFR 801 Subpart D) AND / OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Chellda
on Sign-Off) Collaion of Anesthesiology, General Hospital, Intection Control, Dental Devices
510(k) Number: K070193
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.