SPHINX JR

K132975 · Allmed Systems, Inc. · GEX · Mar 18, 2014 · General, Plastic Surgery

Device Facts

Record IDK132975
Device NameSPHINX JR
ApplicantAllmed Systems, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateMar 18, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sphinx Jr. laser system is intended for use in surgical procedures using open, laparoscopic incision, excision, resection, ablation, vaporization and hemostasis of soft and hard tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gasteroenterology, Arthroscopy, Discectory, Pulmonary, Cynecology, Plastic Surgery and General Surgery.

Device Story

Sphinx Jr. is a 2.1-micron surgical laser system for soft and hard tissue ablation, coagulation, dissection, and resection. System comprises a laser console, internal computer control panel, display, footswitch, and fiber port. Laser energy is delivered via standard silica fibers through a handpiece or endoscopic instrument. Used in OR or clinical settings by physicians across various specialties (Urology, ENT, Orthopedics, etc.). Device enables precise tissue interaction for procedures like lithotripsy, tumor resection, and disc decompression. Healthcare providers operate the console to control laser parameters, viewing output via the display; clinical decision-making relies on real-time visualization of tissue effects. Benefits include minimally invasive surgical options and effective hemostasis.

Clinical Evidence

No clinical data. Safety and effectiveness supported by bench testing, including software and EMC testing. Preclinical performance data was not required as specifications and performance are identical to predicate devices.

Technological Characteristics

2.1-micron wavelength surgical laser; silica fiber optic delivery; internal computer control; footswitch-operated. Class II device (21 CFR 878.4810).

Indications for Use

Indicated for surgical incision, excision, resection, ablation, vaporization, coagulation, and hemostasis of soft and hard tissue across multiple specialties including Urology, Gastroenterology, Arthroscopy, Pulmonary, Gynecology, ENT, Dermatology, Plastic, and General Surgery. Specific applications include lithotripsy, tumor ablation, disc decompression, and dermatological lesion removal. Contraindications: None explicitly stated, though spinal surgery is excluded from general arthroscopy indications.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Attachment V # 510(k) Summary K132 975 | 1.General Information | | |-----------------------|-----------------------------------------------------------------------------------------------------| | Submitter: | AllMed Systems Inc.<br>9232 Klemetson Drive<br>Pleasanton CA 94588 | | Phone: | 925-468-0433 | | Fax | 925-399-5984 | | Contact Person | Peter Allen | | Date Prepared | 17th September 2013 | | 2. Names | | | Device Name | Sphinx Jr Laser | | Common Name | laser surgical instrument | | Classification Name | 79GEX | | CFR Reference | 21 CFR 878.4810 Laser surgical instrument for use in<br>general, plastic and dermatological surgery | | 3. Predicate Device | | Sphinx Family of lasers, and Lumenis - Versapulse # 4. Product Description The Sphinx Jr. system is a surgical laser system operating at a wavelength of 2.1 micron. The purpose of the laser is the ablation, coagulation, dissection and resection of soft tissue. The laser is designed for open surgery and surgical applications in aqueous media and non aqueous medium. The laser power is delivered via standard silica laser fibers. The distal tip is guided by a handpiece endoscopic surgical instrument or similar approved device {1}------------------------------------------------ It consists of: Laser Console with Internal Computer Control Panel and Display Operating Software Fiber port Footswitch #### 5. Indications for Use The Sphinx Jr laser system is intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gasteroenterology, Arthroscopy, Discetomy Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery. #### Urology Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: Urethral Strictures Bladder Neck Incisions (BNI) Ablation and resection of Bladder Tumors. Uretheral Tumors and Ureteral Tumors. Ablation of Benign Prostatic Hypertrophy (BHP), Transurethral incision of the prostate (TUIP) Holmium Laser Resection of the Prostrate (HoLRP) Holmium Laser Enuculeation of the Prostate (HoLEP) Holmium laser Ablation of the Prostate (HoLAP) Condvlomas Lesions of external genitalia Lithotripsy and Percutaneous Urinary Lithotripsy Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones. Endoscopic fragmentation of kidnev calculi Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed. # Gasteroenterlogy Open and endoscopic gasteroenterlogy surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis ) including: Appendectomy Polyps Biopsy {2}------------------------------------------------ Gall Bladder calculi Biliary/Bile duct calculi Ulcers Gastric ulcers Duodenal ulcers Non Bleeding Ulcers Pancreatitas Hemorrhoids Cholecystectomy Benign and Malignant Neoplasm Angiodysplasia Colorectal cancer Telangiectasias Telangiectasias of the Osler-Weber-Renu disease Vascular Malformation Gastritis Esophagitis Esophageal ulcers Varices Colitis Mallory-Weiss tear Gastric Erosions #### Arthroscopy Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but includina: Ligament and tendon Release Contouring and sculpting of articular surfaces Capsulectomy in the Knee Chondreplasty in the Knee Debridement of inflamed synovial tissue Chondromalacia Ablation Chondromalacia and tears Plica Removal Meniscectomy Loose Body Debridement Lateral retinecular release Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including Percutaneous Laser Disc Decompression/Discectomy of the L4-5 and L5-S1 lumbar discs, including Foraminoplasty 3 Lumbar discs, including Foraminoplasty Percutaneous Cervical Disc Decompression/Disce Percutaneous Cervical Disc Decompression/Discectomy Percutaneous Thoracic Disc Decompression/Discectomy ## Thoracic and Pulmonary Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) {3}------------------------------------------------ #### Gynecology Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue ## ENT Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and cartilage) including: Endonasal/sinus Surgery Partial turbinectomy Polypectomy Dacryocystorhinostomy Frontal Sinusotomy Ethmoidectomy Maxillary antrostomy Functional endoscopic sinus surgery #### Dermatology and Plastic Surgery Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft.mucosal, fatty and cartilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including: Basal Cell Carcinomas Lesions of skin and subcutaneous tissue Skin tags Plantar warts Lesions of skin and subcutaneous tissue Port Wine Stains Papillomas #### General Surgery Open, laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including: Appendectomv Skin incision Excision of external and internal lesions Complete of partial resection of internal organs, tumors and lesions Biopsy {4}------------------------------------------------ #### Safety and Effectiveness of the Sphinx Jr. ## Rationale for Substantial Equivalence The Sphinx Jr systems with fiber optic delivery devices share the same intended use, indications for use, similar design features and functional features and therefore are substantially equivalent to the Sphinx Family of Lasers K033437 and Versapulse K011703. # Non Clinical Testing All necessary Software and EMC testing was conducted on the proposed Sphinx Jr to support a determination of safety to the predicate devices. #### Performance Since the specification and performance of the Sphinx laser system and the predicate devices are the same, it is suggested that preclinical performance data is not required #### Conclusion The Sphinx Jr Systems with fiber optic delivery devices was found to be safe and effective and therefore substantially equivalent to the predicate surgical laser systems and delivery devices. {5}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 March 18, 2014 AllMed Systems, Inc. Peter N. Allen 9232 Klemetson Drive Pleasanton, California 94588 Re: K132975 Trade/Device Name: Sphinx Jr. Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: February 10, 2014 Received: February 25, 2014 Dear Mr. Allen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act, The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {6}------------------------------------------------ Page 2 - Peter N. Allen forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Felipe Aguel - Binita Ashar, MD, MBA, FACS for Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use 510(k) Number (if known) K132975 Device Name Sphinx Jr Laser System #### Indications for Use (Describe) The Sphinx Jr. laser system is intended for use in surgical procedures using open, laparoscopic incision, excision, resection, ablation, vaporization and hemostasis of soft and hard tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gasteroenterology, Arthroscopy, Discectory, Pulmonary, Cynecology, Plastic Surgery and General Surgery #### Urology Open and endoscopic surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including: Urethral Strictures Bladder Neck Incisions Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors. Ablation of Benign Prostatic Hypertrophy (BHP) Resection of the Prostrate Condylomas Lesions of external genitalia Lithouripsy and Percutaneous Urinary Lithotripsy Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi Endoscopic fragmentation of kidney calculi Treatment of steinstrasse when guide wire cannot be passed #### Gasteroenterlogy Open and laparoscopic gynecological surgery (incision, resection, ablation, vaporization, cought:ation and hemostasis) including Appendectomy Polyps Biopsy Gall Bladder calculi Biliary/Bile duct calculi Uicers Gastric ulcers Duodenal ulcers Pancreatitas Hemorrhouds Cholecystectomy Benign and Malignant Neoplasm Angiodysplasia FORM FDA 3881 (1/14) {8}------------------------------------------------ Colorectal cancer Telangicctasias Vascular Malformation Gastritis Esophagitis ## Arthroscopy Arthroscopy/Orthopedic surgery (excision, ablation of soft and cartilaginous tissue) excluding the spinc but including: Ligament and tendon Release Contouring and sculpting of articular surfaces Capsulectomy in the Knee Chondreplasty in the Knee Debridement of inflamed synovial tissue Chondromalacia Ablation Chondromalacia and tears Plica Removal Meniscectomy Loose Body Debridement Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including Percutaneous Laser Disc Decompression/Discectomy Foraminoplasty Pulmonary Open and endoscopic surgery (incision, escision, resection, coagulation, coagulation and hemostasis) Gynecology Open and laparoscopic gynecological surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including: Outer female genitals Ablation of condilomata and genital warts Inner genitalia Endometrium ablation Treatment of Uterine polyps Treatment of endometriosis # ENT Endoscopic endonasal surgery (incision, excision, ablation, vaporization, coagulation and hemostasis of soli tissue) including: Endonasal/sinus Surgery Partial turbinectomy Polypectomy Dacryocystorhinostomy Ethmoidectomy {9}------------------------------------------------ Dermatology and Plastic Surgery Incision, excision, resection, ablation, coagulation and hemostasis of soft,mucosal, fatty and carilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including: Basal Cell Carcinomas Lesions of skin and subcutancous tissue Skin tags Plantar warts Lesions of skin and subcutaneous tissue General Surgery Open laparoscopic and cndoscopic surgery (incision, resection, ablation, vaportization, coagulation and hemostasis) including: Appendectomy Skin incision Excision of external and internal lesions Complete of partial resection of internal organs, turnors and lesions Biopsv Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) # Neil R Ogden -S 2014.03.18 13:44:01 -04'00' For BSA This section applies only to requirements of the Paperwork Reduction Act of 1995, # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...