K101940 · Equashield Medical , Ltd. · LHI · Nov 19, 2010 · General Hospital
Device Facts
Record ID
K101940
Device Name
EQUASHIELD SYSTEM PROTECTIVE PLUS
Applicant
Equashield Medical , Ltd.
Product Code
LHI · General Hospital
Decision Date
Nov 19, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The EQUASHIELD™ System reconstitution and transfer system is a contained system to be used by pharmacists or other healthcare professionals to prepare drugs, including cytotoxic drugs for intravenous infusion or injection. The EQUASHIELD™ System has an additional Protective Plug accessory which functions as a feature that mechanically prohibits the entrance of any objects into the EQUASHIELD™ System by physically blocking and protecting the system's connector ports.
Device Story
EQUASHIELD System Protective Plug acts as a mechanical barrier for EQUASHIELD connector ports; prevents accidental damage or penetration by foreign objects during transportation/conveyance of system components. Used in hospital settings by pharmacists or healthcare professionals. Plug is inserted into connector openings to physically block access. Device ensures integrity of closed drug transfer system during handling; protects against contamination or mechanical damage. Benefits include enhanced safety and system protection during routine hospital logistics.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via conveyance simulation under actual use conditions, including visual tests, detachment tests, and fluid flow analysis to ensure system integrity post-assembly.
Technological Characteristics
Mechanical protective plug; acts as a physical barrier/cork for connector ports. Materials, design, and performance are identical to predicate EQUASHIELD components. Non-powered, passive accessory.
Indications for Use
Indicated for pharmacists or healthcare professionals to prepare drugs, including cytotoxic drugs, for intravenous infusion or injection using a contained system.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for EQ Ashield Medical Ltd. The logo is black and white and features the text "EQ" with a curved line underneath. To the right of "EQ" is the text "ASHIELD Medical Ltd."
Bequashield com
## Section 13-510(K) SUMMARY
NOV 1 9 2010
## 510(K) SUMMARY EQUASHIELD™ SYSTEM PROTECTIVE PLUG
Date of summary: October 31, 2010
510(k) Number K101940/S2
- Applicant's Name: EQUASHIELD Medical Ltd. Tefen Industrial Park P.O.B 12, ISRAEL Tel: (972)4-987-3737 Fax: (972)4-987-3001
- Contact Person: Elissa Burg Tefen Industrial Park P.O.B 12, ISRAEL , 24959 Tel: +972-4-987-3737; Fax: +972-4-987-3001 Email: elissa@equashield.com Trade Name: EQUASHIELDTM System Protective Plug
Common name: Closed drug transfer system
Classification: Name: Intravascular administration set Product Code: LHI Regulation No: 880.5440 Class: II Classification Panel: General hospital
New EQUASHIELD™ System Protective Plug is substantially equivalent to Predicate Devices: the original EQUASHIELD™ System for the preparation and administration of parenteral drugs, cleared under 510(k) number K083152 and to the EQUASHIELD™ Luer Lock Connector Pair cleared under 510(k) number K091389.
> EQUASHIELD Medical Ltd. | Protective Plug Special 510k, Appendix D Page 2 of 5 - CONFIDENTIAL -
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for EQ Ashield Medical. The logo consists of the letters "EQ" in bold, followed by a curved line that connects to the word "Ashield" in bold. Below "Ashield" is the word "Medical" in a smaller font size. The logo is simple and professional, and it is likely used to represent a medical company.
Elissa Burg
Tel: +972 4 9873737
Fax: +972 4 9873001
elissa@equashield.com
### Device Description:
The EQUASHIELD™ System Protective Plug is used as a "cork" for blocking EQUASHIELD™ connector openings, thus providing the system with further protection, mainly during the transportation and conveying of the EQUASHIELD™ System together with other accompanied items (used routinely by the hospital), where EQUASHIELD™ components might be squeezed in a transportation container together with other items which could accidently cause damage due to penetration of such items into the EQUASHIELD™ System's connector ports. The Protective Plug is inserted into the EQUASHIELD™ connector ports' openings and mechanically prohibits entrance of any accompanied items.
#### Indication for Use Statement:
The EQUASHIELD™ System reconstitution and transfer system is a contained system to be used by pharmacists or other healthcare professionals to prepare drugs, including cytotoxic drugs for intravenous infusion or injection.
The EQUASHIELD™ System has an additional Protective Plug accessory which functions as a feature that mechanically prohibits the entrance of any objects into the EQUASHIELD™ System by physically blocking and protecting the system's connector ports.
### Technological characteristics and Substantial Equivalence:
The comparison table provided in our original submission's section 10 provides a detailed comparison between our new device and its predicates in order to support substantial equivalence.
### The comparison table refers to:
Intended use; System components; Characteristics; Principles of operation; Technological characteristics; Sharps injury prevention feature; Target users; Intended environment of use; sterilization; materials and biocompatibility.
The table shows that in some features the Protective Plug is equivalent (due to the addition of the Protective Plug) to the predicates and in other features it is identical.
## EQUASHIELD Medical Ltd. | Protective Plug Special 510k, Appendix D Page 3 of 5 - CONFIDENTIAL -
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for EQ Shield Medical. The logo is in black and white and features the text "EQ ASHIELD" in a bold, sans-serif font. The "A" in "ASHIELD" is connected to the "Q" in "EQ" by a curved line. Below "ASHIELD" is the word "Medical" in a smaller font.
Tel: +972 4 98737 Fax: +972 4 9873001 a@equashield.com
The new device differs from the predicates only in the fact that it is used as an additional accessory to the entire system and provides an additional benefit in protecting the system components' ports.
In all other features the new device is identical to the predicates as following described:
Both new and predicate devices are the same in terms of components design, materials, device characteristics, mechanical features, connections type, performances and technology features as the legally marketed EQUASHIELD™ System (K083152) and EQUASHIELD 101 Luer Lock Connector Pair (K091389).
Therefore we have concluded that the EQUASHIELD™ System with Protective Plug accessory is substantially equivalent to the original EQUASHIELD™ System that was previously cleared under 510(k) number K083152 and the EQUASHIELD™ Luer Lock Connector Pair that was cleared under 510(k) number K091389.
The evaluation and tests results of the differences raised, showed that the new EQUASHIELD™ Protective Plug accessory is as safe and as effective as the predicate devices.
### Non clinical performance data:
A conveyance simulation was performed by exposing the new device to the actual use conditions. Then, it was tested with accordance to the specification requirements in order to verify that the EQUASHIELD The Protective Plug device performs as intended and without compromising the safe use of the entire EOUASHIELD™ system (K083152) and EQUASHIELD The Luer Lock Connector Pair (K091389) which were used as predicate devices.
The tests that were conducted included: Visual tests; detachment tests; fluid flow through the entire system.
All functional tests results, post mimicking the actual conditions, met the specification predetermined acceptance criteria and demonstrated that the EQUASHIELD™ System components functioned appropriately after Protective Plug assembly and detachment and that it prevented the penetration of accompanying parts into EQUASHIELD™ ports during conveying.
We therefore believe that the tests results support all our labeling claims, substantial equivalency requirements and device's intended use.
EQUASHIELD Medical Ltd. | Protective Plug Special 510k. Appendix D Page 4 of 5 - CONFIDENTIAL -
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for EQ Ashield Medical Ltd. The logo features the letters "EQ" in a bold, sans-serif font, with a curved line extending from the bottom of the "Q" to the base of the "A" in "Ashield." The words "Medical Ltd." are printed in a smaller font size below "Ashield."
Flieca Rur Tel: +972 4 9873737 Fax: +972 4 9873001 elissa@equashield.com
### Conclusions:
The evaluation of EQUASHIELD Medical's new device's performance tests demonstrated that the device performs as intended and that it is as safe and as effective as the predicate devices.
In the light of the above, we believe it is substantially equivalent to EQUASHIELD Medical's legally marketed devices; EQUASHIELD™ System (K083152) and the EQUASHIELD™ Luer Lock Connector Pair (K091389) and appropriate for its intended use.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, topped by a cross. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Elissa Burg Quality Assurance and Regulatory Affairs Manager Equashield Medical Limited Tefen Industrial Park P.O. B. 12, Israel, 24959
NOV 1 9 2010
Re: K101940
Trade/Device Name: EQUASHIELD™ System with Protective Plug Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: November 3, 2010 Received: November 5, 2010
· Dear Ms. Burg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Ms. Burg
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for EQ Ashield Medical. The logo consists of the letters "EQ" in bold, followed by a curved line that connects to the word "ASHIELD" in bold. Below the word "ASHIELD" is the word "Medical" in a smaller font. The logo is simple and modern, and the use of bold fonts makes it easy to read.
issamenuashield com
# Appendix D - Indications for Use Statement
INDICATIONS FOR USE STATEMENT
NOV 1 9 2010
510(k) Number (if known):KJOl Y YO
Device Name: EQUASHIELD TM System with Protective Plug
Indications for Use:
The EQUASHIELD™ System reconstitution and transfer system is a contained system to be used by pharmacists or other healthcare professionals to prepare drugs, including cytotoxic drugs for intravenous infusion or injection.
The EQUASHIELD™ System has an additional Protective Plug accessory which functions as a feature that mechanically prohibits the entrance of any objects into the EQUASHIELD™ System by physically blocking and protecting the system's connector ports.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-off)
Division of Division of General, Restorative and Neurological Devices
11/17/10
510(k) Number
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Antonio Dental Devices
EQUASHIELD Medical Ltd. | Protective Plug ? nal 510k - CONFIDENTIAL Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.