EQUASHIELD CLOSED SYSTEM DRUG TRANSFER DEVICE (CSTD)
K132899 · Equashield Medical , Ltd. · ONB · May 13, 2014 · General Hospital
Device Facts
Record ID
K132899
Device Name
EQUASHIELD CLOSED SYSTEM DRUG TRANSFER DEVICE (CSTD)
Applicant
Equashield Medical , Ltd.
Product Code
ONB · General Hospital
Decision Date
May 13, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
Closed System drug Transfer Device (CSTD) for preparation, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress.
Device Story
Equashield is a multi-component Closed System drug Transfer Device (CSTD) used in hospitals/clinics by pharmacists and healthcare professionals. System components include a piston syringe unit, vial adaptor, infusion bag spike adaptors, luer lock adaptors, and protective plugs. Operation relies on a double-membrane method; when components join, membranes press together and are pierced by needles to create a leak-proof, drug-residual-free connection. The syringe unit features an airtight, sealed barrel isolating the plunger rod and interior; an integrated closed pressure equalization chamber containing sterile air prevents aerosol/vapor escape. The system requires no priming and is not a gravity-feed system. Accessories include Y-site tubing and secondary tubing with drip chambers. By mechanically prohibiting environmental contaminant transfer and drug/vapor escape, the device minimizes exposure to hazardous drugs and prevents microbial ingress. The system is single-use, non-pyrogenic, and provided sterile.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing and performance data demonstrating compliance with recognized consensus standards, including ISO 10993 for biological evaluation (cytotoxicity, blood interaction), ISO 594 for luer fittings, and ISO 23908 for sharps injury protection. The system was substantiated to prevent microbial ingress for up to 7 days and supports multiple component reconnections up to 10 times.
Technological Characteristics
Multi-component system; components include piston syringe, vial/spike/luer adaptors, and protective plugs. Materials evaluated per ISO 10993-1, -4, -5. Features double-membrane (septum) connection mechanism; integrated closed pressure equalization chamber; needle-stick safe design. Sterile, single-use, EO sterilized (SAL 10^-6). No DEHP plasticizer in tubing. Complies with ISO 594-1/2 (luer tapers), ISO 7886-1 (syringes), ISO 8536-4 (infusion equipment), and ISO 23908 (sharps protection).
Indications for Use
Indicated for use by licensed pharmacists and healthcare professionals for the preparation, compounding, and administration of drugs, including antineoplastic and hazardous drugs, in hospital and clinical settings.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
# Medical Ltd.
POB 12 , Tefen Industrial Park Migdal Tefen 24959 | Israel www.equashieid.com
#### 510(k) Summary K132899
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Date Prepared: May 12, 2014
Applicant Equashield Medical Ltd. POB 12 , Tefen Industrial Park Migdal Tefen 24959 , Israel
Contact Person Raymond Kelly Licensale Inc. 57 Lazy Brook Rd Monroe, CT 06468 USA Phone: (203) 880-4091
Alternate Contact Dorit Valal Projects Manager Equashield Medical Ltd. POB 12 , Tefen Industrial Park Migdal Tefen 24959 . Israel Phone: +972-49873737 (ex. 114) Cell: +972-549952810
Device Information Trade Name: Equashield Closed System drug Transfer Device (CSTD) Model Numbers: Generations 1 & 2: (VA-x, SU-x, SA-x, LL-x, MC-x, FC-x, PP-x) Classification Name: Closed Antineoplastic and Hazardous Drug Reconstitution and Transfer System Review Panel: General Hospital Product Code: ONB Device Class: Class II Regulation: 21 C.F.R. §880.5440
### Predicate Device Information
K101940 Equashield Closed System, cleared November 19, 2010 K 130197 PhaSeal Closed System, cleared February 27, 2013
{1}------------------------------------------------
### Intended Use/Indications for Use
Closed System drug Transfer Device (CSTD) for safe preparation, reconstitution, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress.
### Principle of Operation
Equashield (first and second generation) is a device which has been substantiated to prevent microbial ingress up to a time period of 7 days and has been substantiated for multiple reconnections of components up to 10 times. Equashield is a Closed System drug Transfer Device (CSTD) which does not require priming, does not include an Insertion Tube, and is not a "Gravity Feed" System. Equashield does contain accessories which contain tubing, one accessory uses "drip therapy" but no tubing contains plasticizer DEHP. Equashield is a single use, nontoxic, non-pyrogenic fluid path system which is provided sterile as long as the package is intact, undamaged, and protective caps are secure. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress. System contains filters and needles using a needle stick safe design,
### Device Description
The Equashield system consists of a piston syringe (Syringe Unit), an adaptor to the medication vial (Vial Adaptor), an adaptor for the infusion bag for injection (Spike Adaptor 1), an adaptor for the infusion bag for withdrawal (Spike Adaptor W), an adaptor for female luer lock ports (Luer Lock Adaptor), nursing pair connectors for male and female luer lock ports (Female LL Connector and Male LL Connector), and a protective plug (Protective Plug). The Equashield closed system consists of 8 components that are dedicated to each other to create a system. The components are not intended to be used individually. Vial Adaptor is 48 x 31mm, 23mm at distal end. Plug is 51 x 17mm large. Female luer lock is 50 x 23 mm with a 5.2mm ID. Male luer lock is 42 x 21mm with a 15mm depth. Spike Adaptor is 75 x 42 mm and the height is 22mm.
### Accessories Description
Y-Site Tubing - this accessory consists of a Luer Lock Adaptor permanently attached to a Y-Site. Short pieces of tubing extend from each of the other two Y-Site ports, creating one tubing line. One tube is terminated with a female luer lock and the other is terminated with a male luer lock. The Y-Site Tubing Accessory is intended to be attached in series to an infusion line and the Equashield Luer Lock Adaptor enables safe connection and injection with the Equashield Syringe Unit or Female Connector. Priming volume: 1.9ml, length 25cm, Y-angle 30 degrees.
{2}------------------------------------------------
Spike Adaptor Secondary Tubing - consists of a Spike Adaptor-1 for injection with a drip chamber at its outlet. The drip chamber with the attached tubing, the roller clamp and the male luer lock at the tubing end, all constitute standard secondary infusion tubing sets. The integration of the standard infusion tubing set with the Equashield Spike Adaptor provides a closed system protection during injection of drugs into the infusion bag and during their administration. It reduces the assembly steps and the risk of exposure to drugs associated with such assembly procedures. Length is 100 cm, priming volume 7.5ml.
### Performance
No performance standards or special controls have been developed under Section 514 of the Federal Food, Drug, and Cosmetic Act ("FD&C Act") for Closed System Transfer Devices (CSTD). However, the Equashield System Generation 2, its accessories, and its components follow FDA recognized consensus standards:
- . Recognition Number 6-11; ISO 594-1:1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements
- . Recognition Number 6-129: ISO 594-2:1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
- . Recognition Number 6-170: ISO 7886-1:1993, Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
- Recognition Number 6-245: ISO 8536-4:2010, Infusion equipment for medical use Part . 4: Systems for single use, gravity feed
- Recognition Number 2-156: ISO 10993-1:2009 Biological evaluation of medical devices . -- Part 1: Evaluation and testing within a risk management process
- ISO 10993-4:2002, Biological evaluation of medical devices -- Part 4: Selection of tests . for interactions with blood
- Recognition Number 2-153: ISO 10993-5:2009, Biological evaluation of medical devices . -- Part 5: Tests for In Vitro cytotoxicity
- . Recognition Number 5-40: ISO 14971:2007, Medical devices - Application of risk management to medical devices
- . ISO 22413:2010, Transfer sets for pharmaceutical preparations -- Requirements and test methods
- Recognition Number 6-273: ISO 23908:2011, Sharps injury protection Requirements . and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
### Substantial Equivalence
Equivalence was determined using a side by side tabular comparison between the predicate and proposed devices which included: Features, Intended Use, Labeling, Materials, Specifications, Performance Data, and Technological Aspects
{3}------------------------------------------------
| Predicate Device (K13019 | Same | system and the escape of drum vapor contractions outside of drug system thereby minimizing ndividual and environments<br>System is an airtight and leak-<br>sposure to drug vapor, aeroso and spills. The system also prevents microbial ingress.<br>vironmental contaminants | Class II | 888.544( | ONB |
|---------------------------|-----------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------------|------------|
| Predicate Device (K101940 | Same | nacists or other healthcar<br>fessionals to prepare drugs<br>ned system to be used | Class II | 888.5440 | r H |
| Proposed Device | Same | | | 888.544( | ONB |
| | tended Us | dications for<br>Use | assificatio | Regulation Number | roduct Cod |
SE Table of Attributes
Page 32
{4}------------------------------------------------
| | Proposed Device | Predicate Device (K101940) | Predicate Device (K130197) |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Vial Adaptor | Vial Adaptor | Vial Adaptor ("Protector") |
| | Syringe Unit | Syringe Unit | Syringe connector ("Injector"<br>with female luer lock) |
| | Spike Adaptor I- for injection | Spike Adaptor I- for injection | Spike Adaptor ("Infusion<br>Adaptor") |
| | Spike Adaptor W- for withdrawal | Spike Adaptor W- for withdrawal | |
| | Luer Lock Adaptor | Luer Lock Adaptor | Male "Connector Luer Lock" |
| System<br>components | Male Luer Lock Connector | Male Luer Lock Connector | Male "Connector Luer Lock" |
| | Female Luer Lock Connector | Female Luer Lock Connector | Female luer lock connector<br>("Injector") |
| | Protective Plug | Protective Plug | NA |
| | Y-Site Tubing<br>(Accessory) | NA | Luer Lock Adaptor with Y-site<br>tubing ("Y-Site Connector") |
| | Secondary Tubing<br>(Accessory) | NA | Spike adaptor with secondary<br>tubing ("Secondary Set with Drip<br>Chamber") |
| Characteristics | Closed System used for<br>antineoplastic and hazardous drug<br>reconstitution, transfer and<br>administration, in order to prevent<br>contamination of the surrounding<br>environment and of the drug | Closed System used for<br>antineoplastic and hazardous drug<br>reconstitution, transfer and<br>administration, in order to prevent<br>contamination of the surrounding<br>environment and of the drug | Closed System used for<br>antineoplastic and hazardous drug<br>reconstitution, transfer and<br>administration, in order to prevent<br>contamination of the surrounding<br>environment and of the drug |
| Principles of<br>Operation | Multi-component system.<br>Components are intended to be used<br>as a system. | Multi-component system.<br>Components are intended to be used<br>as a system. | Multi-component system.<br>Components are intended to be<br>used as a system. |
| Technological<br>Characteristics | Proposed Device | Predicate Device (K101940) | Predicate Device (K130197) |
| | A leak-proof connector with a<br>single use syringe permanently<br>attached to it as part of the system. | A leak-proof connector with a<br>single use syringe permanently<br>attached to it as part of the system. | A leak-proof connector<br>("Injector") that needs to be<br>connected to an off the shelf<br>single use syringe. |
| | All system components are sealed<br>with resealing membranes<br>(Septum). When components are<br>joined together the two membranes<br>are pressed together and then<br>pierced by needles. System has<br>integrated closed pressure<br>equalization. | All system components are sealed<br>with resealing membranes<br>(Septum). When components are<br>joined together the two membranes<br>are pressed together and then<br>pierced by needles. System has<br>integrated closed pressure<br>equalization. | All system components are sealed<br>with resealing membranes<br>(Septum). When components are<br>joined together the two<br>membranes are pressed together<br>and then pierced by needles.<br>System has integrated closed<br>pressure equalization. |
| | The system syringe is closed from<br>all sides. The syringe barrel is<br>sealed airtight also at its rear end,<br>thereby isolating the plunger rod<br>and the interior of the barrel. | The system syringe is closed from<br>all sides. The syringe barrel is<br>sealed airtight also at its rear end,<br>thereby isolating the plunger rod<br>and the interior of the barrel. | NA |
| | Proposed Device | Predicate Device (K101940) | Predicate Device (K130197) |
| Secondary Tubing<br>("Secondary Set with Drip Chamber") | Spike Adaptor with drip chamber and secondary tubing attached. | NA | Spike Adaptor ("infusion adaptor") with drip chamber and secondary tubing attached. |
| System | A fully encapsulated Syringe Unit is the active transfer device of this closed system.<br><br>The syringe barrel is sealed airtight also at its rear end, thereby isolating the plunger rod and the interior of the barrel. A leak-proof connector is permanently welded to the syringe. The closed pressure equalization system with a chamber containing sterile air is built-in the Syringe Unit and makes the system airtight consequently containing all aerosols, particles and vapors.<br><br>For transfer of fluids the Syringe Unit connects to the passive Vial Adaptors, infusion bag and infusion tubing adaptors of the system, using the double membrane method to create a leak-proof and drug residual-free connection. | A fully encapsulated Syringe Unit is the active transfer device of this closed system.<br><br>The syringe barrel is sealed airtight also at its rear end, thereby isolating the plunger rod and the interior of the barrel. A leak-proof connector is permanently welded to the syringe. The closed pressure equalization system with a chamber containing sterile air is built-in the Syringe Unit and makes the system airtight consequently containing all aerosols, particles and vapors.<br><br>For transfer of fluids the Syringe Unit connects to the passive Vial Adaptors, infusion bag and infusion tubing adaptors of the system, using the double membrane method to create a leak-proof and drug residual-free connection. | The system uses a double membrane to create a leak-proof connection. The closed expansion chamber makes the system airtight consequently containing all aerosols, particles and vapors. The expansion chamber also equalizes the pressure within the system. It can handle most types of IV sets and syringes to enable a totally safe and convenient administration of drugs. |
| Device Type | Rx/Single Use | Rx/Single Use | Rx/Single Use |
| Target Users | Licensed Pharmacists/Health Care Professionals | Licensed Pharmacists/Health Care Professionals | Licensed Pharmacists/Health Care Professionals |
| Environment | Hospitals and clinics | Hospitals and clinics | Hospitals and clinics |
| Sterilization | EO/SAL 10-6 | EO/SAL 10-6 | EO/SAL 10-6 |
{5}------------------------------------------------
...
.
Page 34
the country of the county of
and the comments of the country of
{6}------------------------------------------------
:
:
:
.
ﺮ
.
Page 35
{7}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### May 13, 2014
Equashield Medical Ltd. c/o Raymond J Kelly IV Arazy Group 57 Lazy Brook Rd MONROE, CT 06468
Re: K132899
Trade/Device Name: Equashield Closed System drug Transfer Device (CSTD) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: II Product Code: ONB Dated: April 14, 2014 Received: April 17, 2014
Dear Mr. Kelly,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2 - Mr. Kelly
forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
· Sincerely yours,
## Mary S. Runner -S
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K 132899
#### Device Name
Equashield Closed System drug Transfer Device (CSTD)
### Indications for Use (Describe)
Closed System drug Transfer Device (CSTD) for preparation, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
. . . .
### Digitally signed by Richard C. Chapman Date: 2014.05.13 14:21:18 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.